  "results": [
    {
      "cfres_id": "190058",
      "product_res_number": "Z-0312-2022",
      "event_date_initiated": "2021-10-11",
      "event_date_created": "2021-11-30",
      "recall_status": "Terminated",
      "event_date_terminated": "2023-08-10",
      "res_event_number": "88934",
      "product_code": "PEN",
      "k_numbers": [
        "K193519"
      ],
      "product_description": "BioFire Blood Culture Identification 2 (BCID2) Panel\nPart No: RFIT-ASY-0147\nUDI: 00815381020338\n\nBlood Culture Catalog No./Description:\n442023/BD BACTEC Plus Aerobic medium in plastic vials\n442021/BD BACTEC Lytic Anaerobic medium in plastic vials\n442192//BD BACTEC Plus Aerobic medium in glass vials\n442022/BD BACTEC Plus Anaerobic medium in plastic vials\n442020/BD BACTEC Peds Plus mediums in plastic vials\n\nAll lots with an expiry date prior to",
      "code_info": "Blood Culture Identification 2 (BCID2) Panel Part Number: RFIT-ASY-0147\nUDI Code: 00815381020338\n\nBlood Culture Media Catalog No./Description:\n442023/BD BACTEC Plus Aerobic medium in plastic vials\n442021/BD BACTEC Lytic Anaerobic medium in plastic vials\n442192//BD BACTEC Plus Aerobic medium in glass vials\n442022/BD BACTEC Plus Anaerobic medium in plastic vials\n442020/BD BACTEC Peds Plus mediums in plastic vials\n\nAll BCID2 Panel lots BD Blood Culture Media vials with expiration date of April 30, 2022 or earlier.",
      "firm_fei_number": "3002773840",
      "recalling_firm": "BioFire Diagnostics, LLC",
      "address_1": "515 S Colorow Dr",
      "city": "Salt Lake City",
      "state": "UT",
      "postal_code": "84108-1248",
      "additional_info_contact": "800-735-6544 Ext. 5",
      "reason_for_recall": "Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results.  The cause may be as a due to non-viable contamination in the blood culture vials.",
      "root_cause_description": "Material/Component Contamination",
      "action": "On or about 10/10/2021, the Recalling Firm sent an \"URGENT: MEDICAL DEVICE CORRECTION\" Notification Letter to U.S. customer via overnight FedEx to inform them that there is an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results when the blood culture identification panels are used with some types of blood culture vials.  The increased risk are potential due to an increased level of nucleic acid from non-viable Acinetobacter calcoaceticus- baumannii complex targets in the blood culture vials.\n\nAffected Blood Culture Media vials are:\n442023 - BD BACTEC Plus Aerobic medium in plastic vials;\n442021 - BD BACTEC Lytic Anaerobic medium in plastic vials;\n442192 - BD BACTEC Plus Aerobic medium in glass vials;\n442022 - BD BACTEC Plus Anaerobic medium in plastic vials;\n442020 - BD BACTEC Peds Plus mediums in plastic vials;\n\nCustomers are instructed to:\n1) If the BCID2 Panel is used to test BD BACTEC\" blood culture vials with an expiration date of April 30, 2022 or earlier, positive results for Acinetobacter calcoaceticus- baumannii complex should be confirmed by another method prior to reporting the test results.\n2) Complete the accompanying Field Action Acknowledgment of Receipt Form and return to the Recalling Firm via one of the listed return methods so that the Recalling Firm may acknowledge their receipt of this notification.\n\nAny questions or concerns, contact customer support department at Support@biofiredx.com or via telephone by dialing 1.800.735.6544 and selecting option 5 for Product Technical Support.",
      "product_quantity": "Total 14,140 Kits [11,155 kits (U.S.) and 2,985 kits (O.U.S.)]",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The Countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dominican Rep/, Ecuador, Egypt, \nFrance, Georgia, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Israel\nItaly, Kuwait, Latvia, Macedonia, Malaysia, Mexico, Netherlands, Panama, Paraguay, \nPeru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, \nShanghai, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey,\nUnited Kingdom, Uruguay, Utd/Arab Emir/.",
      "openfda": {
        "k_number": [
          "K132843",
          "K213236",
          "K130914",
          "K243759",
          "K182619",
          "K160457",
          "K181493",
          "K243013",
          "K193519",
          "K143171",
          "K190341"
        ],
        "registration_number": [
          "3002773840",
          "3010749841",
          "3006028115",
          "3008632402"
        ],
        "fei_number": [
          "3002773840",
          "3010749841",
          "3006028115",
          "3008632402"
        ],
        "device_name": "Gram-Negative Bacteria And Associated Resistance Markers",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.3365",
        "device_class": "2"
      }
    }
  ]
}