{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "189114",
      "product_res_number": "Z-0005-2022",
      "event_date_initiated": "2021-09-02",
      "event_date_posted": "2021-10-14",
      "recall_status": "Terminated",
      "event_date_terminated": "2024-07-23",
      "res_event_number": "88626",
      "product_code": "QJR",
      "product_description": "Alinity m Resp-4-Plex AMP Kit (Emergency Use Authorization, US Assay) and Alinity m Resp-4-Plex Application Specification File. in vitro diagnostic",
      "code_info": "List Number: 09N79-096, UDI 00884999049390;\nLot/Serial, Expiration:\n\n514074, 09/15/2021; \n519243, 05/10/2022; \n519243, 5/10/2022; \n522088, 8/4/2022; \n522128, 8/26/2022; \n523152, 10/10/2022; \n524433, 1/15/2023; \n526710, 1/21/2023; \n527845, 2/8/2023; \n\nList Number 09N79-05D, UDI (01) 00884999050242(240 )09N79-05D(8012)4.oo",
      "firm_fei_number": "3005248192",
      "recalling_firm": "Abbott Molecular, Inc.",
      "address_1": "1300 E Touhy Ave",
      "city": "Des Plaines",
      "state": "IL",
      "postal_code": "60018-3315",
      "additional_info_contact": "Ray Bastian\n224-668-3866",
      "reason_for_recall": "There is a potential for false positive results.",
      "root_cause_description": "Device Design",
      "action": "The firm issued an Urgent Field Safety Notice/Field Correction Recall 09/02/2021 to all impacted customers.  Per the customer letter, all positives are to be considered presumptive until the Alinity m SARS-CoV-2 Application Specification file version 5 or the updated Alinity m Resp-4-Plex Application Specification file is implemented at the customer site.  \nOn 03/30/2022, Abbott Molecular issued a follow-up Urgent Field Safety Notice/Field Correction Recall letter to all customers who ordered Alinity m SARS-CoV-2 AMP Kit Emergency Use Authorization (EUA), List Number 09N78-095 and/or Alinity m Resp-4-Plex AMP Kit EUA, List Number 09N79-096 used with the Alinity m System.  Also to customers who ordered Alinity m SARS-CoV-2 AMP Kit (CE), List Number 09N78-090 or 09N78-091 and/or Alinity m Resp-4-Plex AMP Kit (CE), List Number 09N79-090 used with the Alinity m System.  \nThe Customer Letter explains Abbott has updated the existing Application Specification files for Alinity m SARS-CoV-2 and Resp-4-Plex to further reduce the potential for carryover. FDA has not yet authorized the Alinity m SARS-CoV-2 Application Specification file version 6 update or the Alinity m Resp-4-Plex Application Specification file version 5 update. Implementation of the modification for Alinity m SARS-CoV-2 and Resp-4-Plex will occur when the update to the Application Specification file has been authorized.  \nA Molecular Diagnostics Abbott Representative will be contacting you once the updated Application Specification files are available.  \nIf you have any questions, please contact your local Molecular Diagnostics Abbott representative or call Abbott Technical Support at 1-800-553-7042 Option 2.",
      "product_quantity": "8469 units",
      "distribution_pattern": "Worldwide Distribution. US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, COLOMBIA, CZECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ISRAEL, ITALY, JAPAN, MALAYSIA, MONTENEGRO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, UNITED KINGDOM, and VIETNAM.",
      "openfda": {
        "device_name": "Reagents, 2019-Novel Coronavirus Nucleic Acid",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N",
        "registration_number": [
          "3007420875",
          "2243471",
          "1221359",
          "3016619649",
          "3025663244",
          "2024800",
          "3024756665",
          "3013019728",
          "3016838963",
          "3032562",
          "3017358793",
          "3017485950",
          "3014214006",
          "3009238284",
          "9615778",
          "3038610665",
          "3013540184",
          "3004141078",
          "9610126",
          "1063851",
          "3009014878",
          "3004013603",
          "3010607747",
          "3017638293",
          "3013160357",
          "1528450",
          "3016608638",
          "3004146366",
          "2031468",
          "3016994837",
          "3005248192",
          "3009875001",
          "3004530258",
          "3003902721",
          "3008632402",
          "3014553740",
          "2023365",
          "1119779",
          "3019837962",
          "3009613136",
          "3009207631",
          "3003335080",
          "3024647994",
          "1524213",
          "3008517993",
          "3009976420",
          "3011104381",
          "3012918440",
          "2032682",
          "3002792284",
          "3003795116",
          "3010770794",
          "3010749841",
          "3012758946",
          "3003517954"
        ],
        "fei_number": [
          "3007420875",
          "2243471",
          "1000122536",
          "1221359",
          "3016619649",
          "3025663244",
          "2024800",
          "3024756665",
          "3013019728",
          "3016838963",
          "3017358793",
          "3002806559",
          "3017485950",
          "3014214006",
          "3009238284",
          "3038610665",
          "3013540184",
          "3004141078",
          "3002606748",
          "3009014878",
          "3004013603",
          "3010607747",
          "3017638293",
          "3013160357",
          "1000442352",
          "3016608638",
          "3004146366",
          "3016994837",
          "3005248192",
          "3009875001",
          "3004530258",
          "1000351505",
          "3003902721",
          "3008632402",
          "3014553740",
          "2023365",
          "1119779",
          "3019837962",
          "3009613136",
          "3009207631",
          "3003335080",
          "3024647994",
          "1524213",
          "3008517993",
          "3009976420",
          "3003952975",
          "3011104381",
          "3012918440",
          "3002023342",
          "3002792284",
          "3003795116",
          "3010770794",
          "3010749841",
          "3012758946",
          "3003517954"
        ]
      }
    }
  ]
}