  "results": [
    {
      "cfres_id": "184583",
      "product_res_number": "Z-0600-2021",
      "event_date_initiated": "2020-11-24",
      "event_date_posted": "2020-12-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2022-09-07",
      "res_event_number": "86804",
      "product_code": "DYB",
      "k_numbers": [
        "K142829"
      ],
      "product_description": "Flexor Check-Flo Introducer, RPN KCFW-6.0-35-70-RB-HFANL0-HC, Order Number G56221\n\n Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.",
      "code_info": "Lot Numbers: 13143098, 13143099, 13143100, 13145952, 13145953, 13145955, 13150444, 13150446, 13159696, 13159697, 13159706, 13159708, 13166164, 13166165, 13166168, 13175473, 13175474, 13175480, 13178582, 13186336, 13192068, 13192074, 13211826, 13211827, 13214440, 13217125, 13217126, 13221802, 13223830, 13223831, 13229680, 13231823, 13231824, 13240026, 13240027, 13243203, 13246547, 13246566, 13252353, 13260022, 13260023, 13265330, 13265331, 13265335, 13268038, 13279080, 13281540, 13281800, 13281801, 13286656, 13286658, 13286659, 13286660, 13286662, 13288115, 13304447, 13306333, 13306334, 13307178, 13307189, 13315200, 13315217, 13320585, 13320594, 13329232, 13329248, 13332167, 13334774, 13334775, 13342868, 13349026, 13352485, 13366188, 13366191, 13366194, 13366195, 13366196, 13369415, 13369426, 13380379, 13382712, 13391138, 13391139, 13396128, 13404831, 13404832, 13416497, 13420104, 13420105, 13281802X, NS13315219, NS13315220, NS13329249, NS13334773, NS13349037, NS13352484, NS13352486, NS13401845, NS13426336",
      "firm_fei_number": "1820334",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "city": "Bloomington",
      "state": "IN",
      "postal_code": "47404-9120",
      "reason_for_recall": "There is an increased likelihood of the introducer sheath separating at the proximal bond site.  If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.",
      "root_cause_description": "Process control",
      "action": "The firm, Cook Medical, sent \"URGENT: MEDICAL DEVICE RECALL\" letters dated 11/24/20 to customers. The letter described the product, problem and actions to be taken. The customers were instructed to take the following actions:\n1.\tExamine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product.\n2.\tReturn the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited.\n3.\tPlease complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com).\n4.\tThis medical device recall letter must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. \n5.\tImmediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. \nAction being Taken by Cook Medical\nCook Medical is removing devices potentially impacted from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue.\n\nOther Information\nThis action is being taken with the knowledge of the Food and Drug Administration. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA:\n\"\tVisit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to",
      "product_quantity": "4408",
      "distribution_pattern": "US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belgium, Canary Islands, CN, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Italy, KR, Latvia, Lithuania, MY, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, TW, TH, and United Arabs Emirates.",
      "openfda": {
        "k_number": [
          "K911794",
          "K101604",
          "K122431",
          "K955820",
          "K061015",
          "K062852",
          "K122960",
          "K935077",
          "K855126",
          "K912961",
          "K000313",
          "K082580",
          "K081046",
          "K902477",
          "K050023",
          "K181088",
          "K252309",
          "K081697",
          "K820834",
          "K231828",
          "K911883",
          "K063182",
          "K171275",
          "K020834",
          "K940079",
          "K955785",
          "K001367",
          "K230631",
          "K171853",
          "K112407",
          "K934901",
          "K873394",
          "K863917",
          "K820710",
          "K823262",
          "K140287",
          "K120428",
          "K792131",
          "K770561",
          "K171557",
          "K132789",
          "K031583",
          "K043525",
          "K853100",
          "K926340",
          "K903463",
          "K871208",
          "K895044",
          "K153494",
          "K800356",
          "K121521",
          "K191871",
          "K160254",
          "K112398",
          "K130855",
          "K180913",
          "K162427",
          "K140641",
          "K001135",
          "K121404",
          "K890277",
          "K960445",
          "K001346",
          "K170373",
          "K930194",
          "K034061",
          "K132164",
          "K253741",
          "K233646",
          "K171820",
          "K895580",
          "K914090",
          "K192313",
          "K252508",
          "K940092",
          "K920779",
          "K190628",
          "K120617",
          "K123618",
          "K160272",
          "K162184",
          "K040150",
          "K090114",
          "K020563",
          "K932733",
          "K170668",
          "K923603",
          "K261242",
          "K112887",
          "K102540",
          "K130730",
          "K863139",
          "K860307",
          "K001985",
          "K020090",
          "K912664",
          "K230704",
          "K040713",
          "K093791",
          "K260942",
          "K142445",
          "K191546",
          "K870907",
          "K072719",
          "K003424",
          "K945616",
          "K013202",
          "K111417",
          "K954234",
          "K051468",
          "K855061",
          "K100799",
          "K945396",
          "K152225",
          "K113196",
          "K202330",
          "K891022",
          "K172331",
          "K051515",
          "K792171",
          "K213638",
          "K101399",
          "K894430",
          "K792170",
          "K041506",
          "K222184",
          "K021723",
          "K934285",
          "K210685",
          "K881825",
          "K873727",
          "K970201",
          "K232260",
          "K181817",
          "K852880",
          "K870542",
          "K954595",
          "K062858",
          "K221440",
          "K150257",
          "K260839",
          "K894343",
          "K102008",
          "K170997",
          "K200258",
          "K243061",
          "K910861",
          "K962746",
          "K913801",
          "K191190",
          "K132030",
          "K182243",
          "K173831",
          "K090999",
          "K120212",
          "K122084",
          "K213002",
          "K991647",
          "K240004",
          "K123714",
          "K800620",
          "K981909",
          "K211530",
          "K972570",
          "K011761",
          "K140917",
          "K854831",
          "K972684",
          "K142684",
          "K201489",
          "K861951",
          "K190184",
          "K161659",
          "K973840",
          "K854559",
          "K190098",
          "K923681",
          "K780532",
          "K910933",
          "K885179",
          "K812502",
          "K864705",
          "K210185",
          "K901391",
          "K072130",
          "K771041",
          "K113110",
          "K212708",
          "K022513",
          "K964034",
          "K964814",
          "K240600",
          "K233708",
          "K170775",
          "K101150",
          "K983023",
          "K952828",
          "K240589",
          "K100819",
          "K130687",
          "K884494",
          "K971608",
          "K930057",
          "K131148",
          "K030944",
          "K171999",
          "K131300",
          "K190974",
          "K031922",
          "K130559",
          "K171988",
          "K130633",
          "K800580",
          "K082356",
          "K073434",
          "K043084",
          "K140543",
          "K073100",
          "K861984",
          "K883933",
          "K771040",
          "K091329",
          "K010913",
          "K172980",
          "K152044",
          "K870189",
          "K030905",
          "K023937",
          "K060519",
          "K173399",
          "K052644",
          "K102948",
          "K020964",
          "K852917",
          "K920327",
          "K133114",
          "K920908",
          "K091006",
          "K152498",
          "K844840",
          "K873195",
          "K101784",
          "K984500",
          "K904475",
          "K250468",
          "K141969",
          "K140028",
          "K873100",
          "K061363",
          "K012854",
          "K140513",
          "K181463",
          "K091367",
          "K163268",
          "K963886",
          "K202226",
          "K003889",
          "K111943",
          "K121185",
          "K152054",
          "K042623",
          "K120459",
          "K132720",
          "K000824",
          "K092014",
          "K882221",
          "K903906",
          "K910688",
          "K864352",
          "K261529",
          "K915078",
          "K822834",
          "K984062",
          "K242169",
          "K170020",
          "K122190",
          "K813019",
          "K893658",
          "K131079",
          "K960406",
          "K790034",
          "K203459",
          "K941428",
          "K842259",
          "K871153",
          "K081045",
          "K181855",
          "K140793",
          "K942977",
          "K920251",
          "K911678",
          "K051096",
          "K990975",
          "K843642",
          "K143255",
          "K831751",
          "K852918",
          "K004026",
          "K153533",
          "K070865",
          "K033681",
          "K233432",
          "K082063",
          "K252441",
          "K812964",
          "K961513",
          "K083269",
          "K123974",
          "K830813",
          "K111606",
          "K903840",
          "K241708",
          "K002054",
          "K122926",
          "K070166",
          "K110450",
          "K002847",
          "K002507",
          "K231227",
          "K781954",
          "K844593",
          "K042036",
          "K854941",
          "K951313",
          "K252364",
          "K201116",
          "K181031",
          "K254279",
          "K933522",
          "K254248",
          "K860958",
          "K843118",
          "K894445",
          "K243493",
          "K260091",
          "K081341",
          "K142829",
          "K251838",
          "K223872",
          "K851947",
          "K124046",
          "K821188",
          "K771982",
          "K040728",
          "K142183",
          "K181592",
          "K171866",
          "K152194",
          "K042399",
          "K122854",
          "K231108",
          "K070197",
          "K021004",
          "K884104",
          "K183655",
          "K212392",
          "K052185",
          "K111138",
          "K133296",
          "K161183",
          "K051513",
          "K012922",
          "K924456",
          "K821324",
          "K924491",
          "K240829",
          "K081645",
          "K260626",
          "K070816",
          "K040552",
          "K982797",
          "K902251",
          "K241145",
          "K242106",
          "K230051",
          "K142245",
          "K822593",
          "K030723",
          "K990705",
          "K960625",
          "K861238",
          "K112554",
          "K233503",
          "K140015",
          "K081394",
          "K960310",
          "K060190",
          "K123990",
          "K193125",
          "K132617",
          "K171491",
          "K792172",
          "K103143",
          "K080500",
          "K152116",
          "K180785",
          "K931743",
          "K202584",
          "K901452",
          "K180620",
          "K930801",
          "K233959",
          "K120158",
          "K183066",
          "K131332",
          "K190941",
          "K053400",
          "K771042",
          "K221914",
          "K190369",
          "K812143",
          "K894431",
          "K981782",
          "K032404",
          "K172117",
          "K171609",
          "K897041",
          "K132592",
          "K041708",
          "K012812",
          "K942191",
          "K151936",
          "K060666",
          "K924562",
          "K220047",
          "K062446",
          "K890766",
          "K160716",
          "K792130",
          "K140201",
          "K102728",
          "K241230",
          "K254236",
          "K092097",
          "K903410",
          "K210627",
          "K894446",
          "K851141",
          "K132498",
          "K222113",
          "K140398",
          "K250909",
          "K875158",
          "K002365",
          "K842004",
          "K260459",
          "K083402",
          "K090394",
          "K140420",
          "K153246",
          "K965167",
          "K001139",
          "K261492",
          "K122958",
          "K083855",
          "K791963",
          "K843500",
          "K924607",
          "K983940",
          "K051248",
          "K153197",
          "K010128",
          "K213582",
          "K931489",
          "K093232",
          "K932370",
          "K920528",
          "K111092",
          "K162769",
          "K230283",
          "K203179",
          "K920907",
          "K110051",
          "K014052",
          "K023447",
          "K850820",
          "K150186",
          "K193160",
          "K172099",
          "K843719",
          "K202167",
          "K121748",
          "K992668",
          "K151951",
          "K240957",
          "K962180",
          "K860432",
          "K945524",
          "K845037",
          "K083083",
          "K042308",
          "K011790",
          "K812325",
          "K004025",
          "K900658",
          "K032569",
          "K994252",
          "K853072",
          "K871132",
          "K213492",
          "K051601",
          "K973007",
          "K844693",
          "K992309",
          "K180081",
          "K961135",
          "K843641",
          "K102925",
          "K931662",
          "K000004",
          "K220414",
          "K840923",
          "K980578",
          "K930808",
          "K173750",
          "K851078",
          "K040029",
          "K931218",
          "K173287",
          "K812973",
          "K231412",
          "K101497",
          "K050015",
          "K903442",
          "K915421",
          "K090867",
          "K162097",
          "K170695",
          "K160721",
          "K914330",
          "K842780",
          "K241370",
          "K220604",
          "K870374",
          "K860898",
          "K111603",
          "K122980",
          "K152173",
          "K141275",
          "K192769",
          "K193509",
          "K974448",
          "K171081",
          "K800619",
          "K891121",
          "K915167",
          "K920998",
          "K123430",
          "K162322",
          "K103083",
          "K994334",
          "K140871",
          "K093877",
          "K925633",
          "K253652",
          "K903530",
          "K113849",
          "K032073",
          "K123334",
          "K140406",
          "K153771",
          "K070396",
          "K260606",
          "K212626",
          "K982187",
          "K894157",
          "K253526",
          "K244046",
          "K931102",
          "K242248",
          "K140338",
          "K022170",
          "K914352",
          "K252766",
          "K932323",
          "K190688",
          "K121504",
          "K202141",
          "K852916",
          "K260163",
          "K140327",
          "K984260",
          "K072248",
          "K961780",
          "K101883",
          "K170831",
          "K091137",
          "K062145",
          "K211989",
          "K221655",
          "K914046",
          "K043438",
          "K093023",
          "K902275",
          "K960200",
          "K915553",
          "K252027",
          "K152177",
          "K082847",
          "K110090",
          "K091954",
          "K100615",
          "K031906",
          "K021094",
          "K954317",
          "K252227",
          "K093026",
          "K982740",
          "K243928",
          "K060902",
          "K161102",
          "K192106",
          "K173685",
          "K173068",
          "K092575",
          "K034025",
          "K002973",
          "K132263",
          "K931587",
          "K053092",
          "K152913",
          "K915625",
          "K012489",
          "K022689",
          "K013029",
          "K212152",
          "K233691",
          "K022003",
          "K101422",
          "K003729",
          "K121234",
          "K822247",
          "K961745",
          "K242520",
          "K900515",
          "K001926",
          "K813143",
          "K251051",
          "K071574",
          "K003731",
          "K090372",
          "K123235",
          "K140768",
          "K993191",
          "K082644",
          "K093022",
          "K962978",
          "K983039",
          "K002676",
          "K181237",
          "K153430",
          "K843798",
          "K871140",
          "K033527",
          "K800621",
          "K770427",
          "K030734",
          "K190644",
          "K070417",
          "K111747",
          "K142819",
          "K902932",
          "K172995",
          "K934624",
          "K063658",
          "K791129",
          "K142357",
          "K970229",
          "K840455",
          "K861082",
          "K855060",
          "K911236",
          "K093130",
          "K961466",
          "K964518",
          "K042376",
          "K011085",
          "K171206",
          "K022634",
          "K970264",
          "K243954",
          "K250439",
          "K131627",
          "K954906",
          "K261950",
          "K820581",
          "K834074",
          "K123135",
          "K152381",
          "K884159",
          "K895214",
          "K080415",
          "K131144",
          "K242756",
          "K031087",
          "K203670",
          "K950155",
          "K211405",
          "K882657",
          "K141696",
          "K980504",
          "K123475",
          "K944107",
          "K830829",
          "K900756",
          "K030135"
        ],
        "registration_number": [
          "",
          "1000523114",
          "1650907",
          "2953359",
          "3030446844",
          "3007882731",
          "2648045",
          "3006260740",
          "3004784537",
          "3001124136",
          "3009746061",
          "3004737415",
          "1820334",
          "1824619",
          "3010391198",
          "1000121056",
          "3008853977",
          "1220477",
          "3011564431",
          "3020954746",
          "1828424",
          "3004091615",
          "1018470",
          "9610825",
          "9612164",
          "3007695959",
          "1313046",
          "3013557562",
          "3042279517",
          "3010966701",
          "3032916632",
          "3010141307",
          "1048735",
          "9612086",
          "2134812",
          "3012179728",
          "3016678045",
          "3015225571",
          "3031983053",
          "3011657089",
          "3027366703",
          "1721504",
          "3006082230",
          "3001374820",
          "1220648",
          "3020968442",
          "3037050339",
          "3043337628",
          "3012050423",
          "3015138222",
          "3011610434",
          "3017689643",
          "3011024991",
          "1224960",
          "3022518300",
          "3007697249",
          "1035166",
          "3017567825",
          "3038188187",
          "1423507",
          "3008998256",
          "9680794",
          "2011171",
          "2183870",
          "3007207906",
          "1722746",
          "3017060084",
          "3025341484",
          "2029015",
          "1221019",
          "2025816",
          "3010055973",
          "1526711",
          "3003915875",
          "9614279",
          "3008328690",
          "3008514120",
          "1424263",
          "1450662",
          "3011642792",
          "1928237",
          "3010421104",
          "3010419931",
          "3038966633",
          "2030598",
          "2134244",
          "2030404",
          "3003610571",
          "3013288201",
          "9681834",
          "2939561",
          "1710034",
          "3030733800",
          "3002808467",
          "3033589330",
          "1061124",
          "1417485",
          "2024168",
          "3010200163",
          "1921846",
          "3004193842",
          "3015453963",
          "3035111606",
          "2528981",
          "3046772394",
          "3011396687",
          "3030236028",
          "1647149",
          "3030447506",
          "2134265",
          "3010220595",
          "3045178474",
          "3004111573",
          "9617465",
          "3009605245",
          "1225687",
          "3005941719",
          "1282497",
          "3009437315",
          "3030574705",
          "3007603855",
          "3031282706",
          "3036802499",
          "3014687026",
          "9612126",
          "3002648230",
          "2246552",
          "9616683",
          "3031823122",
          "3005168196",
          "3009350690",
          "1118880",
          "1000563940",
          "3019807891",
          "1016427",
          "3013986346",
          "3017298588",
          "9681477",
          "1625425",
          "1721686",
          "2183787",
          "2518902",
          "3010665433",
          "3011752740",
          "1223004",
          "3010041511",
          "3013162291",
          "9616099",
          "9681260",
          "3006131984",
          "3016904853",
          "3031571797",
          "3013400708",
          "3002601200",
          "1644312",
          "2523676",
          "3008452825",
          "3014272734",
          "1064858",
          "8020785",
          "3008114965",
          "2182269",
          "3005232773",
          "3014656749",
          "3005334138",
          "3011137372",
          "3037047821",
          "8043892",
          "1043214",
          "3020347218",
          "3015859709",
          "3019751610",
          "3027815",
          "3009420598",
          "3008002401",
          "3013666336",
          "3017636737",
          "1000393132",
          "2020394",
          "3010155661",
          "2527072",
          "3004785273",
          "1225698",
          "1061927",
          "3002807314",
          "1319639",
          "3008824097",
          "3022815697",
          "3011689956",
          "3011063223",
          "3009393076",
          "1018233",
          "2015691",
          "2090040",
          "1222140",
          "3013764800",
          "3010432890",
          "3016986313",
          "9611612",
          "3003637635",
          "3011237770",
          "2017865",
          "3008261722",
          "9710641",
          "1526863",
          "1422634",
          "3005580113",
          "2953200",
          "8010026",
          "3009211636",
          "2124215",
          "9616666",
          "3000268902",
          "3007830707",
          "2245270",
          "1055236",
          "3010882708",
          "3003898360",
          "3034300126",
          "1835568",
          "3012536737",
          "9616088",
          "3015309643",
          "3004134316",
          "3041086754",
          "1056553",
          "1319211",
          "2024311",
          "3004519921",
          "3005994106",
          "3010131137",
          "3008361782",
          "1423537",
          "1724474",
          "1123137",
          "3008307705",
          "3009905888",
          "3015615738",
          "1649395",
          "3017168767",
          "3017406698",
          "9673241",
          "3021951050",
          "2248146",
          "3013758550",
          "3006425876",
          "2182208",
          "3008403546",
          "3015970743",
          "3009144177",
          "2030624",
          "1721676",
          "9610847",
          "2521402",
          "3013300026",
          "3039454510",
          "1625685",
          "3010273872",
          "2183744",
          "1054241",
          "2320762",
          "1318694",
          "3007284313",
          "9617604",
          "2032112",
          "3016591327",
          "3025028711",
          "3006950086",
          "3020283264",
          "3006283319",
          "9617592",
          "3008500478",
          "9612433",
          "3009888344",
          "1643817",
          "3015173212",
          "3011623145",
          "3010291427"
        ],
        "fei_number": [
          "",
          "1000523114",
          "3002807439",
          "1000222374",
          "1000518731",
          "3002615969",
          "3030446844",
          "3007882731",
          "2648045",
          "3006260740",
          "3004784537",
          "3001124136",
          "3009746061",
          "3004737415",
          "1820334",
          "1824619",
          "3008853977",
          "3010391198",
          "1000121056",
          "1220477",
          "3011564431",
          "3020954746",
          "1828424",
          "3008397795",
          "3003528016",
          "3004091615",
          "1018470",
          "3007695959",
          "1313046",
          "3013557562",
          "3042279517",
          "3010966701",
          "3032916632",
          "3010141307",
          "1048735",
          "3012179728",
          "3016678045",
          "3015225571",
          "1721504",
          "3031983053",
          "3011657089",
          "3027366703",
          "3003248396",
          "3006082230",
          "3001374820",
          "1220648",
          "3020968442",
          "3001451463",
          "3004970277",
          "3002807111",
          "3037050339",
          "3012050423",
          "3043337628",
          "3015138222",
          "3011610434",
          "3017689643",
          "3011024991",
          "1000160458",
          "3022518300",
          "3007697249",
          "1035166",
          "3017567825",
          "3038188187",
          "3008998256",
          "2011171",
          "2183870",
          "3007207906",
          "3000718472",
          "3004635528",
          "3017060084",
          "3025341484",
          "3002879021",
          "1221019",
          "2025816",
          "1000120555",
          "3010055973",
          "1526711",
          "3003915875",
          "1450662",
          "3008328690",
          "3008514120",
          "1928237",
          "1000138054",
          "3011642792",
          "3010421104",
          "3010419931",
          "3038966633",
          "3003769549",
          "3003610571",
          "3013288201",
          "3002681132",
          "1710034",
          "3030733800",
          "3002808467",
          "3033589330",
          "3002095335",
          "1417485",
          "2024168",
          "3003076831",
          "3008002452",
          "1921846",
          "3010200163",
          "3004193842",
          "3001452571",
          "3003565456",
          "3002558692",
          "3015453963",
          "3035111606",
          "2528981",
          "3046772394",
          "3000204727",
          "3011396687",
          "3030236028",
          "3002806575",
          "3000215205",
          "3030447506",
          "1000112877",
          "3010220595",
          "3045178474",
          "1000519796",
          "3004111573",
          "3009605245",
          "3005846326",
          "1282497",
          "3009437315",
          "3002827704",
          "3030574705",
          "3007603855",
          "3031282706",
          "3036802499",
          "3002807548",
          "3014687026",
          "1000123180",
          "3002648230",
          "1000132370",
          "2246552",
          "3031823122",
          "3003737899",
          "1000306139",
          "3005168196",
          "3009350690",
          "1118880",
          "1000514656",
          "1000563940",
          "3019807891",
          "3003094851",
          "1016427",
          "3013986346",
          "3017298588",
          "1625425",
          "1721686",
          "2183787",
          "2518902",
          "3010665433",
          "3011752740",
          "3010041511",
          "3013162291",
          "1000117748",
          "3006131984",
          "3016904853",
          "3031571797",
          "3013400708",
          "3002601200",
          "2523676",
          "3008452825",
          "3014272734",
          "1000306216",
          "2182269",
          "3008114965",
          "3005232773",
          "3014656749",
          "3005334138",
          "3037047821",
          "1000112968",
          "1043214",
          "3002808327",
          "3009291251",
          "3020347218",
          "3015859709",
          "3019751610",
          "3009420598",
          "3008002401",
          "3013666336",
          "3017636737",
          "1000393132",
          "3003407487",
          "2020394",
          "3010155661",
          "2527072",
          "3000203000",
          "3001236905",
          "3004785273",
          "3002807314",
          "3008824097",
          "3022815697",
          "3011689956",
          "3002806565",
          "1000121130",
          "3011063223",
          "3009393076",
          "1018233",
          "2015691",
          "3013764800",
          "3010432890",
          "3016986313",
          "3003637635",
          "3011237770",
          "2017865",
          "3008261722",
          "1526863",
          "3005580113",
          "3009211636",
          "2124215",
          "3000268902",
          "3007830707",
          "2245270",
          "3004134316",
          "3010882708",
          "1000512168",
          "3003898360",
          "3034300126",
          "3012536737",
          "3002808050",
          "3015309643",
          "3041086754",
          "2828",
          "1319211",
          "2024311",
          "3003015552",
          "3005994106",
          "3010131137",
          "3008361782",
          "3015615738",
          "3002806469",
          "3008307705",
          "3009905888",
          "3008906816",
          "3017168767",
          "3017406698",
          "3021951050",
          "3013758550",
          "1713910",
          "3006425876",
          "2182208",
          "3008403546",
          "3015970743",
          "3009144177",
          "3003767881",
          "1721676",
          "3000206435",
          "3001451473",
          "2521402",
          "3013300026",
          "3039454510",
          "3002807847",
          "1625685",
          "3010273872",
          "2183744",
          "3002806434",
          "1318694",
          "3002902735",
          "3005747797",
          "3001237677",
          "3007284313",
          "3016591327",
          "3025028711",
          "1000127185",
          "3006950086",
          "3002808464",
          "3020283264",
          "3006283319",
          "3002468371",
          "3009888344",
          "1643817",
          "3015173212",
          "3003040625",
          "3011623145",
          "3010291427"
        ],
        "device_name": "Introducer, Catheter",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1340",
        "device_class": "2"
      }
    }
  ]
}