  "results": [
    {
      "cfres_id": "183781",
      "product_res_number": "Z-0159-2021",
      "event_date_initiated": "2020-09-09",
      "event_date_created": "2020-10-21",
      "recall_status": "Open, Classified",
      "res_event_number": "86462",
      "product_code": "MAI",
      "k_numbers": [
        "K151105"
      ],
      "product_description": "TWINFIX ULTRA PLLA HA 4.5MM W/ NEEDLES- Absorbable Suture anchor\nProduct Number: 72202612",
      "code_info": "Batch Numbers: 2027487 2031136 2033257 2034471 2038440 2050957 2053792 2056056",
      "firm_fei_number": "3003604053",
      "recalling_firm": "Smith & Nephew, Inc.",
      "address_1": "150 Minuteman Rd",
      "city": "Andover",
      "state": "MA",
      "postal_code": "01810-1031",
      "reason_for_recall": "Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch",
      "root_cause_description": "Equipment maintenance",
      "action": "Smith & Nephew issued Urgent Medical Device Recall letter (R-2020-19) FSN on September 9, 2020  via email or overnight mail  to customers that received affected products. Letter states reason for recall, health risk, and action to take:\nRequired Actions:\n1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately.\na. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete.\nb. Include the customers account number and name in the table below.\n2. If you have no product to return, please put an X in the appropriate location below.\n3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below.\n4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com.\nPlease Note  even if you have no product to return, this form must be completed, signed and returned.\n5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number.\nIf you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.",
      "product_quantity": "263,874 (total)",
      "distribution_pattern": "Nationwide\nForeign:\nAE, AT, AU, BE, BM, BR, CA, CH, CL, CN, CO, CY,\nCZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GE, GR, HK,\nHU, IE, IL, IN, IT, JP,KR, LV,MY, NL, NO, PA,PE, PL,\nPR, PT, RU, SE, SG, SI, TH, TR, TW, UA,  & ZA",
      "openfda": {
        "k_number": [
          "K012334",
          "K171299",
          "K112249",
          "K973381",
          "K133794",
          "K091041",
          "K990361",
          "K062782",
          "K110660",
          "K180763",
          "K180361",
          "K051219",
          "K031981",
          "K933948",
          "K983186",
          "K140643",
          "K240441",
          "K062466",
          "K003984",
          "K955486",
          "K993061",
          "K051983",
          "K250528",
          "K161174",
          "K252946",
          "K950275",
          "K143172",
          "K233971",
          "K222263",
          "K952687",
          "K063497",
          "K003076",
          "K091844",
          "K955003",
          "K092898",
          "K003970",
          "K960555",
          "K971358",
          "K002406",
          "K210239",
          "K230394",
          "K963369",
          "K243467",
          "K090835",
          "K200725",
          "K970896",
          "K180662",
          "K952471",
          "K180101",
          "K171114",
          "K070178",
          "K140855",
          "K173788",
          "K983435",
          "K992938",
          "K993453",
          "K130716",
          "K000506",
          "K212912",
          "K983843",
          "K973281",
          "K973849",
          "K060522",
          "K994161",
          "K120078",
          "K992567",
          "K201786",
          "K192824",
          "K230892",
          "K170868",
          "K193558",
          "K180768",
          "K191226",
          "K221468",
          "K051250",
          "K103309",
          "K181746",
          "K082999",
          "K103831",
          "K183506",
          "K251680",
          "K260294",
          "K162429",
          "K011985",
          "K173240",
          "K040827",
          "K052799",
          "K192532",
          "K101823",
          "K983577",
          "K192032",
          "K171228",
          "K243760",
          "K120589",
          "K192441",
          "K221099",
          "K981340",
          "K210122",
          "K251309",
          "K170501",
          "K230435",
          "K202535",
          "K090124",
          "K203334",
          "K000616",
          "K220833",
          "K253931",
          "K013057",
          "K181809",
          "K964521",
          "K193157",
          "K240090",
          "K943249",
          "K093844",
          "K120540",
          "K122869",
          "K964805",
          "K131974",
          "K201749",
          "K993074",
          "K011495",
          "K243790",
          "K032197",
          "K231641",
          "K090669",
          "K180960",
          "K993373",
          "K041307",
          "K213415",
          "K121890",
          "K101679",
          "K955768",
          "K993115",
          "K122635",
          "K011299",
          "K191242",
          "K003273",
          "K041128",
          "K100012",
          "K170639",
          "K071456",
          "K123393",
          "K143660",
          "K163672",
          "K991715",
          "K193503",
          "K213958",
          "K202806",
          "K031519",
          "K993339",
          "K231787",
          "K151342",
          "K042966",
          "K122228",
          "K990987",
          "K090186",
          "K002413",
          "K240288",
          "K171141",
          "K991009",
          "K203523",
          "K141259",
          "K003227",
          "K112040",
          "K082810",
          "K982095",
          "K994269",
          "K961789",
          "K201414",
          "K221128",
          "K132241",
          "K151105",
          "K001378",
          "K192484",
          "K231326",
          "K101459",
          "K964013",
          "K053492",
          "K083268",
          "K130917",
          "K254055",
          "K963396",
          "K984490",
          "K070925",
          "K203495",
          "K990999",
          "K974876",
          "K041698",
          "K252022",
          "K972783",
          "K250526",
          "K002020",
          "K102982",
          "K160804",
          "K101100",
          "K173859",
          "K150209",
          "K241912",
          "K113534",
          "K993073",
          "K183236",
          "K142948",
          "K203493",
          "K944544",
          "K965228",
          "K041305",
          "K073412",
          "K232513",
          "K051362",
          "K183501",
          "K262357",
          "K180594",
          "K003743",
          "K982900",
          "K990340",
          "K071176",
          "K250544",
          "K984171",
          "K171592",
          "K241010",
          "K012465"
        ],
        "registration_number": [
          "3015453963",
          "3001592626",
          "3022518322",
          "1226146",
          "1058584",
          "3004983210",
          "3011430871",
          "3006990600",
          "3009905888",
          "3015543569",
          "3015806723",
          "9610612",
          "3003604053",
          "3003755939",
          "1047843",
          "3013700547",
          "3015542154",
          "3008114965",
          "3006946276",
          "3010375065",
          "1825034",
          "3016678045",
          "3017196117",
          "3012883081",
          "8043792",
          "3006498370",
          "3030655570",
          "3016948539",
          "1220246",
          "3022472578",
          "3015780293",
          "9615788",
          "1643264",
          "2936485",
          "1226183",
          "3016851379",
          "3012470322",
          "3007648354",
          "3015718210",
          "3018094310",
          "3019269298",
          "3014554088",
          "9710629",
          "3011530718",
          "3005381997",
          "3021393544",
          "9681465",
          "3005536892",
          "3012358417",
          "1017294",
          "2530154",
          "1048735",
          "3002579136",
          "3004365956",
          "1319660",
          "3014527904",
          "3013086329",
          "3021226419",
          "3038284030",
          "3015539532",
          "2183967",
          "3008868758",
          "3008102042",
          "2024024",
          "3015231789",
          "3014302784",
          "1220477",
          "1018470",
          "3010041511",
          "1450662",
          "1000517406",
          "1061927",
          "3021336182",
          "3008812560",
          "1320894",
          "1219930",
          "3005941719",
          "3025318945",
          "3008793310",
          "3012329552",
          "3027339877",
          "1526534",
          "3008729892",
          "1054811",
          "3007842664",
          "3014286836",
          "1221934",
          "1225798",
          "3038503932",
          "2183946",
          "1834331",
          "3002907620",
          "3007193505",
          "2031917",
          "3011683674",
          "8031010",
          "1721676",
          "3014004349",
          "3014232239",
          "3010331645",
          "2032830",
          "3009732568",
          "3015173212",
          "3014262693",
          "3011324393",
          "3013722959",
          "1649390",
          "8010177",
          "3005180920",
          "3005064037",
          "3006638824",
          "2245304",
          "1836357",
          "3014560508",
          "3019807891",
          "3011230048",
          "3010400367",
          "3016727307",
          "3003139373",
          "3007319107",
          "3027242374",
          "2648666",
          "3006981798",
          "3015974593",
          "3014252644",
          "2246552",
          "3007109793",
          "3008110533",
          "3024021261",
          "1219602",
          "3012818475",
          "1319639",
          "3013176080",
          "3032314868",
          "3015724777",
          "3034676720",
          "3035366890",
          "9610622",
          "3010273872",
          "1221051",
          "3020704423",
          "3010162973",
          "3006017180",
          "3002807310",
          "3015207155",
          "3004549189",
          "3020899789",
          "3014279513",
          "3004638532"
        ],
        "fei_number": [
          "1000116116",
          "3015453963",
          "1000138054",
          "1000219637",
          "3001592626",
          "3022518322",
          "3004983210",
          "3011430871",
          "3006990600",
          "3009905888",
          "3015543569",
          "3015806723",
          "3003604053",
          "3003755939",
          "1047843",
          "3013700547",
          "3015542154",
          "3008114965",
          "3006946276",
          "3002806315",
          "3010375065",
          "1825034",
          "3016678045",
          "3017196117",
          "3003064967",
          "3012883081",
          "3006498370",
          "3002929183",
          "3004051158",
          "3030655570",
          "1000306340",
          "3002807111",
          "1220246",
          "3002821504",
          "3022472578",
          "3015780293",
          "1000307073",
          "3002807041",
          "1643264",
          "2936485",
          "3016851379",
          "3012470322",
          "3007648354",
          "3015718210",
          "3018094310",
          "3019269298",
          "3014554088",
          "3002680121",
          "3011530718",
          "3005381997",
          "3021393544",
          "3005536892",
          "3012358417",
          "1017294",
          "1048735",
          "3002579136",
          "3004365956",
          "3014527904",
          "3013086329",
          "3021226419",
          "3038284030",
          "3015539532",
          "3008066433",
          "3008868758",
          "2024024",
          "3015231789",
          "3014302784",
          "1220477",
          "1018470",
          "3010041511",
          "1450662",
          "1000517406",
          "3001595430",
          "3021336182",
          "3008812560",
          "1000136478",
          "1219930",
          "3025318945",
          "3008793310",
          "3002808269",
          "3012329552",
          "3027339877",
          "1526534",
          "3014286836",
          "3007842664",
          "1221934",
          "3038503932",
          "3002757549",
          "1000114185",
          "3002907620",
          "3007193505",
          "3011683674",
          "1000115735",
          "1721676",
          "1000185904",
          "3014004349",
          "3002807231",
          "3014232239",
          "3010331645",
          "3009732568",
          "3015173212",
          "3014262693",
          "3011324393",
          "3013722959",
          "3005180920",
          "3005064037",
          "3006638824",
          "3000203000",
          "3003407235",
          "3004167986",
          "2245304",
          "3014560508",
          "3019807891",
          "3011230048",
          "3010400367",
          "3016727307",
          "3002924544",
          "3003139373",
          "3007319107",
          "3000718467",
          "3027242374",
          "2648666",
          "3006981798",
          "3015974593",
          "3014252644",
          "2246552",
          "3007109793",
          "3008110533",
          "3024021261",
          "3003854593",
          "1219602",
          "3012818475",
          "3002807113",
          "3013176080",
          "3032314868",
          "3015724777",
          "3034676720",
          "3035366890",
          "3010273872",
          "1221051",
          "3020704423",
          "3006017180",
          "3002807310",
          "3002807555",
          "3015207155",
          "3001451571",
          "3020899789",
          "3014279513",
          "3004638532"
        ],
        "device_name": "Fastener, Fixation, Biodegradable, Soft Tissue",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3030",
        "device_class": "2"
      }
    }
  ]
}