      "cfres_id": "183649",
      "product_res_number": "Z-0115-2021",
      "event_date_initiated": "2020-08-25",
      "event_date_created": "2020-10-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2023-03-10",
      "res_event_number": "86383",
      "product_code": "NCW",
      "k_numbers": [
        "K024017"
      ],
      "product_description": "Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes:\n(01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130\n\nAtellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code\n(01)00630414599168(10)22695A49(17)20201211\n\nADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes :\n(01)00630414471235(10)43157138(17)20210130\n(01)00630414471235(10)81653138(17)20210130\n(01)00630414471235(10)99728138(17)20210130\n(01)00630414471235(10)24043138(17)20210130\n(01)00630414471235(10)32477138(17)20210130\n\nAdvia Centaur Systems HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10308993, UDI Codes:\n(01)00630414471228(10)64661A48(17)20201112\n(01)00630414471228(10)81211A48(17)20201112\n(01)00630414471228(10)91743A48(17)20201112\n(01)00630414471228(10)22124A49(17)20201211     \nProduct Usage: This product is intended to be used for in vitro diagnostic use in calibrating the ADVIA Centaur systems HER-2/neu assay.",
      "code_info": "Atellica IM HER-2/neu Assay 50 Test Kit: Siemens Material Number (SMN) 10995591; EXP Date 2021-01-30; Lot # 98034139 w/ UDI (01)00630414599151(10)98034139(17)20210130; Lot # 32478139 w/ UDI (01)00630414599151(10)32478139(17)20210130  Atellica IM HER-2/neu Calibrator (2 pack): SMN 10995592; EXP Date 2020-12-11;  Lot # 22695A49; UDI (01)00630414599168(10)22695A49(17)20201211  ADVIA Centaur HER-2 Assay 50 Test Kit: SMN 10308994; EXP Date 2021/01/30; Lot # 43157138 w/ UDI (01)00630414471235(10)43157138(17)20210130 ; Lot# 81653138 w/ UDI (01)00630414471235(10)81653138(17)20210130 ; Lot# 99728138 w/ UDI (01)00630414471235(10)99728138(17)20210130 ; Lot # 24043138 w/ UDI (01)00630414471235(10)24043138(17)20210130 ; Lot #32477138 w/ UDI  (01)00630414471235(10)32477138(17)20210130 ;  ADVIA Centaur HER-2/neu Calibrator (2 pack): SMN 10308993; EXP Date 2020-11-12 Lot # 64661A48 w/ UDI (01)0063041447(10)64661A48(17)20201112 Lot # 81211A48 w/ UDI (01)0063041447(10)81211A48(17)20201112 Lot# 91743A48  w/ UDI (01)0063041447(10)91743A48(17)20201112  EXP Date 2020-12-11;  Lot # 22124A49 w/ UDI (01)0063041447(10)22124a49(17)20201112",
      "firm_fei_number": "1219913",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc",
      "address_1": "333 Coney St",
      "city": "East Walpole",
      "state": "MA",
      "postal_code": "02032-1516",
      "additional_info_contact": "Peter Erickson\n508-660-6570",
      "reason_for_recall": "Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results.",
      "root_cause_description": "Process control",
      "action": "All affected customers were sent a letter on 08/25/2020 stating the following: Siemens Healthcare Diagnostics Inc. (Siemens Healthineers) notified customers on August  25th, 2020. Letter states reason for recall, health risk and action to take.  Actions to be taken by their customers include:\n\n- Reviewing the letter with their Medical Director.\n- Customers may continue use of the product lots referenced in Table 1 until they receive replacement product in their laboratory. Refer to the Table 3  for product biasinformation.\n- If you are currently using the products listed in Table1, review your inventory of these products, as well as the associated product QC and MCM, and order replacement products by completing the Field Correction Effectiveness Check Form attached to the letter.\n- Upon acceptance of the replacement lots referenced in Table 2 New Lot Combinations column, discontinue use of and discard the products listed in Table 2 Current Lot Combinations column. Refer to Figures 1 through 4 for expected results with replacement lots.\n- Complete and return the Field Correction Effectiveness Check Form attached to this letter.\n\nRetain the letter with your laboratory records and forward the letter to those who may have received this product.",
      "product_quantity": "Atellica IM H2n Assay =135 kits; Atellica IM H2n Calibrator = 46 kits;  ADVIA Centaur Assay = 1,823 kits; ADVIA Centaur calibrator = 161 kits",
      "distribution_pattern": "US Nationwide distribution.",
      "openfda": {
        "k_number": [
          "K992228",
          "K994112",
          "K024017"
        ],
        "device_name": "System, Test, Her-2/Neu, Monitoring",
        "medical_specialty_description": "Immunology",
        "regulation_number": "866.6010",
        "device_class": "2",
        "registration_number": [
          "2032839",
          "2432235",
          "1219913"
        ],
        "fei_number": [
          "2432235",
          "3004185580",
          "1219913"
        ]
      }
    }
  ]
}