  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "182834",
      "product_res_number": "Z-0257-2021",
      "event_date_initiated": "2020-07-06",
      "event_date_created": "2020-10-22",
      "recall_status": "Terminated",
      "event_date_terminated": "2023-05-11",
      "res_event_number": "86054",
      "product_code": "FPA",
      "k_numbers": [
        "K974589"
      ],
      "product_description": "98 IN (249cm) APPX 8.4 ml, 60 Drop Set, 2 Clave, REMV 2 Clave STOPCOCKS. 1 unit per pouch, 25 pouches per case. UDI:(01)10887709076119(17)250101(30)25(10)4584862.\n\nThe SmartValve is a single use, sterile, non-pyrogenic pressure activated check valve system intended for use as an accessory to Intravascular administration set.",
      "code_info": "Lot Number:4584862.",
      "firm_fei_number": "2025816",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "city": "San Clemente",
      "state": "CA",
      "postal_code": "92673-6212",
      "additional_info_contact": "844-654-7780",
      "reason_for_recall": "Identification of a  potential  manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.",
      "root_cause_description": "Process control",
      "action": "On July 6, 2020, the firm, icumedical, sent an \"URGENT: MEDICAL DEVICE RECALL\"  notification of the market action to direct and indirect consignees of record via UPS. Stericycle, Inc. was contracted to send letters to consignees.\nThe following is included in the \"Required Actions for Users\" section of the medical device recall notification:\n1)   Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility.\n\n2) Inform potential users of the product in your organization of this notification and complete the response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product.\n\n3) Return affected product using the return label provided with the letter. Call 1-866-203-5570 (M F, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not  reproduce. Visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, you will be credited for any product returned. You will only receive credit for product that you return. NOTE: Credits for product purchased through distributor will be credited by the distributor.\n\n4)  If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product I Table 1 sections of this notification and ask them to call 1-866-203-5570 (M-F, 8am-5pm ET) to obtain a response form.\n\nFollow up Actions:\nContact Customer Service using the information provided below for assistance reordering replacement product and for further inquiries:\n\nGlobal Complaint Management -1-844-654-7780 or ProductCompla",
      "product_quantity": "Total of all products (Listed #1 thru 101) = 304735 units",
      "distribution_pattern": "Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.",
      "openfda": {
        "k_number": [
          "K925228",
          "K944619",
          "K032485",
          "K060074",
          "K873894",
          "K925219",
          "K021488",
          "K992072",
          "K862653",
          "K031783",
          "K213004",
          "K926184",
          "K852140",
          "K890489",
          "K242692",
          "K232328",
          "K823845",
          "K801821",
          "K771983",
          "K220301",
          "K770763",
          "K963009",
          "K822408",
          "K950303",
          "K922637",
          "K031923",
          "K002449",
          "K080660",
          "K190233",
          "K860201",
          "K905219",
          "K860605",
          "K133867",
          "K033350",
          "K153257",
          "K870866",
          "K852881",
          "K060580",
          "K041490",
          "K061115",
          "K981683",
          "K021916",
          "K850112",
          "K850506",
          "K864453",
          "K955775",
          "K841717",
          "K123344",
          "K013621",
          "K840410",
          "K951922",
          "K900909",
          "K192681",
          "K070189",
          "K893909",
          "K991932",
          "K970438",
          "K891226",
          "K955821",
          "K992290",
          "K142367",
          "K971293",
          "K925418",
          "K051939",
          "K890340",
          "K896896",
          "K881321",
          "K002492",
          "K801613",
          "K003977",
          "K844917",
          "K041861",
          "K974145",
          "K021480",
          "K853387",
          "K994429",
          "K891193",
          "K832123",
          "K140131",
          "K840409",
          "K001233",
          "K220267",
          "K113471",
          "K981318",
          "K771267",
          "K080544",
          "K120548",
          "K221327",
          "K853086",
          "K830565",
          "K901089",
          "K896291",
          "K150921",
          "K961928",
          "K082306",
          "K914792",
          "K050906",
          "K062308",
          "K140133",
          "K890679",
          "K120443",
          "K841563",
          "K872123",
          "K042693",
          "K103165",
          "K961049",
          "K011071",
          "K963981",
          "K913982",
          "K884053",
          "K990693",
          "K920694",
          "K955585",
          "K853385",
          "K820278",
          "K830891",
          "K902166",
          "K901088",
          "K780481",
          "K123874",
          "K103551",
          "K102405",
          "K970909",
          "K180739",
          "K974589",
          "K051499",
          "K860242",
          "K954283",
          "K940319",
          "K050600",
          "K043328",
          "K053049",
          "K896893",
          "K990388",
          "K790604",
          "K901081",
          "K900336",
          "K872124",
          "K821750",
          "K801872",
          "K964968",
          "K800878",
          "K974571",
          "K023595",
          "K881052",
          "K895367",
          "K864296",
          "K811885",
          "K872945",
          "K972957",
          "K911815",
          "K132734",
          "K821624",
          "K984063",
          "K840878",
          "K991513",
          "K841027",
          "K841562",
          "K953825",
          "K013871",
          "K120375",
          "K202437",
          "K993187",
          "K821995",
          "K895480",
          "K971701",
          "K833740",
          "K251375",
          "K874133",
          "K932637",
          "K942446",
          "K082806",
          "K061285",
          "K021999",
          "K871995",
          "K831088",
          "K960661",
          "K050207",
          "K891146",
          "K012845",
          "K820133",
          "K833178",
          "K991112",
          "K013714",
          "K960431",
          "K250616",
          "K221069",
          "K884102",
          "K840760",
          "K121554",
          "K191362",
          "K940053",
          "K780551",
          "K223645",
          "K970800",
          "K001872",
          "K021395",
          "K780657",
          "K023585",
          "K023917",
          "K051009",
          "K013502",
          "K000017",
          "K940697",
          "K895481",
          "K243392",
          "K060488",
          "K093271",
          "K100946",
          "K982123",
          "K861521",
          "K930751",
          "K941453",
          "K913337",
          "K902330",
          "K931377",
          "K243841",
          "K871732",
          "K834252",
          "K833030",
          "K813083",
          "K884141",
          "K063631",
          "K800403",
          "K884511",
          "K930927",
          "K933326",
          "K810239",
          "K210544",
          "K041545",
          "K891164",
          "K072375",
          "K952242",
          "K051142",
          "K822836",
          "K033892",
          "K900426",
          "K801874",
          "K896895",
          "K921576",
          "K861519",
          "K940867",
          "K882797",
          "K912141",
          "K042234",
          "K060082",
          "K010846",
          "K070532",
          "K953231",
          "K844843",
          "K813259",
          "K933898",
          "K213588",
          "K831311",
          "K840443",
          "K830755",
          "K212262",
          "K921674",
          "K790582",
          "K972493",
          "K951665",
          "K893716",
          "K051264",
          "K090189",
          "K830654",
          "K861518",
          "K963112",
          "K081289",
          "K852606",
          "K822775",
          "K964384",
          "K883764",
          "K173665",
          "K861250",
          "K905208",
          "K072542",
          "K182897",
          "K961225",
          "K952760",
          "K070430",
          "K853019",
          "K852259",
          "K791496",
          "K990496",
          "K761327",
          "K221481",
          "K002689",
          "K112636",
          "K120261",
          "K133351",
          "K142622",
          "K872795",
          "K791173",
          "K800600",
          "K925430",
          "K081210",
          "K072576",
          "K832284",
          "K153731",
          "K202672",
          "K130023",
          "K962802",
          "K113398",
          "K210430",
          "K101196",
          "K020419",
          "K011120",
          "K033099",
          "K944692",
          "K020990",
          "K010404",
          "K901046",
          "K203796",
          "K772036",
          "K161898",
          "K933732",
          "K971266",
          "K983217",
          "K163160",
          "K020530",
          "K830453",
          "K952810",
          "K955709",
          "K924718",
          "K811078",
          "K922558",
          "K090459",
          "K833949",
          "K854365",
          "K821377",
          "K823027",
          "K992145",
          "K041844",
          "K190806",
          "K052432",
          "K151151",
          "K781637",
          "K863814",
          "K953791",
          "K150860",
          "K172544",
          "K223101",
          "K860609",
          "K042129",
          "K941244",
          "K915678",
          "K883638",
          "K841882",
          "K112893",
          "K160648",
          "K981548",
          "K230343",
          "K991759",
          "K833033",
          "K151079",
          "K171101",
          "K850277",
          "K010197",
          "K842088",
          "K973107",
          "K931813",
          "K820453",
          "K041845",
          "K010279",
          "K982047",
          "K902109",
          "K933390",
          "K041495",
          "K871842",
          "K151250",
          "K023668",
          "K121092",
          "K931119",
          "K171118",
          "K232475",
          "K832434",
          "K850533",
          "K070547",
          "K012754",
          "K926448",
          "K200425",
          "K121803",
          "K770744",
          "K100432",
          "K802169",
          "K123729",
          "K982672",
          "K811463",
          "K895046",
          "K930406",
          "K933749",
          "K984060",
          "K062909",
          "K842985",
          "K944293",
          "K040533",
          "K891176",
          "K963659",
          "K925058",
          "K200661",
          "K014061",
          "K954983",
          "K980127",
          "K964812",
          "K020544",
          "K854965",
          "K900651",
          "K171117",
          "K861182",
          "K915571",
          "K911868",
          "K001423",
          "K083482",
          "K013104",
          "K962893",
          "K031917",
          "K222780",
          "K992999",
          "K180843",
          "K811006",
          "K883986",
          "K930750",
          "K083472",
          "K780559",
          "K221075",
          "K033208",
          "K003854",
          "K970485",
          "K903541",
          "K122404",
          "K010888",
          "K800637",
          "K042182",
          "K073050",
          "K853327",
          "K852488",
          "K790901",
          "K973167",
          "K922115",
          "K143446",
          "K854804",
          "K952074",
          "K921860",
          "K993463",
          "K904886",
          "K242126",
          "K950895",
          "K926119",
          "K921429",
          "K031500",
          "K000472",
          "K021565",
          "K923882",
          "K885031",
          "K100576",
          "K193088",
          "K252543",
          "K961327",
          "K213778",
          "K810921",
          "K870995",
          "K010788",
          "K142527",
          "K011187",
          "K792538",
          "K030149",
          "K771453",
          "K020926",
          "K181718",
          "K882007",
          "K790424",
          "K822407",
          "K823346",
          "K870913",
          "K100720",
          "K082027",
          "K940054",
          "K213846",
          "K960280",
          "K011216",
          "K860322",
          "K931659",
          "K130468",
          "K823347",
          "K192375",
          "K170002",
          "K944824",
          "K850339",
          "K834254",
          "K842858",
          "K821578",
          "K162812",
          "K981157",
          "K781188",
          "K891063",
          "K864405",
          "K042172",
          "K090809",
          "K083765",
          "K941190",
          "K852989",
          "K800137",
          "K132841",
          "K961474",
          "K142097",
          "K955410",
          "K972135",
          "K870582",
          "K891072",
          "K000592",
          "K854435",
          "K780970",
          "K970524",
          "K000942",
          "K021487",
          "K112512",
          "K020189",
          "K810317",
          "K162826",
          "K841564",
          "K801871",
          "K953746",
          "K820788",
          "K162304",
          "K791930",
          "K933852",
          "K954969",
          "K940296",
          "K160007",
          "K192647",
          "K850769",
          "K950970",
          "K905374",
          "K953225",
          "K013949",
          "K942349",
          "K893120",
          "K843054",
          "K992268",
          "K781189",
          "K143015",
          "K130416",
          "K801873",
          "K132142",
          "K052252",
          "K841859",
          "K020533",
          "K241058",
          "K844937",
          "K854630",
          "K884657",
          "K910142",
          "K942046",
          "K925927",
          "K925645",
          "K853204",
          "K771593",
          "K161808",
          "K120799",
          "K901949",
          "K873729",
          "K781184",
          "K964067",
          "K161323",
          "K944208",
          "K790062",
          "K162601",
          "K920537",
          "K924010",
          "K042666",
          "K101368",
          "K924207",
          "K840212",
          "K790077",
          "K901729",
          "K171132",
          "K984638",
          "K823214",
          "K900644",
          "K843185",
          "K950302",
          "K820203",
          "K772008",
          "K801004",
          "K111351",
          "K980583",
          "K854446",
          "K924721",
          "K001329",
          "K882347",
          "K952976",
          "K003513",
          "K914113",
          "K012429",
          "K011580",
          "K841185",
          "K912069",
          "K812688",
          "K811110",
          "K950304",
          "K873885",
          "K913076",
          "K171957",
          "K910546",
          "K914074",
          "K800711",
          "K234129",
          "K130879",
          "K831445",
          "K925732",
          "K972854",
          "K140831",
          "K140311",
          "K910695",
          "K023469",
          "K093773",
          "K980994",
          "K861520",
          "K860541",
          "K904685",
          "K913627",
          "K042936",
          "K881975",
          "K973916",
          "K974163",
          "K851336",
          "K925362",
          "K082818",
          "K101385",
          "K141200",
          "K860616",
          "K153293",
          "K984569",
          "K142433",
          "K072894",
          "K894842",
          "K990331",
          "K945617",
          "K092905",
          "K760103",
          "K910159",
          "K822027",
          "K220288",
          "K791529",
          "K902196",
          "K874581",
          "K944320",
          "K102890",
          "K954823",
          "K915390",
          "K961794",
          "K971004",
          "K020424",
          "K073444",
          "K780656",
          "K780655",
          "K021296",
          "K143087",
          "K860157",
          "K930974",
          "K980992",
          "K974013",
          "K896561",
          "K964439",
          "K854628",
          "K140175",
          "K874403",
          "K160870",
          "K801540",
          "K832622",
          "K914048",
          "K895109",
          "K070362",
          "K133894",
          "K871843",
          "K143234",
          "K843090",
          "K832549",
          "K033576",
          "K850061",
          "K841361",
          "K831087",
          "K020912",
          "K900427",
          "K102512",
          "K002138",
          "K121634",
          "K201872",
          "K844820",
          "K981792",
          "K952053",
          "K011003",
          "K061374",
          "K945298",
          "K895291",
          "K781185",
          "K010647",
          "K933848",
          "K041179",
          "K021692",
          "K110282",
          "K001832",
          "K970855",
          "K851516",
          "K880482",
          "K921322",
          "K940073",
          "K782134",
          "K811837",
          "K031361",
          "K895770",
          "K212842",
          "K061771",
          "K010886",
          "K143082",
          "K883318",
          "K771362",
          "K221121",
          "K942988",
          "K870940",
          "K931658",
          "K942181",
          "K041919",
          "K041410",
          "K073396",
          "K921703",
          "K842657",
          "K854636",
          "K173361",
          "K950597",
          "K790108",
          "K210217",
          "K932856",
          "K882798",
          "K930854",
          "K925930",
          "K003283",
          "K931173",
          "K033301",
          "K011405",
          "K810721",
          "K142036",
          "K040527",
          "K841106",
          "K822848",
          "K230528",
          "K022209",
          "K140838",
          "K021005",
          "K792228",
          "K052887",
          "K033515",
          "K871841",
          "K964456",
          "K953584",
          "K811561",
          "K953474",
          "K771339",
          "K882503",
          "K113470",
          "K760110",
          "K853527",
          "K930855",
          "K843249",
          "K832542",
          "K860272",
          "K102073",
          "K972712",
          "K844255",
          "K870188",
          "K896161",
          "K832105",
          "K874467",
          "K942931",
          "K843631",
          "K871086",
          "K904013",
          "K833939",
          "K932446",
          "K182545",
          "K012330",
          "K032484",
          "K121511",
          "K843748",
          "K840273",
          "K934513",
          "K780130",
          "K833034",
          "K982098",
          "K113683",
          "K791672",
          "K984146",
          "K941445",
          "K032767",
          "K101677",
          "K210381",
          "K153158",
          "K911443",
          "K885030",
          "K892440",
          "K781036",
          "K991301",
          "K771204",
          "K150059",
          "K853287",
          "K052722",
          "K864338",
          "K000152",
          "K081123",
          "K802329",
          "K894787",
          "K861430",
          "K850088",
          "K153440",
          "K932638",
          "K021471",
          "K841186",
          "K760111",
          "K001426",
          "K221518",
          "K132880",
          "K071846",
          "K983411",
          "K902810",
          "K854806",
          "K801179",
          "K900353",
          "K830868",
          "K930410",
          "K914026",
          "K811914",
          "K863304",
          "K870581",
          "K202279",
          "K781187",
          "K150513",
          "K013164",
          "K896346",
          "K954870",
          "K992104",
          "K122686",
          "K885264",
          "K830377",
          "K221078",
          "K852361",
          "K812657",
          "K931550",
          "K923560",
          "K854658",
          "K120872",
          "K121032",
          "K900188",
          "K014062",
          "K170595",
          "K251257",
          "K770589",
          "K832104",
          "K882302",
          "K890176",
          "K943038",
          "K032934",
          "K953343",
          "K132413",
          "K120195",
          "K834071",
          "K103562",
          "K950153",
          "K222889",
          "K243529",
          "K040710",
          "K181870",
          "K821764",
          "K880928",
          "K822354",
          "K142011",
          "K971500",
          "K030002",
          "K960771",
          "K970076",
          "K020629",
          "K083723",
          "K080695",
          "K897065",
          "K840081",
          "K925731",
          "K092382",
          "K242763",
          "K812231",
          "K770171",
          "K011888",
          "K830448",
          "K813194",
          "K841028",
          "K123868",
          "K963070",
          "K090012",
          "K011858",
          "K984561",
          "K120685",
          "K770863",
          "K112073",
          "K781813",
          "K003254",
          "K983926",
          "K940832",
          "K862289",
          "K842678",
          "K854436",
          "K001102",
          "K861275",
          "K915828",
          "K092172",
          "K250325",
          "K925733",
          "K041042",
          "K040385",
          "K841772",
          "K934095",
          "K830182",
          "K092527",
          "K971624",
          "K830258",
          "K962801",
          "K854346",
          "K971066",
          "K964918",
          "K781812",
          "K230992",
          "K854627",
          "K874307",
          "K100434",
          "K010538",
          "K021080",
          "K133073",
          "K251854",
          "K781190",
          "K842984",
          "K921594",
          "K961495",
          "K884264",
          "K935811",
          "K921391",
          "K913311",
          "K912856",
          "K030385",
          "K834289",
          "K251814",
          "K001424",
          "K853184",
          "K940056",
          "K892429",
          "K854574",
          "K981547",
          "K873042",
          "K120395",
          "K051943",
          "K032854",
          "K842143",
          "K911931",
          "K011193",
          "K810365",
          "K770743",
          "K800638",
          "K834130",
          "K881234",
          "K881390",
          "K904663",
          "K901151",
          "K861439",
          "K954123",
          "K010373",
          "K852254",
          "K983073",
          "K960391",
          "K093489",
          "K220312",
          "K142974",
          "K990777",
          "K864164",
          "K943770",
          "K033290",
          "K971844",
          "K001106",
          "K912861",
          "K030246",
          "K930926",
          "K023769",
          "K991985",
          "K871753",
          "K781785",
          "K943591",
          "K896892",
          "K913333",
          "K002471",
          "K083687",
          "K121269",
          "K954883",
          "K822406",
          "K052658",
          "K991653",
          "K872953",
          "K021216",
          "K060672",
          "K142590",
          "K972839",
          "K964850",
          "K954841",
          "K911753",
          "K011928",
          "K860952",
          "K242339",
          "K893646",
          "K040636",
          "K071421",
          "K993848",
          "K851299",
          "K834253",
          "K811685",
          "K940689",
          "K103224",
          "K965027",
          "K001599",
          "K192366",
          "K221473",
          "K972663",
          "K101958",
          "K990288",
          "K811107",
          "K010103",
          "K223088",
          "K900187",
          "K952852",
          "K954813",
          "K802170",
          "K964844",
          "K940657",
          "K141306",
          "K982587",
          "K771268",
          "K013950",
          "K843463",
          "K934119",
          "K830194",
          "K914924",
          "K060352",
          "K001827",
          "K092930",
          "K903809",
          "K832329",
          "K112799",
          "K984381",
          "K991255",
          "K852106",
          "K151969",
          "K862491",
          "K945070",
          "K943897",
          "K895290",
          "K012526",
          "K904217",
          "K940072",
          "K080665",
          "K853255",
          "K781186",
          "K873876",
          "K940897",
          "K011336",
          "K173879",
          "K203609",
          "K201626",
          "K213089",
          "K053106",
          "K090134",
          "K040633",
          "K952160",
          "K833741",
          "K162804",
          "K925126",
          "K840619",
          "K771271",
          "K964876",
          "K911136",
          "K261530",
          "K790574",
          "K970259",
          "K991979",
          "K990083",
          "K952664",
          "K222766",
          "K991599",
          "K993147",
          "K191640",
          "K942271",
          "K083576",
          "K915109",
          "K790408",
          "K800358",
          "K012879",
          "K013394",
          "K024363",
          "K043235",
          "K103344",
          "K003225",
          "K141834",
          "K081156",
          "K922469",
          "K081361",
          "K780658",
          "K842351",
          "K970255",
          "K874725",
          "K080303",
          "K221411",
          "K141285",
          "K813092",
          "K843692",
          "K070801",
          "K854437",
          "K955783",
          "K030467",
          "K070630",
          "K053129",
          "K191658",
          "K792560",
          "K896894",
          "K972527",
          "K902858",
          "K952392",
          "K200439",
          "K030137",
          "K770051",
          "K021530",
          "K992266",
          "K841025",
          "K032150",
          "K821515",
          "K161036",
          "K202618",
          "K830152",
          "K781744",
          "K911128",
          "K051437",
          "K000343",
          "K854754",
          "K921899",
          "K002006",
          "K101180",
          "K160651",
          "K932165",
          "K162951",
          "K984441",
          "K821032",
          "K892438",
          "K935601",
          "K905218",
          "K770518",
          "K111016",
          "K781575",
          "K964455",
          "K010676",
          "K852616",
          "K912860",
          "K031279",
          "K955732",
          "K231707",
          "K833398",
          "K883612",
          "K031266",
          "K771270",
          "K850507",
          "K820747",
          "K981386",
          "K830178",
          "K946082"
        ],
        "registration_number": [
          "",
          "3010291427",
          "3023228837",
          "3012494290",
          "3005012805",
          "3012410077",
          "3027327302",
          "3014246639",
          "9611590",
          "3009631669",
          "1724474",
          "1223925",
          "3006194679",
          "3009976527",
          "3005987567",
          "3010621964",
          "9680306",
          "1063851",
          "3017262697",
          "3015142721",
          "2528981",
          "3016661372",
          "3013557562",
          "3010413216",
          "3002807090",
          "3008403546",
          "1928237",
          "3036978618",
          "1000304305",
          "9680787",
          "3024122953",
          "3007392007",
          "1721676",
          "3005951712",
          "3010293992",
          "2016945",
          "3033589330",
          "2011068",
          "3015173212",
          "3007829657",
          "1721686",
          "3010515547",
          "1713468",
          "1000587249",
          "3010131137",
          "2025816",
          "1038185",
          "3030626857",
          "3019807891",
          "1417592",
          "3008448733",
          "3014456066",
          "9616088",
          "3013319212",
          "3005340341",
          "3004215921",
          "3008808123",
          "1423537",
          "1647137",
          "3009456923",
          "3009515470",
          "9680222",
          "2523676",
          "3017567825",
          "3008810357",
          "1062381",
          "1416980",
          "3022964615",
          "3012039263",
          "3003700453",
          "3015453963",
          "3000240707",
          "3010419931",
          "1035907",
          "1000112137",
          "3015276780",
          "3007305485",
          "3019004087",
          "3013696225",
          "3006950086",
          "3043051314",
          "3013187142",
          "9616662",
          "2029275",
          "3005382983",
          "3014579161",
          "3043226252",
          "3011987967",
          "2243072",
          "3005670221",
          "3006425876",
          "3015058854",
          "9614279",
          "1018233",
          "1319639",
          "3012104670",
          "9680794",
          "3020778",
          "9617032",
          "3015060216",
          "3008328679",
          "3008252904",
          "1419937",
          "3030733800",
          "1720734",
          "3014312726",
          "1417485",
          "3013820501",
          "3015983454",
          "1019003",
          "3006534321",
          "9681413",
          "3011944170",
          "3013188547",
          "3004605321",
          "3015620563",
          "3004548776",
          "1030451",
          "3008588205",
          "3010531069",
          "9681572",
          "3003310007",
          "8021545",
          "3002808466",
          "3006942524",
          "3002907620",
          "3003974574",
          "3005778453",
          "3005458990",
          "3009882464",
          "3004569180",
          "3010882065",
          "3014314985",
          "2011171",
          "1833470",
          "1061124",
          "1314417",
          "3005405146",
          "3008733163",
          "3011642792",
          "3007616112",
          "3003674698",
          "2523167",
          "3017636737",
          "3020777588",
          "3007363456",
          "3007836437",
          "3003915875",
          "1037885",
          "3033536319",
          "3012536737",
          "1220973",
          "3012116434",
          "3016938816",
          "3007007790",
          "3008132411",
          "3014936217",
          "3010273872",
          "1000563940",
          "1625425",
          "1043214",
          "3016096098",
          "1643817",
          "3018125942",
          "3010830834",
          "3004140833",
          "3042871620",
          "3042955949",
          "9610993",
          "3006082230",
          "1320894",
          "1047843",
          "1018470",
          "1422078",
          "3009746425",
          "3004911384",
          "3007185216",
          "3044994889",
          "1424263",
          "3015085713",
          "3006271008",
          "9615050",
          "3008650117",
          "3015724078",
          "3015859709",
          "3010966701",
          "9616067",
          "1047429",
          "1055236",
          "3005669815",
          "8021083",
          "1060680",
          "3002806603",
          "9617604",
          "1410248",
          "3006575795",
          "3004143450",
          "3004122598",
          "3031233007",
          "8010026",
          "3014383500",
          "3009746061",
          "9611665",
          "3017509841",
          "3013764800",
          "2520313",
          "3007344957",
          "1125615",
          "1220477",
          "9617594",
          "1220905",
          "3007817041",
          "2244478",
          "1911916",
          "3004194523",
          "1649518",
          "1450662",
          "2030598",
          "3015192982",
          "1526863",
          "1048735",
          "3012429465",
          "2032112",
          "1056186",
          "2320762",
          "3012461127",
          "3003927385",
          "3013445787",
          "3006542380",
          "3007590959",
          "3014162214",
          "1720929",
          "2427500",
          "3012050423",
          "1055927",
          "1319211",
          "2032098",
          "3012445205",
          "3002807314",
          "3013055489",
          "3013162291",
          "3017406698",
          "3003442380",
          "2321034",
          "3013517171",
          "9616066",
          "3002808467",
          "3011270181",
          "8021079",
          "9681835",
          "1450392",
          "3004167615",
          "1024879",
          "3020460367",
          "3007738736",
          "3015309643",
          "1123137",
          "3015997711",
          "3011088743",
          "3010405396",
          "3001124136",
          "3029973819",
          "1000523114",
          "3010934635",
          "3010082909",
          "3012552122",
          "2248588",
          "3008806809",
          "3014662855",
          "3003512059",
          "3006174295",
          "3012337868",
          "3004102031",
          "9616567",
          "8040700",
          "3006942602",
          "3010041511",
          "9610825",
          "3012187635",
          "1721504",
          "9615817",
          "3008692839",
          "3045816907",
          "3006479639",
          "9680448",
          "3034669683",
          "3004519921",
          "9615588",
          "3008307705",
          "3044026004",
          "3045322464",
          "3016472589",
          "9612052",
          "3030419987",
          "3011137372",
          "2024064",
          "3013234390",
          "3013298431",
          "3029706818",
          "2518902",
          "3009656988",
          "3018507299",
          "3012798826",
          "3005609751",
          "1224960",
          "3003678543",
          "8020040",
          "3016727301",
          "3017088037",
          "3041276591",
          "3002296055",
          "3003627840",
          "3003189845",
          "9611446",
          "2245270",
          "3010149039",
          "2648727",
          "3035111606",
          "3010872792",
          "1054241",
          "8022210",
          "2030624",
          "3005910889",
          "3038356900",
          "1217831",
          "8020785",
          "3005934752",
          "2029015",
          "3030989877",
          "1933913",
          "3030447506",
          "3036773433",
          "3017961114",
          "3018229072",
          "9617475",
          "1643059",
          "3004111573",
          "9615908",
          "3015140199",
          "8041145",
          "3008537309",
          "9680425",
          "3010381606",
          "1932668",
          "9612051",
          "3020417125",
          "9617592",
          "3012809732",
          "3016798778",
          "3004538333",
          "1723241",
          "3003704957",
          "3009518015",
          "3006846316",
          "1423662",
          "1000324175",
          "3011237770",
          "1644312",
          "9612152",
          "2032227",
          "3004906872",
          "3032916632",
          "3005232773",
          "9616099",
          "3016904853",
          "1526611",
          "3035642068",
          "3012969104",
          "9611004",
          "1423507",
          "3012421607",
          "3006891665",
          "3007939342",
          "1649914",
          "3008782867",
          "3010047442",
          "3004416208",
          "3006795797",
          "3004735647",
          "3010432890",
          "3011689956",
          "2183650",
          "3016844398",
          "9611423",
          "3008717264",
          "2024024",
          "1710034",
          "3015910259",
          "3002806818",
          "3006158088",
          "9610847",
          "3009328501",
          "3005770916",
          "3014732157",
          "3014498869",
          "3015537049",
          "9612057",
          "3017210488",
          "2437459",
          "3009742443",
          "3005515211",
          "2032521",
          "1063382",
          "3006260740",
          "1316297",
          "2649614",
          "1319660",
          "2433012",
          "1000595987",
          "9681819",
          "8041187",
          "3018071447",
          "3002807350",
          "3005941719",
          "1221765",
          "9617229",
          "2521402",
          "1016427",
          "9616848",
          "1422634",
          "3015225571",
          "1223004",
          "3000247873",
          "3013023255",
          "3000268902",
          "3003895440",
          "1820334",
          "1530671"
        ],
        "fei_number": [
          "",
          "3000718659",
          "3003528016",
          "3010291427",
          "3023228837",
          "3012494290",
          "3005012805",
          "3002826093",
          "3012410077",
          "3027327302",
          "3014246639",
          "3001639729",
          "3009631669",
          "3003282181",
          "3006194679",
          "3005987567",
          "3010621964",
          "1000306431",
          "3017262697",
          "3001451226",
          "3015142721",
          "2528981",
          "3016661372",
          "3013557562",
          "3010413216",
          "3008403546",
          "1928237",
          "3003263857",
          "3036978618",
          "3000204727",
          "1000304305",
          "1000306068",
          "3024122953",
          "3007392007",
          "1721676",
          "3005951712",
          "3010293992",
          "2016945",
          "1000514656",
          "3033589330",
          "2011068",
          "3015173212",
          "3007829657",
          "1721686",
          "3010515547",
          "1713468",
          "3003040625",
          "1000587249",
          "3002806469",
          "3010131137",
          "3003767881",
          "3000215273",
          "2025816",
          "1038185",
          "3002930098",
          "3030626857",
          "3019807891",
          "1417592",
          "3008448733",
          "3014456066",
          "3013319212",
          "3000204695",
          "3005340341",
          "3004215921",
          "3008808123",
          "3009456923",
          "3009515470",
          "3001570557",
          "3000327333",
          "2523676",
          "3017567825",
          "3008810357",
          "1416980",
          "3022964615",
          "3012039263",
          "3003700453",
          "3015453963",
          "1000136478",
          "3000974732",
          "3000240707",
          "3010419931",
          "1035907",
          "1000112137",
          "3015276780",
          "3007305485",
          "3019004087",
          "3013696225",
          "3006950086",
          "3043051314",
          "3013187142",
          "3003166194",
          "3005382983",
          "3014579161",
          "1000305720",
          "3043226252",
          "3011987967",
          "2243072",
          "1000112968",
          "3005670221",
          "3006097687",
          "3006425876",
          "1000120555",
          "3015058854",
          "1018233",
          "3002807111",
          "3012104670",
          "1000136269",
          "3009291251",
          "3020778",
          "3015060216",
          "3000206435",
          "3008252904",
          "3008328679",
          "1419937",
          "3030733800",
          "1720734",
          "3003076831",
          "3003015552",
          "3014312726",
          "3002807615",
          "1417485",
          "3013820501",
          "3015983454",
          "1019003",
          "3006534321",
          "3011944170",
          "3013188547",
          "3004605321",
          "3015620563",
          "3002806460",
          "3004548776",
          "3003154916",
          "1030451",
          "3008588205",
          "3010531069",
          "3003310007",
          "1000117748",
          "3002808466",
          "3006942524",
          "3002907620",
          "1000442352",
          "3003974574",
          "3002806588",
          "3005778453",
          "3005458990",
          "3009882464",
          "3004569180",
          "3014314985",
          "3001436772",
          "3003752255",
          "2011171",
          "1314417",
          "3005405146",
          "3008733163",
          "3011642792",
          "3007616112",
          "3002808374",
          "3003674698",
          "2523167",
          "3017636737",
          "3020777588",
          "3007363456",
          "3007836437",
          "3003094851",
          "3003915875",
          "3006003617",
          "1037885",
          "3033536319",
          "3012536737",
          "1220973",
          "3012116434",
          "3016938816",
          "3007007790",
          "3008132411",
          "3014936217",
          "3001236905",
          "3010273872",
          "1000563940",
          "1625425",
          "1043214",
          "3016096098",
          "1643817",
          "1000517719",
          "3018125942",
          "3004140833",
          "3001665800",
          "3042871620",
          "1000512168",
          "3042955949",
          "3007596547",
          "3006082230",
          "1047843",
          "1018470",
          "3000718854",
          "3009746425",
          "3004911384",
          "3007185216",
          "3044994889",
          "3015085713",
          "3003058448",
          "3006271008",
          "3008650117",
          "3015724078",
          "3015859709",
          "3010966701",
          "3002808169",
          "3000335778",
          "1047429",
          "1000123180",
          "3005669815",
          "1000208624",
          "3002806603",
          "1410248",
          "3006575795",
          "3004143450",
          "1000477110",
          "3004122598",
          "3031233007",
          "3014383500",
          "1000115518",
          "3002963732",
          "3009746061",
          "3017509841",
          "3013764800",
          "2520313",
          "3001236277",
          "3007344957",
          "3002808075",
          "1220477",
          "1220905",
          "3007817041",
          "1000160458",
          "3001451571",
          "1911916",
          "3004194523",
          "3004525092",
          "1450662",
          "3015192982",
          "1526863",
          "1048735",
          "3012429465",
          "3003560965",
          "3012461127",
          "3003927385",
          "3005747797",
          "3013445787",
          "3002807095",
          "3006542380",
          "3007590959",
          "3002806457",
          "3014162214",
          "1720929",
          "2427500",
          "3012050423",
          "1319211",
          "3012445205",
          "3002807314",
          "1000138054",
          "3013055489",
          "1000113359",
          "3013162291",
          "3017406698",
          "3003442380",
          "3013517171",
          "3002808467",
          "3011270181",
          "1450392",
          "3008002452",
          "3002806901",
          "3004167615",
          "1024879",
          "1000132332",
          "3020460367",
          "3005846326",
          "3007738736",
          "3015309643",
          "1000112877",
          "3015997711",
          "3011088743",
          "3010405396",
          "3001124136",
          "3029973819",
          "1000523114",
          "3010934635",
          "3010082909",
          "3012552122",
          "3008806809",
          "3014662855",
          "3003512059",
          "3006174295",
          "3003081151",
          "3012337868",
          "3004102031",
          "3006942602",
          "3010041511",
          "3012187635",
          "1721504",
          "3008692839",
          "3045816907",
          "3006479639",
          "3003807466",
          "3002999813",
          "1000036860",
          "3034669683",
          "3008307705",
          "3044026004",
          "3045322464",
          "3016472589",
          "3002806443",
          "3030419987",
          "2024064",
          "3013234390",
          "3002879021",
          "3013298431",
          "3005622231",
          "3029706818",
          "3003737899",
          "2518902",
          "3009656988",
          "3018507299",
          "3012798826",
          "3003399884",
          "3005609751",
          "3003678543",
          "3002807725",
          "3016727301",
          "3017088037",
          "3041276591",
          "3002296055",
          "3003627840",
          "3003189845",
          "1000174591",
          "2245270",
          "3010149039",
          "3006108784",
          "2648727",
          "3002468371",
          "1000518731",
          "3035111606",
          "3010872792",
          "3005910889",
          "3038356900",
          "1217831",
          "3005934752",
          "3030989877",
          "3003886545",
          "3030447506",
          "3002974635",
          "3036773433",
          "3017961114",
          "3018229072",
          "1643059",
          "3004111573",
          "3015140199",
          "3008537309",
          "3010381606",
          "3020417125",
          "3012809732",
          "3016798778",
          "3004538333",
          "3003704957",
          "3009518015",
          "3006846316",
          "3000147578",
          "1000324175",
          "3011237770",
          "3002669612",
          "3004906872",
          "1000237915",
          "3032916632",
          "3005232773",
          "3016904853",
          "1526611",
          "3035642068",
          "3012969104",
          "3002807099",
          "3012421607",
          "3006891665",
          "3007939342",
          "3008782867",
          "3010047442",
          "3002902735",
          "3004416208",
          "3006795797",
          "3004735647",
          "3010432890",
          "3011689956",
          "1000527169",
          "2183650",
          "3016844398",
          "3932",
          "3003574398",
          "3008717264",
          "2024024",
          "1710034",
          "3015910259",
          "3002806818",
          "3006158088",
          "3009328501",
          "3005770916",
          "3014732157",
          "3014498869",
          "3015537049",
          "3017210488",
          "1000226372",
          "3009742443",
          "3005515211",
          "3006260740",
          "1316297",
          "2649614",
          "2433012",
          "1000595987",
          "3018071447",
          "3002807350",
          "1221765",
          "2521402",
          "3002806485",
          "1016427",
          "1000138417",
          "3015225571",
          "3000247873",
          "3013023255",
          "2828",
          "3000268902",
          "3003895440",
          "3002558692",
          "1820334",
          "3009976527"
        ],
        "device_name": "Set, Administration, Intravascular",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5440",
        "device_class": "2"
      }
    }
  ]
}