  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "178955",
      "product_res_number": "Z-0807-2020",
      "event_date_initiated": "2019-12-20",
      "event_date_created": "2020-01-13",
      "recall_status": "Terminated",
      "event_date_terminated": "2020-05-01",
      "res_event_number": "84648",
      "product_code": "OWB",
      "k_numbers": [
        "K163286"
      ],
      "product_description": "Siemens ARTIS pheno, Model # 10849000.  Interventional Fluoroscopic X-Ray System - Product Usage: Angiography system for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.",
      "code_info": "Serial Numbers 164231, 164234, 164240, 164241, 164242, 164244",
      "firm_fei_number": "2240869",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "city": "Malvern",
      "state": "PA",
      "postal_code": "19355-1418",
      "additional_info_contact": "610-219-4834",
      "reason_for_recall": "The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred.",
      "root_cause_description": "Device Design",
      "action": "Siemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notice to all affected customers via AX070/19/S on 12/20/2019.",
      "product_quantity": "6",
      "distribution_pattern": "US Nationwide distribution in the states of FL, IL, IN, MA, PA, VA.",
      "openfda": {
        "k_number": [
          "K231692",
          "K183109",
          "K251523",
          "K210900",
          "K171829",
          "K232344",
          "K021049",
          "K032761",
          "K240199",
          "K051064",
          "K212523",
          "K092004",
          "K200917",
          "K133278",
          "K202547",
          "K220946",
          "K173736",
          "K181550",
          "K982993",
          "K200074",
          "K212212",
          "K032046",
          "K252099",
          "K041931",
          "K212159",
          "K993037",
          "K160279",
          "K130893",
          "K230871",
          "K181670",
          "K260607",
          "K961105",
          "K182415",
          "K181407",
          "K190768",
          "K172180",
          "K150967",
          "K061173",
          "K212383",
          "K092438",
          "K222998",
          "K141979",
          "K163286",
          "K181830",
          "K251893",
          "K193230",
          "K132094",
          "K171565",
          "K243303",
          "K172863",
          "K061953",
          "K242731",
          "K203551",
          "K133580",
          "K090590",
          "K052412",
          "K213479",
          "K172646",
          "K030061",
          "K152928",
          "K232298",
          "K220311",
          "K170130",
          "K092439",
          "K142126",
          "K223409",
          "K041605",
          "K200218",
          "K961374",
          "K241572",
          "K052202",
          "K024200",
          "K160455",
          "K220627",
          "K042053",
          "K173433",
          "K070834",
          "K142740",
          "K181415",
          "K221281",
          "K222080",
          "K203281",
          "K142243",
          "K190626",
          "K033244",
          "K240224",
          "K222070",
          "K132305",
          "K963952",
          "K201172",
          "K210055",
          "K212336",
          "K210307",
          "K251520",
          "K023178",
          "K033737",
          "K140138",
          "K120073",
          "K243328",
          "K112180",
          "K151000",
          "K091918",
          "K210054",
          "K153232",
          "K070123",
          "K010435",
          "K173639",
          "K122644",
          "K251827",
          "K142736",
          "K243646",
          "K253752",
          "K190064",
          "K202360",
          "K123508",
          "K043076",
          "K181966",
          "K200360",
          "K974355",
          "K253670",
          "K220342",
          "K022114",
          "K061163",
          "K030544",
          "K161563",
          "K133292",
          "K210912",
          "K213768",
          "K152697",
          "K033486",
          "K123529",
          "K093066",
          "K220432",
          "K100102",
          "K202431",
          "K170909",
          "K191503",
          "K102529",
          "K082243",
          "K250660",
          "K042793",
          "K101833",
          "K103416",
          "K131263",
          "K143225",
          "K254186",
          "K243884",
          "K181560",
          "K161976",
          "K012490",
          "K234067",
          "K254173",
          "K243735",
          "K163140",
          "K201743",
          "K200887",
          "K040675",
          "K223410",
          "K203504",
          "K254205",
          "K050489",
          "K183144",
          "K170752",
          "K171194",
          "K152785",
          "K210309",
          "K162148",
          "K240020",
          "K142273",
          "K082781",
          "K251602",
          "K122457",
          "K061203",
          "K052884",
          "K022322",
          "K233230",
          "K173664",
          "K153244",
          "K133535",
          "K132970",
          "K170832",
          "K181834",
          "K111551",
          "K201156",
          "K093688",
          "K222625",
          "K230950",
          "K992660",
          "K982427",
          "K102005",
          "K233748",
          "K041780",
          "K241567",
          "K221230",
          "K201583",
          "K231669",
          "K212557",
          "K200663",
          "K921219",
          "K992103",
          "K252503",
          "K234109",
          "K121444",
          "K083725",
          "K103425",
          "K022069",
          "K140041",
          "K062647",
          "K061534",
          "K000221",
          "K212813",
          "K183040",
          "K250282",
          "K253584",
          "K160147",
          "K170144",
          "K190497",
          "K011311",
          "K170800",
          "K062623",
          "K151904",
          "K031637",
          "K221344",
          "K120282",
          "K011292",
          "K172574",
          "K113403",
          "K172700",
          "K221516",
          "K180473",
          "K252068",
          "K960920",
          "K211777",
          "K130638",
          "K002198",
          "K200022",
          "K122234",
          "K060259",
          "K230949",
          "K101868",
          "K221335",
          "K221922",
          "K970734",
          "K161756",
          "K252500",
          "K193326",
          "K002016",
          "K162859",
          "K231700",
          "K970063",
          "K211064",
          "K192182",
          "K211191",
          "K161839",
          "K151598",
          "K173314",
          "K250010",
          "K181804",
          "K160088",
          "K250241",
          "K143281",
          "K243432",
          "K183101",
          "K172822",
          "K133819",
          "K132027",
          "K243411",
          "K213455",
          "K181943",
          "K182546",
          "K053279",
          "K232420",
          "K132147",
          "K152696",
          "K233200",
          "K172307",
          "K073543",
          "K912448",
          "K121772",
          "K083545",
          "K211782",
          "K240099",
          "K171755",
          "K111344",
          "K203535",
          "K212956",
          "K252229",
          "K091658",
          "K923049",
          "K162614",
          "K132558",
          "K181767",
          "K243376",
          "K182086",
          "K010942",
          "K142708",
          "K252735",
          "K211780",
          "K251199",
          "K041932",
          "K230185",
          "K230787",
          "K243446",
          "K251992",
          "K203623",
          "K113027",
          "K173631",
          "K212442",
          "K080847",
          "K993687",
          "K233107",
          "K123226",
          "K141574",
          "K936180",
          "K231701",
          "K153147",
          "K163715",
          "K220409",
          "K240465",
          "K983069",
          "K081073",
          "K013608",
          "K181403"
        ],
        "registration_number": [
          "",
          "3012624817",
          "3011950073",
          "8043933",
          "3005172103",
          "3026391023",
          "3009001657",
          "3012761544",
          "3010162973",
          "3043048939",
          "3012127758",
          "3013944123",
          "3012957335",
          "3031503890",
          "3003775006",
          "3008632256",
          "3014328942",
          "3011015597",
          "3008284470",
          "3027373189",
          "3023003446",
          "3014391847",
          "1052701",
          "3047842058",
          "3010703925",
          "1054713",
          "3014343100",
          "3036666444",
          "3018077344",
          "3033525987",
          "3035950510",
          "3006128100",
          "1225677",
          "3021559833",
          "3003768277",
          "3001722928",
          "3008044777",
          "3005843418",
          "3012631246",
          "3030691709",
          "9613445",
          "9681146",
          "3016619943",
          "3003202425",
          "3033527282",
          "3006972752",
          "1720753",
          "3011240766",
          "3012911444",
          "3010685285",
          "3004198450",
          "1000188474",
          "3009144915",
          "3011416394",
          "3021059265",
          "3006544299",
          "3010462278",
          "3015223550",
          "3042175844",
          "2247992",
          "1000181430",
          "2939520",
          "3005877899",
          "3018094310",
          "3015232217",
          "3011491794",
          "3033527770",
          "3015437189",
          "3005810333",
          "3000126629",
          "3021060474",
          "3008363989",
          "2027062",
          "9611343",
          "3017375271",
          "3004785967",
          "2243057",
          "3030412764",
          "1835251",
          "3009715136",
          "3013184485",
          "3030089787",
          "3008496839",
          "3016438694",
          "1047843",
          "3008253300",
          "3027339477",
          "3009196578",
          "3009077524",
          "3011713200",
          "3003477135",
          "3009963139",
          "3004754211",
          "3033525986",
          "3004977335",
          "3003935342",
          "3038614464",
          "3018596740",
          "3004026651",
          "3022378058",
          "3025394757",
          "3009973505",
          "1056553",
          "3011347852",
          "3003768251",
          "3036255956",
          "2133772",
          "9614698",
          "2031966",
          "1723170",
          "3019878714",
          "3010398867",
          "3015276088",
          "3012939903",
          "3021027284",
          "3014279513",
          "3005706667",
          "3002466018",
          "8030233",
          "2027467",
          "1037955"
        ],
        "fei_number": [
          "",
          "3012624817",
          "3011950073",
          "3005172103",
          "3026391023",
          "1000284313",
          "3009001657",
          "3012761544",
          "3043048939",
          "3012127758",
          "3013944123",
          "3012957335",
          "3031503890",
          "3002807041",
          "3003775006",
          "3008632256",
          "3014328942",
          "3011015597",
          "3008284470",
          "3027373189",
          "3023003446",
          "3014391847",
          "3047842058",
          "3010703925",
          "3014343100",
          "3002808139",
          "1000517638",
          "3036666444",
          "3018077344",
          "3033525987",
          "3035950510",
          "3006128100",
          "1000582291",
          "3021559833",
          "3001722928",
          "3008044777",
          "3005843418",
          "3012631246",
          "3030691709",
          "3016619943",
          "3003202425",
          "3033527282",
          "3006972752",
          "1720753",
          "3011240766",
          "3012911444",
          "3010685285",
          "3004198450",
          "1000188474",
          "3009144915",
          "1000208827",
          "3000209794",
          "3011416394",
          "3021059265",
          "3006544299",
          "3010462278",
          "3015223550",
          "1000132370",
          "3042175844",
          "1000181430",
          "2939520",
          "3005877899",
          "3018094310",
          "3015232217",
          "3011491794",
          "3002808382",
          "3033527770",
          "3015437189",
          "1000114111",
          "3005810333",
          "3000126629",
          "3021060474",
          "3008363989",
          "2027062",
          "3017375271",
          "3004785967",
          "2243057",
          "3030412764",
          "3009715136",
          "3013184485",
          "3030089787",
          "3008496839",
          "3016438694",
          "1047843",
          "3008253300",
          "3027339477",
          "3009196578",
          "3009077524",
          "3011713200",
          "3003477135",
          "3009963139",
          "3004754211",
          "3033525986",
          "3004977335",
          "3003935342",
          "3038614464",
          "3005430103",
          "3018596740",
          "3000205591",
          "3022378058",
          "3025394757",
          "3009973505",
          "3002175190",
          "3011347852",
          "3003768251",
          "3036255956",
          "3002807880",
          "3002619595",
          "3019878714",
          "1000222289",
          "3010398867",
          "3015276088",
          "3001451453",
          "3002980729",
          "3012939903",
          "3021027284",
          "3014279513",
          "3005706667",
          "3002466018",
          "2027467",
          "1037955"
        ],
        "device_name": "Interventional Fluoroscopic X-Ray System",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1650",
        "device_class": "2"
      }
    }
  ]
}