  "results": [
    {
      "cfres_id": "178032",
      "product_res_number": "Z-0766-2020",
      "event_date_initiated": "2019-11-04",
      "event_date_created": "2019-12-31",
      "recall_status": "Open, Classified",
      "res_event_number": "84309",
      "product_code": "KXJ",
      "k_numbers": [
        "K050190"
      ],
      "product_description": "Apollo EZ, Model Number 9784152036 - Product Usage: The Apollo EZ remote-controlled table is X-ray equipment that can be used to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.",
      "code_info": "Serial Numbers: 13101261",
      "firm_fei_number": "3002808461",
      "recalling_firm": "Villa Sistemi Medicali S.P.A.",
      "address_1": "Via delle Azalee 3 3 via delle azalee",
      "city": "Buccinasco",
      "country": "Italy",
      "reason_for_recall": "An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit.  After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.",
      "root_cause_description": "Device Design",
      "action": "Urgent Field Safety Notice notification letters dated 10/23/19 were sent to customers.\n\nAction to be taken by the customer/user:\nThe user shall take note of amendment/reinforcement of Instructions for Use (IFU) and Service Manual, attached to this FN and, if it is the case, request a service technician to do the periodic preventive maintenance the user shall replace the attached pages in the User Manual and in the Service Manual, prescribing the periodic replacement of the angulation chain.  Fill the table present in the addendum; In the cell \"Next replacement date\" write the date adding 8 years to the replacement date.\n\nAll the examinations performed on the device before the recall remain valid.\n\nThe IFU and the Service Manual need to be amended to prescribe the replacement of the angulation chain after 8 years of service life of the unit.  \n\nThis notice needs to be passed to all users of Apollo, Apollo DRF, Apollo EZ and to all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred.",
      "product_quantity": "1",
      "distribution_pattern": "US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.",
      "openfda": {
        "k_number": [
          "K913121",
          "K811196",
          "K050190",
          "K820111",
          "K951146",
          "K950442",
          "K862144",
          "K922098",
          "K834478",
          "K883155",
          "K790692",
          "K931724",
          "K803120",
          "K950068",
          "K940696",
          "K963905",
          "K860620",
          "K871827",
          "K812791",
          "K854328",
          "K790743",
          "K790783",
          "K881700",
          "K955414",
          "K874934",
          "K790536",
          "K926407",
          "K874639",
          "K811969",
          "K781992",
          "K883885"
        ],
        "registration_number": [
          "3014209556",
          "3042175844",
          "3011491794",
          "1225677",
          "3010326005",
          "3031563141",
          "3014680784",
          "3005778458",
          "3003768277",
          "3003768251",
          "3000126629",
          "8010652",
          "9616047",
          "1418964",
          "3007143268",
          "2183553",
          "3010949642",
          "1066635",
          "3008284470",
          "8021091",
          "3004756607",
          "3013836619",
          "3002466018",
          "8020997",
          "9617251",
          "3004957147",
          "9613183",
          "2921578",
          "3010890274",
          "2247992"
        ],
        "fei_number": [
          "3014209556",
          "3042175844",
          "3011491794",
          "3010326005",
          "3031563141",
          "1841",
          "3005778458",
          "3003768251",
          "3000126629",
          "3002807880",
          "3002495042",
          "1418964",
          "1825560",
          "3003377038",
          "3007143268",
          "2183553",
          "3010949642",
          "3001451453",
          "3008284470",
          "3003184737",
          "1000208827",
          "3004756607",
          "3013836619",
          "3002466018",
          "3002808461",
          "3004957147",
          "2921578",
          "3010890274",
          "3002807092",
          "1000227792"
        ],
        "device_name": "Table, Radiologic",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1980",
        "device_class": "2"
      }
    }
  ]
}