      "cfres_id": "177476",
      "product_res_number": "Z-0567-2020",
      "event_date_initiated": "2016-11-08",
      "event_date_created": "2019-11-25",
      "recall_status": "Terminated",
      "event_date_terminated": "2020-11-03",
      "res_event_number": "84180",
      "product_code": "HWC",
      "product_description": "Dart Fire HEADED SCREW 4.0X 32MM, Part Number D1S40032S",
      "code_info": "Lot Number 1555125",
      "firm_fei_number": "1043534",
      "recalling_firm": "Wright Medical Technology, Inc.",
      "address_1": "11576 Memphis Arlington Rd",
      "city": "Arlington",
      "state": "TN",
      "postal_code": "38002-9497",
      "reason_for_recall": "The firm discovered a lack of adequate documentation confirming sterility on certain products.",
      "root_cause_description": "Process control",
      "action": "The firm initiated the recall on 11/08/2016 by letter and email to the user level.  The firm requested the return of the units.",
      "product_quantity": "77 units",
      "distribution_pattern": "US and Norway, Italy, Germany, France, South Africa, UK, Taiwan, Austria, Japan, Turkey, Denmark, Canada, Switzerland, Belgium, The Netherlands, Spain",
      "openfda": {
        "k_number": [
          "K171407",
          "K110538",
          "K032072",
          "K952831",
          "K130590",
          "K230945",
          "K221645",
          "K021662",
          "K182342",
          "K173710",
          "K062352",
          "K943806",
          "K012001",
          "K180257",
          "K241277",
          "K041936",
          "K163489",
          "K151270",
          "K043281",
          "K203239",
          "K193660",
          "K220359",
          "K130859",
          "K012469",
          "K071092",
          "K211261",
          "K080352",
          "K123362",
          "K935836",
          "K111971",
          "K904809",
          "K023294",
          "K964970",
          "K962011",
          "K232251",
          "K050683",
          "K940207",
          "K153065",
          "K133451",
          "K121148",
          "K121277",
          "K121349",
          "K012798",
          "K131620",
          "K150443",
          "K173556",
          "K962779",
          "K221193",
          "K073426",
          "K190830",
          "K010738",
          "K014304",
          "K090656",
          "K150981",
          "K070507",
          "K892925",
          "K021280",
          "K083792",
          "K960940",
          "K020266",
          "K013903",
          "K870141",
          "K981372",
          "K040656",
          "K053136",
          "K060830",
          "K141857",
          "K233912",
          "K912640",
          "K990194",
          "K922166",
          "K062348",
          "K073054",
          "K032682",
          "K150376",
          "K971800",
          "K895389",
          "K010525",
          "K962236",
          "K031939",
          "K071091",
          "K132244",
          "K223761",
          "K253858",
          "K925872",
          "K954757",
          "K033252",
          "K051309",
          "K110445",
          "K053543",
          "K133636",
          "K170698",
          "K050749",
          "K053394",
          "K021932",
          "K953198",
          "K153180",
          "K213572",
          "K953350",
          "K990804",
          "K080088",
          "K060736",
          "K091202",
          "K991151",
          "K945687",
          "K931524",
          "K083912",
          "K874672",
          "K915424",
          "K960533",
          "K981187",
          "K992301",
          "K102326",
          "K992947",
          "K042030",
          "K070617",
          "K201430",
          "K041749",
          "K092189",
          "K211996",
          "K252699",
          "K183055",
          "K190970",
          "K260934",
          "K021233",
          "K023922",
          "K060710",
          "K203335",
          "K040356",
          "K142776",
          "K931665",
          "K112885",
          "K223847",
          "K180377",
          "K222799",
          "K972233",
          "K003816",
          "K170782",
          "K930980",
          "K122959",
          "K003641",
          "K961413",
          "K070525",
          "K153333",
          "K040740",
          "K101808",
          "K153638",
          "K002070",
          "K023260",
          "K130019",
          "K183026",
          "K103705",
          "K010554",
          "K163038",
          "K213421",
          "K991477",
          "K875050",
          "K111536",
          "K093820",
          "K960692",
          "K200987",
          "K930834",
          "K143188",
          "K161597",
          "K172596",
          "K260361",
          "K896601",
          "K242691",
          "K153154",
          "K130613",
          "K951304",
          "K911398",
          "K080280",
          "K261154",
          "K950704",
          "K161616",
          "K042244",
          "K053344",
          "K213614",
          "K994152",
          "K932862",
          "K931187",
          "K083619",
          "K895219",
          "K132217",
          "K981259",
          "K170889",
          "K923477",
          "K954485",
          "K142707",
          "K140408",
          "K051611",
          "K003077",
          "K093770",
          "K151241",
          "K130627",
          "K030995",
          "K100457",
          "K875053",
          "K160174",
          "K092754",
          "K093055",
          "K013685",
          "K052405",
          "K143460",
          "K081071",
          "K071952",
          "K982222",
          "K112658",
          "K162153",
          "K120493",
          "K992277",
          "K960537",
          "K023365",
          "K013044",
          "K951303",
          "K903254",
          "K113591",
          "K022234",
          "K042078",
          "K142442",
          "K013131",
          "K161334",
          "K012000",
          "K021351",
          "K252019",
          "K021579",
          "K252901",
          "K153759",
          "K041961",
          "K063584",
          "K013954",
          "K910976",
          "K031945",
          "K010783",
          "K030900",
          "K170350",
          "K062980",
          "K893512",
          "K903244",
          "K142658",
          "K984209",
          "K981670",
          "K032894",
          "K912598",
          "K810497",
          "K143165",
          "K222564",
          "K013890",
          "K972217",
          "K031050",
          "K113084",
          "K991298",
          "K090994",
          "K250536",
          "K091798",
          "K020352",
          "K031714",
          "K180932",
          "K915448",
          "K133460",
          "K132910",
          "K021254",
          "K170040",
          "K240815",
          "K970929",
          "K980440",
          "K050388",
          "K963192",
          "K962233",
          "K010595",
          "K180541",
          "K895766",
          "K070347",
          "K073509",
          "K060693",
          "K170594",
          "K050346",
          "K063635",
          "K980841",
          "K051323",
          "K053093",
          "K073159",
          "K963739",
          "K024115",
          "K151708",
          "K993025",
          "K150871",
          "K092909",
          "K030368",
          "K042079",
          "K212547",
          "K032224",
          "K033975",
          "K060249",
          "K041835",
          "K050358",
          "K052576",
          "K051098",
          "K220723",
          "K112237",
          "K112772",
          "K002274",
          "K911768",
          "K122031",
          "K162874",
          "K253042",
          "K923841",
          "K082874",
          "K062103",
          "K203029",
          "K210268",
          "K260809",
          "K894719",
          "K241592",
          "K931155",
          "K023022",
          "K081700",
          "K070906",
          "K073312",
          "K941941",
          "K211290",
          "K012945",
          "K220319",
          "K994202",
          "K000089",
          "K210146",
          "K974785",
          "K140952",
          "K932027",
          "K213036",
          "K923256",
          "K872738",
          "K993762",
          "K991197",
          "K100721",
          "K172383",
          "K130120",
          "K201662",
          "K984331",
          "K123672",
          "K150309",
          "K871037",
          "K033740",
          "K262075",
          "K941866",
          "K180670",
          "K081755",
          "K211783",
          "K931681",
          "K242896",
          "K032830",
          "K011821",
          "K961234",
          "K151418",
          "K153182",
          "K190324",
          "K111265",
          "K960704",
          "K864626",
          "K071541",
          "K950652",
          "K171018",
          "K071035",
          "K021030",
          "K183104",
          "K162353",
          "K220921",
          "K012872",
          "K780235",
          "K043102",
          "K230744",
          "K223184",
          "K132912",
          "K082906",
          "K961320",
          "K021832",
          "K140741",
          "K140283",
          "K160603",
          "K100359",
          "K900132",
          "K033639",
          "K872156",
          "K850024",
          "K133662",
          "K160304",
          "K961157",
          "K170021",
          "K181377",
          "K983006",
          "K071481",
          "K130412",
          "K200259",
          "K180024",
          "K990513",
          "K992945",
          "K955710",
          "K060428",
          "K180573",
          "K211435",
          "K831873",
          "K952296",
          "K041289",
          "K223392",
          "K102324",
          "K131591",
          "K792205",
          "K992106",
          "K061206",
          "K030848",
          "K150188",
          "K183395",
          "K231649",
          "K062820",
          "K111316",
          "K932350",
          "K050407",
          "K043185",
          "K010964",
          "K042584",
          "K193491",
          "K192975",
          "K913269",
          "K140891",
          "K183363",
          "K983508",
          "K934590",
          "K213596",
          "K081392",
          "K081893",
          "K070039",
          "K180500",
          "K002271",
          "K112727",
          "K093676",
          "K102410",
          "K152072",
          "K951656",
          "K163303",
          "K210792",
          "K993166",
          "K954511",
          "K021556",
          "K162825",
          "K020043",
          "K874676",
          "K934321",
          "K182684",
          "K081513",
          "K112672",
          "K070441",
          "K103183",
          "K020056",
          "K071941",
          "K211944",
          "K963172",
          "K143624",
          "K080101",
          "K010633",
          "K990019",
          "K080850",
          "K091577",
          "K955837",
          "K161058",
          "K112737",
          "K940690",
          "K133035",
          "K930970",
          "K153378",
          "K130400",
          "K965073",
          "K022647",
          "K042310",
          "K980364",
          "K934366",
          "K200914",
          "K060026",
          "K102298",
          "K063479",
          "K190921",
          "K190035",
          "K230507",
          "K040331",
          "K926047",
          "K051726",
          "K042778",
          "K182148",
          "K905048",
          "K962706",
          "K894830",
          "K994382",
          "K220839",
          "K153575",
          "K080358",
          "K202817",
          "K161539",
          "K860017",
          "K961905",
          "K162171",
          "K854827",
          "K032717",
          "K050754",
          "K962823",
          "K874130",
          "K023887",
          "K050767",
          "K122596",
          "K061863",
          "K972494",
          "K061262",
          "K945182",
          "K110895",
          "K101254",
          "K083607",
          "K250304",
          "K001581",
          "K030706",
          "K071437",
          "K062863",
          "K061665",
          "K131920",
          "K231504",
          "K182361",
          "K081813",
          "K013572",
          "K023192",
          "K980681",
          "K990622",
          "K021828",
          "K063298",
          "K910193",
          "K190231",
          "K160427",
          "K081428",
          "K983165",
          "K951493",
          "K031751",
          "K101030",
          "K921481",
          "K973586",
          "K150772",
          "K181898",
          "K200291",
          "K961649",
          "K031937",
          "K220511",
          "K050468",
          "K213301",
          "K140148",
          "K913210",
          "K121682",
          "K011910",
          "K131471",
          "K944330",
          "K071157",
          "K170382",
          "K182949",
          "K962987",
          "K063478",
          "K203381",
          "K151021",
          "K211770",
          "K093078",
          "K971654",
          "K110658",
          "K040701",
          "K960962",
          "K111608",
          "K050226",
          "K003700",
          "K160617",
          "K030302",
          "K213199",
          "K023021",
          "K172213",
          "K132248",
          "K950250",
          "K160169",
          "K113388",
          "K042796",
          "K033355",
          "K903811",
          "K091093",
          "K242671",
          "K231819",
          "K251382",
          "K945303",
          "K050497",
          "K190586",
          "K213530",
          "K874670",
          "K881327",
          "K162685",
          "K011102",
          "K201319",
          "K020446",
          "K912021",
          "K203011",
          "K191651",
          "K102528",
          "K931680",
          "K050636",
          "K970549",
          "K171628",
          "K932721",
          "K040475",
          "K140962",
          "K130954",
          "K192578",
          "K261241",
          "K033804",
          "K060914",
          "K990351",
          "K182989",
          "K170012",
          "K253108",
          "K231453",
          "K241359",
          "K152841",
          "K020377",
          "K903259",
          "K973810",
          "K230053",
          "K992367",
          "K143614",
          "K954246",
          "K242348",
          "K014154",
          "K032228",
          "K946347",
          "K970829",
          "K091694",
          "K230397",
          "K080649",
          "K971069",
          "K080259",
          "K220892",
          "K112583",
          "K060664",
          "K984320",
          "K973309",
          "K170688",
          "K152527",
          "K143039",
          "K972403",
          "K913415",
          "K100532",
          "K032770",
          "K143402",
          "K052630",
          "K040594",
          "K222820",
          "K053561",
          "K141401",
          "K112028",
          "K222967",
          "K203467",
          "K903253",
          "K022726",
          "K983592",
          "K012280",
          "K202569",
          "K951545",
          "K071816",
          "K150211",
          "K013786",
          "K040542",
          "K191817",
          "K993975",
          "K151078",
          "K042552",
          "K161449",
          "K171448",
          "K113399",
          "K151146",
          "K163148",
          "K203773",
          "K193497",
          "K862230",
          "K082516",
          "K914004",
          "K915187",
          "K945236",
          "K024233",
          "K150693",
          "K101700",
          "K982497",
          "K110592",
          "K013546",
          "K944213",
          "K943539",
          "K192974",
          "K070393",
          "K951232",
          "K051310",
          "K201850",
          "K060253",
          "K012817",
          "K082949",
          "K141937",
          "K951389",
          "K120449",
          "K041533",
          "K983495",
          "K933719",
          "K014153",
          "K191995",
          "K080943",
          "K203595",
          "K903255",
          "K100325",
          "K240530",
          "K953650",
          "K014185",
          "K203105",
          "K090175",
          "K082282",
          "K931652",
          "K896044",
          "K170624",
          "K193451",
          "K071177",
          "K924514",
          "K111001",
          "K150974",
          "K931790",
          "K060248",
          "K060714",
          "K060271",
          "K990770",
          "K041075",
          "K231236",
          "K043176",
          "K231860",
          "K971068",
          "K230749",
          "K082574",
          "K070389",
          "K895218",
          "K132893",
          "K022981",
          "K011946",
          "K063244",
          "K990454",
          "K924018",
          "K231231",
          "K172617",
          "K254215",
          "K111799",
          "K151173",
          "K072127",
          "K250867",
          "K991024",
          "K060478",
          "K954330",
          "K052607",
          "K172698",
          "K770035",
          "K041189",
          "K925328",
          "K180753",
          "K992487",
          "K050924",
          "K212755",
          "K151701",
          "K243195",
          "K031155",
          "K091686",
          "K050819",
          "K142490",
          "K934044",
          "K111447",
          "K160058",
          "K124027",
          "K850770",
          "K061621",
          "K024060",
          "K191424",
          "K032966",
          "K033233",
          "K042356",
          "K111000",
          "K242442",
          "K955733",
          "K210159",
          "K905357",
          "K191771",
          "K160946",
          "K905333",
          "K080778",
          "K854336",
          "K901616",
          "K152187",
          "K252312",
          "K982662",
          "K122489",
          "K192745",
          "K222505",
          "K061763",
          "K250251",
          "K030388",
          "K032098",
          "K043142",
          "K031685",
          "K971070",
          "K230595",
          "K102162",
          "K003599",
          "K915316",
          "K250613",
          "K992611",
          "K142057",
          "K072848",
          "K142534",
          "K141949",
          "K012818",
          "K113006",
          "K181610",
          "K001231",
          "K170056",
          "K143146",
          "K071115",
          "K943160",
          "K081390",
          "K234040",
          "K212996",
          "K131324",
          "K233198",
          "K853869",
          "K901814",
          "K960652",
          "K872859",
          "K200840",
          "K092440",
          "K151838",
          "K103712",
          "K892214",
          "K041287",
          "K061620",
          "K942340",
          "K063589",
          "K141290",
          "K033749",
          "K934469",
          "K020222",
          "K011719",
          "K111230",
          "K081149",
          "K081122",
          "K252758",
          "K233396",
          "K943474",
          "K033792",
          "K150252",
          "K072291",
          "K014243",
          "K950833",
          "K942341",
          "K202589",
          "K221333",
          "K200124",
          "K980457",
          "K063070",
          "K133842",
          "K092082",
          "K081011",
          "K230118",
          "K792022",
          "K895217",
          "K233567",
          "K990776",
          "K970912",
          "K873251",
          "K183628",
          "K254110",
          "K143229",
          "K061765",
          "K161778",
          "K161894",
          "K011262",
          "K170571",
          "K951472",
          "K894124",
          "K090949",
          "K150126",
          "K060071",
          "K071656",
          "K133713",
          "K170038",
          "K943804",
          "K033524",
          "K190430",
          "K211628",
          "K201556",
          "K952697",
          "K191289",
          "K120645",
          "K934615",
          "K903258",
          "K040265",
          "K963937",
          "K020791",
          "K901144",
          "K142585",
          "K210754",
          "K222409",
          "K915124",
          "K111908",
          "K173286",
          "K152710",
          "K123890",
          "K121417",
          "K973758",
          "K230946",
          "K851214",
          "K934616",
          "K010594",
          "K192417",
          "K895711",
          "K202326",
          "K251555",
          "K062679",
          "K963398",
          "K071464",
          "K032447",
          "K062487",
          "K043337",
          "K010884",
          "K130810",
          "K071498",
          "K210994",
          "K140970",
          "K905358",
          "K230867",
          "K082934",
          "K950875",
          "K922322",
          "K101165",
          "K013107",
          "K061949",
          "K945426",
          "K060393",
          "K973005",
          "K971778",
          "K193156",
          "K254077",
          "K945525",
          "K161259",
          "K002654",
          "K952460",
          "K963350",
          "K933738",
          "K111994",
          "K882934",
          "K221090",
          "K951233",
          "K082320",
          "K160791",
          "K050634",
          "K163710",
          "K122738",
          "K855031",
          "K193366",
          "K922788",
          "K143596",
          "K132911",
          "K991964",
          "K071639",
          "K993630",
          "K972919",
          "K930107",
          "K161904",
          "K153338",
          "K080974",
          "K160548",
          "K171558",
          "K931538",
          "K102343",
          "K230908",
          "K221610",
          "K243943",
          "K992396",
          "K042295",
          "K101696",
          "K013534",
          "K081510",
          "K072785",
          "K974756",
          "K200226",
          "K883910",
          "K233008",
          "K121713",
          "K023013",
          "K972722",
          "K020159",
          "K240175",
          "K941174",
          "K202680",
          "K112240",
          "K133515",
          "K223468",
          "K000080",
          "K041069",
          "K852026",
          "K043583",
          "K142528",
          "K032902",
          "K926579",
          "K110763",
          "K091069",
          "K102903",
          "K874593",
          "K974538",
          "K012572",
          "K083612",
          "K041274",
          "K182955",
          "K150213",
          "K041356",
          "K972599",
          "K915613",
          "K133440",
          "K933352",
          "K952272",
          "K041456",
          "K062617",
          "K183603",
          "K043410",
          "K131264",
          "K963118",
          "K203228",
          "K081934",
          "K032167",
          "K810205",
          "K173135",
          "K864633",
          "K024086",
          "K191342",
          "K022952",
          "K053338",
          "K262344",
          "K092133",
          "K903810",
          "K180050",
          "K091118",
          "K882831",
          "K021883",
          "K210490",
          "K192096",
          "K052080",
          "K860323",
          "K170687",
          "K854826",
          "K130319",
          "K083096",
          "K180348",
          "K120390",
          "K011451",
          "K063588",
          "K221342",
          "K182677",
          "K111834"
        ],
        "registration_number": [
          "",
          "3016851379",
          "3017209758",
          "3008583793",
          "3006791286",
          "3016759112",
          "3004154314",
          "3007289093",
          "8010379",
          "3000203391",
          "3014967149",
          "3009941480",
          "3004464325",
          "3015718210",
          "3013055499",
          "9710312",
          "2532027",
          "3010097171",
          "3007125392",
          "3008812251",
          "3005827567",
          "3016727307",
          "3004358587",
          "3010032903",
          "3007366790",
          "3010560653",
          "9617083",
          "3010120104",
          "3011015572",
          "3022928161",
          "3008870565",
          "8010099",
          "3004638532",
          "3021008900",
          "3014257776",
          "3034525",
          "3007113169",
          "3014314623",
          "3012470322",
          "1221934",
          "3004491689",
          "3021226419",
          "1225838",
          "3020899789",
          "3007441485",
          "3011301313",
          "3042508254",
          "3015075784",
          "3014104093",
          "3010120148",
          "2011171",
          "1423662",
          "3011656326",
          "2133928",
          "1036836",
          "3009882675",
          "3008747271",
          "1043534",
          "3010673777",
          "3041501961",
          "3003898228",
          "3003236702",
          "3006128100",
          "1526711",
          "3026586698",
          "1319660",
          "9681465",
          "3009973699",
          "3008791302",
          "3009189869",
          "2031093",
          "2183650",
          "3015183635",
          "3013561205",
          "3009132077",
          "3008650117",
          "9611813",
          "9615021",
          "3009201148",
          "3010120339",
          "3016247852",
          "3004549189",
          "2528981",
          "3005180920",
          "8010177",
          "1647149",
          "9611390",
          "3009732568",
          "9616944",
          "3016778139",
          "8010944",
          "3006638824",
          "3010057495",
          "3014483751",
          "3012724179",
          "3030514683",
          "3030926959",
          "1835831",
          "3007700286",
          "3017521423",
          "3009417901",
          "3010118938",
          "3010440433",
          "2648666",
          "3014937101",
          "3017528621",
          "3020584246",
          "3013564228",
          "3018094310",
          "3013194153",
          "3016660811",
          "1527105",
          "3009394448",
          "3008951116",
          "3003139373",
          "3005029036",
          "3010123256",
          "3014522470",
          "3038776593",
          "2134285",
          "3013176080",
          "3005381997",
          "1319429",
          "1835296",
          "3014170143",
          "3014315560",
          "3010178296",
          "3010363503",
          "3015869493",
          "3020704423",
          "3016438694",
          "3003604053",
          "3011623994",
          "3022520753",
          "3012755988",
          "3032314868",
          "3022295665",
          "2134470",
          "2531477",
          "2183449",
          "3005536892",
          "3004187715",
          "3008559295",
          "3009532798",
          "3008034554",
          "1644408",
          "3011053036",
          "3010536692",
          "3010470577",
          "3012491394",
          "3015356625",
          "3030714745",
          "3008599979",
          "3031800723",
          "2133433",
          "3006108336",
          "3010707607",
          "1722511",
          "3009158523",
          "3006179046",
          "3015399803",
          "1526439",
          "3009153862",
          "9710629",
          "3010287737",
          "3015542154",
          "3019923333",
          "3008812560",
          "1219602",
          "3005977257",
          "3007143290",
          "1226544",
          "3002962587",
          "2032521",
          "3008729892",
          "3010068573",
          "3004369035",
          "3026593233",
          "3010329663",
          "3012275107",
          "3004365956",
          "3012130008",
          "2249697",
          "3001239363",
          "3014273501",
          "3009141957",
          "3010047402",
          "3015398319",
          "2437731",
          "3015974593",
          "9610921",
          "2953359",
          "3005497913",
          "3017565094",
          "2027148",
          "1221053",
          "3013700547",
          "3009381600",
          "3026311512",
          "2648623",
          "1935627",
          "3009110133",
          "2650143",
          "1724955",
          "3010509633",
          "1835572",
          "3010531069",
          "3010132111",
          "3007738812",
          "3014262693",
          "9611827",
          "1825034",
          "3015200759",
          "3009971621",
          "3011580264",
          "2027754",
          "3013767421",
          "3014655115",
          "3009923578",
          "8043862",
          "3006524618",
          "3009871135",
          "3009631939",
          "3005061536",
          "2242056",
          "3006001176",
          "3004431266",
          "3016261131",
          "1720929",
          "3016844398",
          "3010303097",
          "3014273424",
          "1450662",
          "1219655",
          "3010202439",
          "3012711663",
          "1061124",
          "3011296511",
          "3015231789",
          "3012329552",
          "3031261833",
          "3015453963",
          "3043543260",
          "3010667733",
          "3005290844",
          "3002808145",
          "3010846112",
          "8030965",
          "3012599928",
          "3027827832",
          "3009144915",
          "3006356043",
          "3010622588",
          "3005874553",
          "3007993775",
          "3029933740",
          "3010120326",
          "3012103114",
          "8010665",
          "3007051929",
          "3009504230",
          "1000461041",
          "3005562917",
          "3033509898",
          "3026576110",
          "3025114181",
          "3013846070",
          "2029275",
          "3015882686",
          "1320894",
          "3016669046",
          "3014302784",
          "3014644608",
          "2020601",
          "2031009",
          "3014680795",
          "1064858",
          "3015531529",
          "9680619",
          "3008114965",
          "3024813436",
          "1064017",
          "3036042032",
          "3006639944",
          "1822565",
          "1226425",
          "3004579081",
          "3015151147",
          "3003477135",
          "3002806470",
          "3010041993",
          "3022142628",
          "3008749819",
          "2090040",
          "3013086329",
          "9613910",
          "3010041430",
          "3014232239",
          "3008868758",
          "3013843230",
          "3003935342",
          "3016084569",
          "3005596372",
          "3012086398",
          "3010863450",
          "3010047454",
          "3003761012",
          "3042903229",
          "3015440604",
          "3009998573",
          "9610726",
          "9680825",
          "3010287687",
          "2022435",
          "3009513193",
          "9613369",
          "9611112",
          "3012249461",
          "1222928",
          "3010407203",
          "9613350",
          "2030624",
          "3010155648",
          "1721676",
          "3015780293",
          "3010331645",
          "3021948057",
          "3010273872",
          "9616671",
          "3014279513",
          "1000517406",
          "3012478585",
          "3014976005",
          "1649390",
          "8010257",
          "3021632375",
          "3009337401",
          "3031240334",
          "3013302242",
          "3015173212",
          "3006395932",
          "2320767",
          "3007740680",
          "3007913880",
          "3015895045",
          "3038503572",
          "9614438",
          "1420032",
          "9611281",
          "3022777867",
          "3012358417",
          "3030400539",
          "3010197224",
          "1220477",
          "3006783837",
          "3035366890",
          "3000270450",
          "3008793310",
          "3006498370",
          "2530154",
          "1018470",
          "3004513872",
          "1048735",
          "3009756153",
          "3008668801",
          "3006801265",
          "3001278222",
          "3009217531",
          "3012966183",
          "3008102042",
          "3009996260",
          "3014554088",
          "3043026734",
          "2531195",
          "8044172",
          "3019455",
          "3011127597",
          "1051526",
          "1421101",
          "3017980266",
          "3006215390",
          "3009826895",
          "3030271054",
          "3027553380",
          "1043653",
          "1531050",
          "3009959868",
          "3017210488",
          "1221051",
          "3011354099",
          "1422572",
          "3005064037",
          "3009888740",
          "3036756245",
          "3015131017",
          "3005076207",
          "3027339877",
          "3011295718",
          "3002807310",
          "3027760323",
          "3029894043",
          "9611579",
          "3015259876",
          "3009149199",
          "3017597868",
          "3014527928",
          "3003678543",
          "3024021261",
          "3004748528",
          "3009116370",
          "3010606097",
          "3006946279",
          "3016537375",
          "2936485",
          "3007922509",
          "3009973336",
          "3010041499",
          "3008396464",
          "3015831087",
          "1017294",
          "3003249380",
          "1835444",
          "3011061237",
          "2530808",
          "3030655570",
          "3010462278",
          "2134947",
          "3025603301",
          "2249615",
          "3035785605",
          "8010468",
          "3008262872",
          "9610617",
          "3019807891",
          "1064129",
          "1526534",
          "3002907620",
          "3012883081",
          "3022132808",
          "9611274",
          "3007583504",
          "3006460162",
          "1057425",
          "3015219237",
          "3010123172",
          "9616680",
          "1835251",
          "3025318945",
          "3021010222",
          "3009882462",
          "3012447612",
          "9610622",
          "3023852420",
          "3011230048",
          "3009540749",
          "1061927",
          "3017980495",
          "1649518",
          "1038671",
          "2647346",
          "3008572101",
          "3011187779",
          "1417592",
          "3006513362",
          "3007109793",
          "2028632",
          "3010502940",
          "3010375065",
          "3014315669",
          "3002719998",
          "2183967",
          "3009423112",
          "2649614",
          "3030183061",
          "3008110533",
          "3017435639",
          "3017196117",
          "3006849754",
          "3015207155",
          "3004016606",
          "8043971",
          "3007766601",
          "8031020",
          "2183744",
          "1226146",
          "3007319107",
          "3004983210",
          "3005706667",
          "3011530718",
          "3005004799",
          "3015410662",
          "3010049501",
          "3015167917",
          "3004153896",
          "2183946",
          "3011500362",
          "3016050940",
          "3007420745",
          "3016237080",
          "1643264",
          "1060840",
          "3022518322",
          "3023272531",
          "3002579136",
          "3015806723",
          "3019269298",
          "2183785",
          "3035373797",
          "3011170410",
          "3032588003",
          "1828288",
          "3004049923",
          "2032098",
          "3021393544",
          "3011302692",
          "1833986",
          "3013843055",
          "3010162973",
          "3008696586",
          "3010314800",
          "3027242374",
          "3002807295",
          "2249529",
          "3010287665",
          "3016678045",
          "3014129047",
          "3013820441",
          "3015869492",
          "3006846753",
          "3023265483",
          "3013756169",
          "3010060157",
          "3011388423",
          "3033519208",
          "3010370880",
          "1222274",
          "3015516266",
          "9610612",
          "3022159082",
          "3015724777",
          "1054811",
          "2245304",
          "1032347",
          "1424263",
          "3018077525",
          "1057946",
          "3013365802",
          "3008236846",
          "3005641619",
          "3014128887",
          "3016112537",
          "3004976965",
          "2244478",
          "3015869121",
          "3034532468",
          "3011943495",
          "3007593722",
          "3008837339",
          "2024024",
          "3015877656",
          "3012818475",
          "3038503932",
          "3009812190",
          "3014004349",
          "3034676720",
          "3007334927",
          "3006742481",
          "3042765769",
          "1222773",
          "3003755939",
          "1827096",
          "3005941719",
          "3010120135",
          "2648920",
          "2246552",
          "3013014058",
          "3006017180",
          "3008534770",
          "3003525643",
          "3011683674",
          "3012131184",
          "3010039400",
          "3015876273",
          "3021336182",
          "3004961578",
          "1424434",
          "3010041511",
          "3017191784",
          "3005983700",
          "3015946784",
          "1219930",
          "2431166",
          "3019767615",
          "3030412764",
          "3000327445",
          "3023189283",
          "3011137372",
          "9610905",
          "1834331",
          "1047843",
          "3004635447",
          "1020279",
          "1833824",
          "1833920",
          "3010400367",
          "3011989923",
          "1319639",
          "3008824097",
          "3011302058",
          "3017075357",
          "1833506",
          "3003541440",
          "3004142400",
          "3022472578",
          "3006272282",
          "9614209",
          "3010887276",
          "3010052483",
          "3010673850",
          "3004499989",
          "3012429289",
          "8043792",
          "3008262715",
          "9615128",
          "3030691054",
          "1226183",
          "3014763043",
          "3015539532",
          "3015212339",
          "3009973505",
          "3002807315",
          "3004188066",
          "3006563559",
          "3002949614",
          "3007923096",
          "3006694268",
          "3009656988",
          "8031010",
          "2020550",
          "3014342096",
          "3000170817",
          "9612277",
          "3010383847",
          "1058584",
          "3014162916",
          "3008132398",
          "2031917",
          "3010173425",
          "3004993707",
          "3013540005",
          "3025141",
          "3006217744",
          "1828464",
          "3010158639",
          "3023808054",
          "3013011598",
          "1834379",
          "3008337808",
          "1220246",
          "1818910",
          "3043620689"
        ],
        "fei_number": [
          "",
          "3016851379",
          "3017209758",
          "3008583793",
          "3006791286",
          "3016759112",
          "3004154314",
          "2523865",
          "3007289093",
          "3000203391",
          "3014967149",
          "3009941480",
          "3002807464",
          "3004464325",
          "3008647423",
          "3015718210",
          "3002808268",
          "3013055499",
          "3007125392",
          "3010097171",
          "3005827567",
          "3016727307",
          "3004358587",
          "3010032903",
          "3010560653",
          "1000120314",
          "3010120104",
          "3011015572",
          "3022928161",
          "3008870565",
          "3004638532",
          "3003056094",
          "3002807893",
          "3014257776",
          "3021008900",
          "3007113169",
          "1000121057",
          "3014314623",
          "3012470322",
          "1221934",
          "3001451571",
          "3004491689",
          "3021226419",
          "5870",
          "3020899789",
          "3010871687",
          "3007441485",
          "3011301313",
          "3042508254",
          "3015075784",
          "3014104093",
          "3001814066",
          "3010120148",
          "2011171",
          "3011656326",
          "1036836",
          "3003110073",
          "3008747271",
          "1043534",
          "3010673777",
          "3041501961",
          "3003898228",
          "3003236702",
          "3006128100",
          "1526711",
          "3026586698",
          "1000136478",
          "3009973699",
          "3008791302",
          "3009189869",
          "3002786339",
          "2183650",
          "3015183635",
          "3013561205",
          "3004051158",
          "3009132077",
          "3008650117",
          "3002808293",
          "3002806577",
          "3002807113",
          "3003128516",
          "3009201148",
          "3010120339",
          "3001406084",
          "3016247852",
          "2528981",
          "3002904345",
          "3005180920",
          "3009732568",
          "3016778139",
          "1000615957",
          "3006638824",
          "3010057495",
          "3014483751",
          "3012724179",
          "3030514683",
          "3030926959",
          "3007700286",
          "3017521423",
          "3009417901",
          "3010118938",
          "3010440433",
          "1000272060",
          "3003142829",
          "3008782585",
          "3014937101",
          "3017528621",
          "3020584246",
          "3013564228",
          "2648666",
          "3010813691",
          "3018094310",
          "3013194153",
          "3016660811",
          "3009394448",
          "3008951116",
          "3003139373",
          "3005029036",
          "3010123256",
          "3014522470",
          "3038776593",
          "3013176080",
          "3005381997",
          "3014170143",
          "3014315560",
          "3010363503",
          "3015869493",
          "3020704423",
          "3016438694",
          "3003604053",
          "1000307073",
          "1000120613",
          "3011623994",
          "3022520753",
          "3012755988",
          "3032314868",
          "3022295665",
          "3000203000",
          "2183449",
          "3005536892",
          "3004187715",
          "3002806725",
          "3002806724",
          "3000207350",
          "1000220336",
          "1000306303",
          "3002807555",
          "3008559295",
          "3009532798",
          "3008034554",
          "3011053036",
          "3010536692",
          "3010470577",
          "3001236277",
          "3012491394",
          "3015356625",
          "3001617730",
          "3030714745",
          "3008599979",
          "3031800723",
          "3006108336",
          "1000221916",
          "3010707607",
          "3001642380",
          "3009158523",
          "3003574398",
          "3006179046",
          "3015399803",
          "1000520823",
          "1526439",
          "3003292403",
          "3009153862",
          "1000405819",
          "3000206435",
          "3010287737",
          "3015542154",
          "3019923333",
          "3008812560",
          "1219602",
          "3005977257",
          "3007143290",
          "3002962587",
          "3010068573",
          "3004369035",
          "3026593233",
          "3010329663",
          "3012275107",
          "3004365956",
          "3012130008",
          "2648209",
          "3001239363",
          "3014273501",
          "1000125025",
          "3009141957",
          "3010047402",
          "3015974593",
          "3007932279",
          "3005497913",
          "3017565094",
          "2027148",
          "3002615969",
          "1221053",
          "3013700547",
          "3001452082",
          "3009381600",
          "3026311512",
          "2648623",
          "1935627",
          "3009110133",
          "3010509633",
          "3010531069",
          "3010132111",
          "3007738812",
          "3014262693",
          "1825034",
          "3015200759",
          "3009971621",
          "3011580264",
          "1000143502",
          "2027754",
          "3000718467",
          "3009256700",
          "3002808269",
          "3002807111",
          "3014655115",
          "3009923578",
          "3013767421",
          "3009871135",
          "3009631939",
          "3005061536",
          "2242056",
          "3006001176",
          "3004431266",
          "3016261131",
          "1720929",
          "3016844398",
          "3010303097",
          "3002806685",
          "3014273424",
          "1450662",
          "1219655",
          "3003018882",
          "3012711663",
          "3022270316",
          "3011296511",
          "3000215273",
          "3015231789",
          "3012329552",
          "3031261833",
          "3015453963",
          "3043543260",
          "3010667733",
          "3005290844",
          "3002808145",
          "3010846112",
          "3012599928",
          "3027827832",
          "3009144915",
          "3006356043",
          "3010622588",
          "3005874553",
          "3007993775",
          "3029933740",
          "3010120326",
          "3012103114",
          "1000525271",
          "3007051929",
          "3009504230",
          "1000461041",
          "3005562917",
          "3033509898",
          "3026576110",
          "3025114181",
          "3013846070",
          "3015882686",
          "3016669046",
          "3002806534",
          "3014302784",
          "3014644608",
          "2020601",
          "3014680795",
          "3015531529",
          "3024813436",
          "3008114965",
          "3036042032",
          "3002854251",
          "3006639944",
          "3000190635",
          "3004579081",
          "3015151147",
          "3003477135",
          "3002806470",
          "3010041993",
          "3001236812",
          "3022142628",
          "3008749819",
          "3013086329",
          "3010041430",
          "3014232239",
          "3008868758",
          "3013843230",
          "3003935342",
          "3016084569",
          "3005596372",
          "3012086398",
          "3002757549",
          "3010863450",
          "3010047454",
          "3003761012",
          "3042903229",
          "3015440604",
          "3009998573",
          "3010287687",
          "2022435",
          "3000717687",
          "3009513193",
          "3012249461",
          "3002807231",
          "3010407203",
          "3002807395",
          "3002929183",
          "3010155648",
          "1721676",
          "3015780293",
          "3010331645",
          "3021948057",
          "3010273872",
          "1000185904",
          "3014279513",
          "1000517406",
          "3012478585",
          "3014976005",
          "3021632375",
          "3009337401",
          "3031240334",
          "3013302242",
          "3015173212",
          "3006395932",
          "3007740680",
          "3007913880",
          "3015895045",
          "3038503572",
          "1420032",
          "3022777867",
          "3012358417",
          "3030400539",
          "3010197224",
          "3002755653",
          "1220477",
          "3035366890",
          "3000270450",
          "3008793310",
          "3019827409",
          "3006498370",
          "3002807305",
          "3008397795",
          "1000306431",
          "1018470",
          "1000115735",
          "3005233524",
          "3004513872",
          "3001236549",
          "1000116912",
          "1048735",
          "3002462840",
          "3009756153",
          "3008668801",
          "3006801265",
          "3001278222",
          "3009217531",
          "3002807041",
          "3002807172",
          "3012966183",
          "3009996260",
          "3014554088",
          "1000115331",
          "3043026734",
          "3019455",
          "3011127597",
          "1051526",
          "1421101",
          "3017980266",
          "3006215390",
          "3009826895",
          "3030271054",
          "3027553380",
          "1043653",
          "3009959868",
          "3017210488",
          "1221051",
          "3011354099",
          "3005064037",
          "3009888740",
          "3036756245",
          "3015131017",
          "3005076207",
          "3027339877",
          "3011295718",
          "3002807310",
          "3027760323",
          "3029894043",
          "3015259876",
          "3009149199",
          "3017597868",
          "3014527928",
          "3003678543",
          "3024021261",
          "3004748528",
          "3009116370",
          "3010606097",
          "3006946279",
          "3000215205",
          "3016537375",
          "2936485",
          "3007922509",
          "3009973336",
          "3015831087",
          "3008396464",
          "1017294",
          "3003249380",
          "3002684506",
          "3011061237",
          "3030655570",
          "3010462278",
          "3025603301",
          "3000205819",
          "3035785605",
          "3002806545",
          "3008262872",
          "3019807891",
          "1526534",
          "4498",
          "3002907620",
          "3012883081",
          "3022132808",
          "3007583504",
          "3006460162",
          "3002795782",
          "3002808043",
          "3001644153",
          "3010123172",
          "3015219237",
          "3025318945",
          "3021010222",
          "3009882462",
          "3012447612",
          "3023852420",
          "3011230048",
          "3009540749",
          "1000120768",
          "3017980495",
          "1038671",
          "2647346",
          "3008572101",
          "3009322142",
          "3011187779",
          "1000116116",
          "1417592",
          "3006513362",
          "3002807827",
          "3002807036",
          "3003356728",
          "3007109793",
          "3010502940",
          "3010375065",
          "3014315669",
          "1000512168",
          "3000205591",
          "3002719998",
          "3009423112",
          "1000522098",
          "3008820440",
          "3003015552",
          "2649614",
          "3003058448",
          "3030183061",
          "3002806315",
          "3001049282",
          "3008110533",
          "1000136269",
          "3017435639",
          "3017196117",
          "3006849754",
          "3015207155",
          "3004016606",
          "3007766601",
          "2183744",
          "3002680121",
          "3007319107",
          "1000220733",
          "3004983210",
          "3002807328",
          "3005706667",
          "3011530718",
          "3005004799",
          "3015410662",
          "3010049501",
          "3015167917",
          "3004153896",
          "3011500362",
          "3016050940",
          "3007420745",
          "3003407235",
          "1643264",
          "3016237080",
          "3003414678",
          "3022518322",
          "3023272531",
          "3002579136",
          "3015806723",
          "3019269298",
          "3002807421",
          "2183785",
          "3035373797",
          "1000139987",
          "3011170410",
          "3032588003",
          "1828288",
          "3004049923",
          "3021393544",
          "3007282753",
          "3011302692",
          "3013843055",
          "3008696586",
          "3010314800",
          "3027242374",
          "3002807295",
          "3010287665",
          "3016678045",
          "3002924544",
          "3013820441",
          "3015869492",
          "3006846753",
          "3023265483",
          "3013756169",
          "3010060157",
          "3007475226",
          "3011388423",
          "3033519208",
          "3010370880",
          "3015516266",
          "1000115656",
          "3022159082",
          "3002287318",
          "3015724777",
          "2245304",
          "1032347",
          "3002806653",
          "3018077525",
          "1000138054",
          "3013365802",
          "3008236846",
          "3005641619",
          "3014128887",
          "3016112537",
          "3004976965",
          "3002459190",
          "3015869121",
          "3000210784",
          "3034532468",
          "3011943495",
          "3007593722",
          "3003070421",
          "3008837339",
          "2024024",
          "3015877656",
          "3012818475",
          "3038503932",
          "3014004349",
          "3034676720",
          "3007334927",
          "1000114185",
          "3006742481",
          "3004574120",
          "3042765769",
          "3003755939",
          "1827096",
          "1000219637",
          "3010120135",
          "3002594206",
          "2246552",
          "3013014058",
          "1000306139",
          "3006017180",
          "3008534770",
          "3003525643",
          "3011683674",
          "3012131184",
          "3010039400",
          "3015876273",
          "3021336182",
          "3004961578",
          "3010041511",
          "1000306340",
          "3017191784",
          "3005983700",
          "3015946784",
          "1219930",
          "3019767615",
          "3030412764",
          "3000327445",
          "3023189283",
          "1047843",
          "3004635447",
          "1020279",
          "3010400367",
          "1000069183",
          "3011989923",
          "3008824097",
          "3011302058",
          "3017075357",
          "3003541440",
          "3004142400",
          "3022472578",
          "3006272282",
          "3010887276",
          "3010052483",
          "3010673850",
          "3004499989",
          "3010938185",
          "3002821504",
          "3012429289",
          "3008262715",
          "3030691054",
          "3014763043",
          "3015539532",
          "3015212339",
          "3009973505",
          "3002807315",
          "1000199976",
          "3004188066",
          "3002807091",
          "3006563559",
          "3002949614",
          "3007923096",
          "3006694268",
          "3001026312",
          "1832656",
          "3009656988",
          "2020550",
          "1000132037",
          "3014342096",
          "3000170817",
          "3010383847",
          "3014162916",
          "3000717902",
          "1000125930",
          "3005747797",
          "3013540005",
          "3004993707",
          "3006217744",
          "1828464",
          "3010158639",
          "3002806773",
          "3023808054",
          "3013011598",
          "3008337808",
          "1220246",
          "1818910",
          "3043620689",
          "3002807830"
        ],
        "device_name": "Screw, Fixation, Bone",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3040",
        "device_class": "2"
      }
    }
  ]
}