  "results": [
    {
      "cfres_id": "176461",
      "product_res_number": "Z-0206-2020",
      "event_date_initiated": "2019-09-10",
      "event_date_posted": "2019-10-24",
      "recall_status": "Terminated",
      "event_date_terminated": "2020-10-20",
      "res_event_number": "83839",
      "product_code": "GEX",
      "product_description": "Canister HFC-134a /1000g, ALUM\nProduct Number: 1600-00-0218\nRevision .0C  \nCoolant canister for use with surgical laser.",
      "code_info": "Lot # 1904CC1598 1905CC0007 1905CC0178  1905CC0217  1906CC0196  1906CC0172  1907CC0008  1908CC0025  Additional lots added as of 12/6/2019: 1902CC0043 1902CC0143 1811CC0016 1812CC0074 1812CC0133 1901CC0087 1904CC0126 1802CC0092 1802CC0232 1803CC0169 1804CC0073 1805CC0058 1806CC0068 1807CC0009 1807CC0104 1808CC0112 1809CC0070 1810CC0015 1810CC0117 1907CC0187",
      "firm_fei_number": "1218402",
      "recalling_firm": "Candela Corporation",
      "address_1": "530 Boston Post Rd",
      "city": "Wayland",
      "state": "MA",
      "postal_code": "01778-1833",
      "additional_info_contact": "Sharon Timberlake\n508-358-7400",
      "reason_for_recall": "The canister neck bushing may become loose and separate from the body of the canister during use. The issue occurs only when the canisters are used in Candela legacy devices that utilize  screw-in  or  threaded interface  type DCD modules",
      "root_cause_description": "Other",
      "action": "1. Immediately examine your inventory and quarantine the identified lots of the GentleCool Pro Cryogen Canister.\n\n2. Immediately discontinue use and distribution of the identified lots of the GentleCool Pro Cryogen Canister. \n\n3. Please complete and return the enclosed Acknowledgement Form - GentleCool Pro Canister Recall. The completed form must be scanned and emailed to cryogen.usa@candelamedical.com within 24 hours.\n\n4. If you may have further distributed this product, please identify those locations and notify them at once of this product recall. Your communication should include a copy of this recall notification letter. \n\n5. Please contact Candela Customer Service at 800-733-8550 ext. 0 to make arrangements for replacement of your affected inventory.\n\n6. Information regarding retrieval of your affected inventory will be forthcoming from Candela Medical.\n\nIf you have any questions, please do not hesitate to call our Customer Service team at 800-733-8550 ext. 0, M-F 8:30 am  8 pm EST, or reach them via email at cryogen.usa@candelamedical.com.\n\nAn Urgent Medical Device Expanded Recall notification letter dated 12/6/19 was sent to customers.\n\nTwenty additional lots have now been identified (listed in bold on the following page), since the initial notification was sent mid-September. Thus, we are updating this notice to make you aware of these additional lots. The reason for this expanded recall is unchanged from the initial communications sent.\n\nActions to Be Taken by the Customer/User:\nTo safely and efficiently implement this recall, we are asking that you please take the following actions:\n1. Immediately examine your inventory and quarantine the identified lots of the GentleCool Pro Cryogen Canister.\n2. Immediately discontinue use and distribution of the identified lots of the GentleCool Pro Cryogen Canister.\n3. Please complete and return the enclosed Acknowledgement Form - GentleCool Pro Canister Recall. The completed form must be sca",
      "product_quantity": "in total 7,560 canisters (3240 additional as of 2/5/20)",
      "distribution_pattern": "States: NC\nGA\nCA\nMA\nOH\nFL\nIL\nCO\nSD\nNY\nKS\nWA\nMI\nPA\nAZ\nWI\nTX\nLA\nIA\nMN\nVA\nTN\nSC\n\nOR\nIN\nWY\nMD\nVT\nON\nBC\nQC\nNV\n\n\nOUS: Barbados\n\nCanada\nGermany\nSpain\nFrance\nHR\nPortugal\nIreland\nNorway\nGreece\nNL\nSouth Africa\nVietnam\nHong Kong\nJapan\nNew Zealand\nAustralia\nChina\nKorea",
      "openfda": {
        "k_number": [
          "K091922",
          "K972099",
          "K023547",
          "K855210",
          "K141975",
          "K191942",
          "K915656",
          "K853878",
          "K905072",
          "K890287",
          "K241293",
          "K032221",
          "K013021",
          "K010804",
          "K162755",
          "K010885",
          "K990174",
          "K033962",
          "K083748",
          "K944470",
          "K073048",
          "K013901",
          "K960772",
          "K853627",
          "K032667",
          "K173677",
          "K004039",
          "K252369",
          "K192539",
          "K010834",
          "K133915",
          "K962592",
          "K902058",
          "K952046",
          "K181510",
          "K000693",
          "K981351",
          "K203804",
          "K132976",
          "K181867",
          "K113472",
          "K914519",
          "K073427",
          "K002421",
          "K910341",
          "K990452",
          "K922881",
          "K233090",
          "K033549",
          "K071469",
          "K090618",
          "K122358",
          "K892912",
          "K100098",
          "K031258",
          "K151655",
          "K894977",
          "K940607",
          "K892933",
          "K854469",
          "K912987",
          "K983211",
          "K051399",
          "K131278",
          "K943479",
          "K131857",
          "K253765",
          "K250998",
          "K923050",
          "K901127",
          "K023543",
          "K942934",
          "K063682",
          "K180406",
          "K972872",
          "K240874",
          "K072751",
          "K073022",
          "K922901",
          "K992597",
          "K123168",
          "K030186",
          "K042439",
          "K122922",
          "K011423",
          "K022466",
          "K073313",
          "K050679",
          "K930361",
          "K231514",
          "K210535",
          "K142952",
          "K081578",
          "K183307",
          "K082938",
          "K954934",
          "K170597",
          "K023182",
          "K894210",
          "K895430",
          "K103455",
          "K931051",
          "K131805",
          "K014236",
          "K162363",
          "K904331",
          "K233304",
          "K961748",
          "K081015",
          "K081939",
          "K904342",
          "K883488",
          "K233024",
          "K954931",
          "K960579",
          "K905233",
          "K030147",
          "K254271",
          "K140719",
          "K971821",
          "K913887",
          "K991935",
          "K003460",
          "K053324",
          "K124030",
          "K011041",
          "K131081",
          "K243588",
          "K854387",
          "K963991",
          "K964520",
          "K902871",
          "K934209",
          "K924213",
          "K060787",
          "K945181",
          "K213554",
          "K010928",
          "K133945",
          "K092734",
          "K231876",
          "K040744",
          "K030523",
          "K173359",
          "K990568",
          "K051287",
          "K063141",
          "K033183",
          "K252063",
          "K222862",
          "K912628",
          "K944091",
          "K241406",
          "K070805",
          "K050047",
          "K880908",
          "K890059",
          "K191970",
          "K080121",
          "K884459",
          "K000222",
          "K151763",
          "K213225",
          "K932272",
          "K896711",
          "K935495",
          "K022381",
          "K060448",
          "K090164",
          "K032991",
          "K914175",
          "K071687",
          "K961006",
          "K942786",
          "K853877",
          "K960636",
          "K954727",
          "K071295",
          "K063834",
          "K953658",
          "K213267",
          "K053616",
          "K902470",
          "K060459",
          "K041739",
          "K200786",
          "K870838",
          "K910826",
          "K871206",
          "K981126",
          "K944282",
          "K162659",
          "K232769",
          "K972765",
          "K232709",
          "K955007",
          "K903230",
          "K901205",
          "K182997",
          "K183122",
          "K252599",
          "K923191",
          "K242207",
          "K130344",
          "K901340",
          "K970667",
          "K852190",
          "K152722",
          "K982860",
          "K080463",
          "K003151",
          "K960032",
          "K201594",
          "K882772",
          "K944628",
          "K070138",
          "K043429",
          "K213461",
          "K984601",
          "K952813",
          "K013748",
          "K171009",
          "K955758",
          "K973324",
          "K072925",
          "K896049",
          "K242202",
          "K233803",
          "K931774",
          "K950662",
          "K990902",
          "K241670",
          "K023464",
          "K043319",
          "K172198",
          "K180426",
          "K071896",
          "K261701",
          "K151232",
          "K973498",
          "K041086",
          "K991659",
          "K900463",
          "K940461",
          "K003886",
          "K042968",
          "K920821",
          "K083203",
          "K140458",
          "K910272",
          "K030426",
          "K992230",
          "K033251",
          "K031993",
          "K954013",
          "K974633",
          "K896762",
          "K961682",
          "K171648",
          "K023257",
          "K251339",
          "K063249",
          "K000192",
          "K904423",
          "K141063",
          "K142372",
          "K192735",
          "K912918",
          "K123494",
          "K980840",
          "K992574",
          "K240983",
          "K854472",
          "K792418",
          "K110243",
          "K192569",
          "K011021",
          "K161692",
          "K981350",
          "K123777",
          "K913109",
          "K062303",
          "K071051",
          "K980420",
          "K071363",
          "K073317",
          "K053511",
          "K070466",
          "K172376",
          "K060567",
          "K011207",
          "K052442",
          "K131955",
          "K232133",
          "K023081",
          "K243121",
          "K031680",
          "K933841",
          "K904354",
          "K971903",
          "K090126",
          "K033437",
          "K900345",
          "K082876",
          "K932376",
          "K083207",
          "K260489",
          "K896939",
          "K073436",
          "K231910",
          "K001676",
          "K202258",
          "K220853",
          "K963249",
          "K962446",
          "K921549",
          "K030846",
          "K072331",
          "K202043",
          "K151569",
          "K910422",
          "K250722",
          "K013940",
          "K852252",
          "K002649",
          "K901008",
          "K933017",
          "K091562",
          "K893701",
          "K974039",
          "K864562",
          "K901021",
          "K915578",
          "K233301",
          "K930793",
          "K030897",
          "K884204",
          "K131894",
          "K974900",
          "K092660",
          "K961560",
          "K942204",
          "K131059",
          "K132607",
          "K210925",
          "K221427",
          "K182088",
          "K163103",
          "K860273",
          "K890279",
          "K250504",
          "K933880",
          "K984591",
          "K050824",
          "K110686",
          "K120377",
          "K832550",
          "K072906",
          "K980156",
          "K231548",
          "K982827",
          "K181614",
          "K003046",
          "K121570",
          "K944625",
          "K022378",
          "K071883",
          "K914521",
          "K041095",
          "K063104",
          "K062589",
          "K874948",
          "K062248",
          "K073063",
          "K013623",
          "K003541",
          "K210033",
          "K230580",
          "K982940",
          "K022923",
          "K161162",
          "K060301",
          "K022351",
          "K953794",
          "K103667",
          "K934211",
          "K960867",
          "K980501",
          "K113792",
          "K212270",
          "K980001",
          "K882738",
          "K884182",
          "K131095",
          "K121938",
          "K213669",
          "K890286",
          "K043521",
          "K973576",
          "K004014",
          "K082263",
          "K053457",
          "K910894",
          "K882243",
          "K894591",
          "K162191",
          "K001746",
          "K012682",
          "K051167",
          "K000811",
          "K112600",
          "K113151",
          "K203788",
          "K950831",
          "K102050",
          "K060457",
          "K150140",
          "K941909",
          "K170626",
          "K964211",
          "K896838",
          "K862815",
          "K043164",
          "K992914",
          "K944090",
          "K040005",
          "K934811",
          "K914423",
          "K133895",
          "K882365",
          "K073322",
          "K953174",
          "K041419",
          "K852436",
          "K043255",
          "K072484",
          "K211534",
          "K230990",
          "K963340",
          "K210964",
          "K172818",
          "K251288",
          "K251329",
          "K250206",
          "K884682",
          "K192483",
          "K161632",
          "K042630",
          "K944317",
          "K880061",
          "K963081",
          "K062210",
          "K992765",
          "K032459",
          "K243780",
          "K900128",
          "K203441",
          "K082409",
          "K242035",
          "K043173",
          "K050672",
          "K853620",
          "K042474",
          "K113115",
          "K955653",
          "K961516",
          "K895786",
          "K080163",
          "K231948",
          "K943411",
          "K061436",
          "K924457",
          "K873246",
          "K930543",
          "K191962",
          "K151855",
          "K191611",
          "K063867",
          "K850985",
          "K112141",
          "K945648",
          "K120561",
          "K101555",
          "K915768",
          "K014234",
          "K992773",
          "K113018",
          "K170179",
          "K983524",
          "K030235",
          "K860831",
          "K072779",
          "K083915",
          "K952216",
          "K053628",
          "K081619",
          "K974674",
          "K920708",
          "K212127",
          "K896595",
          "K222673",
          "K991234",
          "K222352",
          "K193656",
          "K011175",
          "K864887",
          "K954316",
          "K030959",
          "K030181",
          "K041879",
          "K051671",
          "K133837",
          "K924258",
          "K170724",
          "K911288",
          "K014040",
          "K171255",
          "K953494",
          "K261674",
          "K993911",
          "K873773",
          "K161670",
          "K170718",
          "K014035",
          "K212492",
          "K897157",
          "K903838",
          "K950313",
          "K921841",
          "K222629",
          "K964353",
          "K921807",
          "K250936",
          "K200736",
          "K181486",
          "K944624",
          "K051255",
          "K063073",
          "K060050",
          "K955131",
          "K013493",
          "K040433",
          "K082230",
          "K920967",
          "K002072",
          "K913244",
          "K944758",
          "K252732",
          "K100194",
          "K882734",
          "K050696",
          "K913917",
          "K923341",
          "K081219",
          "K905676",
          "K963229",
          "K801624",
          "K062321",
          "K203054",
          "K850818",
          "K984486",
          "K192272",
          "K223404",
          "K965128",
          "K034011",
          "K901638",
          "K944558",
          "K895945",
          "K953065",
          "K060216",
          "K052991",
          "K024206",
          "K944704",
          "K912793",
          "K052324",
          "K202601",
          "K080591",
          "K031440",
          "K904632",
          "K081628",
          "K092860",
          "K051876",
          "K904444",
          "K952682",
          "K173169",
          "K162416",
          "K193367",
          "K243630",
          "K151947",
          "K951045",
          "K122567",
          "K070353",
          "K012790",
          "K060653",
          "K061850",
          "K180719",
          "K100415",
          "K083899",
          "K253828",
          "K130681",
          "K020835",
          "K932062",
          "K992757",
          "K884431",
          "K183392",
          "K160513",
          "K014057",
          "K240474",
          "K992707",
          "K252529",
          "K153671",
          "K240644",
          "K192600",
          "K083069",
          "K983020",
          "K905082",
          "K902063",
          "K921671",
          "K190938",
          "K083215",
          "K152667",
          "K971879",
          "K071152",
          "K894066",
          "K910152",
          "K894471",
          "K183566",
          "K223727",
          "K932683",
          "K033237",
          "K933016",
          "K910898",
          "K053047",
          "K896988",
          "K211735",
          "K895518",
          "K870837",
          "K160413",
          "K013300",
          "K902133",
          "K980517",
          "K172819",
          "K020973",
          "K904407",
          "K904420",
          "K900539",
          "K033331",
          "K170936",
          "K873790",
          "K172077",
          "K071943",
          "K200104",
          "K081214",
          "K231894",
          "K883738",
          "K903639",
          "K992814",
          "K962965",
          "K030805",
          "K100746",
          "K072034",
          "K160368",
          "K871933",
          "K250776",
          "K082825",
          "K112031",
          "K891016",
          "K242951",
          "K202503",
          "K082586",
          "K871619",
          "K902940",
          "K896847",
          "K253245",
          "K053608",
          "K030883",
          "K903288",
          "K900630",
          "K901204",
          "K252427",
          "K113709",
          "K251824",
          "K191321",
          "K153566",
          "K971743",
          "K913566",
          "K013047",
          "K042319",
          "K954544",
          "K040617",
          "K982316",
          "K222916",
          "K914520",
          "K964585",
          "K212703",
          "K881619",
          "K943491",
          "K001990",
          "K923605",
          "K101415",
          "K213569",
          "K003202",
          "K962902",
          "K123483",
          "K930159",
          "K925124",
          "K904203",
          "K111851",
          "K973172",
          "K853874",
          "K040659",
          "K100078",
          "K120460",
          "K192331",
          "K062619",
          "K083730",
          "K253829",
          "K023954",
          "K923896",
          "K890922",
          "K901481",
          "K220793",
          "K032805",
          "K072995",
          "K895042",
          "K241358",
          "K051259",
          "K952489",
          "K935305",
          "K012127",
          "K953424",
          "K260100",
          "K090776",
          "K060966",
          "K883642",
          "K942622",
          "K250018",
          "K920673",
          "K170073",
          "K180140",
          "K994010",
          "K140837",
          "K060905",
          "K952974",
          "K220041",
          "K896619",
          "K070476",
          "K211979",
          "K213726",
          "K900346",
          "K900570",
          "K080731",
          "K041969",
          "K030810",
          "K081704",
          "K253920",
          "K980201",
          "K921146",
          "K051268",
          "K100558",
          "K152939",
          "K230460",
          "K063396",
          "K131142",
          "K210183",
          "K033176",
          "K112505",
          "K943445",
          "K911150",
          "K963969",
          "K120926",
          "K020513",
          "K241791",
          "K010104",
          "K172363",
          "K211977",
          "K021616",
          "K131473",
          "K021696",
          "K093087",
          "K873443",
          "K010041",
          "K100893",
          "K260257",
          "K083272",
          "K113504",
          "K120810",
          "K041829",
          "K920553",
          "K052310",
          "K202428",
          "K251076",
          "K971814",
          "K062721",
          "K131987",
          "K902562",
          "K003781",
          "K112669",
          "K212611",
          "K921300",
          "K921380",
          "K052680",
          "K150997",
          "K250165",
          "K922807",
          "K100199",
          "K120622",
          "K032191",
          "K083918",
          "K070388",
          "K903027",
          "K241407",
          "K920596",
          "K011174",
          "K872578",
          "K993209",
          "K885060",
          "K930210",
          "K083142",
          "K071959",
          "K083476",
          "K024260",
          "K051113",
          "K953585",
          "K991258",
          "K192617",
          "K992061",
          "K972548",
          "K973431",
          "K990780",
          "K894074",
          "K912438",
          "K132806",
          "K190936",
          "K082096",
          "K252519",
          "K895581",
          "K060752",
          "K140120",
          "K894396",
          "K972347",
          "K020412",
          "K971193",
          "K952733",
          "K050218",
          "K040294",
          "K003953",
          "K914918",
          "K023193",
          "K070216",
          "K041893",
          "K982230",
          "K080244",
          "K921596",
          "K170970",
          "K132286",
          "K232106",
          "K982119",
          "K980710",
          "K222983",
          "K193266",
          "K203241",
          "K254197",
          "K260307",
          "K934486",
          "K984517",
          "K852122",
          "K950505",
          "K981090",
          "K034030",
          "K031671",
          "K231309",
          "K964684",
          "K240497",
          "K182173",
          "K031224",
          "K202257",
          "K113489",
          "K180224",
          "K062822",
          "K142376",
          "K861179",
          "K013043",
          "K050108",
          "K242349",
          "K203491",
          "K944151",
          "K891130",
          "K854833",
          "K944346",
          "K900592",
          "K993942",
          "K893642",
          "K974878",
          "K854398",
          "K925642",
          "K180423",
          "K953074",
          "K152636",
          "K991285",
          "K261738",
          "K173038",
          "K944928",
          "K842703",
          "K110941",
          "K082039",
          "K082028",
          "K970361",
          "K820327",
          "K925549",
          "K083379",
          "K874811",
          "K915169",
          "K954597",
          "K945178",
          "K013941",
          "K993828",
          "K051434",
          "K061652",
          "K854399",
          "K100678",
          "K192990",
          "K212607",
          "K990851",
          "K133297",
          "K142924",
          "K030654",
          "K240752",
          "K240604",
          "K863086",
          "K910530",
          "K032735",
          "K001704",
          "K250261",
          "K152736",
          "K080248",
          "K160952",
          "K920865",
          "K911925",
          "K933322",
          "K950215",
          "K914568",
          "K881581",
          "K242534",
          "K063715",
          "K252319",
          "K080525",
          "K110079",
          "K251100",
          "K031960",
          "K192528",
          "K121127",
          "K080567",
          "K060033",
          "K210634",
          "K173199",
          "K252971",
          "K041710",
          "K082928",
          "K970243",
          "K241092",
          "K944206",
          "K213761",
          "K140679",
          "K962068",
          "K110907",
          "K073477",
          "K071768",
          "K961710",
          "K032023",
          "K030506",
          "K230268",
          "K920334",
          "K032205",
          "K231061",
          "K882942",
          "K991993",
          "K971539",
          "K980559",
          "K003705",
          "K910726",
          "K902534",
          "K190319",
          "K903689",
          "K932156",
          "K024061",
          "K240611",
          "K901365",
          "K852121",
          "K082721",
          "K143105",
          "K032514",
          "K031819",
          "K240179",
          "K241642",
          "K990417",
          "K050796",
          "K233007",
          "K163196",
          "K853319",
          "K992083",
          "K942343",
          "K920584",
          "K993153",
          "K242183",
          "K160312",
          "K091909",
          "K143666",
          "K082407",
          "K260974",
          "K011382",
          "K072564",
          "K221766",
          "K990119",
          "K894393",
          "K002228",
          "K931286",
          "K150326",
          "K041660",
          "K101152",
          "K152737",
          "K022583",
          "K193609",
          "K910137",
          "K993834",
          "K974057",
          "K212502",
          "K232236",
          "K972028",
          "K083889",
          "K880808",
          "K900837",
          "K926224",
          "K854134",
          "K022838",
          "K955006",
          "K992536",
          "K922213",
          "K852470",
          "K921843",
          "K021222",
          "K854998",
          "K982629",
          "K050216",
          "K063112",
          "K252877",
          "K240816",
          "K990725",
          "K132975",
          "K965124",
          "K932723",
          "K220268",
          "K232858",
          "K980142",
          "K243044",
          "K952162",
          "K130420",
          "K021227",
          "K231901",
          "K925840",
          "K873963",
          "K892516",
          "K033933",
          "K911926",
          "K132661",
          "K061720",
          "K970461",
          "K915352",
          "K032565",
          "K953800",
          "K030539",
          "K073659",
          "K983673",
          "K182930",
          "K172439",
          "K260256",
          "K022181",
          "K253007",
          "K964031",
          "K143154",
          "K925414",
          "K041994",
          "K946304",
          "K252561",
          "K965180",
          "K920659",
          "K112253",
          "K171921",
          "K160012",
          "K941838",
          "K140005",
          "K860974",
          "K072332",
          "K202175",
          "K895837",
          "K031451",
          "K962948",
          "K201455",
          "K951760",
          "K954460",
          "K253951",
          "K963339",
          "K050779",
          "K881584",
          "K020697",
          "K850018",
          "K900253",
          "K992374",
          "K192856",
          "K242859",
          "K904930",
          "K020099",
          "K170331",
          "K251432",
          "K182663",
          "K913430",
          "K841781",
          "K022655",
          "K112299",
          "K914104",
          "K060304",
          "K020958",
          "K030700",
          "K033461",
          "K072344",
          "K063074",
          "K061455",
          "K192583",
          "K222701",
          "K203799",
          "K072833",
          "K214049",
          "K003952",
          "K903154",
          "K100270",
          "K231896",
          "K041262",
          "K955014",
          "K202288",
          "K040563",
          "K082445",
          "K160227",
          "K990947",
          "K040131",
          "K974789",
          "K103288",
          "K232568",
          "K870517",
          "K944631",
          "K062314",
          "K152714",
          "K201801",
          "K844178",
          "K200751",
          "K040156",
          "K071734",
          "K881788",
          "K233143",
          "K855246",
          "K171079",
          "K852696",
          "K253251",
          "K991464",
          "K950120",
          "K924588",
          "K895299",
          "K001031",
          "K884456",
          "K955862",
          "K993154",
          "K000317",
          "K200234",
          "K014130",
          "K173676",
          "K140503",
          "K083682",
          "K022951",
          "K201663",
          "K952113",
          "K181272",
          "K113390",
          "K942182",
          "K931306",
          "K073496",
          "K101817",
          "K111360",
          "K053139",
          "K220381",
          "K111303",
          "K140009",
          "K013461",
          "K261214",
          "K896618",
          "K881582",
          "K863183",
          "K143457",
          "K860083",
          "K230340",
          "K913256",
          "K112442",
          "K983034",
          "K834166",
          "K073074",
          "K033423",
          "K880816",
          "K142223",
          "K904068",
          "K160480",
          "K934501",
          "K910062",
          "K252048",
          "K172096",
          "K910051",
          "K211761",
          "K193478",
          "K993328",
          "K212978",
          "K860087",
          "K952049",
          "K964253",
          "K233804",
          "K161981",
          "K261570",
          "K922496",
          "K931212",
          "K243037",
          "K060372",
          "K080735",
          "K943171",
          "K050021",
          "K973828",
          "K161503",
          "K881586",
          "K012982",
          "K971737",
          "K890865",
          "K952299",
          "K925726",
          "K024064",
          "K162914",
          "K891461",
          "K013366",
          "K181044",
          "K113140",
          "K211821",
          "K897040",
          "K101916",
          "K193486",
          "K935775",
          "K052371",
          "K993002",
          "K032846",
          "K062871",
          "K242251",
          "K984110",
          "K935258",
          "K024093",
          "K062423",
          "K950819",
          "K181431",
          "K915000",
          "K903943",
          "K000990",
          "K143713",
          "K013365",
          "K160607",
          "K021650",
          "K952636",
          "K914679",
          "K161925",
          "K231791",
          "K011769",
          "K050453",
          "K181480",
          "K222485",
          "K883089",
          "K973354",
          "K010316",
          "K090410",
          "K983054",
          "K853626",
          "K926210",
          "K922935",
          "K853900",
          "K202828",
          "K260138",
          "K193375",
          "K933362",
          "K991297",
          "K910823",
          "K081352",
          "K100384",
          "K170500",
          "K961380",
          "K013193",
          "K880807",
          "K993815",
          "K052848",
          "K955662",
          "K201109",
          "K965015",
          "K901227",
          "K081509",
          "K070296",
          "K243993",
          "K111592",
          "K051508",
          "K033586",
          "K070063",
          "K202172",
          "K951562",
          "K900890",
          "K070185",
          "K952277",
          "K022327",
          "K011092",
          "K961079",
          "K183452",
          "K201406",
          "K943543",
          "K213252",
          "K241459",
          "K920597",
          "K853117",
          "K202702",
          "K142860",
          "K073715",
          "K926452",
          "K933318",
          "K862422",
          "K890321",
          "K073583",
          "K242941",
          "K022760",
          "K960413",
          "K120437",
          "K942935",
          "K242206",
          "K141101",
          "K234004",
          "K022807",
          "K060680",
          "K091320",
          "K161669",
          "K021946",
          "K253100",
          "K092735",
          "K140378",
          "K983215",
          "K061898",
          "K964818",
          "K141973",
          "K953917",
          "K872711",
          "K110304",
          "K853904",
          "K883379",
          "K130671",
          "K950725",
          "K882037",
          "K905424",
          "K080104",
          "K151473",
          "K030033",
          "K193477",
          "K051996",
          "K081652",
          "K251981",
          "K913127",
          "K970948",
          "K031488",
          "K233757",
          "K212478",
          "K924720",
          "K031954",
          "K072106",
          "K181602",
          "K890775",
          "K180518",
          "K152417",
          "K123443",
          "K981952",
          "K891088",
          "K921038",
          "K032218",
          "K140732",
          "K861688",
          "K083123",
          "K974381",
          "K894962",
          "K241771",
          "K931499",
          "K992508",
          "K940636",
          "K913513",
          "K930582",
          "K192160",
          "K153527",
          "K902947",
          "K091796",
          "K053604",
          "K973470",
          "K142614",
          "K011747",
          "K910027",
          "K900153",
          "K101783",
          "K112987",
          "K851527",
          "K960086",
          "K040055",
          "K140386",
          "K170215",
          "K041943",
          "K020847",
          "K900751",
          "K894164",
          "K890322",
          "K913564",
          "K140366",
          "K934089",
          "K170738",
          "K970757",
          "K152856",
          "K241930",
          "K991628",
          "K050165",
          "K210142",
          "K882926",
          "K893206",
          "K251230",
          "K946002",
          "K060114",
          "K232532",
          "K904067",
          "K912284",
          "K113502",
          "K032383",
          "K082004",
          "K943716",
          "K070298",
          "K221571",
          "K043032",
          "K041323",
          "K933918",
          "K895656",
          "K060310",
          "K120932",
          "K012071",
          "K260690",
          "K091746",
          "K931490",
          "K902939",
          "K170409",
          "K102036",
          "K001589",
          "K101006",
          "K123573",
          "K902061",
          "K060326",
          "K212793",
          "K061178",
          "K874648",
          "K073411",
          "K914678",
          "K875338",
          "K233224",
          "K983977",
          "K962242",
          "K133891",
          "K050738",
          "K221597",
          "K030423",
          "K192269",
          "K902148",
          "K944437",
          "K141851",
          "K070284",
          "K053087",
          "K931776",
          "K091800",
          "K971207",
          "K250431",
          "K101601",
          "K180350",
          "K032204",
          "K132989",
          "K900077",
          "K241860",
          "K933715",
          "K212734",
          "K910037",
          "K892211",
          "K872179",
          "K944965",
          "K240520",
          "K895347",
          "K920120",
          "K210663",
          "K873760",
          "K951117",
          "K862514",
          "K010689",
          "K043329",
          "K201466",
          "K970804",
          "K920632",
          "K953933",
          "K090297",
          "K152898",
          "K964076",
          "K202758",
          "K093502",
          "K223647",
          "K080530",
          "K020673",
          "K841234",
          "K943638",
          "K023967",
          "K905511",
          "K904700",
          "K240543",
          "K024189",
          "K222800",
          "K083478",
          "K782030",
          "K972561",
          "K904841",
          "K881848",
          "K170121",
          "K250183",
          "K231718",
          "K033172",
          "K243445",
          "K980361",
          "K003440",
          "K991891",
          "K133364",
          "K020071",
          "K925908",
          "K220189",
          "K990043",
          "K894241",
          "K140727",
          "K895746",
          "K913037",
          "K080697",
          "K200331",
          "K082911",
          "K980187",
          "K972002",
          "K072023",
          "K060069",
          "K950661",
          "K003147",
          "K183563",
          "K004006",
          "K862862",
          "K063828",
          "K903418",
          "K945143",
          "K992013",
          "K951775",
          "K093251",
          "K895918",
          "K941419",
          "K024088",
          "K200525",
          "K862864",
          "K121580",
          "K162837",
          "K885181",
          "K162765",
          "K250189",
          "K123385",
          "K024371",
          "K250465",
          "K895874",
          "K032797",
          "K924166",
          "K102359",
          "K840739",
          "K141511",
          "K953851",
          "K980131",
          "K931101",
          "K160488",
          "K920106",
          "K213350",
          "K881940",
          "K931264",
          "K061489",
          "K945228",
          "K974896",
          "K243141",
          "K990748",
          "K933140",
          "K120235",
          "K895263",
          "K221083",
          "K152220",
          "K100726",
          "K980398",
          "K041554",
          "K932468",
          "K150027",
          "K022839",
          "K063072",
          "K130413",
          "K063698",
          "K121327",
          "K241643",
          "K944623",
          "K081181",
          "K023624",
          "K952661",
          "K943201",
          "K050274",
          "K904628",
          "K880216",
          "K251094",
          "K961269",
          "K001784",
          "K222915",
          "K041011",
          "K894841",
          "K912792",
          "K180353",
          "K220426",
          "K001211",
          "K932883",
          "K141986",
          "K031425",
          "K921708",
          "K250656",
          "K022060",
          "K113858",
          "K994273",
          "K102639",
          "K933333",
          "K023962",
          "K133283",
          "K202250",
          "K933007",
          "K161771",
          "K972915",
          "K242946",
          "K072159",
          "K971118",
          "K222221",
          "K150664",
          "K061656",
          "K051428",
          "K912063",
          "K110364",
          "K885170",
          "K182831",
          "K030510",
          "K032085",
          "K990680",
          "K883541",
          "K912332",
          "K952772",
          "K152012",
          "K103654",
          "K083616",
          "K843915",
          "K952754",
          "K073534",
          "K020729",
          "K982031",
          "K941694",
          "K213261",
          "K103539",
          "K920529",
          "K162169",
          "K141658",
          "K914197",
          "K011703",
          "K935034",
          "K063001",
          "K093691",
          "K911408",
          "K960475",
          "K854386",
          "K962974",
          "K945614",
          "K895029",
          "K173067",
          "K030338",
          "K200693",
          "K780585",
          "K952118",
          "K240159",
          "K072514",
          "K063060",
          "K943409",
          "K242943",
          "K904126",
          "K240222",
          "K994014",
          "K014048",
          "K230371",
          "K260632",
          "K220467",
          "K901016",
          "K090312",
          "K871512",
          "K042000",
          "K130193",
          "K122020",
          "K112013",
          "K173305",
          "K210847",
          "K923855",
          "K974460",
          "K002296",
          "K032944",
          "K040081",
          "K141983",
          "K181662",
          "K881824",
          "K133158",
          "K090281",
          "K003993",
          "K890174",
          "K032599",
          "K203728",
          "K253342",
          "K081744",
          "K903444",
          "K062762",
          "K090235",
          "K110080",
          "K233416",
          "K240747",
          "K933057",
          "K951876",
          "K900166",
          "K943292",
          "K140122",
          "K073000",
          "K885291",
          "K970017",
          "K895143",
          "K010046",
          "K994437",
          "K944749",
          "K001527",
          "K060407",
          "K010284",
          "K150587",
          "K981143",
          "K122493",
          "K182036",
          "K211401",
          "K123004",
          "K101306",
          "K992866",
          "K951933",
          "K060977",
          "K003318",
          "K073617",
          "K991994",
          "K110502",
          "K921995",
          "K180044",
          "K230274",
          "K193446",
          "K972368",
          "K041767",
          "K950351",
          "K920589",
          "K232117",
          "K904402",
          "K851000",
          "K942217",
          "K952554",
          "K913191",
          "K904149",
          "K031795",
          "K902241",
          "K952340",
          "K151306",
          "K221789",
          "K081612",
          "K971404",
          "K173827",
          "K151331",
          "K911830",
          "K952755",
          "K951215",
          "K113588",
          "K241656",
          "K242293",
          "K161185",
          "K002853",
          "K914334",
          "K023881",
          "K101712",
          "K022837",
          "K904631",
          "K844643",
          "K983963",
          "K981438",
          "K031988",
          "K052874",
          "K253219",
          "K011677",
          "K112715",
          "K951204",
          "K200116",
          "K172325",
          "K062363",
          "K031125",
          "K943045",
          "K880603",
          "K091059",
          "K080932",
          "K203395",
          "K060411",
          "K190392",
          "K912067",
          "K091323",
          "K234085",
          "K032260",
          "K102749",
          "K171945",
          "K152529",
          "K182924",
          "K011960",
          "K213105",
          "K050452",
          "K980560",
          "K081095",
          "K022266",
          "K210226",
          "K944632",
          "K092964",
          "K033799",
          "K083900",
          "K884029",
          "K931070",
          "K081541",
          "K101321",
          "K950493",
          "K201171",
          "K241270",
          "K946336",
          "K250782",
          "K062963",
          "K980547",
          "K993671",
          "K910496",
          "K053421",
          "K021550",
          "K884313",
          "K140964",
          "K010580",
          "K020072",
          "K022607",
          "K910233",
          "K082225",
          "K983543",
          "K210951",
          "K992482",
          "K130256",
          "K073332",
          "K971250",
          "K172025",
          "K892346",
          "K013028",
          "K904629",
          "K102954",
          "K983100",
          "K070494",
          "K252658",
          "K952006",
          "K120737",
          "K901786",
          "K200659",
          "K030453",
          "K011939",
          "K211362",
          "K082423",
          "K202985",
          "K895559",
          "K974346",
          "K915354",
          "K072868",
          "K212790",
          "K902990",
          "K980538",
          "K083034",
          "K960521",
          "K062719",
          "K153718",
          "K945156",
          "K053527",
          "K233344",
          "K951034",
          "K043486",
          "K991798",
          "K945560",
          "K033072",
          "K201773",
          "K170187",
          "K894709",
          "K930789",
          "K952047",
          "K000158",
          "K052526",
          "K922375",
          "K000498",
          "K072682",
          "K251303",
          "K900076",
          "K193661",
          "K042785",
          "K893832",
          "K142638",
          "K882160",
          "K904162",
          "K910114",
          "K042211",
          "K914827",
          "K982113",
          "K233157",
          "K942729",
          "K083500",
          "K061470",
          "K153124",
          "K150907",
          "K973867",
          "K171227",
          "K944622",
          "K191349",
          "K974732",
          "K052688",
          "K972767",
          "K974142",
          "K851405",
          "K981627",
          "K895348",
          "K041930",
          "K897099",
          "K141838",
          "K903936",
          "K082809",
          "K090962",
          "K912703",
          "K080959",
          "K921298",
          "K180508",
          "K021548",
          "K063800",
          "K171443",
          "K881587",
          "K000805",
          "K942439",
          "K163128",
          "K091625",
          "K172413",
          "K942759",
          "K902946",
          "K930098",
          "K251801",
          "K192430",
          "K905802",
          "K021724",
          "K850628",
          "K894186",
          "K083082",
          "K022228",
          "K942152",
          "K960821",
          "K962791",
          "K990903",
          "K191685",
          "K063808",
          "K990243",
          "K000929",
          "K913465",
          "K942865",
          "K032357",
          "K241547",
          "K000503",
          "K133006",
          "K022569",
          "K854468",
          "K935631",
          "K070536",
          "K014253",
          "K173002",
          "K080406",
          "K163197",
          "K030480",
          "K924644",
          "K921096",
          "K974641",
          "K072767",
          "K990014",
          "K914991",
          "K944474",
          "K852860",
          "K932421",
          "K060234",
          "K223685",
          "K903079",
          "K911523",
          "K221306",
          "K951812",
          "K924647",
          "K871334",
          "K896478",
          "K931305",
          "K923575",
          "K894211",
          "K050841",
          "K010005",
          "K053161",
          "K222064",
          "K895279",
          "K003885",
          "K093793",
          "K012398",
          "K945054",
          "K123037",
          "K881585",
          "K882208",
          "K251464",
          "K882951",
          "K030290",
          "K161356",
          "K970143",
          "K162781",
          "K202953",
          "K221712",
          "K944621",
          "K052129",
          "K893956",
          "K924537",
          "K151890",
          "K941559",
          "K091666",
          "K003955",
          "K201056",
          "K980561",
          "K915651",
          "K211722",
          "K172193",
          "K070392",
          "K193464",
          "K923953",
          "K902211",
          "K932981",
          "K200110",
          "K123257",
          "K844644",
          "K870161",
          "K153229",
          "K053540",
          "K981447",
          "K183647",
          "K172908",
          "K032704",
          "K043012",
          "K182669",
          "K172881",
          "K083165",
          "K162114",
          "K222109",
          "K944977",
          "K083258",
          "K934865",
          "K921320",
          "K241498",
          "K930879",
          "K904266",
          "K913606",
          "K980753",
          "K884994",
          "K013402",
          "K242191",
          "K905708",
          "K040921",
          "K113212",
          "K090762",
          "K964831",
          "K131837",
          "K190007",
          "K934208",
          "K163655",
          "K151109",
          "K905006",
          "K001975",
          "K950825",
          "K163009",
          "K924160",
          "K964486",
          "K080118",
          "K980160",
          "K913569",
          "K020339",
          "K070157",
          "K090084",
          "K234057",
          "K252515",
          "K080638",
          "K954834",
          "K910200",
          "K904630",
          "K063571",
          "K133319",
          "K791895",
          "K100474",
          "K012545",
          "K885257",
          "K252745",
          "K895239",
          "K934935",
          "K071815",
          "K952796",
          "K141403",
          "K242987",
          "K052604",
          "K090181",
          "K220020",
          "K951714",
          "K211637",
          "K873571",
          "K971648",
          "K253123",
          "K060395",
          "K111359",
          "K962109",
          "K162851",
          "K182045",
          "K914677",
          "K142226",
          "K120231",
          "K251919",
          "K121346",
          "K913758",
          "K022464",
          "K904627",
          "K211228",
          "K021686",
          "K231950",
          "K191501",
          "K944630",
          "K963094",
          "K022341",
          "K070378",
          "K910825",
          "K991634",
          "K911852",
          "K213594",
          "K896412",
          "K201111",
          "K930981",
          "K050900",
          "K081416",
          "K261841",
          "K964760",
          "K970475",
          "K083613",
          "K903131",
          "K870971",
          "K150516",
          "K920625",
          "K912538",
          "K112166",
          "K243957",
          "K102716",
          "K942542",
          "K111144",
          "K133254",
          "K912029",
          "K041957",
          "K032863",
          "K081408",
          "K962993",
          "K944626",
          "K220103",
          "K190542",
          "K140388",
          "K933245",
          "K142216",
          "K954195",
          "K050950",
          "K172362",
          "K935310",
          "K093451",
          "K163123",
          "K011828",
          "K896183",
          "K131838",
          "K101658",
          "K121471",
          "K072182",
          "K091169",
          "K122368",
          "K022226",
          "K191162",
          "K951935",
          "K032703",
          "K121031",
          "K972325",
          "K180597",
          "K020941",
          "K881953",
          "K981718",
          "K000490",
          "K143723",
          "K850633",
          "K901943",
          "K900858",
          "K924158",
          "K941312",
          "K070071",
          "K142225",
          "K221312",
          "K912618",
          "K023734",
          "K122251",
          "K243701",
          "K950826",
          "K895073",
          "K063038",
          "K913918",
          "K951910",
          "K080496",
          "K020839",
          "K882952",
          "K945564",
          "K971065",
          "K073549",
          "K123031",
          "K983965",
          "K252345",
          "K212082",
          "K260153",
          "K193500",
          "K853901",
          "K923274",
          "K900066",
          "K971874",
          "K971912",
          "K915497",
          "K220531",
          "K964074",
          "K904155",
          "K090571",
          "K180922",
          "K090820",
          "K183199",
          "K973719",
          "K872589",
          "K073394",
          "K202998",
          "K915561",
          "K261693",
          "K023041",
          "K950019",
          "K062369",
          "K082232",
          "K181019",
          "K211517",
          "K973603",
          "K983058",
          "K921840",
          "K230660",
          "K070924",
          "K010715",
          "K000982",
          "K162556",
          "K900511",
          "K061366",
          "K212573",
          "K000023",
          "K896517",
          "K100143",
          "K233756",
          "K924943",
          "K031389",
          "K992630",
          "K060338",
          "K090195",
          "K032539",
          "K915370",
          "K862472",
          "K911987",
          "K180158",
          "K113868",
          "K931762",
          "K863433",
          "K930975",
          "K980475",
          "K910602",
          "K171711",
          "K961136",
          "K171986",
          "K984349",
          "K922577",
          "K040384",
          "K882212",
          "K893013",
          "K932730",
          "K050293",
          "K915583",
          "K960905",
          "K991632",
          "K250071",
          "K993146",
          "K022884",
          "K981746",
          "K141425",
          "K110685",
          "K120870",
          "K971459",
          "K102755",
          "K945587",
          "K982295",
          "K094049",
          "K022709",
          "K926543",
          "K943526",
          "K073572",
          "K963947",
          "K080006",
          "K972575",
          "K895093",
          "K041569",
          "K132185",
          "K180993",
          "K954911",
          "K233237",
          "K081324",
          "K013691",
          "K980685",
          "K130164",
          "K042139",
          "K952301",
          "K073158",
          "K163638",
          "K991183",
          "K944153",
          "K010927",
          "K030342",
          "K133561",
          "K102796",
          "K183156",
          "K260017",
          "K181523",
          "K032864",
          "K050069",
          "K010113",
          "K010285",
          "K892345",
          "K083927",
          "K930611",
          "K100228",
          "K251077",
          "K041242",
          "K161257",
          "K013499",
          "K161403",
          "K083848",
          "K983950",
          "K031233",
          "K141908",
          "K014272",
          "K882516",
          "K994225",
          "K241951",
          "K994116",
          "K895247",
          "K914726",
          "K961279",
          "K140470",
          "K920062",
          "K162762",
          "K960768",
          "K944472",
          "K023899",
          "K914396",
          "K923599",
          "K073300",
          "K121162",
          "K080975",
          "K990914",
          "K895663",
          "K102669",
          "K882180",
          "K963683",
          "K992084",
          "K923683",
          "K212663",
          "K052336",
          "K935563",
          "K990718",
          "K083253",
          "K050412",
          "K040424",
          "K895774",
          "K070870",
          "K963124",
          "K163222",
          "K013825",
          "K180593",
          "K922497",
          "K241699",
          "K904911",
          "K101573",
          "K081457",
          "K050254",
          "K222369",
          "K915840",
          "K012552",
          "K152153",
          "K952045",
          "K170366",
          "K994260",
          "K951165",
          "K021848",
          "K191693",
          "K173700",
          "K040289",
          "K052777",
          "K152591",
          "K043251",
          "K984424",
          "K052806",
          "K170850",
          "K962354",
          "K121286",
          "K223680",
          "K251727",
          "K073464",
          "K130444",
          "K242142",
          "K011409",
          "K082298",
          "K250113",
          "K893529",
          "K063152",
          "K911457",
          "K213250",
          "K021975",
          "K781453",
          "K050562",
          "K241144",
          "K020849",
          "K093545",
          "K213484",
          "K011070",
          "K250809",
          "K954904",
          "K061909",
          "K994427",
          "K901710",
          "K150336",
          "K003765",
          "K980838",
          "K062048",
          "K201687",
          "K950710",
          "K081424",
          "K041103",
          "K230308",
          "K982548",
          "K052354",
          "K020021",
          "K944627",
          "K251149",
          "K230373",
          "K954905",
          "K970637",
          "K002890",
          "K110897",
          "K170060",
          "K950684",
          "K040270",
          "K913990",
          "K130465",
          "K964497",
          "K180967",
          "K133788",
          "K230819",
          "K252787",
          "K251699",
          "K172639",
          "K031690",
          "K011322",
          "K032397",
          "K180193",
          "K212733",
          "K013974",
          "K915223",
          "K020856",
          "K001243",
          "K950248",
          "K862863",
          "K936109",
          "K162454",
          "K873664",
          "K080915",
          "K142093",
          "K033181",
          "K202991",
          "K211335",
          "K124003",
          "K955861",
          "K163449",
          "K092197",
          "K031184",
          "K920107",
          "K883563",
          "K912879",
          "K983654",
          "K913839",
          "K970157",
          "K021444",
          "K243683",
          "K944204",
          "K983868",
          "K091551",
          "K013216",
          "K963193",
          "K110225",
          "K941841",
          "K240396",
          "K962632",
          "K992472",
          "K223778",
          "K241646",
          "K912813",
          "K103753",
          "K233018",
          "K901211",
          "K061966",
          "K253344",
          "K051359",
          "K243465",
          "K202848",
          "K083733",
          "K230535",
          "K093162",
          "K070355",
          "K152032",
          "K222751",
          "K050317",
          "K881588",
          "K160549",
          "K850804",
          "K943527",
          "K904229",
          "K191554",
          "K980537",
          "K971024",
          "K232716",
          "K944629",
          "K974586",
          "K972272",
          "K142186",
          "K021149",
          "K934706",
          "K202980",
          "K162334",
          "K970808",
          "K930490",
          "K010771",
          "K103118",
          "K081589",
          "K900165",
          "K915631",
          "K961935",
          "K024204",
          "K131839",
          "K002884",
          "K901636",
          "K082033",
          "K964128",
          "K140249",
          "K963389",
          "K850805",
          "K182604",
          "K192951",
          "K861044",
          "K021180",
          "K896314",
          "K261094",
          "K181601",
          "K193426",
          "K944471",
          "K004021",
          "K110434",
          "K251049",
          "K914136",
          "K895363",
          "K213889",
          "K050382",
          "K970422",
          "K904227",
          "K071621",
          "K900144",
          "K141555",
          "K913488",
          "K042496",
          "K914952",
          "K030527",
          "K014100",
          "K032220",
          "K921990",
          "K002899",
          "K162886",
          "K234104",
          "K973764",
          "K905737",
          "K050673",
          "K093852",
          "K003614",
          "K915236",
          "K143519",
          "K062312",
          "K914518",
          "K022338",
          "K161926",
          "K223540",
          "K870970",
          "K071451",
          "K032460",
          "K992824",
          "K033259",
          "K083795",
          "K840406",
          "K000106",
          "K924214",
          "K950996",
          "K881952",
          "K832983",
          "K062071",
          "K974536",
          "K252447",
          "K191842",
          "K881870",
          "K071328",
          "K142286",
          "K090094",
          "K915117",
          "K945126",
          "K930099",
          "K031805",
          "K012806",
          "K862408",
          "K223588",
          "K002308",
          "K001322",
          "K153564",
          "K860278",
          "K974229",
          "K891126",
          "K131757",
          "K022568",
          "K945888",
          "K896617",
          "K072424",
          "K955612",
          "K960820",
          "K230077",
          "K192755",
          "K080382",
          "K903883",
          "K024259",
          "K980784",
          "K933211",
          "K884092",
          "K122159",
          "K242132",
          "K971843",
          "K001511",
          "K922917",
          "K932065",
          "K073409",
          "K051442",
          "K980389",
          "K022817",
          "K955621",
          "K971830",
          "K043575",
          "K042918",
          "K963996",
          "K000072",
          "K955734",
          "K111475",
          "K252564",
          "K875115",
          "K913290",
          "K200118",
          "K983463",
          "K003385",
          "K240132",
          "K062258",
          "K864563",
          "K130250",
          "K024335",
          "K231945",
          "K881580",
          "K970069",
          "K052495",
          "K923539",
          "K121475",
          "K133774",
          "K063427",
          "K915085",
          "K080374",
          "K030508",
          "K931445",
          "K001093",
          "K191518",
          "K251759",
          "K240118",
          "K850779",
          "K000727",
          "K111213",
          "K241527",
          "K020453",
          "K231054",
          "K881583",
          "K061618",
          "K123682",
          "K120573",
          "K040727",
          "K003715",
          "K160677",
          "K903849",
          "K924664",
          "K043377",
          "K901541",
          "K172771",
          "K880809",
          "K033354",
          "K222555",
          "K980313",
          "K904610",
          "K253633",
          "K082101",
          "K900925",
          "K020107",
          "K031489",
          "K261221",
          "K924120",
          "K961705",
          "K061117",
          "K060516",
          "K905232",
          "K934389",
          "K983984",
          "K041721",
          "K001263",
          "K970934",
          "K082247",
          "K240537",
          "K230342",
          "K101904",
          "K953670",
          "K202041",
          "K160392",
          "K151629",
          "K093048",
          "K914089",
          "K953412",
          "K112324",
          "K243147",
          "K924424",
          "K252771",
          "K884311",
          "K931978",
          "K002347",
          "K252753",
          "K040103",
          "K022544",
          "K992620",
          "K945563",
          "K984339",
          "K915679",
          "K925178",
          "K233077",
          "K123293",
          "K241600",
          "K922634",
          "K890040",
          "K924729",
          "K250195",
          "K250041",
          "K922663",
          "K192516",
          "K251457",
          "K251632",
          "K030834",
          "K113688",
          "K862423",
          "K943693",
          "K923991",
          "K943311",
          "K933525",
          "K021508",
          "K892359",
          "K042114",
          "K981149",
          "K914849",
          "K173285",
          "K952598",
          "K130199",
          "K020463",
          "K141852",
          "K073404",
          "K221623",
          "K091115",
          "K041959",
          "K901128",
          "K884060",
          "K905637",
          "K920004",
          "K914455",
          "K002032",
          "K100035",
          "K010618",
          "K181632",
          "K072262",
          "K992298",
          "K022610",
          "K942868",
          "K921842",
          "K981379",
          "K982980",
          "K060455",
          "K240386",
          "K210168",
          "K021304",
          "K946386",
          "K014128",
          "K213557",
          "K063384",
          "K103501",
          "K062336",
          "K091537",
          "K934008",
          "K170734",
          "K993285",
          "K212073",
          "K260375",
          "K111689",
          "K152611",
          "K920934",
          "K943444",
          "K031140",
          "K241672",
          "K013864",
          "K182270",
          "K043046",
          "K881887",
          "K820101",
          "K211605",
          "K940173",
          "K040234",
          "K203544",
          "K854473",
          "K031317",
          "K200042",
          "K013201",
          "K891909"
        ],
        "registration_number": [
          "",
          "3009444646",
          "9610617",
          "3005987567",
          "3019906161",
          "1058584",
          "3025482817",
          "1222993",
          "3016181563",
          "3011310540",
          "3043012240",
          "3043140693",
          "3011657089",
          "2032092",
          "3007048256",
          "3003616071",
          "9612147",
          "3013789508",
          "2124215",
          "3040278150",
          "3004505220",
          "3020770009",
          "8010720",
          "1000391004",
          "3012772112",
          "3027619554",
          "3040278125",
          "2938889",
          "1419951",
          "3017144492",
          "3014222368",
          "3009736547",
          "2529392",
          "3043657394",
          "3016678045",
          "3009298127",
          "3013697074",
          "3003168101",
          "3002808227",
          "3026540840",
          "2011171",
          "3012669557",
          "3011555967",
          "3008156177",
          "3017181358",
          "9611388",
          "3004096429",
          "9612448",
          "3003015433",
          "3017400436",
          "3020900954",
          "3007120647",
          "3002736133",
          "3014737458",
          "3006247712",
          "3009383334",
          "3012301874",
          "8010349",
          "3004167969",
          "3013376016",
          "3023344928",
          "3019559090",
          "3030495752",
          "3006368133",
          "3036240802",
          "3010314781",
          "3042255006",
          "3045892905",
          "3008729892",
          "3023850562",
          "3004940677",
          "3033273466",
          "3006546082",
          "1222773",
          "8040564",
          "3003288808",
          "3016453096",
          "2954354",
          "3026400760",
          "3014577013",
          "3013041594",
          "3020611964",
          "2936921",
          "3008252121",
          "1424263",
          "3009970070",
          "3017786046",
          "3012980277",
          "3007731000",
          "3017638059",
          "8021543",
          "9611516",
          "3004069780",
          "3027851853",
          "1420032",
          "3010912900",
          "3012133441",
          "3006807694",
          "3007535734",
          "3006849754",
          "3001222017",
          "9710340",
          "3027605507",
          "3003683459",
          "3037874486",
          "3015262665",
          "3004497289",
          "3018318562",
          "3011514163",
          "2018957",
          "3042261306",
          "1216677",
          "3014253920",
          "3011137372",
          "3005350457",
          "3007689716",
          "3035375056",
          "3003610374",
          "3011743768",
          "3009019578",
          "3010707607",
          "3030418454",
          "3005581016",
          "3034246188",
          "3015730775",
          "3010281590",
          "3001605696",
          "3026038437",
          "2031934",
          "1423662",
          "3003610527",
          "3005855240",
          "3015976667",
          "3003899230",
          "3003082988",
          "3017954802",
          "3013602304",
          "3037047060",
          "1221275",
          "1226005",
          "3021159346",
          "3004374705",
          "3043729119",
          "3038373471",
          "3005205318",
          "3036425107",
          "3014980461",
          "3005972890",
          "1226486",
          "1225450",
          "3002807636",
          "3033482500",
          "2247224",
          "3004642415",
          "2032112",
          "1043653",
          "9616684",
          "1048735",
          "3017060146",
          "3014225468",
          "1064858",
          "3025268687",
          "3008925277",
          "3014384451",
          "1220477",
          "3017200975",
          "3002807715",
          "1018470",
          "3003340844",
          "3044995573",
          "3006680097",
          "3011423170",
          "3033905238",
          "3009593803",
          "3015907115",
          "1219930",
          "3008720474",
          "3030733800",
          "3007181527",
          "3012294783",
          "3017132875",
          "3021380980",
          "3005282635",
          "3001431138",
          "3010363436",
          "3016661372",
          "3005730446",
          "3032147807",
          "3042953831",
          "3004858034",
          "3009462521",
          "3032916575",
          "3004450661",
          "3025379728",
          "3017124433",
          "3012104022",
          "3041275224",
          "3042641021",
          "1820334",
          "8044098",
          "1018233",
          "1225072",
          "3011644607",
          "3008805497",
          "3004483538",
          "1038823",
          "3008206114",
          "2032513",
          "3010933841",
          "3033844600",
          "3006546131",
          "3006555617",
          "2246552",
          "3010683226",
          "3011782009",
          "3013397886",
          "3005545844",
          "1928237",
          "3017929928",
          "3042351542",
          "3012911444",
          "3013503862",
          "3004361445",
          "3034995492",
          "3005695724",
          "3013121972",
          "3027605646",
          "3002806603",
          "9681260",
          "3015779311",
          "3004633032",
          "2011068",
          "3023344923",
          "3012236936",
          "3021334828",
          "3025104654",
          "3017065670",
          "3015151147",
          "3021315370",
          "3007712511",
          "3017173524",
          "3010627091",
          "3013404959",
          "3005099803",
          "3018188770",
          "3010902869",
          "3009887511",
          "3004710757",
          "3021349626",
          "3017125161",
          "3016746323",
          "3016496662",
          "3043012074",
          "3013515803",
          "2125050",
          "3010202439",
          "3043127589",
          "3017264244",
          "3006985208",
          "3025861981",
          "3016413636",
          "3006695864",
          "3023808190",
          "3033511105",
          "2031763",
          "3012799115",
          "3010511300",
          "3005739661",
          "1319211",
          "3039396569",
          "2134812",
          "1000121056",
          "1221543",
          "3007549439",
          "3033884002",
          "3007156625",
          "3043047259",
          "8030115",
          "9611036",
          "3015658550",
          "3038189188",
          "3042228064",
          "3033891767",
          "3031944951",
          "3031233199",
          "3010175098",
          "3003790304",
          "3010972758",
          "8043564",
          "3031582430",
          "3010822033",
          "9613462",
          "3015529593",
          "3043654467",
          "2939653",
          "2020550",
          "3038372243",
          "3010364538",
          "3005941719",
          "3012125201",
          "3010053203",
          "3043011436",
          "3030420005",
          "3039935725",
          "1000523114",
          "3004135191",
          "1424785",
          "3000313549",
          "3033126947",
          "3009040934",
          "8030987",
          "3020221089",
          "3038356864",
          "3006617841",
          "3005791808",
          "3005801723",
          "3002991496",
          "3016405041",
          "3004635398",
          "2916735",
          "3016830208",
          "3008442971",
          "3007336385",
          "3011670002",
          "3006786907",
          "3007085226",
          "3013532143",
          "3007113239",
          "1000393132",
          "3011877721",
          "3042262496",
          "3012452798",
          "3004529369",
          "3044905918",
          "3030090167",
          "3014980477",
          "2953696",
          "3024753061",
          "1032227",
          "3010060157",
          "3044175602",
          "3016076077",
          "3012639861",
          "3005580113",
          "3006082230",
          "3033715699",
          "3003708273",
          "3033524191",
          "3003477135",
          "2183744",
          "3027725665",
          "3014561425",
          "3016096097",
          "3011302631",
          "3013613043",
          "1037955",
          "3010035831",
          "3038197172",
          "3030543531",
          "2438558",
          "3009380063",
          "3030233762",
          "3005238200",
          "3015077548",
          "3037966040",
          "3006985163",
          "3010359222",
          "3015313684",
          "1825146",
          "3004378299",
          "3015392301",
          "3009241253",
          "3008915316",
          "3019705154",
          "3011050570",
          "3010273372",
          "3008965332",
          "3034456223",
          "1045254",
          "3016445722",
          "3009649442",
          "3007689729",
          "3002807314",
          "8043792",
          "3010027944",
          "3013529876",
          "3005180774",
          "3012157621",
          "2647580",
          "3005515469",
          "3004475754",
          "3015807001",
          "3016170453",
          "3007674911",
          "3030230686",
          "3005403788",
          "3009746061",
          "3018094310",
          "3020756",
          "1644312",
          "1647149",
          "2938071",
          "3009078366",
          "3004111573",
          "2183911",
          "3013619029",
          "2027755",
          "3021011208",
          "3014342096",
          "3013734919",
          "3017140525",
          "3041918479",
          "3026360847",
          "3014310779",
          "9680671",
          "3013917867",
          "3004884150",
          "9614977",
          "1000548922",
          "3030125922",
          "3043153468",
          "3008627763",
          "3007583504",
          "3023343409",
          "3016076698",
          "3008776882",
          "3005210579",
          "1035236",
          "3026294985",
          "3014144061",
          "2953684",
          "3016453004",
          "3021550489",
          "1313525",
          "3014226003",
          "3022327202",
          "3038715747",
          "1721676",
          "1318319",
          "3010724743",
          "3012289707",
          "3018116888",
          "3015452693",
          "9615030",
          "3013557562",
          "1054713",
          "3033274371",
          "8030392",
          "3011024620",
          "3009039068",
          "1419489",
          "1060848",
          "3046005941",
          "2245304",
          "3030979156",
          "3026668625",
          "3019252665",
          "3009488834",
          "3034630389",
          "3013446079",
          "2521877",
          "3008872264",
          "1723170",
          "3008393409",
          "3012440264",
          "3043149902",
          "3006716777",
          "3010220595",
          "3008085016",
          "3015167792",
          "3015433394",
          "3004485166",
          "3006427656",
          "1725006",
          "3006580954",
          "3003542872",
          "3014388884",
          "3012322232",
          "1319639",
          "3009453258",
          "3014132468",
          "3010273872",
          "3030979014",
          "3011964664",
          "2528981",
          "3031233197",
          "3029897009",
          "8030885",
          "3007084556"
        ],
        "fei_number": [
          "3002942014",
          "3009444646",
          "",
          "1000138054",
          "3005747797",
          "3005987567",
          "3019906161",
          "3025482817",
          "3016181563",
          "3011310540",
          "3043012240",
          "3043140693",
          "3011657089",
          "3007048256",
          "3003616071",
          "3013789508",
          "2124215",
          "3040278150",
          "3004505220",
          "3020770009",
          "3004028763",
          "1000391004",
          "3012772112",
          "3027619554",
          "3000219468",
          "3040278125",
          "2938889",
          "1419951",
          "3017144492",
          "3000204219",
          "3014222368",
          "3009736547",
          "3043657394",
          "3016678045",
          "3009298127",
          "3013697074",
          "3003168101",
          "3002808227",
          "3026540840",
          "3002807111",
          "1000114111",
          "2011171",
          "3012669557",
          "3011555967",
          "1000597018",
          "3017181358",
          "3004096429",
          "3003015433",
          "3017400436",
          "3002690217",
          "3020900954",
          "3007120647",
          "3002736133",
          "3014737458",
          "3002806565",
          "3003244961",
          "3006247712",
          "3009383334",
          "3012301874",
          "3004167969",
          "3013376016",
          "3023344928",
          "3019559090",
          "1000122089",
          "3030495752",
          "3006368133",
          "3036240802",
          "3010314781",
          "3042255006",
          "3045892905",
          "3023850562",
          "3002808268",
          "3033273466",
          "3006546082",
          "3007475226",
          "3003288808",
          "3016453096",
          "3026400760",
          "3014577013",
          "3013041594",
          "3002900199",
          "2936921",
          "3008252121",
          "1000517638",
          "3010666513",
          "3009970070",
          "3003640911",
          "3017786046",
          "3012980277",
          "3007731000",
          "3017638059",
          "3004069780",
          "3027851853",
          "1420032",
          "3010912900",
          "3012133441",
          "3006807694",
          "3007535734",
          "3006849754",
          "3001222017",
          "3027605507",
          "3003683459",
          "3037874486",
          "3015262665",
          "3004497289",
          "3018318562",
          "7759",
          "3011514163",
          "2018957",
          "3005616124",
          "3042261306",
          "1216677",
          "3014253920",
          "3005350457",
          "3007689716",
          "3035375056",
          "3002807555",
          "3003610374",
          "3011743768",
          "3009019578",
          "1000219637",
          "3010707607",
          "3030418454",
          "3005581016",
          "3034246188",
          "1720381",
          "3015730775",
          "3010281590",
          "1000120578",
          "3001605696",
          "1000120768",
          "3026038437",
          "3003610527",
          "3000205157",
          "3005855240",
          "3015976667",
          "3003899230",
          "3003082988",
          "3017954802",
          "3013602304",
          "3037047060",
          "3002806519",
          "1221275",
          "3002806902",
          "3021159346",
          "3004374705",
          "3043729119",
          "3038373471",
          "3005205318",
          "3036425107",
          "3014980461",
          "3005972890",
          "3002807636",
          "3033482500",
          "3004642415",
          "3003067608",
          "3008397795",
          "1043653",
          "3002806913",
          "1048735",
          "3017060146",
          "3014225468",
          "3025268687",
          "3014384451",
          "3017200975",
          "1220477",
          "3002807715",
          "1018470",
          "3003340844",
          "3044995573",
          "3006680097",
          "3033905238",
          "3009593803",
          "3015907115",
          "1219930",
          "3008720474",
          "3002808269",
          "3030733800",
          "3007181527",
          "3012294783",
          "3017132875",
          "3000718272",
          "3021380980",
          "3005282635",
          "3001431138",
          "3010363436",
          "3016661372",
          "3005730446",
          "3032147807",
          "3042953831",
          "3004858034",
          "3009462521",
          "3032916575",
          "3004450661",
          "3025379728",
          "3017124433",
          "3000215273",
          "3012104022",
          "3041275224",
          "3042641021",
          "1820334",
          "1000117748",
          "1018233",
          "3011644607",
          "3008805497",
          "3004483538",
          "1038823",
          "3008206114",
          "3010933841",
          "3033844600",
          "3006546131",
          "3006555617",
          "2246552",
          "3010683226",
          "3013397886",
          "3005545844",
          "3003015552",
          "3017929928",
          "1928237",
          "3042351542",
          "3012911444",
          "1000306139",
          "3013503862",
          "3004361445",
          "3034995492",
          "3005695724",
          "3013121972",
          "3027605646",
          "3002806603",
          "3015779311",
          "3004633032",
          "2011068",
          "3023344923",
          "3012236936",
          "3021334828",
          "3025104654",
          "3017065670",
          "3015151147",
          "3021315370",
          "3007712511",
          "3017173524",
          "3010627091",
          "3004106598",
          "3005099803",
          "3018188770",
          "3010902869",
          "1000623228",
          "1000279065",
          "3009887511",
          "3004710757",
          "3021349626",
          "3017125161",
          "3016746323",
          "3016496662",
          "3002806833",
          "3043012074",
          "3002809599",
          "2125050",
          "3043127589",
          "3017264244",
          "3006985208",
          "3025861981",
          "3016413636",
          "3006695864",
          "3023808190",
          "3033511105",
          "3012799115",
          "3010511300",
          "1000121498",
          "3005739661",
          "1319211",
          "3039396569",
          "1000121056",
          "1221543",
          "3007549439",
          "3033884002",
          "3043047259",
          "3015658550",
          "3038189188",
          "3042228064",
          "3033891767",
          "3031944951",
          "3031233199",
          "3010175098",
          "3003790304",
          "3010972758",
          "3031582430",
          "3010822033",
          "3015529593",
          "3043654467",
          "2939653",
          "2020550",
          "3038372243",
          "3008133910",
          "3010364538",
          "3012125201",
          "3003258642",
          "3010053203",
          "3043011436",
          "3030420005",
          "3039935725",
          "1000523114",
          "3004135191",
          "3000313549",
          "3033126947",
          "3009040934",
          "3020221089",
          "3038356864",
          "3006617841",
          "3002913962",
          "3005791808",
          "3005801723",
          "3008697515",
          "3002991496",
          "3016405041",
          "3004635398",
          "2916735",
          "3016830208",
          "3001362763",
          "3008442971",
          "3007336385",
          "3011670002",
          "3006786907",
          "3007085226",
          "3013532143",
          "3007113239",
          "1000393132",
          "3011877721",
          "3042262496",
          "3012452798",
          "3004529369",
          "3044905918",
          "3030090167",
          "3014980477",
          "3024753061",
          "1032227",
          "3003719498",
          "3010060157",
          "3044175602",
          "3016076077",
          "3012639861",
          "3003465293",
          "3005580113",
          "3006082230",
          "3008108972",
          "3033715699",
          "3003708273",
          "3033524191",
          "3003477135",
          "2183744",
          "3027725665",
          "3014561425",
          "3016096097",
          "1000116142",
          "3011302631",
          "3013613043",
          "1037955",
          "3010035831",
          "3038197172",
          "3030543531",
          "6137",
          "3009380063",
          "3030233762",
          "3005238200",
          "3015077548",
          "3037966040",
          "3006985163",
          "3010359222",
          "3003462036",
          "3002757272",
          "3015313684",
          "1825146",
          "3004378299",
          "3002483854",
          "3015392301",
          "3009241253",
          "3008915316",
          "3019705154",
          "3002827704",
          "3011050570",
          "3010273372",
          "3008965332",
          "3003559308",
          "3034456223",
          "1045254",
          "3016445722",
          "3009649442",
          "3002807314",
          "3010027944",
          "3013529876",
          "3005180774",
          "3012157621",
          "2647580",
          "3003473212",
          "3005515469",
          "3004475754",
          "3015807001",
          "3016170453",
          "3007674911",
          "3030230686",
          "3005403788",
          "3003076831",
          "3009746061",
          "3018094310",
          "3020756",
          "3001415984",
          "2938071",
          "3009078366",
          "3004111573",
          "3013619029",
          "2027755",
          "3021011208",
          "3014342096",
          "3013734919",
          "3017140525",
          "3041918479",
          "3026360847",
          "3014310779",
          "3013917867",
          "3004884150",
          "1000548922",
          "3030125922",
          "3043153468",
          "3008627763",
          "3004143769",
          "3007583504",
          "3023343409",
          "3016076698",
          "3008776882",
          "3005210579",
          "1035236",
          "3026294985",
          "3014144061",
          "3016453004",
          "3021550489",
          "3014226003",
          "3022327202",
          "3038715747",
          "1721676",
          "1318319",
          "3010724743",
          "3012289707",
          "3018116888",
          "3015452693",
          "3013557562",
          "3033274371",
          "3011024620",
          "3009039068",
          "1419489",
          "3046005941",
          "2245304",
          "3030979156",
          "3026668625",
          "3019252665",
          "3009488834",
          "3034630389",
          "3013446079",
          "2521877",
          "3008872264",
          "3008393409",
          "3012440264",
          "3043149902",
          "3006716777",
          "3010220595",
          "3008085016",
          "3015167792",
          "3015433394",
          "3004485166",
          "3006427656",
          "3002806876",
          "3006580954",
          "3003542872",
          "3014388884",
          "3012322232",
          "3009453258",
          "3014132468",
          "3002806910",
          "3010273872",
          "3030979014",
          "1317628",
          "3011964664",
          "2528981",
          "3031233197",
          "3029897009",
          "3002806863",
          "3001611958",
          "3007084556"
        ],
        "device_name": "Powered Laser Surgical Instrument",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4810",
        "device_class": "2"
      }
    }
  ]
}