  "results": [
    {
      "cfres_id": "176420",
      "product_res_number": "Z-0150-2020",
      "event_date_initiated": "2019-09-19",
      "event_date_posted": "2019-10-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2024-03-04",
      "res_event_number": "83819",
      "product_code": "GEI",
      "k_numbers": [
        "K053433"
      ],
      "product_description": "Electrosurgical Tip Cleaner\nREF: 62-8499-001\n\nProduct Usage:  electrosurgical accessory",
      "code_info": "Lot # 16DEC19\n17JAN06\n17FEB10\n17MAR27\n17OCT09\n17OCT09A\n17OCT23\n17OCT23A\n18JAN05\n18JAN05A\n18JAN19\n18MAR02\n18MAR19\n18MAR19A\n\nUDI: Case   20787551000584 and Each - 00787551000580\n\nAdditional lots as of 6/8/20:\n18APR09\n18MAY01\n18MAY11\n18MAY25\n18MAY25A\n18MAY25B\n18JUN27\n18AUG14\n18SEP12\n18OCT03\n\nAdditional lots as of 9/9/2021:\n18DEC05\n18DEC26\n18DEC26A\n19JAN30\n19JAN30A",
      "firm_fei_number": "1000221205",
      "recalling_firm": "Xodus Medical Inc",
      "address_1": "702 Prominence Dr",
      "city": "New Kensington",
      "state": "PA",
      "postal_code": "15068-7052",
      "additional_info_contact": "Paul Lloyd\n724-337-5500",
      "reason_for_recall": "The sterile barrier of some devices may have been affected",
      "root_cause_description": "Packaging",
      "action": "Urgent Medical Device Voluntary Recall notification letters dated 9/20/19 were sent to customers.  An Urgent Medical Device Voluntary Recall REVISED RECALL NOTICE notification letters were sent to customers on 9/27/19.\n\n1.  Immediately check your inventory and quarantine affected stock on hand.\n\n2. The following process shall be undertaken in order to segregate any affected\ninventory. The existing inventory shall be inspected and any affected product\nquarantined and destroyed.\n   A. Inspect the Tyvek pouch seal integrity.\n   B. Secondly, the product contained within the pouch may be partially sealed\ninto the Tyvek seal itself.\n\n3. unaffected product shall be distributed with no futher action necessary.\n\n4. For any product that is quarantined and must be destroyed:\n\nA. Xodus Medical requires documentation of the affected inventory that is\n found to be affected and destroyed. Information regarding the following\n  shall be documented and provided to Xodus Medical.\n\nB. Product Number; Quantity Affected; Method of Destruction;                          Evidence   of  Destruction. It can be photos or videos of product being\ndestroyed\n\n5. Please immediately complete the attached Acknowledgement form even if you do not have any affected stock, and return the acknowledgement to Xodus\nMedical at the following: Email   PLloyd@XodusMedical.com or Fax +1 724-337-1131\n\nFor further information and question,  please call Paul Lloyd, Vice President, Global QA/RA & Technology, Xodus Medical; +1 724-337-5500; PLloyd@XodusMedical.com\n\nUrgent: Medical Device Voluntary Recall REVISED RECALL NOTICE dated 6/8/20 were sent to customers.\n\nXodus Medical has received complaints of some devices where the sterile barrier of some devices may have been affected. Xodus Medical has confirmed instances where the pouch seal was compromised on some of the reported product. In no instance has it been reported to Xodus Medical that a compromise in the sterile barrier has resul",
      "product_quantity": "460,000 eaches",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the States of : CA, ID, IL, LA, MD, NY, OH, OK, TX, VA, WI and OUS: Malaysia, Kuwait, Australia, Canada, UAE, Ireland, Korea, Saudi Arabia, Japan, Mexico",
      "openfda": {
        "k_number": [
          "K961087",
          "K243713",
          "K160602",
          "K233120",
          "K200300",
          "K974279",
          "K081559",
          "K041968",
          "K013348",
          "K972510",
          "K944012",
          "K964176",
          "K100258",
          "K213127",
          "K192728",
          "K191528",
          "K113545",
          "K014216",
          "K951912",
          "K062539",
          "K070101",
          "K113407",
          "K033592",
          "K203640",
          "K920071",
          "K982744",
          "K032132",
          "K001382",
          "K971674",
          "K942183",
          "K911365",
          "K040135",
          "K955467",
          "K955346",
          "K052449",
          "K041383",
          "K000911",
          "K781510",
          "K962690",
          "K003092",
          "K001442",
          "K111262",
          "K841483",
          "K972273",
          "K112033",
          "K905529",
          "K982743",
          "K896839",
          "K081431",
          "K942710",
          "K252211",
          "K222429",
          "K060117",
          "K210569",
          "K081634",
          "K083528",
          "K203299",
          "K960554",
          "K912182",
          "K062935",
          "K955166",
          "K063025",
          "K162810",
          "K101080",
          "K911904",
          "K013730",
          "K083856",
          "K124029",
          "K232505",
          "K971310",
          "K872705",
          "K062723",
          "K903812",
          "K881427",
          "K060823",
          "K050223",
          "K143306",
          "K230663",
          "K943991",
          "K061174",
          "K170721",
          "K944634",
          "K001150",
          "K924469",
          "K813071",
          "K992406",
          "K260522",
          "K964329",
          "K180848",
          "K233604",
          "K233879",
          "K992542",
          "K820857",
          "K974793",
          "K024059",
          "K942607",
          "K141122",
          "K080741",
          "K172856",
          "K935970",
          "K041633",
          "K994428",
          "K231216",
          "K110924",
          "K972335",
          "K003005",
          "K231405",
          "K951032",
          "K211579",
          "K000496",
          "K202415",
          "K092191",
          "K070004",
          "K030028",
          "K202199",
          "K844004",
          "K182244",
          "K010662",
          "K081644",
          "K191947",
          "K251492",
          "K791415",
          "K072493",
          "K201150",
          "K142910",
          "K963965",
          "K020919",
          "K943542",
          "K143536",
          "K922999",
          "K040699",
          "K053365",
          "K170296",
          "K983852",
          "K001330",
          "K122425",
          "K023986",
          "K990474",
          "K111202",
          "K122092",
          "K870397",
          "K980170",
          "K964602",
          "K781429",
          "K251836",
          "K212329",
          "K992045",
          "K081398",
          "K232407",
          "K250657",
          "K233705",
          "K942627",
          "K942215",
          "K091290",
          "K250852",
          "K940192",
          "K040584",
          "K102275",
          "K992760",
          "K083306",
          "K823924",
          "K171391",
          "K934656",
          "K211639",
          "K033205",
          "K173405",
          "K122716",
          "K240017",
          "K062047",
          "K925266",
          "K800022",
          "K232033",
          "K873595",
          "K914131",
          "K942499",
          "K210423",
          "K963930",
          "K010060",
          "K921921",
          "K192518",
          "K130744",
          "K143175",
          "K914910",
          "K183112",
          "K912070",
          "K073057",
          "K132754",
          "K061001",
          "K241099",
          "K173877",
          "K132665",
          "K161804",
          "K222556",
          "K932817",
          "K992972",
          "K190052",
          "K183610",
          "K212472",
          "K082890",
          "K812203",
          "K023311",
          "K091228",
          "K960918",
          "K932632",
          "K103282",
          "K231495",
          "K011634",
          "K961812",
          "K151696",
          "K080287",
          "K911341",
          "K821356",
          "K062949",
          "K160103",
          "K974250",
          "K032088",
          "K861594",
          "K160360",
          "K014222",
          "K082954",
          "K050102",
          "K152317",
          "K183009",
          "K910184",
          "K000480",
          "K972269",
          "K981888",
          "K904993",
          "K885172",
          "K013369",
          "K072789",
          "K013641",
          "K250065",
          "K191583",
          "K840377",
          "K944433",
          "K180872",
          "K200931",
          "K062072",
          "K094009",
          "K132224",
          "K071543",
          "K870149",
          "K030557",
          "K942577",
          "K140629",
          "K903855",
          "K203150",
          "K951732",
          "K874775",
          "K980393",
          "K252083",
          "K992931",
          "K024047",
          "K950347",
          "K152933",
          "K052857",
          "K052936",
          "K051749",
          "K242286",
          "K791752",
          "K984275",
          "K181071",
          "K133273",
          "K123444",
          "K162547",
          "K000936",
          "K983788",
          "K021402",
          "K253504",
          "K073455",
          "K082786",
          "K211444",
          "K152731",
          "K033232",
          "K945759",
          "K254056",
          "K180403",
          "K150735",
          "K954253",
          "K073450",
          "K170908",
          "K990586",
          "K973890",
          "K130689",
          "K221407",
          "K905108",
          "K926203",
          "K140624",
          "K932392",
          "K102913",
          "K001159",
          "K211217",
          "K982556",
          "K980177",
          "K222793",
          "K170369",
          "K192813",
          "K150297",
          "K926297",
          "K955502",
          "K063748",
          "K230432",
          "K230621",
          "K964736",
          "K182653",
          "K970541",
          "K221907",
          "K955681",
          "K011929",
          "K854231",
          "K140006",
          "K912781",
          "K152860",
          "K100816",
          "K960476",
          "K253656",
          "K083737",
          "K162490",
          "K003773",
          "K220138",
          "K822070",
          "K082391",
          "K150364",
          "K952204",
          "K013639",
          "K944796",
          "K230586",
          "K032584",
          "K023291",
          "K900947",
          "K130027",
          "K221830",
          "K010661",
          "K910510",
          "K233830",
          "K952698",
          "K791541",
          "K032965",
          "K231872",
          "K984276",
          "K142350",
          "K960694",
          "K201637",
          "K093075",
          "K910334",
          "K193601",
          "K211946",
          "K041599",
          "K964621",
          "K972519",
          "K942957",
          "K895737",
          "K924524",
          "K191064",
          "K131696",
          "K905605",
          "K092634",
          "K001029",
          "K010815",
          "K980758",
          "K933157",
          "K944045",
          "K093511",
          "K170990",
          "K033095",
          "K192545",
          "K972071",
          "K831508",
          "K942319",
          "K191601",
          "K955035",
          "K842596",
          "K993854",
          "K200598",
          "K230499",
          "K253484",
          "K002680",
          "K912614",
          "K991190",
          "K012033",
          "K780924",
          "K946104",
          "K240897",
          "K250048",
          "K993991",
          "K001253",
          "K790367",
          "K241055",
          "K900347",
          "K161825",
          "K241008",
          "K921770",
          "K153675",
          "K961505",
          "K926306",
          "K040989",
          "K981781",
          "K945995",
          "K994173",
          "K110824",
          "K973234",
          "K001273",
          "K082568",
          "K200169",
          "K122383",
          "K140926",
          "K830926",
          "K911377",
          "K954871",
          "K141507",
          "K222997",
          "K261202",
          "K181257",
          "K000961",
          "K964071",
          "K061635",
          "K170116",
          "K183594",
          "K002257",
          "K111784",
          "K122652",
          "K771331",
          "K992485",
          "K032047",
          "K212666",
          "K153265",
          "K822572",
          "K110462",
          "K033421",
          "K133726",
          "K820983",
          "K823491",
          "K011722",
          "K915608",
          "K023966",
          "K972562",
          "K203580",
          "K122797",
          "K252576",
          "K183564",
          "K904808",
          "K150029",
          "K980171",
          "K063630",
          "K982212",
          "K882742",
          "K122949",
          "K033003",
          "K081312",
          "K920166",
          "K182812",
          "K254290",
          "K974553",
          "K955020",
          "K250217",
          "K902307",
          "K042170",
          "K964849",
          "K861595",
          "K002021",
          "K020557",
          "K092879",
          "K010210",
          "K243067",
          "K861614",
          "K183611",
          "K180945",
          "K941160",
          "K850297",
          "K941337",
          "K143145",
          "K882049",
          "K905752",
          "K800317",
          "K994333",
          "K152286",
          "K834329",
          "K252845",
          "K162126",
          "K924028",
          "K982705",
          "K981614",
          "K902735",
          "K000348",
          "K022718",
          "K032452",
          "K092009",
          "K780298",
          "K082955",
          "K983782",
          "K952587",
          "K050161",
          "K193145",
          "K161994",
          "K051956",
          "K023765",
          "K233351",
          "K171094",
          "K083895",
          "K940750",
          "K211806",
          "K771751",
          "K223932",
          "K812644",
          "K230467",
          "K823184",
          "K072089",
          "K072126",
          "K812760",
          "K913623",
          "K082956",
          "K060232",
          "K945914",
          "K140521",
          "K790296",
          "K020047",
          "K083711",
          "K781148",
          "K003624",
          "K903006",
          "K013416",
          "K823083",
          "K251566",
          "K971540",
          "K952231",
          "K091733",
          "K935971",
          "K050550",
          "K973159",
          "K771329",
          "K963765",
          "K103665",
          "K081647",
          "K060183",
          "K182773",
          "K955623",
          "K931885",
          "K933699",
          "K923845",
          "K061417",
          "K932938",
          "K130760",
          "K193070",
          "K182365",
          "K081766",
          "K031926",
          "K010265",
          "K072865",
          "K061225",
          "K761231",
          "K070275",
          "K211652",
          "K925647",
          "K182451",
          "K931959",
          "K823689",
          "K954171",
          "K061835",
          "K971805",
          "K172368",
          "K190422",
          "K792343",
          "K933131",
          "K033888",
          "K030697",
          "K012986",
          "K123662",
          "K240070",
          "K960808",
          "K222425",
          "K945805",
          "K162469",
          "K925984",
          "K002422",
          "K042710",
          "K171717",
          "K922289",
          "K220563",
          "K150024",
          "K253915",
          "K001593",
          "K201221",
          "K990146",
          "K071859",
          "K170758",
          "K992460",
          "K070711",
          "K231327",
          "K030049",
          "K150915",
          "K963338",
          "K240311",
          "K023886",
          "K141225",
          "K922802",
          "K955836",
          "K021804",
          "K220996",
          "K192471",
          "K160539",
          "K253165",
          "K040240",
          "K013463",
          "K991669",
          "K234147",
          "K152031",
          "K972311",
          "K951328",
          "K954088",
          "K822062",
          "K212093",
          "K103375",
          "K160041",
          "K940909",
          "K861742",
          "K062171",
          "K080145",
          "K013873",
          "K052104",
          "K160053",
          "K020031",
          "K223303",
          "K250688",
          "K970838",
          "K191732",
          "K972497",
          "K952792",
          "K180189",
          "K061425",
          "K926242",
          "K944898",
          "K060170",
          "K982742",
          "K070162",
          "K160706",
          "K933728",
          "K031285",
          "K001262",
          "K001161",
          "K042098",
          "K210526",
          "K810101",
          "K152570",
          "K972350",
          "K140896",
          "K021117",
          "K900332",
          "K792600",
          "K254122",
          "K070958",
          "K093622",
          "K830613",
          "K030967",
          "K091674",
          "K910779",
          "K971677",
          "K983871",
          "K221522",
          "K881687",
          "K103032",
          "K190168",
          "K834434",
          "K242204",
          "K802827",
          "K760722",
          "K100711",
          "K041981",
          "K032721",
          "K202938",
          "K022296",
          "K955764",
          "K830277",
          "K191484",
          "K991218",
          "K031081",
          "K260255",
          "K102470",
          "K203183",
          "K963189",
          "K944706",
          "K191847",
          "K972267",
          "K182340",
          "K944578",
          "K984352",
          "K883774",
          "K112454",
          "K031247",
          "K010956",
          "K850129",
          "K032149",
          "K120476",
          "K981981",
          "K014224",
          "K942077",
          "K210338",
          "K063097",
          "K861495",
          "K946177",
          "K001723",
          "K251899",
          "K030819",
          "K143109",
          "K772199",
          "K172624",
          "K023482",
          "K023924",
          "K972558",
          "K802942",
          "K002752",
          "K894602",
          "K923416",
          "K120416",
          "K971532",
          "K992714",
          "K151273",
          "K023280",
          "K863412",
          "K770969",
          "K191542",
          "K983236",
          "K935490",
          "K031046",
          "K102368",
          "K913625",
          "K924918",
          "K963985",
          "K945531",
          "K942915",
          "K032969",
          "K072394",
          "K261179",
          "K120510",
          "K963783",
          "K951905",
          "K061975",
          "K240657",
          "K854592",
          "K012315",
          "K151649",
          "K120527",
          "K030720",
          "K111026",
          "K904506",
          "K922576",
          "K010010",
          "K023813",
          "K912637",
          "K955261",
          "K191065",
          "K241022",
          "K973158",
          "K071080",
          "K220384",
          "K050737",
          "K251334",
          "K980366",
          "K203013",
          "K191599",
          "K133180",
          "K965199",
          "K190909",
          "K093008",
          "K161305",
          "K972901",
          "K881413",
          "K101711",
          "K991140",
          "K052274",
          "K881460",
          "K970826",
          "K923563",
          "K111285",
          "K202962",
          "K160469",
          "K081954",
          "K963157",
          "K252150",
          "K954229",
          "K080093",
          "K050364",
          "K040763",
          "K242983",
          "K142759",
          "K153288",
          "K945975",
          "K053297",
          "K020408",
          "K955618",
          "K923928",
          "K831489",
          "K232903",
          "K070997",
          "K915300",
          "K000511",
          "K994336",
          "K162770",
          "K831299",
          "K142868",
          "K874779",
          "K063639",
          "K993655",
          "K020579",
          "K984018",
          "K170377",
          "K250400",
          "K973341",
          "K041347",
          "K120322",
          "K001151",
          "K192124",
          "K132703",
          "K250483",
          "K072937",
          "K080482",
          "K812637",
          "K102114",
          "K974521",
          "K863413",
          "K082584",
          "K031079",
          "K101990",
          "K833441",
          "K934281",
          "K072124",
          "K223039",
          "K133405",
          "K030142",
          "K944775",
          "K103508",
          "K003060",
          "K062525",
          "K243176",
          "K924188",
          "K940552",
          "K934784",
          "K234140",
          "K013668",
          "K945030",
          "K992149",
          "K020334",
          "K193356",
          "K912337",
          "K170054",
          "K781101",
          "K102372",
          "K222284",
          "K070374",
          "K093108",
          "K162676",
          "K150824",
          "K043402",
          "K014290",
          "K123310",
          "K974022",
          "K183538",
          "K253134",
          "K152702",
          "K070851",
          "K072849",
          "K083754",
          "K012631",
          "K844816",
          "K860236",
          "K213786",
          "K121150",
          "K202880",
          "K940955",
          "K924264",
          "K083415",
          "K052506",
          "K134006",
          "K232947",
          "K001438",
          "K143103",
          "K232805",
          "K140500",
          "K132136",
          "K853968",
          "K022475",
          "K982677",
          "K031950",
          "K172671",
          "K140384",
          "K103840",
          "K950507",
          "K000170",
          "K113265",
          "K910074",
          "K223110",
          "K982971",
          "K210942",
          "K180584",
          "K213386",
          "K200455",
          "K960569",
          "K183240",
          "K162802",
          "K232072",
          "K230837",
          "K101108",
          "K945688",
          "K780977",
          "K822069",
          "K781799",
          "K955001",
          "K080451",
          "K103775",
          "K163548",
          "K150112",
          "K950663",
          "K981219",
          "K970797",
          "K955109",
          "K971758",
          "K082962",
          "K991960",
          "K053433",
          "K214058",
          "K992218",
          "K952973",
          "K173582",
          "K031080",
          "K033981",
          "K192867",
          "K042274",
          "K201129",
          "K863701",
          "K942944",
          "K172012",
          "K820971",
          "K230547",
          "K991859",
          "K970837",
          "K170869",
          "K252412",
          "K973478",
          "K100686",
          "K232809",
          "K880064",
          "K832157",
          "K981908",
          "K960695",
          "K061526",
          "K861372",
          "K082505",
          "K982749",
          "K833449",
          "K061307",
          "K050831",
          "K230914",
          "K842622",
          "K993412",
          "K912968",
          "K193587",
          "K914193",
          "K960255",
          "K050361",
          "K063485",
          "K202114",
          "K132771",
          "K000245",
          "K052759",
          "K791137",
          "K253917",
          "K862753",
          "K950877",
          "K012695",
          "K813072",
          "K203795",
          "K970130",
          "K092761",
          "K982816",
          "K062082",
          "K982717",
          "K072261",
          "K932102",
          "K030551",
          "K191612",
          "K881403",
          "K082980",
          "K812702",
          "K121290",
          "K251254",
          "K170325",
          "K100353",
          "K133167",
          "K091259",
          "K935105",
          "K192027",
          "K132431",
          "K000720",
          "K200472",
          "K171869",
          "K952072",
          "K971829",
          "K885199",
          "K811820",
          "K160554",
          "K090306",
          "K242044",
          "K981870",
          "K936249",
          "K821505",
          "K193201",
          "K912656",
          "K860479",
          "K130669",
          "K143235",
          "K092487",
          "K193611",
          "K211562",
          "K131058",
          "K841591",
          "K061884",
          "K982176",
          "K210488",
          "K963772",
          "K172535",
          "K002877",
          "K851305",
          "K962119",
          "K141587",
          "K162034",
          "K963499",
          "K062441",
          "K931480",
          "K083452",
          "K002582",
          "K053282",
          "K854413",
          "K770646",
          "K241635",
          "K920217",
          "K192128",
          "K221873",
          "K051831",
          "K123172",
          "K023492",
          "K993647",
          "K240248",
          "K142929",
          "K926481",
          "K033027",
          "K122086",
          "K963180",
          "K822092",
          "K051710",
          "K092149",
          "K120909",
          "K935237",
          "K961132",
          "K060948",
          "K183526",
          "K970809",
          "K863731",
          "K112233",
          "K043291",
          "K000228",
          "K981389",
          "K243929",
          "K133338",
          "K161581",
          "K993944",
          "K233002",
          "K970184",
          "K140603",
          "K021077",
          "K892502",
          "K052893",
          "K163373",
          "K900953",
          "K822513",
          "K080715",
          "K011279",
          "K242370",
          "K792315",
          "K851040",
          "K243451",
          "K253634",
          "K140008",
          "K902232",
          "K132374",
          "K840434",
          "K021577",
          "K141484",
          "K110672",
          "K033257",
          "K925619",
          "K000501",
          "K973297",
          "K953195",
          "K030954",
          "K213042",
          "K010444",
          "K253405",
          "K051168",
          "K010025",
          "K802972",
          "K944602",
          "K101520",
          "K893659",
          "K253750",
          "K973295",
          "K203522",
          "K040279",
          "K941908",
          "K142208",
          "K982854",
          "K252367",
          "K260287",
          "K083608",
          "K041587",
          "K101274",
          "K132974",
          "K134036",
          "K211019",
          "K211173",
          "K971040",
          "K201738",
          "K803055",
          "K253803",
          "K040338",
          "K781949",
          "K953271",
          "K854864",
          "K852683",
          "K014260",
          "K944363",
          "K011220",
          "K061878",
          "K780206",
          "K202106",
          "K952577",
          "K222075",
          "K955501",
          "K940466",
          "K931982",
          "K122966",
          "K252215",
          "K903127",
          "K083461",
          "K984236",
          "K951035",
          "K953889",
          "K152057",
          "K914400",
          "K220970",
          "K031078",
          "K153194",
          "K201063",
          "K182355",
          "K252043",
          "K812769",
          "K861139",
          "K942080",
          "K962935",
          "K032702",
          "K161378",
          "K994075",
          "K912611",
          "K800597",
          "K123061",
          "K060255",
          "K063161",
          "K896626",
          "K153017",
          "K203277",
          "K143517",
          "K241621",
          "K965054",
          "K760168",
          "K134054",
          "K103017",
          "K242241",
          "K914132",
          "K791639",
          "K254220",
          "K905315",
          "K132099",
          "K002000",
          "K900954",
          "K883060",
          "K210315",
          "K981188",
          "K253777",
          "K014172",
          "K053326",
          "K933002",
          "K944244",
          "K941721",
          "K031082",
          "K030890",
          "K982748",
          "K130596",
          "K912597",
          "K992581",
          "K982130",
          "K141153",
          "K171359",
          "K903610",
          "K931028",
          "K965061",
          "K920185",
          "K231732",
          "K023812",
          "K032207",
          "K031133",
          "K162656",
          "K183445",
          "K955792",
          "K915731",
          "K945694",
          "K143662",
          "K911476",
          "K992693",
          "K922423",
          "K981430",
          "K970576",
          "K153055",
          "K931537",
          "K955168",
          "K230387",
          "K993112",
          "K180654",
          "K071728",
          "K822355",
          "K864402",
          "K170107",
          "K973682",
          "K170105",
          "K972415",
          "K812198",
          "K970140",
          "K160290",
          "K945765",
          "K193355",
          "K941968",
          "K955163",
          "K913281",
          "K915826",
          "K071976",
          "K123260",
          "K945254",
          "K120431",
          "K944030",
          "K840585",
          "K992382",
          "K223158",
          "K901145",
          "K052907",
          "K171262",
          "K201224",
          "K080282",
          "K992769",
          "K081503",
          "K151044",
          "K013139",
          "K881407",
          "K101111",
          "K091824",
          "K071709",
          "K961647",
          "K925850",
          "K910831",
          "K120116",
          "K965243",
          "K082643",
          "K091310",
          "K142248",
          "K183284",
          "K800557",
          "K201735",
          "K910978",
          "K033547",
          "K924410",
          "K051627",
          "K002981",
          "K182325",
          "K221989",
          "K190336",
          "K071343",
          "K992572",
          "K080197",
          "K010013",
          "K171825",
          "K001302",
          "K003471",
          "K230272",
          "K241832",
          "K021299",
          "K232301",
          "K993265",
          "K014207",
          "K970987",
          "K974484",
          "K030831",
          "K925707",
          "K102029",
          "K211107",
          "K000036",
          "K182354",
          "K910939",
          "K053447",
          "K160754",
          "K012625",
          "K972299",
          "K950932",
          "K172529",
          "K080557",
          "K951731",
          "K052203",
          "K981724",
          "K213696",
          "K252740",
          "K903724",
          "K033089",
          "K071786",
          "K061829",
          "K212107",
          "K052702",
          "K090393",
          "K962692",
          "K153727",
          "K970993",
          "K965007",
          "K925851",
          "K991529",
          "K162053",
          "K093086",
          "K132797",
          "K233374",
          "K023288",
          "K915204",
          "K033584",
          "K070443",
          "K180575",
          "K844565",
          "K945317",
          "K232223",
          "K161661",
          "K052035",
          "K110268",
          "K870981",
          "K974066",
          "K822405",
          "K023111",
          "K900767",
          "K970066",
          "K024138",
          "K053290",
          "K130793",
          "K221277",
          "K090580",
          "K223784",
          "K021329",
          "K223292",
          "K253918",
          "K950375",
          "K010487",
          "K033942",
          "K925130",
          "K961771",
          "K180031",
          "K840376",
          "K251235",
          "K823670",
          "K021079",
          "K202917",
          "K062337",
          "K240758",
          "K213354",
          "K791767",
          "K792044",
          "K945492",
          "K000032",
          "K852942",
          "K844403",
          "K973173",
          "K851487",
          "K021046",
          "K911787",
          "K232763",
          "K943055",
          "K152777",
          "K841604",
          "K090025",
          "K961197",
          "K220124",
          "K911274",
          "K052141",
          "K994365",
          "K945260",
          "K922090",
          "K142757",
          "K030362",
          "K252833",
          "K844428",
          "K991855",
          "K911316",
          "K021519",
          "K854109",
          "K180359",
          "K983743",
          "K040780",
          "K142975",
          "K951245",
          "K182605",
          "K121481",
          "K242630",
          "K162074",
          "K070446",
          "K201018",
          "K052859",
          "K031869",
          "K230650",
          "K070165",
          "K962593",
          "K992408",
          "K965086",
          "K111672",
          "K981570",
          "K030380",
          "K042608",
          "K062225",
          "K964175",
          "K213610",
          "K950188",
          "K171684",
          "K971446",
          "K980913",
          "K771330",
          "K984240",
          "K983986",
          "K882903",
          "K013034",
          "K942892",
          "K890648",
          "K871910",
          "K052686",
          "K250126",
          "K203461",
          "K153063",
          "K992805",
          "K012669",
          "K801158",
          "K953696",
          "K032327",
          "K041980",
          "K042216",
          "K260397",
          "K880863",
          "K911312",
          "K802330",
          "K012129",
          "K223856",
          "K944947",
          "K043273",
          "K240572",
          "K080187",
          "K902642",
          "K153679",
          "K032223",
          "K905654",
          "K951659",
          "K052003",
          "K842595",
          "K132199",
          "K042909",
          "K140255",
          "K033652",
          "K143475",
          "K983761",
          "K083281",
          "K831290",
          "K771007",
          "K904560",
          "K873591",
          "K840387",
          "K170188",
          "K041328",
          "K000250",
          "K231146",
          "K111412",
          "K982746",
          "K072177",
          "K163615",
          "K061258",
          "K841509",
          "K010112",
          "K946105",
          "K223135",
          "K992186",
          "K170582",
          "K172367",
          "K161038",
          "K914883",
          "K013255",
          "K791868",
          "K241535",
          "K001904",
          "K063172",
          "K963884",
          "K024286",
          "K190643",
          "K180908",
          "K003249",
          "K944617",
          "K925903",
          "K051214",
          "K051644",
          "K082054",
          "K183126",
          "K860687",
          "K140495",
          "K944823",
          "K211000",
          "K252704",
          "K141357",
          "K123478",
          "K221574",
          "K992764",
          "K113572",
          "K930243",
          "K173759",
          "K001407",
          "K201164",
          "K083162",
          "K220725",
          "K072699",
          "K082666",
          "K944784",
          "K955167",
          "K261833",
          "K780976",
          "K923199",
          "K952161",
          "K002002",
          "K031011",
          "K201066",
          "K173503",
          "K922002",
          "K243918",
          "K160206",
          "K991814",
          "K060419",
          "K102461",
          "K120095",
          "K141288",
          "K073498",
          "K993967",
          "K972072",
          "K023601",
          "K111732",
          "K201833",
          "K221074",
          "K790293",
          "K936246",
          "K881408",
          "K921271",
          "K230255",
          "K033880",
          "K202818",
          "K200841",
          "K081553",
          "K230194",
          "K791362",
          "K062712",
          "K911160",
          "K970039",
          "K050898",
          "K050460",
          "K850668",
          "K801153",
          "K180372",
          "K131012",
          "K083590",
          "K111670",
          "K071759",
          "K882646",
          "K212678",
          "K182801",
          "K022532",
          "K772235",
          "K231985",
          "K181085",
          "K934828",
          "K250200",
          "K983208",
          "K081678",
          "K885200",
          "K182497",
          "K080844",
          "K050579",
          "K812125",
          "K123353",
          "K171752",
          "K981361",
          "K024266",
          "K821263",
          "K003183",
          "K193292",
          "K192224",
          "K943818",
          "K031675",
          "K981310",
          "K010465",
          "K862221",
          "K091074",
          "K171593",
          "K150251",
          "K213612",
          "K904545",
          "K120537",
          "K894166",
          "K874798",
          "K965241",
          "K940936",
          "K021777",
          "K231790",
          "K954633",
          "K954965",
          "K841459",
          "K945758",
          "K926415",
          "K833364",
          "K192621",
          "K921696",
          "K003587",
          "K791224",
          "K984552",
          "K090309",
          "K192695",
          "K864771",
          "K052246",
          "K082323",
          "K853168",
          "K053128",
          "K080610",
          "K243217",
          "K191878",
          "K150993",
          "K053510",
          "K003676",
          "K121426",
          "K941426",
          "K021817",
          "K151480",
          "K820902",
          "K032062",
          "K121085",
          "K231615",
          "K231126",
          "K884641",
          "K043036",
          "K233123",
          "K864851",
          "K941897",
          "K840433",
          "K140459",
          "K821169",
          "K241157",
          "K000241",
          "K844371",
          "K952726",
          "K981416",
          "K062350",
          "K113500",
          "K163450",
          "K770647",
          "K973190",
          "K900714",
          "K070311",
          "K955144",
          "K251108",
          "K823905",
          "K081580",
          "K972441",
          "K142999",
          "K240932",
          "K810277",
          "K231212",
          "K160639",
          "K973701",
          "K831236",
          "K233147",
          "K102698",
          "K844780",
          "K253680",
          "K943231",
          "K132300",
          "K030473",
          "K142364",
          "K232290",
          "K820156",
          "K831291",
          "K193579",
          "K003126",
          "K961616",
          "K850943",
          "K240975",
          "K964636",
          "K830925",
          "K935222",
          "K100224",
          "K170170",
          "K173112",
          "K260466",
          "K122580",
          "K003050",
          "K180200",
          "K221945",
          "K780978",
          "K160050",
          "K945299",
          "K014201",
          "K981321",
          "K944530",
          "K922523",
          "K121977",
          "K070896",
          "K926414",
          "K110143",
          "K030339",
          "K012679",
          "K944265",
          "K761221",
          "K033499",
          "K841603",
          "K974464",
          "K192265",
          "K102983",
          "K233425",
          "K082109",
          "K171707",
          "K955352",
          "K180999",
          "K140658",
          "K093855",
          "K091440",
          "K120791",
          "K103707",
          "K990859",
          "K040855",
          "K072559",
          "K052778",
          "K233118",
          "K250275",
          "K243097",
          "K984185",
          "K123141",
          "K181389",
          "K150637",
          "K110695",
          "K883823",
          "K002898",
          "K200372",
          "K120454",
          "K883396",
          "K143654",
          "K790187",
          "K161134",
          "K190678",
          "K790770",
          "K193018",
          "K151743",
          "K060169",
          "K854814",
          "K183346",
          "K200639",
          "K012519",
          "K840375",
          "K990723",
          "K203494",
          "K201696",
          "K003674",
          "K833015",
          "K160128",
          "K101057",
          "K182772",
          "K012279",
          "K851304",
          "K954056",
          "K971728",
          "K031846",
          "K914252",
          "K003190",
          "K002269",
          "K162941",
          "K962044",
          "K182587",
          "K040104",
          "K123201",
          "K042604",
          "K030158",
          "K260636",
          "K032504",
          "K210693",
          "K031085",
          "K213344",
          "K221743",
          "K243737",
          "K900629",
          "K233493",
          "K954439",
          "K935815",
          "K053100",
          "K010710",
          "K191234",
          "K100669",
          "K212643",
          "K020832",
          "K771752",
          "K183180",
          "K151858",
          "K844372",
          "K070812",
          "K083013",
          "K942194",
          "K022856",
          "K232460",
          "K002987",
          "K913799",
          "K071963",
          "K080717",
          "K113597",
          "K963901",
          "K212561",
          "K142154",
          "K771440",
          "K960455",
          "K981589",
          "K936036",
          "K992180",
          "K943865",
          "K030108",
          "K961133",
          "K974170",
          "K203682",
          "K014251",
          "K910889",
          "K973618",
          "K003717",
          "K102752",
          "K051420",
          "K913108",
          "K213674",
          "K981220",
          "K001936",
          "K003897",
          "K863414",
          "K093203",
          "K172757",
          "K190075",
          "K021467",
          "K223805",
          "K230539",
          "K251435",
          "K953479",
          "K984239",
          "K062464",
          "K192542",
          "K072041",
          "K200875",
          "K171523",
          "K934075",
          "K222709",
          "K230968",
          "K001955",
          "K883762",
          "K973197",
          "K002841",
          "K031951",
          "K233099",
          "K180699",
          "K950953",
          "K771009",
          "K901302",
          "K021364",
          "K071103",
          "K954321",
          "K122200",
          "K142538",
          "K240313",
          "K950406",
          "K163415",
          "K000074",
          "K073495",
          "K924917",
          "K122278",
          "K252316",
          "K930497",
          "K880073",
          "K143559",
          "K013622",
          "K842574",
          "K222958",
          "K936304",
          "K843974",
          "K163442",
          "K896055",
          "K771352",
          "K951269",
          "K233642",
          "K190823",
          "K211170",
          "K925150",
          "K240780",
          "K192103",
          "K142480",
          "K150835",
          "K223488",
          "K860718",
          "K951985",
          "K100047",
          "K012725",
          "K210727",
          "K952432",
          "K072478",
          "K171066",
          "K013680",
          "K822072",
          "K210492",
          "K091223",
          "K242907",
          "K913514",
          "K073207",
          "K140839",
          "K100586",
          "K102921",
          "K973384",
          "K946237",
          "K962678",
          "K963646",
          "K932818",
          "K172023",
          "K023767",
          "K170777",
          "K992345",
          "K003462",
          "K242996",
          "K200146",
          "K940299",
          "K193591",
          "K020778",
          "K024304",
          "K031564",
          "K021852",
          "K023846",
          "K242469",
          "K061544",
          "K152219",
          "K100896",
          "K031257",
          "K243139",
          "K141858",
          "K231287",
          "K200501",
          "K896541",
          "K233615",
          "K251238",
          "K834048",
          "K162047",
          "K221671",
          "K192235",
          "K093165",
          "K870269",
          "K930097",
          "K132837",
          "K020711",
          "K151793",
          "K922425",
          "K954122",
          "K945893",
          "K924057",
          "K900633",
          "K081129",
          "K932518",
          "K855084",
          "K092732",
          "K142289",
          "K951230",
          "K000296",
          "K914928",
          "K140578",
          "K944370",
          "K250068",
          "K043008",
          "K200559",
          "K153499",
          "K243120",
          "K953992",
          "K915788",
          "K971450",
          "K123366",
          "K200253",
          "K100333",
          "K922509",
          "K962030",
          "K933332",
          "K061373",
          "K013904",
          "K212891",
          "K802605",
          "K922062",
          "K213317",
          "K182013",
          "K914066",
          "K944642",
          "K830545",
          "K760359",
          "K242773",
          "K220071",
          "K960296",
          "K961264",
          "K982884",
          "K100390",
          "K781810",
          "K211172",
          "K100638",
          "K981672",
          "K020574",
          "K972008",
          "K202988",
          "K151222",
          "K012488",
          "K971341",
          "K982309",
          "K771639",
          "K203745",
          "K023733",
          "K151325",
          "K221815",
          "K020067",
          "K163106",
          "K945898",
          "K073360",
          "K141790",
          "K935874",
          "K973110",
          "K120581",
          "K233185",
          "K173281",
          "K991291",
          "K091107",
          "K780975",
          "K113380",
          "K230593",
          "K251813",
          "K983116",
          "K902996",
          "K023413",
          "K800617",
          "K210084",
          "K092383",
          "K030996",
          "K023493",
          "K152703",
          "K161949",
          "K170381",
          "K852935",
          "K953159",
          "K222032",
          "K820494",
          "K951353",
          "K040638",
          "K820713",
          "K020135",
          "K994319",
          "K252487",
          "K924283",
          "K943014",
          "K953007",
          "K052297",
          "K881406",
          "K240964",
          "K981916",
          "K133017",
          "K771332",
          "K990362",
          "K864675",
          "K101967",
          "K100641",
          "K223033",
          "K973648",
          "K934985",
          "K964143",
          "K251996",
          "K073466",
          "K912780",
          "K230498",
          "K101534",
          "K200185",
          "K953446",
          "K791638",
          "K160160",
          "K792150",
          "K142229",
          "K150959",
          "K992996",
          "K251327",
          "K242688",
          "K232703",
          "K181249",
          "K161481",
          "K232992",
          "K946246",
          "K040219",
          "K072911",
          "K893732",
          "K100543",
          "K233849",
          "K952524",
          "K981919",
          "K900533",
          "K930666",
          "K871077",
          "K131949",
          "K961334",
          "K823923",
          "K111523",
          "K183148",
          "K102709",
          "K934427",
          "K031371",
          "K063480",
          "K220481",
          "K953738",
          "K240999",
          "K170358",
          "K231749",
          "K963088",
          "K925830",
          "K081791",
          "K172610",
          "K934800",
          "K212222",
          "K240857",
          "K151229",
          "K113090",
          "K091672",
          "K801323",
          "K142410",
          "K231777",
          "K790642",
          "K982826",
          "K863360",
          "K131758",
          "K963123",
          "K934035",
          "K834410",
          "K961473",
          "K130110",
          "K951589",
          "K060400",
          "K040785",
          "K945380",
          "K971711",
          "K970493",
          "K942489",
          "K140422",
          "K926103",
          "K082202",
          "K012314",
          "K200947",
          "K884656",
          "K884836",
          "K193004",
          "K233346",
          "K032838",
          "K041285",
          "K241085",
          "K895252",
          "K093695",
          "K050671",
          "K012799",
          "K111887",
          "K172580",
          "K223222",
          "K241138",
          "K954851",
          "K213814",
          "K061660",
          "K041868",
          "K852355",
          "K094041",
          "K830532",
          "K041999",
          "K111986",
          "K051429",
          "K944681",
          "K082834",
          "K150161",
          "K942579",
          "K915296",
          "K052009",
          "K110439",
          "K895841",
          "K131205",
          "K202006",
          "K063195",
          "K945757",
          "K913473",
          "K960009",
          "K952110",
          "K213831",
          "K082197",
          "K231012",
          "K201520",
          "K972161",
          "K073111",
          "K061259",
          "K984417",
          "K052796",
          "K141904",
          "K972628",
          "K093025",
          "K010192",
          "K202558",
          "K853972",
          "K810273",
          "K033748",
          "K181852",
          "K791230",
          "K250327",
          "K982984",
          "K945717",
          "K201884",
          "K160693",
          "K930355",
          "K945760",
          "K131435",
          "K000778",
          "K902035",
          "K242266",
          "K961097",
          "K160331",
          "K942821",
          "K924881",
          "K233644",
          "K973346",
          "K964779",
          "K251988",
          "K930528",
          "K060484",
          "K861036",
          "K001986",
          "K012004",
          "K162075",
          "K172805",
          "K082902",
          "K924960",
          "K152059",
          "K041810",
          "K946377",
          "K953737",
          "K143161",
          "K012018",
          "K123429",
          "K192715",
          "K935091",
          "K020622",
          "K073616",
          "K020585",
          "K202562",
          "K861112",
          "K161307",
          "K223262",
          "K000944",
          "K900949",
          "K912229",
          "K952832",
          "K941435",
          "K073687",
          "K923332",
          "K062257",
          "K972358",
          "K012700",
          "K781511",
          "K953081",
          "K170161",
          "K063538",
          "K073046",
          "K220505",
          "K201212",
          "K161558",
          "K163190",
          "K170610",
          "K874825",
          "K020186",
          "K812722",
          "K910663",
          "K971092",
          "K936075",
          "K000044",
          "K130983",
          "K023566",
          "K132743",
          "K240782",
          "K944817",
          "K760675",
          "K193019",
          "K090586",
          "K201685",
          "K911491",
          "K993055",
          "K173889",
          "K955834",
          "K042077",
          "K940219",
          "K932654"
        ],
        "registration_number": [
          "",
          "3013886523",
          "3010432890",
          "9610617",
          "3005382983",
          "3010594949",
          "3029895042",
          "3007456622",
          "3004777218",
          "3004837686",
          "3016675013",
          "1225492",
          "3005528784",
          "1058584",
          "3014663814",
          "2029275",
          "1222993",
          "3006990600",
          "3017820152",
          "3009905888",
          "3015183635",
          "3008439467",
          "3011586151",
          "3014193078",
          "3007048256",
          "8043396",
          "3005110411",
          "3009423112",
          "3003678543",
          "3012497308",
          "2124215",
          "3013789508",
          "3013466775",
          "1047843",
          "3010402752",
          "3014328942",
          "3010701628",
          "3044777128",
          "3007648309",
          "2133928",
          "3008791915",
          "3009881188",
          "3009736547",
          "3013652305",
          "3016678045",
          "3015540957",
          "2648623",
          "3016420624",
          "2242450",
          "3007593944",
          "9611112",
          "3002808227",
          "3006696607",
          "3009166798",
          "3012204368",
          "3009212837",
          "3011706110",
          "3012447612",
          "3009513193",
          "2183870",
          "9710058",
          "3003687489",
          "1643264",
          "3006059917",
          "2011171",
          "3008753275",
          "3010164891",
          "3019269298",
          "1721686",
          "1222313",
          "3013438730",
          "3012627460",
          "3009762450",
          "3004519921",
          "3013820501",
          "3043226252",
          "3012639025",
          "3015192261",
          "3012358417",
          "1017294",
          "3045816907",
          "3007499195",
          "9611102",
          "3002579136",
          "3004365956",
          "3015988948",
          "3027713941",
          "3013421458",
          "3004044817",
          "3015539532",
          "1926681",
          "3006247712",
          "2184009",
          "3013505983",
          "3008998256",
          "3009350690",
          "3006142527",
          "9616088",
          "3008868758",
          "8010257",
          "3007616088",
          "2024024",
          "3004167969",
          "3006636961",
          "3013376016",
          "3004608878",
          "3013011598",
          "3004491689",
          "3030495752",
          "9611665",
          "3016659819",
          "3007589150",
          "3012582478",
          "3012329552",
          "3002808022",
          "8010456",
          "3027339877",
          "3008729892",
          "3004940677",
          "3016438694",
          "3012421607",
          "3001124136",
          "3006546082",
          "1722040",
          "3002907620",
          "3006524618",
          "3007645866",
          "2954354",
          "9613793",
          "3013563737",
          "3005646850",
          "8043398",
          "3008132398",
          "3014334038",
          "1836161",
          "3043527620",
          "1066427",
          "3014004349",
          "3005677568",
          "1424263",
          "2182269",
          "3022626486",
          "3007593903",
          "3008770252",
          "2020394",
          "2126666",
          "3009732568",
          "3018242561",
          "3007113251",
          "1650907",
          "3004060780",
          "3003767927",
          "8010177",
          "3011120183",
          "3006266622",
          "3030447506",
          "3021008900",
          "2183446",
          "3007867778",
          "9611516",
          "3005064943",
          "2183613",
          "3014656892",
          "1032347",
          "1055236",
          "1824199",
          "1420032",
          "3006807694",
          "3024712917",
          "3015515531",
          "9612086",
          "3007535734",
          "3009495876",
          "1643817",
          "3016904853",
          "1317547",
          "3027242374",
          "2511556",
          "3006849754",
          "3028977843",
          "3015974593",
          "3017185164",
          "3002250546",
          "8010372",
          "3006639944",
          "3000268902",
          "3006747129",
          "2244478",
          "3013176080",
          "3013111980",
          "2530138",
          "2916714",
          "3003174284",
          "3015967359",
          "3030446844",
          "3007699067",
          "3020704423",
          "3015460875",
          "9613563",
          "3026606306",
          "3011137372",
          "1216677",
          "2242056",
          "3005350457",
          "3005064998",
          "2953359",
          "3004601121",
          "2955842",
          "3010069773",
          "1222928",
          "3014615697",
          "3014279513",
          "3015219237",
          "3006183537",
          "3015453963",
          "3007829657",
          "9612278",
          "2182291",
          "1060680",
          "3010707607",
          "3045753434",
          "3001592626",
          "3007029079",
          "3014982808",
          "3044854930",
          "3013298431",
          "3009217531",
          "3003120897",
          "3003418325",
          "3042940623",
          "9611827",
          "1057358",
          "3014931612",
          "3011024991",
          "3007007790",
          "3004105270",
          "3003604053",
          "3003755939",
          "1037007",
          "1423662",
          "3013247477",
          "2027111",
          "3011559089",
          "3015976667",
          "3010684280",
          "3006950086",
          "3008114965",
          "3009763019",
          "1651252",
          "3013868968",
          "2087572",
          "3013602304",
          "9610849",
          "3008720584",
          "3017196117",
          "3015471899",
          "3012883081",
          "9616662",
          "3004215117",
          "2650143",
          "3003455467",
          "3011063460",
          "3047841690",
          "2242352",
          "2649614",
          "3009103001",
          "3009882464",
          "3012263546",
          "1220246",
          "8020045",
          "3007200016",
          "1526711",
          "3036425107",
          "3014980461",
          "2936485",
          "3012790575",
          "3019924",
          "3011277660",
          "3043585137",
          "2521453",
          "3005562917",
          "2032112",
          "3014939461",
          "2433012",
          "3012428212",
          "3045010542",
          "3000236920",
          "3030645016",
          "3008595179",
          "1221541",
          "9616684",
          "1048735",
          "2133936",
          "1056128",
          "9616666",
          "3015309643",
          "3014461810",
          "3030176940",
          "3008933393",
          "9611612",
          "3020142587",
          "1064858",
          "3013596024",
          "3010288205",
          "3009609358",
          "3004748528",
          "3027454635",
          "1220477",
          "1018470",
          "3005467144",
          "3016099036",
          "3011423170",
          "3010353867",
          "3008309062",
          "2212018",
          "3012453745",
          "3004122598",
          "2134265",
          "1421101",
          "3007024186",
          "3010739934",
          "1219930",
          "3008040402",
          "3013665681",
          "9710582",
          "3005369776",
          "3030733800",
          "3001003398",
          "3010687973",
          "3007842664",
          "3009756327",
          "3012494290",
          "3006561161",
          "8043746",
          "3027645761",
          "9616696",
          "3020968442",
          "3038988287",
          "9680721",
          "3009241989",
          "3011221476",
          "3001614464",
          "3001431138",
          "3032391",
          "3016087306",
          "3010363436",
          "3020163307",
          "3011683674",
          "3031240334",
          "1721504",
          "3022320321",
          "3013505638",
          "2649622",
          "3015895045",
          "3042953831",
          "2133772",
          "3012940012",
          "3010305275",
          "2133810",
          "3025268689",
          "3015972753",
          "3011986317",
          "3009420598",
          "8010273",
          "3015449142",
          "3030926959",
          "2085947",
          "9680260",
          "2183715",
          "3015173212",
          "9680340",
          "1062741",
          "3015506965",
          "9614279",
          "3010653929",
          "1820334",
          "3008837339",
          "8030607",
          "3007960282",
          "3029754081",
          "3007353760",
          "2030598",
          "1721251",
          "9614969",
          "1526611",
          "3030153385",
          "1018233",
          "3008805497",
          "3035374386",
          "3004483538",
          "3016727307",
          "3016050940",
          "3010601992",
          "3027988352",
          "3033844600",
          "1030451",
          "1930126",
          "3006555617",
          "3009206941",
          "2246552",
          "9617070",
          "3011782009",
          "1423714",
          "1219602",
          "3008816935",
          "2022435",
          "1928237",
          "3042351542",
          "3012460907",
          "3015612252",
          "3017892510",
          "3010173425",
          "3004754983",
          "3004361445",
          "3004878380",
          "3009555600",
          "9710649",
          "3005695724",
          "3040993693",
          "3002806603",
          "8010591",
          "2245590",
          "3004784537",
          "3015241782",
          "3015779311",
          "9681260",
          "1221826",
          "3007185043",
          "3012236936",
          "3017262662",
          "3015177648",
          "3014579161",
          "3007728276",
          "3003500271",
          "3001675293",
          "3010836847",
          "3015151147",
          "3004638532",
          "3007712511",
          "3042272797",
          "3026612481",
          "2029015",
          "3015224777",
          "3005099803",
          "3011241266",
          "3002807315",
          "3006897899",
          "3039167637",
          "9680357",
          "3007675948",
          "1417592",
          "3013783488",
          "3014479313",
          "3035387162",
          "9610877",
          "3007014520",
          "9680515",
          "3010361165",
          "3009973505",
          "3043532708",
          "3016746323",
          "8040233",
          "3016496662",
          "3020468145",
          "3006411077",
          "3003955307",
          "2031966",
          "1828464",
          "3010157088",
          "9680660",
          "3016096100",
          "3010601984",
          "1825034",
          "3010202439",
          "3006985208",
          "1222780",
          "3007305485",
          "3010383847",
          "3006498370",
          "9615857",
          "3033511105",
          "3009822485",
          "3004636440",
          "3010155648",
          "3010511300",
          "3015531529",
          "3008849319",
          "3004681601",
          "9680837",
          "3012783431",
          "3003008746",
          "1721194",
          "2084346",
          "3010493171",
          "3011625895",
          "1000121056",
          "3016020790",
          "3012470322",
          "3012259264",
          "3021192541",
          "3003898360",
          "2182799",
          "3011989923",
          "9710524",
          "3012606445",
          "3013500228",
          "3005083075",
          "1219619",
          "1319660",
          "3031944951",
          "9614641",
          "2523676",
          "3021226419",
          "3008747667",
          "3007185216",
          "3003790304",
          "3010326740",
          "3015231789",
          "9610773",
          "3006554912",
          "9610614",
          "3015136673",
          "3015895145",
          "3005061536",
          "9681129",
          "3010041511",
          "3010280625",
          "1450662",
          "1061927",
          "1721983",
          "3014023545",
          "9680952",
          "1030489",
          "2020550",
          "3005941719",
          "3027339887",
          "3007544634",
          "3017202975",
          "3014579024",
          "3030839824",
          "9616675",
          "3009051888",
          "3004598644",
          "1526534",
          "1221934",
          "3010805941",
          "3009719225",
          "3017210488",
          "2182536",
          "3040004547",
          "3009903460",
          "3005770916",
          "3007031059",
          "1000523114",
          "3015489752",
          "1422634",
          "3011879048",
          "1519132",
          "8030967",
          "3011270181",
          "3005334138",
          "1423537",
          "3013858269",
          "3010380518",
          "3005857264",
          "1825527",
          "8010312",
          "3008571754",
          "3010131137",
          "3016832577",
          "3007047889",
          "3010188783",
          "3015775837",
          "3014635169",
          "3005809810",
          "1061124",
          "9611503",
          "3009486890",
          "1223925",
          "1124841",
          "3031571464",
          "3017450375",
          "3004903538",
          "1211998",
          "3031916226",
          "1223422",
          "2182318",
          "3013188547",
          "3011230048",
          "1718850",
          "2248148",
          "2032098",
          "3007766590",
          "1051614",
          "3011989678",
          "3012995405",
          "1717344",
          "3016245399",
          "3033715699",
          "3024021261",
          "8044057",
          "3015225571",
          "3020623326",
          "3033524168",
          "2183744",
          "3019751610",
          "1221485",
          "3002807090",
          "3035642068",
          "3011307265",
          "8044091",
          "3006451981",
          "3014925413",
          "3014891416",
          "3017619652",
          "2219920",
          "3007515069",
          "3033536319",
          "3039357913",
          "1054241",
          "1123137",
          "3017448139",
          "1056553",
          "3006404071",
          "3005012805",
          "8010047",
          "3010885402",
          "2529846",
          "3009380063",
          "3012616488",
          "3016248634",
          "3015153469",
          "3020331900",
          "2027460",
          "3006789154",
          "1222274",
          "1222140",
          "1219659",
          "3005595283",
          "3008911114",
          "3005173255",
          "3010806467",
          "3008361498",
          "2134285",
          "3006985163",
          "3036773433",
          "3013551587",
          "3007487007",
          "1650927",
          "3013557681",
          "3023155873",
          "1825146",
          "3008864603",
          "3030412764",
          "3015806723",
          "2183319",
          "9610612",
          "3010226575",
          "1828288",
          "3003072803",
          "3015497424",
          "3016607092",
          "3008915316",
          "3017406274",
          "3015257340",
          "2150060",
          "3009183442",
          "3011050570",
          "1720747",
          "2518897",
          "3008215723",
          "3013700547",
          "3015542154",
          "2028523",
          "1828132",
          "3021355451",
          "3011743372",
          "3014421917",
          "3013440400",
          "3007137643",
          "3002807314",
          "8043792",
          "3014769442",
          "3007421149",
          "2647580",
          "3027267119",
          "3010230734",
          "1418975",
          "1929938",
          "3008843439",
          "3025606176",
          "3020904959",
          "3030230686",
          "3041276883",
          "1526012",
          "2424366",
          "2024311",
          "3009746061",
          "3010155661",
          "3016851379",
          "1718873",
          "9680001",
          "1720929",
          "3018094310",
          "3008824097",
          "1824037",
          "3015286855",
          "2938071",
          "3021393544",
          "3006891611",
          "1056350",
          "3004111573",
          "3004911384",
          "3009941480",
          "3044464950",
          "3003724334",
          "3007791595",
          "3010187826",
          "3011944170",
          "3036803067",
          "1423395",
          "3007738736",
          "1527736",
          "3005075853",
          "3014342096",
          "2031093",
          "1649518",
          "3023146633",
          "3013734919",
          "3041918479",
          "9612074",
          "3010126772",
          "3008102042",
          "3029916132",
          "2183785",
          "3008693678",
          "3013917867",
          "3035482729",
          "2435119",
          "1724474",
          "3016555232",
          "3008264458",
          "3010291427",
          "1320894",
          "3021519357",
          "3007583504",
          "3015227849",
          "3040335671",
          "3009599228",
          "3016813690",
          "3008573567",
          "3031841433",
          "3011070163",
          "8010877",
          "3038675429",
          "3003935342",
          "3016076698",
          "2133786",
          "3010816547",
          "3013517171",
          "1054811",
          "3012757453",
          "8040510",
          "1225798",
          "2434839",
          "1000122786",
          "3021632375",
          "1313525",
          "3004904811",
          "1063285",
          "2954783",
          "1528319",
          "3013686892",
          "3010621964",
          "3010377594",
          "3002808148",
          "3013069546",
          "3003537036",
          "8043548",
          "3013302242",
          "1043214",
          "3038715747",
          "8010871",
          "1721676",
          "3011987935",
          "3004464325",
          "9680718",
          "8040537",
          "8010890",
          "3030977796",
          "9614434",
          "3010331645",
          "3016168848",
          "3004530184",
          "3007748817",
          "3014848720",
          "9681191",
          "3013557562",
          "3014262693",
          "1923569",
          "3032109",
          "1412854",
          "9681383",
          "3006638824",
          "3044156918",
          "3013283620",
          "3010131466",
          "3007710878",
          "2245304",
          "3013077751",
          "3023706404",
          "3010356635",
          "3014122529",
          "3004605321",
          "1932180",
          "8010298",
          "3010400367",
          "3008870565",
          "3008808042",
          "3008786200",
          "3010952064",
          "3033589330",
          "3012510750",
          "9617429",
          "3003782610",
          "3008889904",
          "3012536737",
          "3019569846",
          "1723170",
          "3017405993",
          "2648666",
          "1421879",
          "3006716777",
          "3008085016",
          "3010132111",
          "3015433394",
          "3018126215",
          "2245574",
          "3003542872",
          "3012322232",
          "3012818475",
          "3021477808",
          "1319639",
          "2531415",
          "3013875599",
          "3012507533",
          "3035366890",
          "3010273872",
          "3027669735",
          "3007748793",
          "3013944123",
          "2320762",
          "3008929787",
          "1418479",
          "3006017180",
          "3030626857",
          "1218017",
          "2528981",
          "3002807310",
          "3015207155",
          "2431166",
          "8010418",
          "3020899789",
          "3006891627",
          "3022327202"
        ],
        "fei_number": [
          "",
          "3013886523",
          "3010432890",
          "3005382983",
          "3010594949",
          "1000138054",
          "3007513035",
          "3029895042",
          "3002902735",
          "3005747797",
          "3007456622",
          "3004837686",
          "3003221505",
          "3016675013",
          "3004777218",
          "3005528784",
          "3014663814",
          "3006990600",
          "3017820152",
          "3009905888",
          "3015183635",
          "3008439467",
          "3011586151",
          "3014193078",
          "3007048256",
          "3005110411",
          "3009423112",
          "3003678543",
          "3012497308",
          "2124215",
          "3013789508",
          "3013466775",
          "1047843",
          "3010402752",
          "3014328942",
          "3010701628",
          "3044777128",
          "3007648309",
          "3008791915",
          "3009881188",
          "3009736547",
          "3013652305",
          "3016678045",
          "3015540957",
          "2648623",
          "3016420624",
          "2242450",
          "3007593944",
          "3002695476",
          "3002808227",
          "3006696607",
          "3009166798",
          "3012204368",
          "3009212837",
          "3011706110",
          "3002807111",
          "3002821504",
          "3012447612",
          "3009513193",
          "2183870",
          "3003687489",
          "1643264",
          "1000137656",
          "3006059917",
          "2011171",
          "3008753275",
          "3010164891",
          "2523865",
          "3019269298",
          "1721686",
          "3013438730",
          "3004857377",
          "3012627460",
          "3009762450",
          "3013820501",
          "3001236549",
          "3043226252",
          "3012639025",
          "3015192261",
          "3012358417",
          "1017294",
          "3045816907",
          "3007499195",
          "3002579136",
          "3004365956",
          "3015988948",
          "3027713941",
          "3013421458",
          "3004044817",
          "3015539532",
          "3002806565",
          "1926681",
          "1000305720",
          "3006247712",
          "3006142527",
          "3013505983",
          "3008998256",
          "3009350690",
          "3008868758",
          "3002808273",
          "3007616088",
          "2024024",
          "3004167969",
          "3006636961",
          "3013376016",
          "3003944071",
          "3004608878",
          "3013011598",
          "3004491689",
          "3030495752",
          "3001595430",
          "3002681132",
          "3007130719",
          "3016659819",
          "3007589150",
          "3012582478",
          "3008207730",
          "1000434163",
          "3012329552",
          "3002808022",
          "3027339877",
          "1000121057",
          "3003565435",
          "3002808268",
          "3016438694",
          "3012421607",
          "3001124136",
          "3002806863",
          "3006546082",
          "1000114185",
          "3007475226",
          "3002907620",
          "2024311",
          "3007645866",
          "3013563737",
          "3009438986",
          "3005646850",
          "3002900199",
          "3014334038",
          "3043527620",
          "3014004349",
          "3005677568",
          "1000517638",
          "3006003617",
          "1000512168",
          "3010666513",
          "3022626486",
          "2182269",
          "3007593903",
          "3008770252",
          "2020394",
          "2126666",
          "3009732568",
          "3018242561",
          "3007113251",
          "3004060780",
          "3003767927",
          "3011120183",
          "3006266622",
          "3030447506",
          "3021008900",
          "2183446",
          "3000335778",
          "3007867778",
          "3003186357",
          "1000221205",
          "3005064943",
          "2183613",
          "3014656892",
          "1032347",
          "1824199",
          "1420032",
          "3000204692",
          "3006807694",
          "3024712917",
          "3015515531",
          "3007535734",
          "3009495876",
          "1643817",
          "3016904853",
          "1317547",
          "3027242374",
          "2511556",
          "3006849754",
          "3028977843",
          "3015974593",
          "3017185164",
          "3002250546",
          "1000122656",
          "3006469671",
          "3006639944",
          "3000268902",
          "3006747129",
          "1000403400",
          "1000244733",
          "1000120743",
          "3013176080",
          "3013111980",
          "2916714",
          "3002807038",
          "3003174284",
          "3015967359",
          "3007699067",
          "3030446844",
          "3007573441",
          "3020704423",
          "3015460875",
          "1216677",
          "3026606306",
          "2242056",
          "3005350457",
          "3005064998",
          "3004601121",
          "3002807555",
          "3002808001",
          "3010069773",
          "3014615697",
          "3014279513",
          "3015219237",
          "3006183537",
          "3002807795",
          "3015453963",
          "3007829657",
          "2182291",
          "3005649458",
          "1000219637",
          "3010707607",
          "3045753434",
          "3001592626",
          "3014982808",
          "3008185830",
          "3044854930",
          "3007029079",
          "3001236905",
          "3000204839",
          "3013298431",
          "3000140642",
          "1000120578",
          "3009217531",
          "3003120897",
          "3003418325",
          "3000717902",
          "3042940623",
          "3014931612",
          "3011024991",
          "3007007790",
          "3002477156",
          "3004105270",
          "3003604053",
          "3003755939",
          "1037007",
          "3013247477",
          "4498",
          "2027111",
          "1000136191",
          "1000514656",
          "3011559089",
          "3015976667",
          "3010684280",
          "3006950086",
          "3008114965",
          "3009763019",
          "3013868968",
          "3013602304",
          "3008720584",
          "3017196117",
          "3015471899",
          "3012883081",
          "2242352",
          "3004215117",
          "3003455467",
          "3011063460",
          "1000121130",
          "3047841690",
          "3942",
          "1000080301",
          "2649614",
          "3009103001",
          "3009882464",
          "3012263546",
          "1220246",
          "3002807847",
          "3007200016",
          "1526711",
          "3005461046",
          "3036425107",
          "3014980461",
          "2936485",
          "3012790575",
          "3019924",
          "3011277660",
          "3000302153",
          "3002806920",
          "3043585137",
          "2521453",
          "3005562917",
          "3014939461",
          "2433012",
          "3012428212",
          "3008397795",
          "3045010542",
          "3030645016",
          "3008595179",
          "1221541",
          "3008782585",
          "1048735",
          "3003499455",
          "3015309643",
          "3001319476",
          "3014461810",
          "3030176940",
          "3008933393",
          "3020142587",
          "3013596024",
          "3010288205",
          "3009609358",
          "3004748528",
          "3002807798",
          "3027454635",
          "1220477",
          "1018470",
          "1000518731",
          "3005467144",
          "3016099036",
          "3010353867",
          "3008309062",
          "2212018",
          "3012453745",
          "3004122598",
          "1421101",
          "1000136478",
          "3007024186",
          "3010739934",
          "1219930",
          "3008040402",
          "3013665681",
          "1000160458",
          "3002808269",
          "3005369776",
          "3030733800",
          "3001003398",
          "3010687973",
          "3007842664",
          "3012494290",
          "3006561161",
          "3027645761",
          "3020968442",
          "3038988287",
          "3011221476",
          "3009241989",
          "3001614464",
          "3002806862",
          "3001431138",
          "3010813691",
          "3016087306",
          "3010363436",
          "3020163307",
          "3011683674",
          "3031240334",
          "1721504",
          "3022320321",
          "3013505638",
          "2649622",
          "3002287318",
          "3015895045",
          "3042953831",
          "3012940012",
          "1000132370",
          "3010305275",
          "1000119305",
          "3025268689",
          "3002807584",
          "3015972753",
          "3011986317",
          "3009420598",
          "3015449142",
          "3000215273",
          "3030926959",
          "2183715",
          "3015173212",
          "3004829691",
          "3015506965",
          "3010653929",
          "3002806575",
          "1820334",
          "3008837339",
          "3007960282",
          "3003995201",
          "3029754081",
          "3007353760",
          "3002980729",
          "1721251",
          "1526611",
          "3030153385",
          "1018233",
          "3008805497",
          "3035374386",
          "3004483538",
          "3016727307",
          "3016050940",
          "3010601992",
          "3027988352",
          "3033844600",
          "1030451",
          "1930126",
          "3006555617",
          "3009206941",
          "2246552",
          "1219602",
          "2022435",
          "3003015552",
          "1928237",
          "3012460907",
          "3042351542",
          "3015612252",
          "3017892510",
          "3004754983",
          "3004361445",
          "3004878380",
          "3009555600",
          "3005695724",
          "3040993693",
          "1000123180",
          "3002806603",
          "2245590",
          "3004784537",
          "3002808445",
          "3015241782",
          "3015779311",
          "1221826",
          "3007185043",
          "3012236936",
          "3017262662",
          "3015177648",
          "3002409980",
          "3007264034",
          "1000132435",
          "3014579161",
          "3007728276",
          "3001451571",
          "3003500271",
          "3001675293",
          "3010836847",
          "3015151147",
          "3004638532",
          "3007712511",
          "3042272797",
          "3026612481",
          "3015224777",
          "3004106598",
          "3005099803",
          "3011241266",
          "3002807315",
          "3006897899",
          "3039167637",
          "1000208624",
          "3007675948",
          "1417592",
          "3013783488",
          "3014479313",
          "3035387162",
          "3007014520",
          "3003630373",
          "3010361165",
          "3009973505",
          "1000499348",
          "3043532708",
          "3016746323",
          "3002807391",
          "3016496662",
          "3020468145",
          "3006411077",
          "3003955307",
          "1828464",
          "3010157088",
          "3016096100",
          "3010601984",
          "1825034",
          "1000120613",
          "3006985208",
          "3007305485",
          "3010383847",
          "3006498370",
          "3002807036",
          "3033511105",
          "1000064530",
          "3009822485",
          "3004636440",
          "3010155648",
          "3010511300",
          "3015531529",
          "3008849319",
          "3004681601",
          "3012783431",
          "3000280067",
          "3008906816",
          "3003008746",
          "1721194",
          "1000222372",
          "3010493171",
          "3011625895",
          "1000121056",
          "3016020790",
          "3012470322",
          "1000117361",
          "3012259264",
          "3021192541",
          "3003898360",
          "2182799",
          "3011989923",
          "3006509795",
          "1000124817",
          "3012606445",
          "3013500228",
          "3005083075",
          "1219619",
          "3003105967",
          "3031944951",
          "2523676",
          "3021226419",
          "3008747667",
          "3007185216",
          "3000209794",
          "3003790304",
          "3010326740",
          "3015231789",
          "3006554912",
          "3015136673",
          "3015895145",
          "3005061536",
          "3010041511",
          "3010280625",
          "1450662",
          "3003058448",
          "3014023545",
          "2020550",
          "3027339887",
          "3007544634",
          "3017202975",
          "3014579024",
          "3030839824",
          "3009051888",
          "3004598644",
          "1526534",
          "3002808325",
          "1221934",
          "3010805941",
          "3009719225",
          "3017210488",
          "2182536",
          "3040004547",
          "3009903460",
          "3005770916",
          "3007031059",
          "1000523114",
          "3015489752",
          "3001452522",
          "3011879048",
          "1519132",
          "3011270181",
          "3005334138",
          "3003445992",
          "3002095335",
          "3013858269",
          "3010380518",
          "3005857264",
          "3002913962",
          "1825527",
          "3008571754",
          "3002807091",
          "3010131137",
          "1000125147",
          "3016832577",
          "3007047889",
          "3010188783",
          "3015775837",
          "3014635169",
          "3003164255",
          "3005809810",
          "3009486890",
          "3031571464",
          "3017450375",
          "3004903538",
          "3002770040",
          "1211998",
          "1000123848",
          "3000717925",
          "3031916226",
          "2182318",
          "3013188547",
          "3011230048",
          "1718850",
          "3007290840",
          "1000140363",
          "3007766590",
          "1051614",
          "3011989678",
          "6736",
          "3002808173",
          "1717344",
          "3012995405",
          "3016245399",
          "3008108972",
          "3033715699",
          "3024021261",
          "3015225571",
          "3020623326",
          "3033524168",
          "2183744",
          "3003767881",
          "3019751610",
          "1221485",
          "3035642068",
          "3011307265",
          "3006451981",
          "3014925413",
          "3014891416",
          "3017619652",
          "2219920",
          "3007515069",
          "3033536319",
          "3039357913",
          "3006404071",
          "3017448139",
          "3005012805",
          "3010885402",
          "3009380063",
          "3012616488",
          "3016248634",
          "3015153469",
          "3020331900",
          "2027460",
          "3006789154",
          "1219659",
          "3005595283",
          "1000115048",
          "1000116158",
          "3008911114",
          "3005173255",
          "3010806467",
          "3008361498",
          "3002315844",
          "3006985163",
          "3036773433",
          "3013551587",
          "3007487007",
          "3013557681",
          "3023155873",
          "1825146",
          "3008864603",
          "3030412764",
          "3015806723",
          "2183319",
          "3000719482",
          "3010226575",
          "1828288",
          "3008237808",
          "3003072803",
          "3015497424",
          "3016607092",
          "3008915316",
          "3017406274",
          "3015257340",
          "3003040625",
          "2648209",
          "2150060",
          "3009183442",
          "3011050570",
          "1720747",
          "3008215723",
          "3013700547",
          "3015542154",
          "3004595684",
          "3003448211",
          "3002806315",
          "1828132",
          "3021355451",
          "3011743372",
          "3014421917",
          "3013440400",
          "3005846326",
          "3007137643",
          "3002807314",
          "3014769442",
          "3007421149",
          "2647580",
          "3027267119",
          "3010230734",
          "1418975",
          "1929938",
          "3008843439",
          "3025606176",
          "3020904959",
          "3030230686",
          "3041276883",
          "1526012",
          "3003076831",
          "2424366",
          "3009746061",
          "1000208608",
          "3010155661",
          "3016851379",
          "3001236277",
          "3002750084",
          "1718873",
          "1720929",
          "3018094310",
          "3008824097",
          "1824037",
          "3002807134",
          "1000614294",
          "3015286855",
          "2938071",
          "3021393544",
          "3006891611",
          "1000306431",
          "3001237338",
          "3003249645",
          "3004111573",
          "3004911384",
          "1000136269",
          "3009941480",
          "1000477302",
          "3044464950",
          "3003724334",
          "3007791595",
          "3010187826",
          "3011944170",
          "3036803067",
          "3005075853",
          "3007738736",
          "1527736",
          "1000121050",
          "3014342096",
          "3000215205",
          "3013734919",
          "3041918479",
          "3023146633",
          "3010126772",
          "3029916132",
          "2183785",
          "3008693678",
          "3013917867",
          "3035482729",
          "2435119",
          "3002807095",
          "3016555232",
          "3008264458",
          "3010291427",
          "3021519357",
          "3007583504",
          "3015227849",
          "3040335671",
          "3009599228",
          "3016813690",
          "3008573567",
          "3031841433",
          "1000640985",
          "3011070163",
          "3038675429",
          "3003935342",
          "3002807421",
          "3016076698",
          "3010816547",
          "3013517171",
          "3012757453",
          "3005969827",
          "2434839",
          "1000122786",
          "3021632375",
          "3002808335",
          "3004904811",
          "3008615487",
          "4486",
          "3013686892",
          "3010621964",
          "3010377594",
          "3002808148",
          "3013069546",
          "3003537036",
          "3002808473",
          "3003514332",
          "3013302242",
          "1043214",
          "1000112968",
          "3038715747",
          "3002807072",
          "3000204191",
          "1721676",
          "3011987935",
          "3004464325",
          "3030977796",
          "3010331645",
          "3016168848",
          "3004530184",
          "3007748817",
          "3014848720",
          "3013557562",
          "3014262693",
          "1923569",
          "1412854",
          "3006638824",
          "3044156918",
          "3000203000",
          "3010131466",
          "3013283620",
          "3005276402",
          "3007710878",
          "2245304",
          "3013077751",
          "3023706404",
          "3000203066",
          "1000086912",
          "3014122529",
          "3004605321",
          "1000112877",
          "3010400367",
          "3008870565",
          "3008808042",
          "3008786200",
          "3002924544",
          "3010952064",
          "3033589330",
          "3012510750",
          "3003782610",
          "3008889904",
          "3012536737",
          "3019569846",
          "6166",
          "3017405993",
          "2648666",
          "3006716777",
          "3008085016",
          "3010132111",
          "3015433394",
          "3018126215",
          "3002299608",
          "2245574",
          "3003542872",
          "3012322232",
          "3003094851",
          "3012818475",
          "3021477808",
          "3000718675",
          "3013875599",
          "3012507533",
          "3035366890",
          "3010273872",
          "1317628",
          "3027669735",
          "3007748793",
          "3001238683",
          "3013944123",
          "3008929787",
          "1418479",
          "3006017180",
          "3030626857",
          "1218017",
          "2528981",
          "3002807310",
          "3015207155",
          "3005154439",
          "3020899789",
          "3006891627",
          "3022327202"
        ],
        "device_name": "Electrosurgical, Cutting & Coagulation & Accessories",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4400",
        "device_class": "2"
      }
    }
  ]
}