  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "170203",
      "product_res_number": "Z-0765-2019",
      "event_date_initiated": "2018-12-21",
      "event_date_posted": "2019-01-31",
      "recall_status": "Terminated",
      "event_date_terminated": "2023-11-02",
      "res_event_number": "81878",
      "product_code": "DYG",
      "k_numbers": [
        "K160084"
      ],
      "product_description": "Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7\n\nProduct Usage:\nThe Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.",
      "code_info": "Lot Numbers: 61176367, 61176370",
      "firm_fei_number": "2015691",
      "recalling_firm": "Edwards Lifesciences, LLC",
      "address_1": "1 Edwards Way",
      "city": "Irvine",
      "state": "CA",
      "postal_code": "92614-5688",
      "additional_info_contact": "Linnette Torres\n949-756-4582",
      "reason_for_recall": "These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "An Urgent Product  Recall letter was sent via Fed-Ex to affected customers on 12/21/2018. A second communication was sent via Fed-Ex on 01/10/19.  The letter identified the affected product, problem and actions to be taken.   Customers were instructed to the return any unused units that are currently in their inventory.  Once returned, replacement product will be shipped at no charge.  Customers were asked to complete the attached acknowledgement form.  For questions contact Technical Support at 1-800-822-8837 option 1.",
      "product_quantity": "459 units",
      "distribution_pattern": "Worldwide Distribution",
      "openfda": {
        "k_number": [
          "K081052",
          "K934742",
          "K834037",
          "K822350",
          "K834036",
          "K831945",
          "K904287",
          "K782071",
          "K821057",
          "K813578",
          "K770522",
          "K843411",
          "K832324",
          "K940795",
          "K883948",
          "K823007",
          "K832380",
          "K901247",
          "K920869",
          "K003118",
          "K771043",
          "K833572",
          "K802338",
          "K900879",
          "K935404",
          "K771823",
          "K830604",
          "K160084",
          "K193466",
          "K822723",
          "K874465",
          "K951566",
          "K851948",
          "K820883",
          "K012225",
          "K820229",
          "K832287",
          "K963336",
          "K841639",
          "K821099",
          "K242451",
          "K772059",
          "K963248",
          "K812563",
          "K955816",
          "K993245",
          "K821235",
          "K832957",
          "K231248",
          "K911840",
          "K831854",
          "K233824",
          "K823433",
          "K801865",
          "K893474",
          "K833836",
          "K822806",
          "K820222",
          "K852294",
          "K841876",
          "K850952",
          "K823484",
          "K821996",
          "K831185",
          "K843409",
          "K820702",
          "K094048",
          "K802481",
          "K812111",
          "K810352",
          "K853383",
          "K001063",
          "K841251",
          "K924452",
          "K872770",
          "K771306",
          "K833918",
          "K010253",
          "K821004",
          "K892530",
          "K896466",
          "K874117",
          "K841640",
          "K233983",
          "K802288",
          "K925444",
          "K841705",
          "K822034",
          "K811276",
          "K841317",
          "K940547",
          "K812324",
          "K893435",
          "K862119",
          "K760533",
          "K961999",
          "K810124",
          "K901287",
          "K222117"
        ],
        "pma_number": [
          "P810046"
        ],
        "registration_number": [
          "",
          "1720929",
          "1000563940",
          "2025816",
          "3045042524",
          "2649622",
          "1625425",
          "2648045",
          "3004784537",
          "9680260",
          "2024311",
          "3013162291",
          "3013319212",
          "1713468",
          "3004091615",
          "9615817",
          "2939561",
          "9673241",
          "2523676",
          "1061124",
          "3010532612",
          "3005718570",
          "2024168",
          "3014656749",
          "3032916632",
          "3006425876",
          "2182208",
          "1721676",
          "8043983",
          "3015859709",
          "2521402",
          "3013300026",
          "3006082230",
          "1061927",
          "3002807314",
          "1317547",
          "2133641",
          "1319639",
          "1220948",
          "3004832480",
          "3004111573",
          "1018233",
          "3013764800",
          "1035166",
          "9680794",
          "3008496528",
          "2011171",
          "3038791046",
          "3017060084"
        ],
        "fei_number": [
          "",
          "1720929",
          "1000512168",
          "1000563940",
          "2025816",
          "3000203466",
          "3045042524",
          "1625425",
          "2649622",
          "2648045",
          "3004784537",
          "2024311",
          "3013162291",
          "1000138054",
          "3013319212",
          "1713468",
          "3004091615",
          "2523676",
          "3010532612",
          "3005718570",
          "2024168",
          "3014656749",
          "3032916632",
          "3016261803",
          "3006425876",
          "2182208",
          "3002315844",
          "1721676",
          "3015859709",
          "2521402",
          "3013300026",
          "3003248396",
          "3006082230",
          "3000203000",
          "3002807314",
          "1317547",
          "1220948",
          "3004832480",
          "3004111573",
          "1018233",
          "3013764800",
          "3003081151",
          "1035166",
          "3008496528",
          "2011171",
          "3003737899",
          "3004635528",
          "3038791046",
          "3017060084"
        ],
        "device_name": "Catheter, Flow Directed",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1240",
        "device_class": "2"
      }
    }
  ]
}