      "cfres_id": "167424",
      "product_res_number": "Z-0170-2019",
      "event_date_initiated": "2018-08-30",
      "event_date_created": "2018-10-17",
      "recall_status": "Terminated",
      "event_date_terminated": "2018-12-20",
      "res_event_number": "81084",
      "product_code": "MHX",
      "k_numbers": [
        "K162607"
      ],
      "product_description": "Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00",
      "code_info": "This potential issue affects BeneVision DMS with software version 03.01.00 to 03.03.00",
      "firm_fei_number": "2221819",
      "recalling_firm": "Mindray DS USA, Inc. dba Mindray North America",
      "address_1": "800 Macarthur Blvd",
      "city": "Mahwah",
      "state": "NJ",
      "postal_code": "07430-2001",
      "additional_info_contact": "Ms. Diane Arpino\n201-995-8407",
      "reason_for_recall": "The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around\nbetween multiple central stations in the same care area.",
      "root_cause_description": "Software design",
      "action": "The firm notified affected consignees on August 30, 2018, via Urgent Medical Device Recall letter. The letter advised customers of the software issue when more than 64 telemetry transmitters have been connected to a central station.\n\nCustomers were advised to use the system normally, but informed that a Mindray Service Representative would be in contact to arrange for a system software update to be performed at the customer site.\n\nMindrays Technical Support team may be reached at (877) 913-9663 (Option 1), Monday\nthrough Friday, 8:30am - 5:30pm ET should you have technical questions.\n\nCustomers were also asked to complete and return an attached Acknowledgement and Receipt Form.",
      "product_quantity": "5",
      "distribution_pattern": "Distributed to accounts in CA, CO, FL, MA, NY, PA, PR, TX, and WV.",
      "openfda": {
        "k_number": [
          "K014213",
          "K992273",
          "K062098",
          "K050742",
          "K152902",
          "K023698",
          "K173036",
          "K092369",
          "K081793",
          "K043551",
          "K123472",
          "K024108",
          "K130584",
          "K113657",
          "K192514",
          "K221904",
          "K923054",
          "K023674",
          "K033492",
          "K031320",
          "K043439",
          "K022834",
          "K033444",
          "K100939",
          "K252589",
          "K153160",
          "K050141",
          "K070588",
          "K150975",
          "K250349",
          "K170876",
          "K033694",
          "K942700",
          "K030431",
          "K030105",
          "K082045",
          "K061675",
          "K030038",
          "K220989",
          "K062576",
          "K151063",
          "K043549",
          "K250453",
          "K123074",
          "K011093",
          "K981576",
          "K240320",
          "K062283",
          "K251146",
          "K112803",
          "K152407",
          "K945385",
          "K213490",
          "K221348",
          "K122146",
          "K112962",
          "K073550",
          "K973474",
          "K041772",
          "K192972",
          "K990068",
          "K132533",
          "K922668",
          "K092027",
          "K220308",
          "K141811",
          "K153702",
          "K060039",
          "K052707",
          "K011000",
          "K042656",
          "K171801",
          "K911646",
          "K013199",
          "K163584",
          "K091834",
          "K031403",
          "K221879",
          "K113623",
          "K141015",
          "K030613",
          "K950536",
          "K980619",
          "K972535",
          "K933624",
          "K163459",
          "K003243",
          "K062095",
          "K970445",
          "K082583",
          "K213363",
          "K032369",
          "K102040",
          "K120631",
          "K132026",
          "K213336",
          "K151027",
          "K063690",
          "K980411",
          "K153448",
          "K881643",
          "K101619",
          "K143195",
          "K003248",
          "K073462",
          "K000517",
          "K142601",
          "K240312",
          "K171901",
          "K090172",
          "K031760",
          "K113653",
          "K173768",
          "K090443",
          "K161056",
          "K070566",
          "K132662",
          "K213181",
          "K202336",
          "K103646",
          "K023871",
          "K914838",
          "K172226",
          "K161531",
          "K250757",
          "K213234",
          "K251508",
          "K040246",
          "K013311",
          "K221056",
          "K113054",
          "K100046",
          "K192137",
          "K190008",
          "K090702",
          "K062324",
          "K072346",
          "K231477",
          "K113187",
          "K200859",
          "K001433",
          "K231671",
          "K113441",
          "K053522",
          "K223498",
          "K001348",
          "K152973",
          "K102004",
          "K030313",
          "K152552",
          "K060135",
          "K921014",
          "K061610",
          "K083750",
          "K013624",
          "K040304",
          "K945134",
          "K122253",
          "K100040",
          "K131414",
          "K030931",
          "K041087",
          "K052801",
          "K002158",
          "K173454",
          "K252840",
          "K211619",
          "K133222",
          "K210906",
          "K033365",
          "K021279",
          "K102495",
          "K071020",
          "K002478",
          "K033964",
          "K113798",
          "K121480",
          "K223531",
          "K031781",
          "K242562",
          "K073203",
          "K253170",
          "K000815",
          "K970182",
          "K062392",
          "K001722",
          "K131032",
          "K071164",
          "K182075",
          "K011540",
          "K182979",
          "K213799",
          "K113045",
          "K132075",
          "K130849",
          "K071426",
          "K130711",
          "K180482",
          "K052756",
          "K091395",
          "K131223",
          "K001693",
          "K983996",
          "K223574",
          "K260098",
          "K080157",
          "K060221",
          "K990563",
          "K060223",
          "K993608",
          "K040357",
          "K071073",
          "K180472",
          "K131929",
          "K181831",
          "K160685",
          "K133576",
          "K061993",
          "K142827",
          "K131763",
          "K033957",
          "K982091",
          "K081983",
          "K131872",
          "K063315",
          "K041741",
          "K123048",
          "K250854",
          "K012915",
          "K130208",
          "K031124",
          "K033711",
          "K202405",
          "K872476",
          "K151860",
          "K021325",
          "K944495",
          "K103652",
          "K003621",
          "K191149",
          "K183387",
          "K252250",
          "K100217",
          "K101521",
          "K051658",
          "K950264",
          "K080342",
          "K093501",
          "K911051",
          "K030812",
          "K060541",
          "K030738",
          "K083271",
          "K153707",
          "K123900",
          "K130401",
          "K232694",
          "K080495",
          "K131829",
          "K982279",
          "K063725",
          "K102422",
          "K061379",
          "K992674",
          "K040395",
          "K112637",
          "K920445",
          "K924575",
          "K030575",
          "K001664",
          "K120173",
          "K031849",
          "K053484",
          "K112190",
          "K990476",
          "K101692",
          "K000854",
          "K913263",
          "K150352",
          "K191769",
          "K061185",
          "K053207",
          "K201941",
          "K010923",
          "K042904",
          "K231010",
          "K002209",
          "K051628",
          "K043276",
          "K190011",
          "K181314",
          "K922058",
          "K022889",
          "K173765",
          "K051400",
          "K141156",
          "K110953",
          "K992929",
          "K091594",
          "K150082",
          "K201764",
          "K922210",
          "K090556",
          "K151080",
          "K141963",
          "K103134",
          "K014261",
          "K011918",
          "K922974",
          "K082280",
          "K211900",
          "K120727",
          "K191323",
          "K151812",
          "K041235",
          "K012016",
          "K102239",
          "K080251",
          "K150329",
          "K131971",
          "K092449",
          "K040955",
          "K093788",
          "K021422",
          "K161517",
          "K121367",
          "K120366",
          "K013427",
          "K030657",
          "K172510",
          "K000882",
          "K110015",
          "K200494",
          "K033304",
          "K110922",
          "K993974",
          "K962424",
          "K051367",
          "K110622",
          "K041790",
          "K041607",
          "K023475",
          "K201746",
          "K143676",
          "K251702",
          "K031376",
          "K020531",
          "K203435",
          "K180017",
          "K102376",
          "K120598",
          "K132077",
          "K200015",
          "K050762",
          "K081140",
          "K083228",
          "K061052",
          "K131876",
          "K252726",
          "K102562",
          "K253277",
          "K101067",
          "K083697",
          "K060833",
          "K242962",
          "K042642",
          "K023569",
          "K191106",
          "K051217",
          "K250179",
          "K093268",
          "K060281",
          "K073280",
          "K000168",
          "K945099",
          "K971840",
          "K950436",
          "K203088",
          "K070791",
          "K172966",
          "K001268",
          "K912188",
          "K022485",
          "K150030",
          "K122439",
          "K062194",
          "K222586",
          "K254209",
          "K912835",
          "K190939",
          "K160951",
          "K072070",
          "K170567",
          "K062271",
          "K031481",
          "K191249",
          "K020290",
          "K221616",
          "K912627",
          "K992595",
          "K944860",
          "K082785",
          "K123556",
          "K101449",
          "K053121",
          "K134046",
          "K030722",
          "K232962",
          "K032290",
          "K153275",
          "K171580",
          "K170514",
          "K230604",
          "K170155",
          "K033048",
          "K081891",
          "K062144",
          "K012226",
          "K000563",
          "K160981",
          "K031340",
          "K120616",
          "K171621",
          "K092680",
          "K040458",
          "K033600",
          "K092970"
        ],
        "registration_number": [
          "3030597679",
          "8022257",
          "3009077524",
          "3003589956",
          "3001570235",
          "3020115",
          "3005783425",
          "3021219789",
          "3014848734",
          "8030978",
          "3009499478",
          "3010552191",
          "3007305624",
          "3010698831",
          "3009490946",
          "3008642652",
          "3010611950",
          "9611031",
          "3015328411",
          "3012238587",
          "9681377",
          "3016618143",
          "3013557562",
          "1000484830",
          "3007481893",
          "3022123357",
          "2032233",
          "3013314298",
          "3014633893",
          "3018783526",
          "3009673645",
          "8030229",
          "2221819",
          "1218950",
          "3008729547",
          "3012972516",
          "3013679558",
          "1651104",
          "3016852448",
          "1066270",
          "3021559257",
          "3014263632",
          "1316463",
          "3013501110",
          "1054713",
          "2080783",
          "3010390468",
          "3000162901",
          "3003591740",
          "3005906887",
          "3004832480",
          "2025816",
          "3019545235",
          "1930027",
          "3007836437",
          "3000126629",
          "3015997711",
          "3011120183",
          "3013896610",
          "3021736577",
          "3008363989",
          "3012421607",
          "3009973336",
          "3004035727",
          "3011353843",
          "3008255748",
          "3006147705",
          "3016674095",
          "3031158",
          "3008716327",
          "3043479346",
          "3006979678",
          "3004526608",
          "3030446844",
          "3007734888",
          "3016931915",
          "9611252",
          "3009156722",
          "3009607734",
          "3010882708",
          "3008483389",
          "3004196561",
          "3017446092",
          "3009443693",
          "1219324",
          "3006606827",
          "3006087789",
          "3009831876",
          "3009715978",
          "3015181391",
          "3019388613",
          "3010536822",
          "9610816",
          "3007413079",
          "3014579161",
          "3004977335",
          "3005877899",
          "3010157426",
          "3007048291",
          "3023272766",
          "3003263092",
          "3008058135",
          "9710602",
          "3012359877",
          "3016701404",
          "3003832357",
          "3012528160",
          "3013596024",
          "3012104670",
          "3014541700",
          "3007603826",
          "3032109181",
          "3013500228",
          "1000122786",
          "3004415095",
          "3030516450",
          "9610105",
          "3005521864",
          "3010703925",
          "3003971136"
        ],
        "fei_number": [
          "3003757840",
          "3030597679",
          "3009077524",
          "3003589956",
          "3002807720",
          "3001570235",
          "3020115",
          "3005783425",
          "3002807739",
          "3021219789",
          "3014848734",
          "3009499478",
          "3007305624",
          "3010698831",
          "3009490946",
          "3008642652",
          "3010611950",
          "3015328411",
          "3012238587",
          "3016618143",
          "3013557562",
          "1000484830",
          "3007481893",
          "3014633893",
          "3018783526",
          "3009673645",
          "1000300598",
          "2221819",
          "1218950",
          "3008729547",
          "3012972516",
          "3013679558",
          "3016852448",
          "3003513129",
          "3021559257",
          "3014263632",
          "1316463",
          "3013501110",
          "2080783",
          "3002806783",
          "3010390468",
          "3000162901",
          "3003591740",
          "3005906887",
          "3004832480",
          "2025816",
          "3019545235",
          "1930027",
          "3007836437",
          "3000126629",
          "3000215024",
          "3015997711",
          "3011120183",
          "1051088",
          "1000114111",
          "3013896610",
          "3021736577",
          "3008363989",
          "3012421607",
          "3009973336",
          "3004035727",
          "3011353843",
          "3010830833",
          "3008255748",
          "3006147705",
          "3016674095",
          "3008716327",
          "3043479346",
          "3006979678",
          "3002807019",
          "3004526608",
          "3030446844",
          "3007734888",
          "3016931915",
          "3009156722",
          "3009607734",
          "3010882708",
          "3008483389",
          "3004196561",
          "3017446092",
          "3009443693",
          "1219324",
          "3006606827",
          "3006087789",
          "3009831876",
          "3015191855",
          "3002808182",
          "3009715978",
          "3002807167",
          "3015181391",
          "3019388613",
          "3010536822",
          "3007413079",
          "3014579161",
          "3004977335",
          "3005877899",
          "3010157426",
          "3007048291",
          "3000957797",
          "3023272766",
          "3003263092",
          "3008058135",
          "3012359877",
          "3016701404",
          "3003142870",
          "3003832357",
          "3012528160",
          "3013596024",
          "3012104670",
          "3014541700",
          "3007603826",
          "3002737156",
          "3013500228",
          "3032109181",
          "1000122786",
          "3004415095",
          "3030516450",
          "3005521864",
          "3010703925",
          "3003971136"
        ],
        "device_name": "Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1025",
        "device_class": "2"
      }
    }
  ]
}