{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "165490",
      "product_res_number": "Z-0353-2019",
      "event_date_initiated": "2018-05-24",
      "event_date_created": "2018-10-31",
      "recall_status": "Terminated",
      "event_date_terminated": "2022-12-08",
      "res_event_number": "80434",
      "product_code": "JQP",
      "product_description": "Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc.\n\n The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client for specimen tracking and workload statistics or a physical workstation for manual sorting and label printing.",
      "code_info": "v4.2.1 (and 4.2, discontinued)",
      "firm_fei_number": "2050012",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "city": "Brea",
      "state": "CA",
      "postal_code": "92821-6232",
      "additional_info_contact": "469-961-4941",
      "reason_for_recall": "Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. Sorting Drive software version 4.2 was replaced by 4.2.1 (both versions contain the same option for manual barcode printing).\n\nWhen printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.",
      "root_cause_description": "Software change control",
      "action": "On 6/4/2018  Beckman Coulter (BEC) reported that the issue will be corrected via a field modification performed by Field Service. BEC reported that they  do not intend to send a customer letter for this issue.US customers can contact the Technical Support Troubleshooting Hotline at (800) 854-3633.",
      "product_quantity": "283",
      "distribution_pattern": "Worldwide Distribution:  US (nationwide) and countries of: Algeria, Argentina,  Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland, Israel, Italy ,Republic of  Korea,  Lebanon, Lithuania ,Malaysia, Netherlands, Poland, Portugal , Qatar, Romania, Russian Federation, Saudi Arabia ,South Africa , Spain, Sweden, Taiwan, Turkey , and United Kingdom.",
      "openfda": {
        "k_number": [
          "K833353",
          "K844907",
          "K802471",
          "K780881",
          "K800724",
          "K801167",
          "K761063",
          "K861779",
          "K893838",
          "K831830",
          "K810968",
          "K790543",
          "K830296",
          "K792129",
          "K983468",
          "K821531",
          "K113007",
          "K812604",
          "K801438",
          "K760739",
          "K800241",
          "K810631",
          "K770671",
          "K871339",
          "K823785",
          "K851321",
          "K782052",
          "K802251",
          "K822146",
          "K892060",
          "K953652",
          "K820427",
          "K801131",
          "K821058",
          "K801091",
          "K864557",
          "K934111",
          "K850659",
          "K790336",
          "K801210",
          "K822932",
          "K832746",
          "K822381",
          "K821807",
          "K861778",
          "K810063",
          "K881194",
          "K792158",
          "K801664",
          "K854265",
          "K834490",
          "K861844",
          "K823110",
          "K843365",
          "K871853",
          "K834300",
          "K802594",
          "K885292",
          "K822860",
          "K831574",
          "K831838",
          "K853026",
          "K770877",
          "K803312",
          "K822778",
          "K964460",
          "K781217",
          "K923316",
          "K820721",
          "K781173",
          "K861412",
          "K832760",
          "K800282",
          "K823679",
          "K840502",
          "K842401",
          "K781081",
          "K800242"
        ],
        "registration_number": [
          "3020978007",
          "3008789114",
          "3015967359",
          "3031570369",
          "9616936",
          "3012104009",
          "9710492",
          "3015969360",
          "3003618481",
          "3010141591",
          "2029302",
          "3002807715",
          "2245451",
          "3010258456",
          "3008802064",
          "1220649",
          "3004145393",
          "3003453957",
          "1061932",
          "3003995437",
          "3002889627",
          "1225673",
          "3006543086",
          "8043723",
          "3042228064",
          "3011194542",
          "9615754",
          "3043012074",
          "3011806823",
          "3016764932",
          "1810909",
          "3006677029",
          "3009542936",
          "2243471",
          "9610126",
          "3027486032",
          "3021394862",
          "9615507",
          "3012436569",
          "8020888",
          "9616530",
          "3010097171",
          "2936921",
          "2431530",
          "3043012240",
          "3000126629",
          "3009404844",
          "3009731198",
          "3013896610",
          "3005798890",
          "3042989733",
          "3004959793",
          "9610865",
          "3003825930",
          "8030392",
          "3009711478",
          "3015426626",
          "3023268435",
          "8020030",
          "3038419886",
          "3007134734",
          "3015738702",
          "3002809144",
          "2531491",
          "3017086896",
          "3003559191",
          "3012118659",
          "3030413550",
          "9614673",
          "1058332",
          "3011393376",
          "3000148879",
          "3002710220",
          "2517506",
          "3005509212",
          "3037000637",
          "3011121740",
          "3003795116",
          "3003600763",
          "3002637618",
          "3010825766",
          "3006371704",
          "2432235",
          "1119779",
          "3015633428",
          "3006755541",
          "3004838836",
          "1950204",
          "3005862821",
          "9610816",
          "3013922172",
          "3013646951",
          "3035708936",
          "3024512705",
          "3030714692",
          "3009129531",
          "3011187792",
          "3025376092",
          "3010900203",
          "3008514395",
          "3040278125",
          "1000113657",
          "3016438761",
          "3002821232",
          "3015058854",
          "3004032053",
          "3009288740",
          "3013298431",
          "3003902721",
          "2954323",
          "8043909",
          "3027496319",
          "3015347724",
          "3005333358",
          "3004973408",
          "3014840402",
          "1219145"
        ],
        "fei_number": [
          "3020978007",
          "3008789114",
          "3003516723",
          "3015967359",
          "3031570369",
          "3010700165",
          "3012104009",
          "3015969360",
          "3003618481",
          "3010141591",
          "3002807715",
          "2245451",
          "3001451449",
          "3010258456",
          "3008237827",
          "3008802064",
          "1220649",
          "3002803444",
          "3004145393",
          "3003453957",
          "3003116995",
          "3003559576",
          "3003995437",
          "3002889627",
          "3006461677",
          "3006543086",
          "3042228064",
          "1000306472",
          "3011194542",
          "3043012074",
          "3000203293",
          "3011806823",
          "3016764932",
          "1810909",
          "3006677029",
          "3009542936",
          "2243471",
          "3002829675",
          "3027486032",
          "3021394862",
          "3002806559",
          "3012436569",
          "1000307651",
          "3010097171",
          "2936921",
          "2431530",
          "3043012240",
          "3000126629",
          "3009404844",
          "3009731198",
          "3013896610",
          "3005798890",
          "3042989733",
          "3004959793",
          "3003825930",
          "3009711478",
          "3015426626",
          "3023268435",
          "3038419886",
          "3007134734",
          "3003500908",
          "3015738702",
          "3002809144",
          "3017086896",
          "3003559191",
          "3012118659",
          "3030413550",
          "3011393376",
          "3000148879",
          "3002710220",
          "2517506",
          "3005509212",
          "3037000637",
          "3003795116",
          "3003600763",
          "3002637618",
          "3010825766",
          "3006371704",
          "3007019572",
          "2432235",
          "1119779",
          "3015633428",
          "3002807167",
          "3006755541",
          "3004838836",
          "1950204",
          "3005862821",
          "3013922172",
          "3013646951",
          "3001573851",
          "3035708936",
          "3024512705",
          "3002671736",
          "3030714692",
          "3009129531",
          "3011187792",
          "3025376092",
          "3010900203",
          "3008514395",
          "3040278125",
          "1000113657",
          "3016438761",
          "3002821232",
          "3015058854",
          "3004032053",
          "3002806350",
          "3009288740",
          "3013298431",
          "3003902721",
          "3027496319",
          "3015347724",
          "3005333358",
          "3002808498",
          "3004973408",
          "3014840402",
          "1219145"
        ],
        "device_name": "Calculator/Data Processing Module, For Clinical Use",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.2100",
        "device_class": "1"
      }
    }
  ]
}