{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "160805",
      "product_res_number": "Z-0379-2018",
      "event_date_initiated": "2017-11-14",
      "event_date_created": "2018-01-26",
      "recall_status": "Open, Classified",
      "res_event_number": "78885",
      "product_code": "RCE",
      "product_description": "Metrotom 800 (130kV CT scanner)\n\nIndustrial CT scanner for measuring and inspecting complete components made of plastic or light metal.",
      "code_info": "Serial Numbers:  152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342",
      "firm_fei_number": "3003902825",
      "recalling_firm": "Carl Zeiss Metrology Inc",
      "address_1": "6250 Sycamore Ln N",
      "city": "Maple Grove",
      "state": "MN",
      "postal_code": "55369-6310",
      "additional_info_contact": "Scott Chapman\n763-203-4267",
      "reason_for_recall": "Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Zeiss sent an Safety Notice Letter  dated September 28, 2017, to the customers advising them of the recall. The affected device was identified as well as the issue involved. Customers are to complete and return the confirmation- waiver form. During the next scheduled service event, a ZEISS field service technician will update the Metrotom operating system. The corrections are to be made free of charge, and will be completed by March 2018.\n\nCDRH approves the CAP subject to the following conditions:\n\nNotification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22.  This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification.  For further questions, please call  (763)  203-4267.",
      "product_quantity": "13 US",
      "distribution_pattern": "Worldwide Distribution - US, including the states of :  WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.",
      "openfda": {
        "device_name": "Cabinet X-Ray, Industrial",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}