  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "151111",
      "product_res_number": "Z-0721-2017",
      "event_date_initiated": "2016-11-09",
      "event_date_created": "2016-12-08",
      "recall_status": "Terminated",
      "event_date_terminated": "2019-06-11",
      "res_event_number": "75731",
      "product_code": "MUJ",
      "k_numbers": [
        "K160093"
      ],
      "product_description": "RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System\n\nProduct Usage:\nRayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.",
      "code_info": "Build numbers: 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11 or 5.0.2.35",
      "firm_fei_number": "3007774465",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavaegen 9 9 sveavaegen",
      "city": "Stockholm",
      "country": "Sweden",
      "reason_for_recall": "Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0.  For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.",
      "root_cause_description": "Software design",
      "action": "Ray Search sent an Urgent Field Safety Notice Medical Device Correction letter (dated 11/8/2016) via e-mailed to customers.  The letter identified the affected product, problem and  actions to be taken.  Customers were instructed to Inspect their product and identify all installed units with the above software version number, then confirm you have read and understood this notice (contact information included).",
      "product_quantity": "US: 66 units",
      "distribution_pattern": "US Nationwide Distribution in the states of California, Florida, Illinois, Louisiana, Michigan, New Jersey, Tennessee, Texas, Washington",
      "openfda": {
        "k_number": [
          "K132971",
          "K091179",
          "K982696",
          "K915691",
          "K211379",
          "K232400",
          "K171885",
          "K201444",
          "K982791",
          "K252304",
          "K012017",
          "K242992",
          "K211034",
          "K142108",
          "K041403",
          "K011246",
          "K041719",
          "K100056",
          "K160093",
          "K983343",
          "K180379",
          "K123263",
          "K051497",
          "K022036",
          "K991815",
          "K211841",
          "K260308",
          "K042269",
          "K100417",
          "K220783",
          "K082221",
          "K051941",
          "K100552",
          "K002312",
          "K201607",
          "K223279",
          "K151666",
          "K032724",
          "K210645",
          "K984532",
          "K150547",
          "K993675",
          "K133331",
          "K112723",
          "K081281",
          "K013112",
          "K211511",
          "K253078",
          "K994413",
          "K202789",
          "K013673",
          "K992762",
          "K090972",
          "K240671",
          "K182665",
          "K083264",
          "K133247",
          "K041971",
          "K192065",
          "K010682",
          "K022411",
          "K090531",
          "K151560",
          "K170086",
          "K090182",
          "K151887",
          "K002147",
          "K172163",
          "K252109",
          "K203469",
          "K032345",
          "K190664",
          "K063482",
          "K002509",
          "K080888",
          "K133655",
          "K111311",
          "K151380",
          "K213438",
          "K162370",
          "K220582",
          "K101076",
          "K991237",
          "K061097",
          "K200569",
          "K213787",
          "K250963",
          "K091146",
          "K193109",
          "K974770",
          "K042739",
          "K202284",
          "K092132",
          "K112060",
          "K081112",
          "K141283",
          "K091492",
          "K113102",
          "K022307",
          "K102216",
          "K042861",
          "K063230",
          "K020027",
          "K170750",
          "K061440",
          "K120363",
          "K200608",
          "K220141",
          "K241318",
          "K012575",
          "K123152",
          "K242748",
          "K232799",
          "K252002",
          "K022741",
          "K031349",
          "K062967",
          "K121448",
          "K152393",
          "K141860",
          "K173940",
          "K992864",
          "K013644",
          "K222312",
          "K100092",
          "K203766",
          "K242378",
          "K193640",
          "K252863",
          "K243057",
          "K222803",
          "K990475",
          "K032456",
          "K040474",
          "K253050",
          "K993766",
          "K140187",
          "K161136",
          "K131891",
          "K141900",
          "K200487",
          "K050296",
          "K103093",
          "K071938",
          "K082005",
          "K083414",
          "K203681",
          "K001295",
          "K993923",
          "K182888",
          "K021268",
          "K123357",
          "K993895",
          "K151435",
          "K151369",
          "K233236",
          "K093588",
          "K092928",
          "K101220",
          "K080601",
          "K141248",
          "K214081",
          "K240398",
          "K013661",
          "K061893",
          "K211881",
          "K211867",
          "K990071",
          "K071760",
          "K002132",
          "K242822",
          "K143514",
          "K120387",
          "K250064",
          "K082642",
          "K223724",
          "K231825",
          "K222728",
          "K992434",
          "K170969",
          "K001700",
          "K062514",
          "K153368",
          "K982821",
          "K080929",
          "K071873",
          "K151469",
          "K031088",
          "K260207",
          "K053584",
          "K010784",
          "K191384",
          "K992753",
          "K093391",
          "K170355",
          "K030821",
          "K041997",
          "K130617",
          "K151233",
          "K192554",
          "K252784",
          "K132816",
          "K131428",
          "K013535",
          "K171536",
          "K031281",
          "K083672",
          "K110730",
          "K002237",
          "K103246",
          "K040688",
          "K193381",
          "K103146",
          "K082606",
          "K190178",
          "K052325",
          "K223233",
          "K260750",
          "K984196",
          "K212274",
          "K020971",
          "K022961",
          "K032209",
          "K181145",
          "K102915",
          "K981678",
          "K183037",
          "K230557",
          "K992303",
          "K112446",
          "K120551",
          "K201350",
          "K241876",
          "K102011",
          "K143269",
          "K090808",
          "K190387",
          "K002240",
          "K112776",
          "K234068",
          "K250440",
          "K180308",
          "K232479",
          "K083502",
          "K090893",
          "K000922",
          "K242463",
          "K091145",
          "K173635",
          "K101119",
          "K103515",
          "K073020",
          "K180289",
          "K062032",
          "K253091",
          "K193459",
          "K152488"
        ],
        "registration_number": [
          "1038814",
          "2030598",
          "8043933",
          "3019834893",
          "3012039133",
          "3014425747",
          "3016686005",
          "3003873069",
          "3009529630",
          "2183553",
          "3004555507",
          "3011716550",
          "3025309924",
          "3003418325",
          "3000126629",
          "3007791541",
          "9617978",
          "3044735888",
          "1651102",
          "3002808157",
          "3008156050",
          "3003586672",
          "3003793371",
          "3009090417",
          "3004363352",
          "3005136058",
          "3014873236",
          "3025765259",
          "3006579682",
          "3019543601",
          "3006894636",
          "3015777306",
          "3004561814",
          "3010034862",
          "3013410552",
          "3009302839",
          "9611894",
          "3010123525",
          "3003094912",
          "3007037519",
          "3004580128",
          "3022719374",
          "2916710",
          "3022745617",
          "3018004899",
          "3004608878",
          "2939248",
          "3004977335",
          "3030564015",
          "3007774465",
          "3009077524",
          "3016602874",
          "3042177665",
          "3015232217",
          "3023344262",
          "3033527281",
          "2434141",
          "3003768277",
          "3004526608"
        ],
        "fei_number": [
          "1038814",
          "3019834893",
          "3012039133",
          "3014425747",
          "3016686005",
          "3003873069",
          "3002619595",
          "3009529630",
          "2183553",
          "3004555507",
          "3011716550",
          "3025309924",
          "3003418325",
          "3000126629",
          "3007791541",
          "3044735888",
          "3002807741",
          "3002808157",
          "3008156050",
          "3003586672",
          "3003793371",
          "3009090417",
          "3004363352",
          "3005136058",
          "3014873236",
          "3002807880",
          "3025765259",
          "3006579682",
          "3019543601",
          "3006894636",
          "3015777306",
          "3002802883",
          "3004561814",
          "3010034862",
          "3013410552",
          "3009302839",
          "3003094912",
          "3010123525",
          "3007037519",
          "3004580128",
          "3022719374",
          "2916710",
          "3022745617",
          "3018004899",
          "3004608878",
          "2939248",
          "3004977335",
          "3030564015",
          "3007774465",
          "1000518731",
          "3009077524",
          "3016602874",
          "3008581796",
          "3042177665",
          "3015232217",
          "3023344262",
          "3033527281",
          "2434141",
          "3004526608"
        ],
        "device_name": "System, Planning, Radiation Therapy Treatment",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.5050",
        "device_class": "2"
      }
    }
  ]
}