  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "143625",
      "product_res_number": "Z-0803-2016",
      "event_date_initiated": "2016-01-13",
      "event_date_created": "2016-02-19",
      "recall_status": "Terminated",
      "event_date_terminated": "2021-05-18",
      "res_event_number": "73251",
      "product_code": "RBY",
      "product_description": "Olympus Scientific Solutions Americas Corporation ( OSSA ) Delta XRF Analyzer .  This is a  Analytical X-ray system",
      "code_info": "All models of the Delta XRF Analyzer",
      "firm_fei_number": "3006241029",
      "recalling_firm": "Olympus Scientific Solutions Americas",
      "address_1": "48 Woerd Ave",
      "city": "Waltham",
      "state": "MA",
      "postal_code": "02453-3824",
      "additional_info_contact": "866-446-6689",
      "reason_for_recall": "The Firm has discovered a Software bug.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Olympus Scientific Solutions Americas ( OSSSA) planned action to bring product into compliance: \n\nNotification will be given to purchasers per 21 CFR 1003.21.  A software modification will be made available to all users contacted.  The letter to affected purchasers makes clear that the product will be brought into compliance without charge and offers the purchaser several options for applying the software update.\n\nCDRH approves the CAP subject to the following conditions:\n\n1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR\n1003.21 and 1003.22.  This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification.\n2.  Tracking of customer/owner self-installation(s) of the software update should be done to the maximum possible extent via serial-number- controlled downloads, follow-up email reporting, return-mail card or some other similarly effective means. For further questions you may call : (866) 446-6689.",
      "product_quantity": "US - 5000 units",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) Distribution",
      "openfda": {
        "device_name": "X-Ray Fluorescence",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}