{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "142252",
      "product_res_number": "Z-0633-2016",
      "event_date_initiated": "2015-12-08",
      "event_date_posted": "2016-01-12",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-08-04",
      "res_event_number": "72808",
      "product_code": "JJJ",
      "product_description": "Perkin Elmer WIZARD2 2-detector, 550 samples Product Code: 2470-0020\nThe Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples",
      "code_info": "Serial Numbers: DG02117524 DG09118363 DG08118156 DG08118195 DG02128922 DG10118375 DG12107235 DG08118222 DG02117537 DG08118184 DG09106872 DG11107116 DG03117626 DG03117611 DG01117425 DG12107293 DG09118332 DG06106503 DG06118006 DG01128820 DG12118707 DG05106427 DG03106166 DG11118630 DG01117434 DG09106779 DG02117500 DG12118742 DG06106477 DG03106208 DG09106800 DG10106903 DG05117853 DG03117671 DG06106473 DG04117754 DG04117698 DG11118538 DG11118617 DG01117392 DG03117644 DG05117882 DG03117655 DG04117683 DG02106106 DG05106347 DG06118043 DG10118385 DG08106671 DG01128869 DG06118023 DG05106328 DG02117516 DG11107142 DG05106354 DG08118254 DG03106181 DG03106197 DG02117505 DG11107105",
      "firm_fei_number": "3006159931",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "city": "Waltham",
      "state": "MA",
      "postal_code": "02451",
      "reason_for_recall": "The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect.  The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed.  If the error is undetected, the Gamma Counter may produce erroneous results.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "Perkin Elmer sent an Urgent Medical Device Correction letter dated December 7, 2015 to affected customers.  The letter identified the affected products, problems and actions to be taken.  Customers were instructed to destroy all affected products in their possession. Customers were asked to complete the\nenclosed Correction Response Form and return it by scan and e-mail to QMresponse.SG@perkinelmer.com . Corrected barcode labels ID #023 will be provided upon request and return of the completed Correction Response Form to PerkinElmer.  For further information contact your local PerkinElmer representatives..",
      "product_quantity": "60 units",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of \nARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.",
      "openfda": {
        "k_number": [
          "K842902",
          "K894902",
          "K962762",
          "K771180",
          "K830365",
          "K832209",
          "K832734",
          "K771369",
          "K832736",
          "K946387",
          "K820137",
          "K884269",
          "K801245",
          "K800388",
          "K820376",
          "K761214",
          "K802578",
          "K771614",
          "K920559",
          "K834508",
          "K813472",
          "K810345",
          "K770088",
          "K870115",
          "K821892",
          "K790905",
          "K780383",
          "K851689",
          "K771613",
          "K900087",
          "K801263",
          "K884091",
          "K872429",
          "K790961",
          "K801264",
          "K780179",
          "K771612",
          "K872430",
          "K760900",
          "K832210",
          "K780341",
          "K772099",
          "K760424",
          "K832002",
          "K780344",
          "K760470",
          "K823240",
          "K800601",
          "K771129",
          "K822293",
          "K896202",
          "K841352",
          "K874094",
          "K852434",
          "K760828",
          "K821494",
          "K770740",
          "K832735",
          "K770572",
          "K862206",
          "K830751",
          "K851164",
          "K780790"
        ],
        "device_name": "Counter (Beta, Gamma) For Clinical Use",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.2320",
        "device_class": "1",
        "registration_number": [
          "",
          "1219913",
          "3004049804",
          "3003529145"
        ],
        "fei_number": [
          "",
          "3003857949",
          "1219913",
          "3004049804"
        ]
      }
    }
  ]
}