      "cfres_id": "141684",
      "product_res_number": "Z-0338-2016",
      "event_date_initiated": "2015-10-28",
      "event_date_posted": "2015-11-25",
      "recall_status": "Terminated",
      "event_date_terminated": "2016-05-18",
      "res_event_number": "72626",
      "product_code": "DQA",
      "k_numbers": [
        "K080238"
      ],
      "product_description": "rainbow Reusable Sensors.  RAINBOW DCI-DC3, SpO2/SpCO/SpMet, Part No. 2201;  RAINBOW DCI-dc8, SpO2/SpCO/SpMet, Part No. 2407;  RAINBOW DCI, SpO2/SpCO/SpMet, Part No. 2696;  RAINBOW DCIP, PED, SpO2/SpCO/SpMet, Part No. 2697.  rainbow reusable sensors are indicated for spot check or continuous non-invasive monitoring of arterial saturation (SpO2), arterial carboxyhemoglobin saturation (SpCO), arterial methemoglobin (SpMet), and pulse rate.",
      "code_info": "2201 2407 2696 2697",
      "firm_fei_number": "3001452206",
      "recalling_firm": "Masimo Corporation",
      "address_1": "Forty Parker",
      "city": "Irvine",
      "state": "CA",
      "postal_code": "92618",
      "additional_info_contact": "949-297-7000",
      "reason_for_recall": "These sensors were manufactured with incompatible configurations. This could result in sensors that will provide either no readings or inaccurate readings for the SpCO and SpMet parameters.",
      "root_cause_description": "Finished device change control",
      "action": "Masimo sent an Urgent Product Recall letter on October 28, 2015, to all affected customers to inform them that Masimo has identified specific lots of sensors that were manufactured with incompatiable configurations.  The letter provides the customers with the problems identified and the actions to be taken.  Customers are instructed to complete the tracking/verification form and email to Masimo.  Customers with questions were instructed to contact Masimo Technical Services at http://Service.Masimo.com.  For questions regarding this recall call 949-297-7000.",
      "product_quantity": "3476 units total (3322 units in US)",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to  Australia, Costa Rica, France, Germany, Kuwait, Martinique, Netherlands, Norway, South Korea, and Turkey.",
      "openfda": {
        "k_number": [
          "K230172",
          "K001930",
          "K230587",
          "K142687",
          "K190399",
          "K921519",
          "K033876",
          "K910629",
          "K982331",
          "K081870",
          "K191574",
          "K010413",
          "K203812",
          "K231288",
          "K021090",
          "K183067",
          "K033298",
          "K142209",
          "K123797",
          "K002957",
          "K111617",
          "K073454",
          "K191420",
          "K910077",
          "K040589",
          "K050844",
          "K012891",
          "K941289",
          "K971628",
          "K053381",
          "K021246",
          "K982255",
          "K002751",
          "K952315",
          "K914620",
          "K962127",
          "K083010",
          "K012626",
          "K052755",
          "K061112",
          "K220351",
          "K960251",
          "K941351",
          "K943404",
          "K992211",
          "K151287",
          "K962156",
          "K902060",
          "K940463",
          "K040581",
          "K010463",
          "K051107",
          "K183277",
          "K124041",
          "K972837",
          "K083079",
          "K884050",
          "K895987",
          "K051439",
          "K061204",
          "K883877",
          "K974292",
          "K945754",
          "K092641",
          "K060065",
          "K892217",
          "K203854",
          "K900843",
          "K863857",
          "K971946",
          "K872772",
          "K874132",
          "K884536",
          "K011156",
          "K971828",
          "K072581",
          "K141518",
          "K940933",
          "K053420",
          "K103685",
          "K231243",
          "K071024",
          "K003244",
          "K172492",
          "K873078",
          "K890033",
          "K000126",
          "K053419",
          "K221798",
          "K942842",
          "K971273",
          "K062338",
          "K094046",
          "K140682",
          "K090165",
          "K050056",
          "K052669",
          "K040214",
          "K200069",
          "K970098",
          "K083448",
          "K011974",
          "K923682",
          "K100203",
          "K172981",
          "K172141",
          "K994372",
          "K874492",
          "K030668",
          "K202776",
          "K181956",
          "K012706",
          "K201384",
          "K000794",
          "K172482",
          "K234110",
          "K041204",
          "K971235",
          "K901504",
          "K092652",
          "K030973",
          "K080348",
          "K032831",
          "K875031",
          "K833826",
          "K153021",
          "K091241",
          "K021959",
          "K893867",
          "K051736",
          "K833973",
          "K913790",
          "K122927",
          "K092101",
          "K063641",
          "K230277",
          "K013273",
          "K022766",
          "K190792",
          "K010451",
          "K181503",
          "K050252",
          "K071415",
          "K092950",
          "K871864",
          "K132402",
          "K001995",
          "K072295",
          "K952316",
          "K011670",
          "K972829",
          "K971827",
          "K051271",
          "K043359",
          "K862617",
          "K071672",
          "K061442",
          "K212300",
          "K925542",
          "K822909",
          "K070732",
          "K070408",
          "K092324",
          "K844256",
          "K151024",
          "K993637",
          "K871353",
          "K061931",
          "K890299",
          "K081285",
          "K991819",
          "K032971",
          "K200414",
          "K241086",
          "K061896",
          "K053269",
          "K181738",
          "K855027",
          "K180046",
          "K921454",
          "K924752",
          "K984570",
          "K242876",
          "K242580",
          "K103700",
          "K123871",
          "K062605",
          "K091572",
          "K020350",
          "K903523",
          "K213676",
          "K160268",
          "K083351",
          "K131047",
          "K924285",
          "K160582",
          "K201468",
          "K160231",
          "K122996",
          "K232495",
          "K963496",
          "K090599",
          "K874198",
          "K150093",
          "K113596",
          "K052693",
          "K083821",
          "K080255",
          "K000831",
          "K881883",
          "K003542",
          "K041644",
          "K121806",
          "K901774",
          "K101031",
          "K211632",
          "K241090",
          "K150314",
          "K223721",
          "K062455",
          "K222213",
          "K913695",
          "K123801",
          "K041674",
          "K103285",
          "K252655",
          "K871826",
          "K211140",
          "K910521",
          "K002939",
          "K032947",
          "K021778",
          "K113865",
          "K022168",
          "K252821",
          "K014218",
          "K213515",
          "K212153",
          "K022084",
          "K970769",
          "K141542",
          "K191279",
          "K902192",
          "K060675",
          "K061329",
          "K973147",
          "K152563",
          "K111621",
          "K991823",
          "K991410",
          "K212752",
          "K880756",
          "K032537",
          "K042705",
          "K133064",
          "K873081",
          "K973991",
          "K112843",
          "K890753",
          "K972786",
          "K960614",
          "K120984",
          "K981939",
          "K092838",
          "K013862",
          "K240808",
          "K881136",
          "K884329",
          "K040380",
          "K993512",
          "K091787",
          "K251691",
          "K932203",
          "K012770",
          "K220101",
          "K101871",
          "K132516",
          "K896738",
          "K992092",
          "K192608",
          "K233827",
          "K933755",
          "K030930",
          "K040178",
          "K014294",
          "K160940",
          "K221361",
          "K013792",
          "K920083",
          "K881820",
          "K102350",
          "K141362",
          "K023044",
          "K090662",
          "K113727",
          "K091057",
          "K100617",
          "K141024",
          "K140582",
          "K912880",
          "K911124",
          "K212385",
          "K032949",
          "K212555",
          "K082052",
          "K013319",
          "K943842",
          "K881307",
          "K161938",
          "K963707",
          "K243049",
          "K992455",
          "K001842",
          "K253109",
          "K902779",
          "K874104",
          "K082641",
          "K131111",
          "K901060",
          "K992238",
          "K980299",
          "K083892",
          "K923418",
          "K082487",
          "K911191",
          "K030952",
          "K070371",
          "K162675",
          "K251695",
          "K901819",
          "K910770",
          "K031465",
          "K161560",
          "K874692",
          "K003167",
          "K071610",
          "K954168",
          "K202173",
          "K240092",
          "K233956",
          "K081033",
          "K142959",
          "K162580",
          "K011365",
          "K020739",
          "K901612",
          "K153184",
          "K893587",
          "K102442",
          "K191207",
          "K935510",
          "K001688",
          "K090438",
          "K875298",
          "K080238",
          "K202324",
          "K180174",
          "K081659",
          "K211407",
          "K050018",
          "K102270",
          "K250837",
          "K021469",
          "K151206",
          "K910001",
          "K082888",
          "K192343",
          "K945290",
          "K945464",
          "K071285",
          "K900001",
          "K251693",
          "K101995",
          "K952222",
          "K032979",
          "K073249",
          "K913332",
          "K033715",
          "K232389",
          "K052829",
          "K960058",
          "K100403",
          "K980714",
          "K900757",
          "K011320",
          "K002036",
          "K210032",
          "K901209",
          "K090205",
          "K081712",
          "K052619",
          "K012989",
          "K172366",
          "K052544",
          "K905140",
          "K943817",
          "K903702",
          "K213431",
          "K022163",
          "K082268",
          "K051212",
          "K011851",
          "K202743",
          "K093853",
          "K863186",
          "K884614",
          "K030629",
          "K895425",
          "K152089",
          "K024235",
          "K001708",
          "K081800",
          "K111494",
          "K944608",
          "K112465",
          "K946185",
          "K061191",
          "K953126",
          "K982907",
          "K001770",
          "K920066",
          "K182822",
          "K100225",
          "K062149",
          "K100077",
          "K060576",
          "K050068",
          "K090360",
          "K200537",
          "K172792",
          "K082546",
          "K202851",
          "K073024",
          "K092678",
          "K101012",
          "K101600",
          "K062268",
          "K883297",
          "K884822",
          "K100428",
          "K870163",
          "K130947",
          "K191088",
          "K093306",
          "K132696",
          "K222597",
          "K981655",
          "K181352",
          "K083325",
          "K982776",
          "K893470",
          "K092549",
          "K010419",
          "K260931",
          "K942347",
          "K121914",
          "K964978",
          "K192900",
          "K101896",
          "K193350",
          "K093016",
          "K973887",
          "K173368",
          "K201179",
          "K990082",
          "K022819",
          "K914538",
          "K032755",
          "K884121",
          "K902730",
          "K062724",
          "K142832",
          "K020075",
          "K142865",
          "K873896",
          "K182887",
          "K083622",
          "K231979",
          "K862798",
          "K251751",
          "K133090",
          "K042536",
          "K982728",
          "K872520",
          "K952327",
          "K864613",
          "K874947",
          "K152390",
          "K082789",
          "K004044",
          "K031515",
          "K082846",
          "K203208",
          "K252805",
          "K913516",
          "K063783",
          "K212876",
          "K922075",
          "K201360",
          "K882664",
          "K071642",
          "K120657",
          "K181270",
          "K141128",
          "K002223",
          "K964055",
          "K993555",
          "K152091",
          "K933519",
          "K213430",
          "K942248",
          "K160865",
          "K970763",
          "K221282",
          "K042501",
          "K171241",
          "K042704",
          "K142888",
          "K121912",
          "K063434",
          "K965172",
          "K041647",
          "K982405",
          "K103584",
          "K021955",
          "K240305",
          "K191430",
          "K242455",
          "K893221",
          "K212665",
          "K201887",
          "K083787",
          "K121697",
          "K210274",
          "K081572",
          "K063140",
          "K253887",
          "K904088",
          "K141105",
          "K920966",
          "K043463",
          "K924282",
          "K051352",
          "K191403",
          "K911761",
          "K100815",
          "K032551",
          "K130320",
          "K863784",
          "K143216",
          "K041964",
          "K060143",
          "K024300",
          "K911207",
          "K211143",
          "K221953",
          "K954267",
          "K140785",
          "K073036",
          "K863855",
          "K252448",
          "K863918",
          "K043162",
          "K171423",
          "K872626",
          "K131021",
          "K864279",
          "K884891",
          "K083705",
          "K180685",
          "K031740",
          "K112804",
          "K101694",
          "K033998",
          "K191417",
          "K031742",
          "K072235",
          "K060616",
          "K101539",
          "K905310",
          "K061706",
          "K020479",
          "K962603",
          "K873859",
          "K973222",
          "K090671",
          "K864730",
          "K910811",
          "K123581",
          "K901071",
          "K041156",
          "K780625",
          "K013171",
          "K231531",
          "K934091",
          "K781879",
          "K213984",
          "K964821",
          "K042346",
          "K183556",
          "K873065",
          "K991661",
          "K992215",
          "K913489",
          "K000814",
          "K902758",
          "K101577",
          "K122046",
          "K232975",
          "K120773",
          "K091564",
          "K915699",
          "K904988",
          "K081125",
          "K990966",
          "K923343",
          "K031330",
          "K873624",
          "K973970",
          "K243086",
          "K022193",
          "K002682",
          "K162014",
          "K173123",
          "K103802",
          "K915773",
          "K870621",
          "K931304",
          "K970061",
          "K232520",
          "K953817",
          "K001085",
          "K030640",
          "K221979",
          "K210305",
          "K053477",
          "K211400",
          "K970368",
          "K014025",
          "K172625",
          "K033296",
          "K081937",
          "K001308",
          "K232989",
          "K894100",
          "K041815",
          "K002787",
          "K012533",
          "K961450",
          "K900866",
          "K163135",
          "K901399",
          "K820110",
          "K901066",
          "K002848",
          "K251696",
          "K041522",
          "K131818",
          "K993979",
          "K120502",
          "K873199",
          "K943456",
          "K992340",
          "K880672",
          "K972468",
          "K093881",
          "K924626",
          "K201294",
          "K223399",
          "K980879",
          "K173045",
          "K083557",
          "K862161",
          "K051257",
          "K072825",
          "K111403",
          "K092727",
          "K021089",
          "K110028",
          "K180380",
          "K201770",
          "K083466",
          "K022677",
          "K053130",
          "K233963",
          "K893643",
          "K092620",
          "K892120",
          "K040259",
          "K002574",
          "K250931",
          "K010966",
          "K070193",
          "K100020",
          "K190869",
          "K983684",
          "K860051",
          "K052377",
          "K120052",
          "K111888",
          "K163382",
          "K874442",
          "K060843",
          "K033349",
          "K093757",
          "K170965",
          "K002690",
          "K002477",
          "K060524",
          "K110893",
          "K011518",
          "K063752",
          "K052186",
          "K904039",
          "K172616",
          "K920082",
          "K033307",
          "K101568",
          "K012992",
          "K181006",
          "K102835",
          "K222010",
          "K042306",
          "K920379",
          "K862812",
          "K882273"
        ],
        "registration_number": [
          "",
          "3022418645",
          "3008935567",
          "3012494290",
          "3018188848",
          "3033715634",
          "3010984687",
          "3016171841",
          "3012243264",
          "3021131596",
          "3032948391",
          "3037007069",
          "3009888765",
          "3008361498",
          "3013652261",
          "9611031",
          "3003965134",
          "3012797688",
          "3013557562",
          "3016618143",
          "3011713222",
          "3018783526",
          "2134752",
          "3018685978",
          "3017953522",
          "3015286931",
          "3005569927",
          "3012733231",
          "3009673645",
          "3016859940",
          "3009096682",
          "2221819",
          "3008729547",
          "1651104",
          "1713468",
          "3004091281",
          "3003591740",
          "2025816",
          "3019737388",
          "1063925",
          "1417592",
          "3000126629",
          "3013319212",
          "3008820594",
          "3013896610",
          "3015529350",
          "3014766741",
          "3006733253",
          "1423537",
          "3011353843",
          "3006648320",
          "3026199641",
          "3012933969",
          "3024722701",
          "3011790297",
          "3015453963",
          "3038878471",
          "3010419931",
          "3003800159",
          "3015276780",
          "3015109923",
          "3013187142",
          "3026585241",
          "9611500",
          "3015181391",
          "1058397",
          "3030153385",
          "3009001657",
          "3013545943",
          "3029871637",
          "3014579161",
          "8020893",
          "2518422",
          "2030745",
          "3016595211",
          "9610861",
          "3000206696",
          "3023272766",
          "3004727197",
          "3006636961",
          "3007420856",
          "3010822231",
          "8044004",
          "3009241523",
          "3012104670",
          "3014541700",
          "3006125362",
          "3012302888",
          "3008505660",
          "3015455735",
          "3008700497",
          "1000122786",
          "9610105",
          "3016426850",
          "3030733800",
          "3015422337",
          "3016670149",
          "3001650535",
          "9610849",
          "8022257",
          "3010526986",
          "3001570235",
          "3004485927",
          "3027222147",
          "3030149975",
          "3010552191",
          "3013188547",
          "3024491912",
          "3002736372",
          "3018940143",
          "2081040",
          "3037341611",
          "9681377",
          "3010288146",
          "3008573045",
          "3012348571",
          "3013513377",
          "1650907",
          "3012404027",
          "2011171",
          "3003803088",
          "3005873568",
          "3016998644",
          "3010817335",
          "1314417",
          "8010482",
          "3012834717",
          "1218950",
          "3013679558",
          "3008733163",
          "3003879246",
          "3014015501",
          "3010317211",
          "3016661538",
          "3003674698",
          "2936999",
          "3006793151",
          "3013418500",
          "9617543",
          "3021632375",
          "3007836437",
          "3017520431",
          "3008830133",
          "3017953405",
          "3015616170",
          "3003741676",
          "3019944216",
          "9680215",
          "3007134734",
          "1218058",
          "2434008",
          "3031158",
          "3008716327",
          "1120986",
          "3007699081",
          "3011200334",
          "3006979678",
          "3011651103",
          "3009247326",
          "3009607734",
          "3008483389",
          "3017422354",
          "1320894",
          "3009831876",
          "3017825461",
          "9614860",
          "3044754336",
          "3026998289",
          "3012251275",
          "3011369688",
          "3019388613",
          "3015146106",
          "3032391",
          "3024512705",
          "3006182632",
          "3007048291",
          "3003622639",
          "1036863",
          "3003587684",
          "3008058135",
          "3038761440",
          "9710602",
          "1217183",
          "3014652024",
          "3010838917",
          "3016678045",
          "2939821",
          "3012528160",
          "3013596024",
          "3020163307",
          "3011898560",
          "2030633",
          "3007858728",
          "9614683",
          "3032109181",
          "3013500228",
          "3013116677",
          "3005798905",
          "3024379416",
          "3011420269",
          "3007358963",
          "3006552505",
          "2183613",
          "3014506106",
          "3003971136",
          "3014774380",
          "3043668257",
          "3009077524",
          "3018094310",
          "3015272694",
          "3014848734",
          "3013729860",
          "3007307487",
          "9617566",
          "2016493",
          "3006713849",
          "3027344406",
          "2032112",
          "3044675072",
          "3008164770",
          "3015328411",
          "3034405508",
          "3015697152",
          "3004145393",
          "3009191551",
          "3007583504",
          "1220908",
          "3026740087",
          "3003294644",
          "3003687189",
          "3043513123",
          "1062836",
          "3013517171",
          "3004016520",
          "3011721071",
          "3002807968",
          "3010834934",
          "3004835896",
          "1066270",
          "1054713",
          "3020460367",
          "3015876",
          "3000279201",
          "3010852191",
          "1720839",
          "3017406171",
          "3015997711",
          "3008363989",
          "3006807694",
          "3042156936",
          "3009551875",
          "1523530",
          "3010082909",
          "3009973336",
          "3010071455",
          "3012678246",
          "3036754877",
          "3014235520",
          "3012804266",
          "3044156918",
          "3012337868",
          "1063285",
          "1924066",
          "3012977166",
          "3001104093",
          "3043479346",
          "3014341624",
          "3001421318",
          "3030446844",
          "9617286",
          "3005380088",
          "3016931915",
          "3027256461",
          "3042941992",
          "3009156722",
          "3007607999",
          "3017446092",
          "1219324",
          "3034603279",
          "3011353016",
          "3010536822",
          "3012312061",
          "3008570547",
          "3020924771",
          "3007413079",
          "1419898",
          "3027510187",
          "3004977335",
          "3007662958",
          "3021244152",
          "3032013079",
          "3010157426",
          "3004149774",
          "1320360",
          "3004009431",
          "3018783553",
          "3012359877",
          "3013298431",
          "9615102",
          "3003832357",
          "3017894657",
          "3012725363",
          "3009485437",
          "3011359729",
          "2123774",
          "3007603826",
          "3011200896",
          "3017500903",
          "3004415095",
          "3018514222",
          "3010703925",
          "3009490946",
          "3006240053",
          "3009142030",
          "3041276591",
          "3003589956",
          "2183646",
          "3005783425",
          "3006395447",
          "8030978",
          "3026960443",
          "3035111606",
          "3009499478",
          "3007305624",
          "8021084",
          "3013783488",
          "1430016",
          "3010698831",
          "2030624",
          "3012238587",
          "3038429306",
          "2029015",
          "3011191255",
          "1722070",
          "3030447506",
          "3021343210",
          "1066427",
          "3016939722",
          "3036773433",
          "3003899230",
          "3013250653",
          "3014461810",
          "3015967131",
          "3010381606",
          "3010427901",
          "3021962893",
          "3006463573",
          "3015142815",
          "2022648",
          "3015542362",
          "3014233600",
          "3007490412",
          "3011065399",
          "3021559257",
          "1316463",
          "3004906872",
          "3010390468",
          "3009327812",
          "3013426279",
          "3014856521",
          "3011782009",
          "3008973759",
          "3015134004",
          "3023412068",
          "3007389703",
          "3006147705",
          "3016674095",
          "3010827261",
          "3033959233",
          "3008395508",
          "3002721930",
          "2112020",
          "3008717264",
          "8020045",
          "3003630387",
          "3014389131",
          "3029906224",
          "3010882708",
          "3008595074",
          "3004594267",
          "3021691064",
          "9614470",
          "3019470217",
          "9614500",
          "9610816",
          "3038185755",
          "3012263546",
          "3004102403",
          "3011167030",
          "3013689228",
          "3005877899",
          "3013826848",
          "3003263092",
          "1650347",
          "3016701404",
          "3015007456",
          "1000589001",
          "3016170326",
          "3020018",
          "3009139884",
          "3009617411",
          "3009350690",
          "3038259592",
          "3005004021"
        ],
        "fei_number": [
          "",
          "3022418645",
          "3008935567",
          "3012494290",
          "3018188848",
          "3033715634",
          "3010984687",
          "3016171841",
          "3012243264",
          "3021131596",
          "3032948391",
          "3037007069",
          "3009888765",
          "3008361498",
          "3013652261",
          "3003965134",
          "3012797688",
          "3013557562",
          "3016618143",
          "3011713222",
          "3018783526",
          "3018685978",
          "3017953522",
          "3015286931",
          "3005569927",
          "3012733231",
          "3009673645",
          "3016859940",
          "3003208775",
          "3009096682",
          "2221819",
          "3008729547",
          "1000403400",
          "1713468",
          "3004612204",
          "1000271197",
          "3003591740",
          "3002663379",
          "2025816",
          "3019737388",
          "1417592",
          "3000126629",
          "3013319212",
          "3008820594",
          "3013896610",
          "3015529350",
          "3014766741",
          "3006733253",
          "3001084824",
          "3002807391",
          "3011353843",
          "3006648320",
          "3026199641",
          "3012933969",
          "3024722701",
          "3007513035",
          "3011790297",
          "3015453963",
          "3038878471",
          "3010419931",
          "3003800159",
          "3015276780",
          "3015109923",
          "3013187142",
          "3026585241",
          "3015181391",
          "3030153385",
          "3009001657",
          "3013545943",
          "3029871637",
          "3014579161",
          "2518422",
          "3016595211",
          "2250033",
          "3000957797",
          "3000206696",
          "3023272766",
          "3004727197",
          "3006636961",
          "3007420856",
          "3010822231",
          "3009241523",
          "3012104670",
          "3014541700",
          "3006125362",
          "3012302888",
          "3008505660",
          "3015455735",
          "3008700497",
          "1000122786",
          "3003794634",
          "3016426850",
          "3000206435",
          "3030733800",
          "3015422337",
          "3016670149",
          "3001650535",
          "3003076831",
          "3010526986",
          "3001570235",
          "3004485927",
          "3027222147",
          "3030149975",
          "3013188547",
          "3003007822",
          "3024491912",
          "3018940143",
          "2081040",
          "3037341611",
          "3010288146",
          "3008573045",
          "3012348571",
          "1000065475",
          "3013513377",
          "3012404027",
          "2011171",
          "3003803088",
          "3005873568",
          "3016998644",
          "3010817335",
          "1314417",
          "3012834717",
          "1218950",
          "3013679558",
          "3008733163",
          "3003879246",
          "3014015501",
          "3010317211",
          "3016661538",
          "3003674698",
          "2936999",
          "3006793151",
          "3002806783",
          "3013418500",
          "3021632375",
          "3007836437",
          "3000215024",
          "3017520431",
          "3008830133",
          "3017953405",
          "3015616170",
          "3003741676",
          "3019944216",
          "1000117260",
          "1000158532",
          "3007134734",
          "1218058",
          "2434008",
          "3001236905",
          "3008716327",
          "1120986",
          "3007699081",
          "3011200334",
          "3006979678",
          "3002807019",
          "3011651103",
          "3009247326",
          "3009607734",
          "3008483389",
          "3017422354",
          "3009831876",
          "3002807167",
          "3017825461",
          "3044754336",
          "3026998289",
          "3012251275",
          "3011369688",
          "3019388613",
          "3015146106",
          "3024512705",
          "3006182632",
          "3007048291",
          "3003622639",
          "1036863",
          "3003587684",
          "3008058135",
          "3038761440",
          "1217183",
          "3014652024",
          "3010838917",
          "3016678045",
          "3012528160",
          "3013596024",
          "3020163307",
          "3011898560",
          "3007858728",
          "3032109181",
          "3013500228",
          "3013116677",
          "3005798905",
          "3024379416",
          "3011420269",
          "3007358963",
          "3006552505",
          "2183613",
          "3014506106",
          "3003971136",
          "3014774380",
          "3043668257",
          "3003757840",
          "3009077524",
          "1000160458",
          "3001451571",
          "3018094310",
          "3015272694",
          "3014848734",
          "3013729860",
          "3007307487",
          "1000124207",
          "2016493",
          "3006713849",
          "3027344406",
          "3044675072",
          "3008164770",
          "3015328411",
          "3034405508",
          "3015697152",
          "3004145393",
          "3009191551",
          "3007583504",
          "1220908",
          "3026740087",
          "3003294644",
          "3003687189",
          "3003244771",
          "3043513123",
          "3013517171",
          "3004016520",
          "3002807836",
          "3011721071",
          "3002807968",
          "2995",
          "3010834934",
          "3004835896",
          "6736",
          "3020460367",
          "3015876",
          "3000279201",
          "3010852191",
          "3017406171",
          "3015997711",
          "3008363989",
          "3006807694",
          "3042156936",
          "3003272793",
          "3000210620",
          "3009551875",
          "1523530",
          "3010082909",
          "3009973336",
          "3010071455",
          "3012678246",
          "3036754877",
          "3014235520",
          "3012804266",
          "3044156918",
          "3012337868",
          "1924066",
          "3012977166",
          "3001104093",
          "3043479346",
          "3014341624",
          "3001421318",
          "3010666513",
          "3030446844",
          "3005380088",
          "3016931915",
          "3027256461",
          "3042941992",
          "3009156722",
          "3007607999",
          "3017446092",
          "1219324",
          "3034603279",
          "3004631140",
          "3011353016",
          "3010536822",
          "3012312061",
          "3008570547",
          "3020924771",
          "3007413079",
          "1419898",
          "3027510187",
          "3004977335",
          "3007662958",
          "3021244152",
          "3032013079",
          "3010157426",
          "3004149774",
          "3002253377",
          "3004009431",
          "3018783553",
          "3012359877",
          "3013298431",
          "3003832357",
          "3017894657",
          "3012725363",
          "3009485437",
          "3011359729",
          "2123774",
          "3007603826",
          "3011200896",
          "3000204983",
          "3017500903",
          "3004415095",
          "3018514222",
          "3010703925",
          "3009490946",
          "3006240053",
          "3009142030",
          "3041276591",
          "3003589956",
          "2183646",
          "3005783425",
          "3006395447",
          "3026960443",
          "3035111606",
          "3009499478",
          "3007305624",
          "3013783488",
          "1430016",
          "3006254866",
          "3010698831",
          "3012238587",
          "3038429306",
          "3011191255",
          "3030447506",
          "3002459124",
          "3021343210",
          "3016939722",
          "3036773433",
          "3003899230",
          "3013250653",
          "3002807850",
          "3014461810",
          "3015967131",
          "3010381606",
          "3010427901",
          "3021962893",
          "3006463573",
          "3015142815",
          "1000300598",
          "2022648",
          "3000210521",
          "3015542362",
          "3014233600",
          "3007490412",
          "3011065399",
          "3003513129",
          "3021559257",
          "1316463",
          "3004906872",
          "3010390468",
          "3009327812",
          "3000718682",
          "3013426279",
          "3014856521",
          "1000114111",
          "3008973759",
          "3015134004",
          "3023412068",
          "3002299608",
          "1000292048",
          "3007389703",
          "3006147705",
          "3016674095",
          "3002681132",
          "3010827261",
          "3033959233",
          "3008395508",
          "2112020",
          "3008717264",
          "3003630387",
          "3014389131",
          "3029906224",
          "3010882708",
          "3023530182",
          "3008595074",
          "3015191855",
          "3002808182",
          "3004594267",
          "3021691064",
          "3000225562",
          "3019470217",
          "3038185755",
          "3012263546",
          "3004102403",
          "3011167030",
          "3013689228",
          "3005877899",
          "3013826848",
          "3003263092",
          "3016701404",
          "3015007456",
          "1000589001",
          "3016170326",
          "3001227743",
          "3020018",
          "3002737156",
          "3009139884",
          "3009617411",
          "3009350690",
          "3038259592",
          "3005004021"
        ],
        "device_name": "Oximeter",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.2700",
        "device_class": "2"
      }
    }
  ]
}