  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "141482",
      "product_res_number": "Z-0326-2016",
      "event_date_initiated": "2015-10-29",
      "event_date_created": "2015-11-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2019-04-29",
      "res_event_number": "72543",
      "product_code": "KDI",
      "k_numbers": [
        "K133547"
      ],
      "product_description": "NxStage System One S Cycler (High Permeability Hemodialysis System)\nModel no. NX1000-5",
      "code_info": "Software versions 4.9 and 4.10",
      "firm_fei_number": "3003464075",
      "recalling_firm": "NxStage Medical, Inc.",
      "address_1": "350 Merrimack Street",
      "city": "Lawrence",
      "state": "MA",
      "postal_code": "01843-1748",
      "additional_info_contact": "978-687-4700",
      "reason_for_recall": "Ultrafiltration (UF) Volume software error inaccurate fluid removal",
      "root_cause_description": "Software Design Change",
      "action": "NxStage issued  issued a Field Safety Notice (FSN) on 10/29/15 to consignees who have  NX1000-5 or a NX1000-5-A cycler. The FSN included Instructions for steps to be taken so that the cycler and treatments are not affected by the software error.\nNxStage will be sending a representative to the facility to attach detailed instructions on each cycler as a user reference. Return the attached reply form to acknowledge receipt of this letter.\nQuestions or comments, please contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243).",
      "product_quantity": "323 units",
      "distribution_pattern": "Worldwide Distribution.  US Nationwide, Netherlands, and Sweden.",
      "openfda": {
        "k_number": [
          "K003498",
          "K923727",
          "K811830",
          "K071854",
          "K834589",
          "K062079",
          "K946279",
          "K823881",
          "K040255",
          "K994198",
          "K062922",
          "K041424",
          "K914020",
          "K063183",
          "K820243",
          "K120017",
          "K053539",
          "K902837",
          "K991520",
          "K050603",
          "K210575",
          "K242155",
          "K162488",
          "K140191",
          "K173972",
          "K971180",
          "K232171",
          "K193482",
          "K023224",
          "K050609",
          "K830284",
          "K980386",
          "K261267",
          "K082414",
          "K012510",
          "K140526",
          "K101715",
          "K895759",
          "K040696",
          "K031378",
          "K231125",
          "K002277",
          "K090682",
          "K003023",
          "K091978",
          "K002210",
          "K933883",
          "K243237",
          "K133299",
          "K080650",
          "K864169",
          "K926568",
          "K003425",
          "K001156",
          "K080519",
          "K190793",
          "K093379",
          "K133547",
          "K982414",
          "K023509",
          "K043244",
          "K955384",
          "K945596",
          "K170790",
          "K840449",
          "K013733",
          "K131050",
          "K034064",
          "K964922",
          "K192707",
          "K923139",
          "K900018",
          "K041726",
          "K131908",
          "K830648",
          "K970253",
          "K830595",
          "K832547",
          "K100334",
          "K840680",
          "K981681",
          "K801479",
          "K081313",
          "K170469",
          "K243505",
          "K150880",
          "K963440",
          "K813591",
          "K041005",
          "K013562",
          "K060195",
          "K070512",
          "K013550",
          "K852251",
          "K843014",
          "K982926",
          "K860938",
          "K112314",
          "K853510",
          "K926379",
          "K252377",
          "K872767",
          "K945542",
          "K250508",
          "K062090",
          "K983085",
          "K010667",
          "K992573",
          "K853333",
          "K220721",
          "K911603",
          "K854586",
          "K951311",
          "K884365",
          "K970739",
          "K170574",
          "K242053",
          "K010131",
          "K030050",
          "K071518",
          "K883307",
          "K083158",
          "K811944",
          "K872456",
          "K801303",
          "K253518",
          "K921134",
          "K193005",
          "K020858",
          "K121409",
          "K171671",
          "K131516",
          "K970446",
          "K071298",
          "K001422",
          "K213639",
          "K992894",
          "K232467",
          "K973516",
          "K032356",
          "K945747",
          "K970591",
          "K121341",
          "K032431",
          "K030099",
          "K823589",
          "K021615",
          "K832659",
          "K780527",
          "K160881",
          "K070222",
          "K190910",
          "K023664",
          "K884326",
          "K942681",
          "K082515",
          "K140866",
          "K110823",
          "K071263",
          "K011221",
          "K130039",
          "K955592",
          "K001283",
          "K962427",
          "K990643",
          "K931720",
          "K243874",
          "K222952",
          "K960474",
          "K890315",
          "K103832",
          "K933818",
          "K183158",
          "K252459",
          "K812575",
          "K952983",
          "K171331",
          "K233335",
          "K070643",
          "K113023",
          "K801731",
          "K992563",
          "K143313",
          "K992564",
          "K153449",
          "K980656",
          "K910236",
          "K830596",
          "K892262",
          "K070320",
          "K855251",
          "K841152",
          "K083460",
          "K122347",
          "K131935",
          "K926373",
          "K800536",
          "K011537",
          "K080964",
          "K023874",
          "K823288",
          "K030974",
          "K901286",
          "K253412",
          "K992275",
          "K870724",
          "K001250",
          "K150708",
          "K123395",
          "K050952",
          "K061519",
          "K123262",
          "K201809",
          "K233798",
          "K955487",
          "K883407",
          "K002512",
          "K863823",
          "K914363",
          "K844790",
          "K010985",
          "K013418",
          "K240394",
          "K993064",
          "K051187",
          "K951344",
          "K030975",
          "K952982",
          "K021732",
          "K854011",
          "K924437",
          "K982760",
          "K871108",
          "K050525",
          "K883195",
          "K850979",
          "K113427",
          "K232776",
          "K152367",
          "K193542",
          "K192928",
          "K000028",
          "K211370",
          "K212216",
          "K973322",
          "K040489",
          "K051520",
          "K030592",
          "K093069",
          "K120505",
          "K982134",
          "K854309",
          "K050727",
          "K961465",
          "K832645",
          "K103564",
          "K093120",
          "K011269",
          "K863961",
          "K231406",
          "K230632",
          "K200741",
          "K945621",
          "K192756",
          "K243920",
          "K911094",
          "K061837",
          "K963203",
          "K893742",
          "K950522",
          "K953996",
          "K962707",
          "K203062",
          "K801480",
          "K900918",
          "K994267",
          "K920757",
          "K143482",
          "K823338",
          "K223248",
          "K952342",
          "K160558",
          "K024124",
          "K122051",
          "K222067",
          "K923312",
          "K841585",
          "K830647",
          "K060008",
          "K072232",
          "K000724",
          "K232803",
          "K850866",
          "K093902",
          "K844663",
          "K974584",
          "K982413",
          "K831784",
          "K122343",
          "K861106",
          "K123288",
          "K072093",
          "K162248",
          "K913950",
          "K163530",
          "K790827",
          "K945597",
          "K093964",
          "K153344",
          "K864587",
          "K830487",
          "K160444",
          "K210661",
          "K840543",
          "K954987",
          "K220281",
          "K152938",
          "K945625",
          "K980859",
          "K873887",
          "K945620",
          "K121497",
          "K132602",
          "K260533",
          "K231534",
          "K952978",
          "K233557",
          "K243264",
          "K861776",
          "K974652",
          "K895405",
          "K884353",
          "K111639",
          "K042938",
          "K131381",
          "K945595",
          "K083778",
          "K852885",
          "K903641",
          "K014152",
          "K781464",
          "K010805",
          "K992566",
          "K182940",
          "K193670"
        ],
        "device_name": "Dialyzer, High Permeability With Or Without Sealed Dialysate System",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.5860",
        "device_class": "2",
        "registration_number": [
          "9615942",
          "1225714",
          "3008307705",
          "3017552925",
          "3004201412",
          "8030665",
          "3014272786",
          "3031565407",
          "3003504604",
          "3012102437",
          "3027592507",
          "3010377171",
          "3020689844",
          "3005941719",
          "3016250252",
          "3010355846",
          "3006174295",
          "3005695808",
          "1721676",
          "1649518",
          "9610987",
          "3007137787",
          "3009128476",
          "8010182",
          "3040314808",
          "3000247873",
          "2032098",
          "9610270",
          "3038195011",
          "3004111573",
          "9680579",
          "3013764800",
          "9610825",
          "3015335038",
          "9612052",
          "3013536188",
          "3015136750",
          "3002807408",
          "8041145",
          "3013557562",
          "1713747",
          "3005578748",
          "1000286794",
          "3003589956",
          "3008102042",
          "3013188547",
          "3003661577",
          "3015472042",
          "8031561",
          "3014300273",
          "3006755813",
          "2183744",
          "3033903155",
          "3035708507",
          "9680841",
          "3015349756",
          "9681465",
          "3012149633",
          "3002879653",
          "9616074",
          "3004530184",
          "3003337893",
          "3014479313",
          "3010291427",
          "3038138522",
          "3032828299",
          "9611823",
          "9611824",
          "8030639",
          "3012494290",
          "1047843",
          "9611369",
          "3004176176",
          "3001744144",
          "3014579161",
          "3043014136",
          "3016096143",
          "1056186",
          "3010383847",
          "3014279513",
          "3021632375",
          "3008229990",
          "3005162618",
          "3011554160",
          "3003464075",
          "3016678045",
          "1000312731",
          "9616240",
          "3043226252"
        ],
        "fei_number": [
          "3008307705",
          "3017552925",
          "3001236277",
          "3004201412",
          "3014272786",
          "3031565407",
          "3012394756",
          "3002807710",
          "3002807035",
          "3003504604",
          "3002807111",
          "3012102437",
          "3027592507",
          "3010377171",
          "3002807204",
          "3002654698",
          "1000191953",
          "3002806960",
          "3020689844",
          "3016250252",
          "3010355846",
          "3006174295",
          "3005695808",
          "3002806443",
          "1721676",
          "3002924544",
          "3007137787",
          "3009128476",
          "3040314808",
          "3002807005",
          "3000247873",
          "3038195011",
          "3004111573",
          "1000203464",
          "3013764800",
          "3015335038",
          "3013536188",
          "3015136750",
          "3002807408",
          "3013557562",
          "1713747",
          "3005578748",
          "1000286794",
          "3003589956",
          "3013188547",
          "3002807113",
          "3003661577",
          "3015472042",
          "3002807744",
          "3014300273",
          "6898",
          "3006755813",
          "2183744",
          "3033903155",
          "3001451489",
          "3035708507",
          "3015349756",
          "3012149633",
          "3002879653",
          "3004530184",
          "3003337893",
          "3014479313",
          "1000132332",
          "3010291427",
          "3038138522",
          "3032828299",
          "3002807721",
          "3012494290",
          "1047843",
          "3004176176",
          "3001744144",
          "3014579161",
          "3002974635",
          "3043014136",
          "3002806469",
          "3016096143",
          "3010383847",
          "3014279513",
          "3021632375",
          "3008229990",
          "3011554160",
          "3003464075",
          "3016678045",
          "1000312731",
          "1527681",
          "3043226252",
          "3002807196",
          "3005426550",
          "3003058448"
        ]
      }
    }
  ]
}