{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "138520",
      "product_res_number": "Z-0810-2016",
      "event_date_initiated": "2015-07-06",
      "event_date_posted": "2016-02-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-02-03",
      "res_event_number": "71647",
      "product_code": "IZI",
      "k_numbers": [
        "K033737",
        "K041949",
        "K130842"
      ],
      "product_description": "Actuator for angiographic x-ray monitor ceiling suspension system.  Actuator short Assy: THC8AWDS-121 and Actuator long Assy: THC8AWDS-122 used with the following products: 722001  Allura Xper FD 10 C; 722003 Allura Xper FD10; 722005 Allura Xper FD10/10; 722006 Allura Xper FD20; 722008 Allura Xper FD20 Biplane; 722010 Allura Xper FD10; 722011 Allura Xper FD10/10; 722012 Allura Xper FD20; 722013 Allura XPER FD20 BIPLANE; 722015 Allura Xper FD20 OR Table; 722020 Allura Xper FD20 Biplane OR Table; 722022 Allura Xper FD10 OR Table; 722023 Allura Xper FD20 OR Table; 722025 Allura Xper FD20 Biplane OR Table; 722026 Allura Xper FD10; 722027 Allura Xper FD10/10; 722028 Allura Xper FD20; 722029 Allura Xper FD20/10; 722035 Allura Xper FD20 OR Table; 722038 Allura Xper FD20/20; 722039 Allura Xper FD20/20 OR Table; 722058 Allura Xper FD20/15; 722123 Field extensions Xper cardio R7.6; 722124 Field extensions Xper vascular R7.6; 722133 Field ext. Xper cardio systems R7.6; 722134 Field ext. Xper vascular systems R7.6; 72249 INTEGRIS BV5000; 722043 INTEGRIS Allura 15-12 (mono); 722121 Field extensions INTEGRIS cardio systems",
      "code_info": "Actuator short Assy: THC8AWDS-121 and Actuator long Assy: THC8AWDS-122",
      "firm_fei_number": "1218950",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "city": "Andover",
      "state": "MA",
      "postal_code": "01810-1032",
      "additional_info_contact": "800-722-9377",
      "reason_for_recall": "The actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest position.",
      "root_cause_description": "Process control",
      "action": "Philips sent a Field Safety Notice to all affected customers on July 6, 2015.  The notice informs consignees of the reason for recall and actions to be taken by the customer/user. On September 15, 2015, Philips plans to issued a mandatory Field Safety Corrective action (FCO72200280) via Certified Return Receipt.  This corrective action is the replacement of the affected MCS actuator. For further information, contact your local representative: Technical Support Line: 1-800-722-9377.",
      "product_quantity": "3158 devices",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of : Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, Netherlands Antilles, New Zealand, Norway, Oman, Pakistan, occupied Palestinian territory, Panama, PERU, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom and Viet Nam.",
      "openfda": {
        "k_number": [
          "K972136",
          "K130003",
          "K850455",
          "K915199",
          "K955519",
          "K875376",
          "K945400",
          "K875374",
          "K897115",
          "K875184",
          "K913685",
          "K182907",
          "K955179",
          "K984545",
          "K954581",
          "K991231",
          "K972655",
          "K810183",
          "K812238",
          "K222240",
          "K875375",
          "K880690",
          "K100371",
          "K892849",
          "K935026",
          "K170239",
          "K932410",
          "K910654",
          "K931184",
          "K151780",
          "K933058",
          "K062360",
          "K925961",
          "K180019",
          "K190772",
          "K993765",
          "K945449",
          "K945540",
          "K073290",
          "K842494",
          "K041027",
          "K905653",
          "K953563",
          "K230727",
          "K073088",
          "K082907",
          "K960575",
          "K974367",
          "K010721",
          "K934496",
          "K253181",
          "K903941",
          "K130842",
          "K973482",
          "K922222",
          "K042646",
          "K993947",
          "K052076",
          "K920434",
          "K945375",
          "K920705",
          "K993763",
          "K033184",
          "K041949",
          "K945459",
          "K903371",
          "K133993",
          "K090745",
          "K955395",
          "K970738",
          "K072591",
          "K073328",
          "K875380",
          "K261037",
          "K971489",
          "K982203",
          "K883860",
          "K010391",
          "K931170",
          "K926082",
          "K946378",
          "K903493",
          "K942120",
          "K883831",
          "K961451",
          "K875379",
          "K234090",
          "K945483",
          "K883072",
          "K923813",
          "K781991",
          "K083745",
          "K912416",
          "K113034",
          "K983877",
          "K896088",
          "K060749",
          "K141136",
          "K974679",
          "K936002",
          "K943355",
          "K904447",
          "K944952",
          "K903373",
          "K031333",
          "K042961",
          "K940484",
          "K214097",
          "K884008",
          "K921101",
          "K875372",
          "K913120",
          "K875373",
          "K905518",
          "K133719",
          "K941946",
          "K924708",
          "K924828",
          "K973368",
          "K080612",
          "K202391",
          "K941177",
          "K914376",
          "K100468",
          "K914313",
          "K923112",
          "K940172",
          "K941979",
          "K231986",
          "K945293",
          "K002840",
          "K970525",
          "K061871",
          "K991922",
          "K933664",
          "K072222",
          "K890352",
          "K971181",
          "K920862",
          "K162991",
          "K922804",
          "K063345",
          "K071037",
          "K111209",
          "K231075",
          "K942503",
          "K113594",
          "K903375",
          "K972229",
          "K052039",
          "K073130",
          "K930635",
          "K911732",
          "K812721",
          "K071717",
          "K924719",
          "K933460",
          "K915232",
          "K060867",
          "K192761",
          "K890348",
          "K852331",
          "K903372",
          "K874910",
          "K853044",
          "K933073",
          "K110256",
          "K243077",
          "K904540",
          "K071619",
          "K243645",
          "K971365",
          "K954309",
          "K834496",
          "K971176",
          "K882358",
          "K032443",
          "K181537",
          "K023441",
          "K023970",
          "K030139",
          "K926258",
          "K884912",
          "K072006",
          "K052500",
          "K811149",
          "K973270",
          "K100849",
          "K813588",
          "K943545",
          "K842495",
          "K973342",
          "K812913",
          "K910370",
          "K924137",
          "K922622",
          "K923654",
          "K902542",
          "K952704"
        ],
        "registration_number": [
          "3015224777",
          "3014663814",
          "3007487007",
          "3006990600",
          "3009057691",
          "3017820152",
          "3042940623",
          "3015806723",
          "9610612",
          "3003755939",
          "3016607092",
          "3014321789",
          "3036666444",
          "3013700547",
          "3010684280",
          "3009963139",
          "3015180671",
          "3010157088",
          "3013868968",
          "3007138831",
          "8043903",
          "3015532525",
          "3012883081",
          "3020746799",
          "3009077524",
          "3011063460",
          "3006498370",
          "3031594238",
          "3003202425",
          "3007055152",
          "1220246",
          "3010193531",
          "1625425",
          "3010493171",
          "2936485",
          "3042243691",
          "3016851379",
          "3012470322",
          "3019269298",
          "3012627460",
          "1644312",
          "3031503890",
          "9710122",
          "3004519921",
          "3021393544",
          "2648727",
          "8030233",
          "3004111573",
          "3045816907",
          "3004026651",
          "3002579136",
          "3026934906",
          "3003974370",
          "1048735",
          "2523676",
          "3021226419",
          "2247992",
          "1000181430",
          "3011416394",
          "2521402",
          "9615010",
          "3008632256",
          "3015231789",
          "3015895145",
          "3015276088",
          "1018470",
          "3008284470",
          "3016555232",
          "3008309062",
          "3021336182",
          "3004122598",
          "3013767421",
          "3012127758",
          "3023003446",
          "3009715136",
          "3013665681",
          "3005369776",
          "3012329552",
          "3006972752",
          "3012624817",
          "3003637092",
          "3027339877",
          "3010326005",
          "3009719225",
          "1422634",
          "3000126629",
          "3027373189",
          "1043214",
          "3012075008",
          "9614683",
          "9613445",
          "3005810333",
          "3014004349",
          "3029968414",
          "3016832577",
          "3008496839",
          "3022626486",
          "3030926959",
          "3010331645",
          "3004977335",
          "3014635169",
          "9614698",
          "3015173212",
          "1061124",
          "3014262693",
          "3015232217",
          "1054713",
          "1037885",
          "3014644276",
          "3012761544",
          "3005877899",
          "3007289408",
          "3013283620",
          "3013077751",
          "1055236",
          "3011230048",
          "3010400367",
          "3020331900",
          "3016727307",
          "3020899789",
          "1036836",
          "3042175844",
          "3021059265",
          "3035950510",
          "3015910259",
          "3015974593",
          "3013764800",
          "3014343100",
          "3012345110",
          "8020997",
          "3033527282",
          "3038614464",
          "1319639",
          "3015612252",
          "3000976525",
          "3017619652",
          "2916714",
          "3012939903",
          "3020704423",
          "1054241",
          "9611343",
          "3026606306",
          "8010047",
          "3010885402",
          "1000204638",
          "3026391023",
          "3001675293",
          "3015153469",
          "3003768277"
        ],
        "fei_number": [
          "3002807795",
          "1000138054",
          "3015224777",
          "3014663814",
          "3007487007",
          "3006990600",
          "3009057691",
          "3017820152",
          "3042940623",
          "3015806723",
          "3003755939",
          "3016607092",
          "3014321789",
          "3036666444",
          "1000514656",
          "3013700547",
          "3010684280",
          "3009963139",
          "3015180671",
          "3010157088",
          "3002806315",
          "3013868968",
          "3007138831",
          "3005846326",
          "3015532525",
          "3012883081",
          "3020746799",
          "3009077524",
          "3011063460",
          "3006498370",
          "3031594238",
          "3003202425",
          "3007055152",
          "1220246",
          "3010193531",
          "1625425",
          "3010493171",
          "2936485",
          "1000114111",
          "3042243691",
          "1000582291",
          "3016851379",
          "3012470322",
          "3019269298",
          "3012627460",
          "3031503890",
          "3021393544",
          "2648727",
          "3004111573",
          "3045816907",
          "1048735",
          "3002579136",
          "3026934906",
          "3003974370",
          "1000284313",
          "2523676",
          "3021226419",
          "1000181430",
          "3011416394",
          "2521402",
          "3008632256",
          "3015231789",
          "3015895145",
          "3006149253",
          "3015276088",
          "1018470",
          "3008284470",
          "3016555232",
          "3008309062",
          "3021336182",
          "3004122598",
          "3013767421",
          "3012127758",
          "3023003446",
          "3009715136",
          "3013665681",
          "3005369776",
          "3012329552",
          "3006972752",
          "3012624817",
          "3003637092",
          "3027339877",
          "3010326005",
          "3009719225",
          "3003794634",
          "3000126629",
          "3003481798",
          "3027373189",
          "1043214",
          "3002808139",
          "1000112968",
          "3012075008",
          "3002808088",
          "3005810333",
          "3014004349",
          "3029968414",
          "3016832577",
          "3008496839",
          "1000512168",
          "3022626486",
          "3030926959",
          "3010331645",
          "3004977335",
          "3014635169",
          "3015173212",
          "3014262693",
          "3015232217",
          "1037885",
          "3012761544",
          "3005877899",
          "3007289408",
          "3013283620",
          "1000117748",
          "3013077751",
          "1000208827",
          "3011230048",
          "1000112877",
          "3010400367",
          "3020331900",
          "3016727307",
          "3020899789",
          "1036836",
          "3042175844",
          "3021059265",
          "3035950510",
          "3003595527",
          "3015910259",
          "3015974593",
          "3013764800",
          "3014343100",
          "3002807880",
          "3012345110",
          "3033527282",
          "3038614464",
          "3015612252",
          "3000976525",
          "3002807092",
          "3017619652",
          "3002175190",
          "2916714",
          "3012939903",
          "3020704423",
          "3026606306",
          "3010885402",
          "1000204638",
          "3026391023",
          "3002808382",
          "3001675293",
          "3015153469"
        ],
        "device_name": "System, X-Ray, Angiographic",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1600",
        "device_class": "2"
      }
    }
  ]
}