      "cfres_id": "131680",
      "product_res_number": "Z-0659-2015",
      "event_date_initiated": "2014-11-17",
      "event_date_posted": "2014-12-12",
      "recall_status": "Terminated",
      "event_date_terminated": "2015-07-17",
      "res_event_number": "69838",
      "product_code": "KOG",
      "product_description": "PKG, BIOPSY FORCEPS, PUNCH, P/N 0250080290. \nLaparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "code_info": "842133 940165 941352 942845 944667 944703 944720 944832 1040108 1042637 1042643 1042645 1046592 1140812 1141975 1142916 1143079 1144213 1144312 1144396 1145815 1146345 1146677 1146688 1246120 1246127 1343126 1344291 1345039 1345098",
      "firm_fei_number": "2936485",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "city": "San Jose",
      "state": "CA",
      "postal_code": "95138-1400",
      "additional_info_contact": "408-754-2000",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "root_cause_description": "Error in labeling",
      "action": "Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers.   The letter informed the customers of the potential hazards and actions required.  Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories.  Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below:\nwww.stryker.com/IFU1000401070\nCustomers were instructed to complete and sign the enclosed \"Acknowledgement of Receipt\" form and email them to endorecall@stryker.com or fax to (408) 855-6314.\nCustomers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com.\nFor questions regarding this recall call 408-754-2000.",
      "product_quantity": "180,573 units total",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",