      "cfres_id": "131654",
      "product_res_number": "Z-0633-2015",
      "event_date_initiated": "2014-11-17",
      "event_date_posted": "2014-12-12",
      "recall_status": "Terminated",
      "event_date_terminated": "2015-07-17",
      "res_event_number": "69838",
      "product_code": "KOG",
      "product_description": "PKG, CLAW FORCEPS, DOUBLE ACTION, P/N 0250080245. \n\nLaparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "code_info": "43240 143822 646341 841071 845057 847851 848110 940791 941351 943551 943748 944835 945745 1040099 1040976 1041991 1042472 1042687 1042702 1042706 1141976 1142463 1143133 1144329 1144624 1145286 1145818 1146347 1146686 1240540 1241942 1241943 1242475 1243542 1243543 1244257 1341592 1341593 1341597 1345522 1346223 1346539 1440407 1440672 1440673 1440892 1440893",
      "firm_fei_number": "2936485",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "city": "San Jose",
      "state": "CA",
      "postal_code": "95138-1400",
      "additional_info_contact": "408-754-2000",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "root_cause_description": "Error in labeling",
      "action": "Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers.   The letter informed the customers of the potential hazards and actions required.  Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories.  Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below:\nwww.stryker.com/IFU1000401070\nCustomers were instructed to complete and sign the enclosed \"Acknowledgement of Receipt\" form and email them to endorecall@stryker.com or fax to (408) 855-6314.\nCustomers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com.\nFor questions regarding this recall call 408-754-2000.",
      "product_quantity": "180,573 units total",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "openfda": {}
    }
  ]
}