  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "128639",
      "product_res_number": "Z-0037-2015",
      "event_date_initiated": "2014-07-09",
      "event_date_posted": "2014-11-06",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-04-24",
      "res_event_number": "68777",
      "product_code": "KNH",
      "pma_numbers": [
        "P870080"
      ],
      "product_description": "The Hulka Clip is a sterile packaged medical device that carries a two year expiration date. It is packaged two clips per procedural tray. Theses procedural trays are then sealed in a Tyvek/Mylar pouch and packaged in a cardboard shipping carton with ten pouches in each shipping carton.",
      "code_info": "Product Code: 4986.09;  Lot: 51000232",
      "firm_fei_number": "1418479",
      "recalling_firm": "Richard Wolf Medical Instruments Corp.",
      "address_1": "353 Corporate Woods Pkwy",
      "city": "Vernon Hills",
      "state": "IL",
      "postal_code": "60061-3110",
      "additional_info_contact": "Ms. Lisa Williams\n847-913-1113 Ext. 225",
      "reason_for_recall": "Richard Wolf Medical Instrument Corporation is recalling Hulka Clip Tubal Occluding Device lot 51000232. The devices are being recalled as a part of an investigation into a biocompatibility test discrepancy where the recalled lot (lot 51000232) passed final release testing, but the subsequent lot failed biocompatibility testing on a component that was also used to manufacture lot 51000232.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "\"Urgent: Medical Device Recall\" letters dated July 9, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to: 1) inspect their inventories for the recalled products; 2) contact Richard Wolf Medical Instruments at 800-323-9653 x 225 or 380 to notify Richard Wolf Medical Instruments if you have any devices in inventory; 3) review your records and report the number, if any, of Hulka Clips from lot 51000232 that were implanted in your patients on the Recall Response Form; 4) complete and return the Recall Response Form attached to the letter; and, 5) a RMA number will be provided to customers returning the recalled products and those customers will receive a credit. All returned product should be sent to the attention of Ms. Lisa Williams, Regulatory Specialist. Customers with questions can contact Richard Wolf Medical Instruments at 800-323-9653 x 225 or 380.",
      "product_quantity": "73 cartons (1,460 clips)",
      "distribution_pattern": "Worldwide distribution: US (Nationwide) including states of: (AK, AL, CO, CT, FL, GA, IA, IN, KY, LA, MI, MO, MS, NC, NY, OH, OR, PA, SC, TN, TX, VA, WA, WV;  and country of: Germany.",
      "openfda": {
        "pma_number": [
          "P870076",
          "P920046",
          "P870080",
          "P070022"
        ],
        "registration_number": [
          "3008812251",
          "3002807314",
          "1314417",
          "3011050570",
          "8021955",
          "9616675"
        ],
        "fei_number": [
          "3003105967",
          "3002807314",
          "1314417",
          "3002806976",
          "3011050570",
          "3002808293"
        ],
        "device_name": "Laparoscopic Contraceptive Tubal Occlusion Device",
        "medical_specialty_description": "Obstetrics/Gynecology",
        "regulation_number": "884.5380",
        "device_class": "3"
      }
    }
  ]
}