      "cfres_id": "125015",
      "product_res_number": "Z-0906-2014",
      "event_date_initiated": "2013-09-11",
      "event_date_posted": "2014-02-03",
      "recall_status": "Terminated",
      "event_date_terminated": "2014-08-04",
      "res_event_number": "67215",
      "product_code": "LPH",
      "k_numbers": [
        "K101761",
        "K113609"
      ],
      "product_description": "***REF 186-01-08***InteGrip ACETABULAR AUGMENT***SMALL 8mm. Use with 48/50mm Shell.***EXACTECH, Gainesville, FL 32653-1630.\n\nThe Exactech Novation InteGrip Acetabular Augments are indicated for use in skeletally mature individuals undergoing primary or revision surgery for hip replacement and whose orthopedic surgeon desires a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies.",
      "code_info": "Catalog Numbers: 186-01-08, 186-01-11, 186-01-13, 186-02-08, 186-02-11, 186-02-13, 186-03-11, 186-03-08, 186-03-11, 186-03-13, 186-04-08, 186-04-11, 186-04-13, 186-05-08, 186-05-11, 186-05-13.",
      "firm_fei_number": "1038671",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "city": "Gainesville",
      "state": "FL",
      "postal_code": "32653-1630",
      "additional_info_contact": "Kaya Davis\n352-377-1140",
      "reason_for_recall": "Exactech is recalling the InteGrip Acetabular Augments due to an out of range condition for an in-vitro biological evaluation standard.",
      "root_cause_description": "Employee error",
      "action": "Exactech sent an \"Important Product Recall Notice\" dated September 11, 2013,  to all affected customers. The letter identified the product and the action needed to be taken by the customers.\n\n Customers were instructed to cease distribution or use of the products. Extend this information to your accounts that may have this product in their possession. Verify if you have any of the subject InteGrip Acetabular Augments (catalog numbers 186-01-08 to 186-05-13) in the specified lots. Complete and fax back the attached form. Further questions, please call 1-800-392-2832.",
      "product_quantity": "235",
      "distribution_pattern": "USA Distribution including the states of : FL, VA, NY, OH, ME, TX, CO, and GA.",
      "openfda": {
        "k_number": [
          "K173951",
          "K951674",
          "K100933",
          "K031495",
          "K954466",
          "K261624",
          "K973302",
          "K180798",
          "K003237",
          "K141001",
          "K042774",
          "K172501",
          "K110555",
          "K042989",
          "K953694",
          "K133228",
          "K251718",
          "K000321",
          "K001039",
          "K252846",
          "K932755",
          "K132697",
          "K112779",
          "K963946",
          "K994070",
          "K921301",
          "K232667",
          "K073150",
          "K182109",
          "K242084",
          "K123012",
          "K022556",
          "K141797",
          "K073068",
          "K143009",
          "K990003",
          "K111472",
          "K172456",
          "K934765",
          "K031110",
          "K172684",
          "K170444",
          "K003199",
          "K934412",
          "K072888",
          "K852198",
          "K001991",
          "K010170",
          "K161184",
          "K921274",
          "K964882",
          "K921113",
          "K960660",
          "K070278",
          "K020153",
          "K061833",
          "K191358",
          "K192450",
          "K021911",
          "K991026",
          "K974506",
          "K030236",
          "K935875",
          "K151448",
          "K981211",
          "K000306",
          "K953664",
          "K003875",
          "K132626",
          "K191622",
          "K202568",
          "K121297",
          "K200011",
          "K001320",
          "K033893",
          "K190123",
          "K130063",
          "K190656",
          "K973119",
          "K182678",
          "K000788",
          "K963998",
          "K941612",
          "K993027",
          "K053557",
          "K052996",
          "K222436",
          "K030733",
          "K110542",
          "K022382",
          "K933787",
          "K172551",
          "K951000",
          "K964924",
          "K932923",
          "K113242",
          "K003429",
          "K973163",
          "K010367",
          "K951356",
          "K935484",
          "K243456",
          "K933102",
          "K172806",
          "K171934",
          "K060581",
          "K984408",
          "K021346",
          "K122798",
          "K033273",
          "K960303",
          "K951114",
          "K003181",
          "K152084",
          "K934515",
          "K131454",
          "K860442",
          "K030048",
          "K113609",
          "K011081",
          "K970706",
          "K133370",
          "K101072",
          "K032385",
          "K101229",
          "K142462",
          "K241636",
          "K961939",
          "K961625",
          "K040191",
          "K980020",
          "K092098",
          "K212815",
          "K112898",
          "K072154",
          "K973675",
          "K004032",
          "K112158",
          "K020580",
          "K250989",
          "K103721",
          "K990135",
          "K183457",
          "K931333",
          "K110245",
          "K140547",
          "K190057",
          "K991807",
          "K220468",
          "K101976",
          "K982367",
          "K172857",
          "K960172",
          "K171768",
          "K221636",
          "K061226",
          "K130149",
          "K241059",
          "K971429",
          "K991485",
          "K094045",
          "K182468",
          "K142119",
          "K052255",
          "K994286",
          "K200607",
          "K061124",
          "K042051",
          "K964650",
          "K032726",
          "K242744",
          "K061434",
          "K964227",
          "K862326",
          "K141395",
          "K954698",
          "K052321",
          "K223223",
          "K042252",
          "K955271",
          "K921525",
          "K222066",
          "K974093",
          "K143453",
          "K151440",
          "K993082",
          "K896637",
          "K142606",
          "K100269",
          "K000335",
          "K213842",
          "K103256",
          "K182099",
          "K252623",
          "K123550",
          "K033580",
          "K153345",
          "K994038",
          "K212761",
          "K000760",
          "K984492",
          "K022520",
          "K926156",
          "K926107",
          "K221675",
          "K182502",
          "K030055",
          "K972596",
          "K921114",
          "K212187",
          "K060635",
          "K050327",
          "K050937",
          "K062148",
          "K122512",
          "K002747",
          "K140669",
          "K242045",
          "K955871",
          "K953490",
          "K243192",
          "K103384",
          "K103124",
          "K210614",
          "K984296",
          "K233498",
          "K202472",
          "K142746",
          "K121263",
          "K070369",
          "K131279",
          "K001534",
          "K102399",
          "K922159",
          "K000521",
          "K161569",
          "K972393",
          "K021109",
          "K983509",
          "K071160",
          "K010171",
          "K120924",
          "K190276",
          "K240046",
          "K062775",
          "K182221",
          "K120599",
          "K993259",
          "K072101",
          "K200391",
          "K121308",
          "K043123",
          "K041906",
          "K150790",
          "K833439",
          "K043537",
          "K231522",
          "K921400",
          "K160266",
          "K013475",
          "K132879",
          "K030501",
          "K040383",
          "K063822",
          "K963266",
          "K201348",
          "K973124",
          "K931642",
          "K162127",
          "K974740",
          "K042776",
          "K162957",
          "K143314",
          "K141960",
          "K022463",
          "K123019",
          "K002149",
          "K053417",
          "K031402",
          "K934334",
          "K951095",
          "K101336",
          "K103431",
          "K030979",
          "K061547",
          "K933882",
          "K193233",
          "K961186",
          "K010120",
          "K020497",
          "K031572",
          "K113296",
          "K122099",
          "K000538",
          "K040412",
          "K971854",
          "K031693",
          "K141043",
          "K941141",
          "K980981",
          "K121777",
          "K964509",
          "K150007",
          "K010241",
          "K200854",
          "K963509",
          "K163193",
          "K942115",
          "K100555",
          "K920955",
          "K060072",
          "K242375",
          "K090897",
          "K955033",
          "K033338",
          "K080625",
          "K070050",
          "K971206",
          "K951741",
          "K131022",
          "K030861",
          "K022902",
          "K183582",
          "K992667",
          "K042871",
          "K041950",
          "K923452",
          "K072417",
          "K990830",
          "K192919",
          "K936292",
          "K121935",
          "K061786",
          "K023409",
          "K833464",
          "K862702",
          "K060630",
          "K171073",
          "K132959",
          "K983834",
          "K172251",
          "K220376",
          "K000334",
          "K973714",
          "K851421",
          "K954288",
          "K243784",
          "K132207",
          "K926201",
          "K991123",
          "K991325",
          "K150522",
          "K120030",
          "K003447",
          "K921639",
          "K102019",
          "K023375",
          "K021878",
          "K983491",
          "K110947",
          "K963100",
          "K965164",
          "K101761",
          "K232556",
          "K161080",
          "K252067",
          "K954417",
          "K203557",
          "K974095",
          "K162719",
          "K001277",
          "K060040",
          "K061168",
          "K951115",
          "K122504",
          "K200656",
          "K141676",
          "K021891",
          "K053293",
          "K190904",
          "K102423",
          "K955511",
          "K103662",
          "K192198",
          "K171962",
          "K971718",
          "K121034",
          "K936310",
          "K962769",
          "K952740",
          "K962923",
          "K924492",
          "K012364",
          "K002960",
          "K970501",
          "K081566",
          "K970705",
          "K922500",
          "K183129",
          "K191826",
          "K971824",
          "K980711",
          "K101575",
          "K142201",
          "K190660",
          "K921277",
          "K953337",
          "K940715",
          "K922071",
          "K223348",
          "K860635",
          "K191936",
          "K051556",
          "K955563",
          "K081679",
          "K992109",
          "K960180",
          "K202484",
          "K041109",
          "K020170",
          "K060031",
          "K992154",
          "K061461",
          "K990849",
          "K021466",
          "K102975",
          "K833681",
          "K944592",
          "K964769",
          "K122244",
          "K142882",
          "K983502",
          "K100445",
          "K191002",
          "K952397",
          "K994078",
          "K121024",
          "K014070",
          "K932924",
          "K971020",
          "K111546",
          "K191708",
          "K051491",
          "K001982",
          "K012787",
          "K211176",
          "K943054",
          "K012507",
          "K251043",
          "K220336",
          "K950827",
          "K955170",
          "K032964",
          "K061067",
          "K121636",
          "K022711",
          "K030079",
          "K071784",
          "K132158",
          "K033716",
          "K954213",
          "K920831",
          "K131237",
          "K161073",
          "K973530",
          "K081171",
          "K251270",
          "K181366",
          "K220959",
          "K130099",
          "K993736",
          "K963101",
          "K092443",
          "K223779",
          "K151603",
          "K970300",
          "K921557",
          "K001778",
          "K072817",
          "K040544",
          "K133832",
          "K161190",
          "K983837",
          "K052026",
          "K133386",
          "K110135",
          "K021970",
          "K121684",
          "K943230",
          "K192071",
          "K955739",
          "K934060",
          "K211866",
          "K132572",
          "K003800",
          "K964616",
          "K032581",
          "K093561",
          "K994415",
          "K062419",
          "K211047",
          "K150503",
          "K072963",
          "K010282",
          "K151424",
          "K003478",
          "K954811",
          "K013658",
          "K971018",
          "K983035",
          "K132312",
          "K042337",
          "K921308",
          "K230011",
          "K984459",
          "K961526",
          "K254043",
          "K062489",
          "K210893",
          "K151264",
          "K052089",
          "K251833",
          "K201733",
          "K242833",
          "K041850",
          "K020145",
          "K951301",
          "K120241",
          "K963309",
          "K935731",
          "K190847",
          "K974294",
          "K042029",
          "K983014",
          "K211330",
          "K243977",
          "K242232",
          "K012739",
          "K970241",
          "K982248",
          "K023357",
          "K143085",
          "K032729",
          "K033034",
          "K001660",
          "K931641",
          "K990523",
          "K925172",
          "K063257",
          "K011489",
          "K013490",
          "K171966",
          "K000207",
          "K953703",
          "K974331",
          "K973685",
          "K122636",
          "K130353",
          "K042565",
          "K954935",
          "K091508",
          "K151316",
          "K993352"
        ],
        "pma_number": [
          "P880076",
          "P880025",
          "P850061"
        ],
        "registration_number": [
          "",
          "3010287737",
          "1450662",
          "3025603301",
          "1721676",
          "3010536692",
          "1828288",
          "3006946276",
          "3002807315",
          "3014302784",
          "2249697",
          "3007353762",
          "3002807314",
          "3000236920",
          "3006536454",
          "2134285",
          "3002807090",
          "3013014058",
          "3008262872",
          "3004748528",
          "3009241418",
          "3000264985",
          "3020162561",
          "2528981",
          "9616944",
          "1644408",
          "9611390",
          "2032521",
          "3007311878",
          "3011170410",
          "3013176080",
          "3009882462",
          "3005061536",
          "3003477135",
          "2133928",
          "1526534",
          "9681465",
          "3004188066",
          "3005913274",
          "3019807891",
          "3007539489",
          "3011295718",
          "1827096",
          "3008658204",
          "3010288357",
          "3013557690",
          "3010846112",
          "3014833750",
          "1000547811",
          "3013413393",
          "3013302242",
          "1000517406",
          "3005718816",
          "9613369",
          "3009871135",
          "3010375065",
          "3005562917",
          "1221053",
          "9614209",
          "9617840",
          "1064858",
          "3009546990",
          "3011189525",
          "2244478",
          "3010132111",
          "1833986",
          "3007923096",
          "3012755988",
          "2531477",
          "3011274144",
          "3010576282",
          "3010057495",
          "1531174",
          "3010123256",
          "3006801265",
          "3020282716",
          "3021386586",
          "3009106092",
          "3002498892",
          "1720929",
          "9613350",
          "3010120135",
          "3012329926",
          "3004371426",
          "3013450937",
          "3043543260",
          "3006395932",
          "1526711",
          "3012425419",
          "3003761012",
          "3010173425",
          "3012725451",
          "3016438694",
          "1043653",
          "3021008900",
          "1220477",
          "3023852420",
          "3006946279",
          "2245304",
          "3017397663",
          "3010065126",
          "3011015572",
          "3012119238",
          "2183744",
          "3003061090",
          "3010097171",
          "3035366890",
          "3004638532",
          "1835444",
          "3009813841",
          "3030412764",
          "3005581234",
          "8044172",
          "3009475821",
          "3007740680",
          "1825034",
          "3008021110",
          "1424263",
          "3010041993",
          "3007366790",
          "3006085220",
          "1643264",
          "3014763043",
          "3009760038",
          "3024890263",
          "9610726",
          "8010379",
          "1221763",
          "1218882",
          "3006017180",
          "1061927",
          "3007200272",
          "3013086329",
          "1226544",
          "3005323511",
          "3008572101",
          "3008395353",
          "3002806535",
          "3012392319",
          "1219655",
          "1722511",
          "3007420745",
          "3015207155",
          "3010363503",
          "3002806603",
          "3010009693",
          "3008668801",
          "1645151",
          "1225675",
          "3011683674",
          "2134470",
          "3027484613",
          "3004635447",
          "3018052045",
          "3004142400",
          "9616680",
          "3008812251",
          "3014273644",
          "3010047454",
          "2246552",
          "3017410889",
          "1828464",
          "3009973336",
          "9681642",
          "3000258338",
          "3036773273",
          "3007189546",
          "3009959868",
          "2242737",
          "1057425",
          "3015398319",
          "1047843",
          "3002807310",
          "3009973505",
          "3000196378",
          "3002807112",
          "3003595343",
          "1530390",
          "3009732568",
          "3011277306",
          "9610612",
          "2249615",
          "3004641308",
          "3015542154",
          "3003541440",
          "3006809628",
          "3014315560",
          "3008544874",
          "3002808022",
          "3005852519",
          "3010531069",
          "3010039400",
          "3008868758",
          "1833506",
          "3005751028",
          "1018470",
          "3009998573",
          "1054811",
          "3006128100",
          "1649518",
          "3012319330",
          "1822565",
          "3005551626",
          "3001284227",
          "3008281151",
          "1818910",
          "1529009",
          "3006356043",
          "3014252644",
          "3005180920",
          "3008556682",
          "8043792",
          "2648920",
          "3013791180",
          "3036756245",
          "3020967480",
          "3004153896",
          "3012130008",
          "3008421730",
          "1423662",
          "1038671",
          "3010386387",
          "3026771806",
          "3002807408",
          "3003639920",
          "1000550978",
          "3010178296",
          "1834379",
          "3010220595",
          "1643817",
          "1048735",
          "3007334927",
          "1836357",
          "3022142628",
          "3008110533",
          "3038503932",
          "3002806470",
          "3007913880",
          "9616671",
          "3012275107",
          "3009173317",
          "1834331",
          "3001278222",
          "3015516266",
          "3009582259",
          "3016778562",
          "3003387384",
          "3005360851",
          "3020155054",
          "8010916",
          "3002907620",
          "1833920",
          "3010163695",
          "3010041693",
          "3003998208",
          "3010388970",
          "3013055499",
          "3026776",
          "1835251",
          "2030624",
          "3004513872",
          "3004153240",
          "3009888740",
          "1835572",
          "1020279",
          "1226188",
          "3033509898"
        ],
        "fei_number": [
          "",
          "3010287737",
          "1000136269",
          "1450662",
          "3025603301",
          "1721676",
          "3010536692",
          "1828288",
          "3006946276",
          "3002808269",
          "3002807315",
          "3014302784",
          "3007353762",
          "1000139987",
          "3002807314",
          "3003128516",
          "3006536454",
          "3013014058",
          "3008262872",
          "3004748528",
          "3009241418",
          "3000264985",
          "3020162561",
          "2528981",
          "3007311878",
          "3011170410",
          "3013176080",
          "3009882462",
          "1000118610",
          "3005061536",
          "3003477135",
          "1000114185",
          "3000203000",
          "1526534",
          "1000135253",
          "3007282753",
          "3004188066",
          "3005913274",
          "3019807891",
          "3007539489",
          "3011295718",
          "1827096",
          "3002993297",
          "3008658204",
          "3010288357",
          "3013557690",
          "3010846112",
          "3014833750",
          "1000547811",
          "3013413393",
          "3013302242",
          "1000517406",
          "3005718816",
          "3009871135",
          "3010375065",
          "3002807172",
          "3005562917",
          "1221053",
          "3009546990",
          "3011189525",
          "1000199976",
          "3010132111",
          "3007923096",
          "3012755988",
          "3002808293",
          "3011274144",
          "3010576282",
          "3010057495",
          "3010123256",
          "3006801265",
          "3020282716",
          "3021386586",
          "3009106092",
          "3001452082",
          "3000717687",
          "3002498892",
          "1720929",
          "3002462840",
          "3010120135",
          "3012329926",
          "3013450937",
          "3007932279",
          "3043543260",
          "3003574398",
          "3006395932",
          "1526711",
          "3002854251",
          "3003569173",
          "3003714383",
          "3003761012",
          "3012725451",
          "3016438694",
          "1043653",
          "3002287318",
          "3000196378",
          "3021008900",
          "1220477",
          "3023852420",
          "3006946279",
          "2245304",
          "1000115656",
          "1832656",
          "3017397663",
          "3010065126",
          "3011015572",
          "3012119238",
          "2183744",
          "3002459190",
          "3003061090",
          "3010097171",
          "3035366890",
          "3004638532",
          "3009813841",
          "3030412764",
          "3005581234",
          "3009475821",
          "3007740680",
          "1825034",
          "3008021110",
          "3003110073",
          "3001451451",
          "3010041993",
          "3006085220",
          "3003414678",
          "1643264",
          "3014763043",
          "3009760038",
          "1000220733",
          "3024890263",
          "3002807113",
          "1221763",
          "1218882",
          "3006017180",
          "3001548076",
          "3007200272",
          "3013086329",
          "3001049282",
          "3005323511",
          "3008572101",
          "3008395353",
          "3002806535",
          "3012392319",
          "1219655",
          "3007420745",
          "3015207155",
          "3010363503",
          "3002806603",
          "3010009693",
          "3008668801",
          "3001642380",
          "3011683674",
          "3027484613",
          "3004635447",
          "3018052045",
          "1000132037",
          "3004142400",
          "3014273644",
          "3001236277",
          "3010047454",
          "2246552",
          "1828464",
          "1000143502",
          "3009973336",
          "3036773273",
          "3000258338",
          "3007189546",
          "3009959868",
          "1047843",
          "3002807310",
          "3009973505",
          "1000221916",
          "3002807893",
          "3002807112",
          "3003595343",
          "3010566635",
          "3005233524",
          "3009732568",
          "3011277306",
          "3002806773",
          "1000306303",
          "3002806653",
          "3003386935",
          "3003015552",
          "3015542154",
          "3004641308",
          "3003541440",
          "3000205591",
          "3006809628",
          "3014315560",
          "3008544874",
          "3002808022",
          "3005852519",
          "3000215273",
          "3000215103",
          "3010531069",
          "1000116912",
          "3010039400",
          "3008868758",
          "3002807036",
          "3005751028",
          "3000717902",
          "1018470",
          "3009998573",
          "3006128100",
          "3012319330",
          "3005551626",
          "3001284227",
          "3008281151",
          "1818910",
          "3002806725",
          "1000306340",
          "3014252644",
          "3006356043",
          "3008397795",
          "3005180920",
          "3008556682",
          "3036756245",
          "3002806315",
          "3013791180",
          "3020967480",
          "3004153896",
          "3012130008",
          "3008421730",
          "3005985723",
          "3006003617",
          "3010386387",
          "1038671",
          "3026771806",
          "3003639920",
          "1000550978",
          "3002807408",
          "3010220595",
          "1643817",
          "1048735",
          "3007334927",
          "3022142628",
          "1000395711",
          "3008110533",
          "3002806470",
          "3007913880",
          "3038503932",
          "3012275107",
          "3009173317",
          "3001278222",
          "3015516266",
          "1000086912",
          "3008342029",
          "3016778562",
          "3003387384",
          "3005360851",
          "3020155054",
          "3003070421",
          "3002907620",
          "3003998208",
          "3010163695",
          "3010041693",
          "3010388970",
          "3013055499",
          "3003064967",
          "3004513872",
          "1000149660",
          "3004153240",
          "3009888740",
          "3009582259",
          "1020279",
          "3000206435",
          "3033509898"
        ],
        "device_name": "Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3358",
        "device_class": "2"
      }
    }
  ]
}