      "cfres_id": "124749",
      "product_res_number": "Z-0785-2014",
      "event_date_initiated": "2013-12-04",
      "event_date_posted": "2014-01-17",
      "recall_status": "Terminated",
      "event_date_terminated": "2015-11-06",
      "res_event_number": "67130",
      "product_code": "IZJ",
      "k_numbers": [
        "K913471"
      ],
      "product_description": "ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems.  Distributed by Philips Medical Systems, Cleveland, OH",
      "code_info": "ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005, 9510027, 9510028, 9603042, 9603043, 9604044, 9604047, 9605011, 9607060, 9609065, 9610072, 9704081, 9709090, 9710097, 9711103, 9712105, 9712106, 9801107, 9801109, 9802112, 9803001, 9803116, 9804126, 9804129, 9806136, 9807138, 9807422,9810527, 9901156, 9901160, 9901162, 9908167, 9909172, 9910176, 00709B, 0101003A, 01310B, 01525B, 01920B, 01921B, 01922B, 03041B,  03094B, 03338B, 03368B, 2145-30074A, 7718B, A0103015, A0103213, A0105217, A02100230, A02100231, A02110248, A02110249, A02110250, A02110251, A02120254, A02120256, A03010262, A03010264, A03010265, A03010266, A03010270, A03020274, A03020275, A03030282, A03070283, A03120294, B0003196, B0007200, B0309014, ID3935B, M@10639, M@10645, M@10774, M@10777, M@10877, M@11239, M@11568, M@11640, M@11684, M@11685, M@11729, M@11769, M@11813, M@11947, M@12378-1, M@12462, M@12765, M@13725, M@24470, M@4750028851, M@4750028853, M@5348B, M@5807B, M@5874B, M@5894B, M@5932B, M@5941B, M@5963B, M@6041B, M@6062B, M@6229B, M@6354B, M@6477B, M@6485B, M@6546B, M@6555B, M@6692B, M@6754B, M@6809B, M@6815B, M@6948B, M@7007B, M@7023B, M@7093B, M@7194B, M@7198B, M@7207B, M@7214B, M@7280B, M@7285B, M@7286B, M@7408B, M@7540B, M@7596B, M@7675B, M@7704B, M@8156B, M@8324B, M@8344B, M@8429B, M@8523B, M@8578B, M@8588B, M@8637B, M@8639RB, M@8648B, M@8808B, M@8813B, M@8898B, M@8977B, M@9050B, M@9261B, M@9276B, M@9355RB, M@9475B, M@9544B, M@9647B, M@9656B, M@9733B & M@9796B.  ADAC ARGUS: Model #88295; Serial #: 9211014, 9301005, 9304019, 9304024, 9304028, 9401028, 9401031, 9401037, 9505014, 9910033, 93-10678- 1292, 9910028P, M@10225AMT, M@3732B, M@3903B, M@3954B, M@3956B, M@3960B, M@4249B, M@4415B, M@4646B, M@4868B, M@5172B, M@5321B, M@7341B & M@AMT1154.  ADAC CIRRUS: Model #88294; 34, 718558, 9206009, 9207003, 9207004, 9207005, 9207007, 9211006, 9300113, 9304013, 9309012, 9404022, 9407028, 9509042, 9604098, 9910030, 93-10735-0393, 9621RQ, M@10110AMT, M@10324AMT, M@10402AMT, M@3599G-08, M@3905Q, M@4029Q, M@4129Q, M@4142Q, M@4668Q, M@4743Q, M@5003Q, M@5668Q, M@5895Q, M@AMT126R, M@AMT1282 & M@MVS3227A.",
      "firm_fei_number": "1525965",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "city": "Cleveland",
      "state": "OH",
      "postal_code": "44143-2131",
      "additional_info_contact": "Mrs. Kumudini J. Carter\n440-483-3032",
      "reason_for_recall": "The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "On October 30, 2013 & December 4, 2013, the firm sent Urgent Medical Device Recall letters to their customers.",
      "product_quantity": "255 units",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of:  AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great Britain, Italy, Netherlands, Norway, Pakistan, Russia, Singapore & South Korea.",
      "openfda": {
        "k_number": [
          "K802132",
          "K881688",
          "K972130",
          "K945580",
          "K853983",
          "K812613",
          "K024085",
          "K792394",
          "K896615",
          "K810072",
          "K920793",
          "K810191"
        ],
        "device_name": "Camera, X-Ray, Fluorographic, Cine Or Spot",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1620",