  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "122168",
      "product_res_number": "Z-0305-2014",
      "event_date_initiated": "2013-06-12",
      "event_date_posted": "2013-11-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-04-11",
      "res_event_number": "65572",
      "product_code": "JHS",
      "k_numbers": [
        "K922072"
      ],
      "product_description": "VITROS Chemistry Products CK Slides packaged as 300 slides/pack catalog number 8479396 and packaged as 90 slides/pack catalog number 8478034\r\nVITROS CK Slides quantitatively measure creatine kinase (CK) activity in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.",
      "code_info": "Multiple lots affected, with expiry dates through 22-Mar-2015",
      "firm_fei_number": "1000136573",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "city": "Rochester",
      "state": "NY",
      "postal_code": "14626-5101",
      "reason_for_recall": "Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.",
      "root_cause_description": "Equipment maintenance",
      "action": "Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides.  VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue.  OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs.  OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue.  In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges.  Complete and return the Confirmation of Receipt by June 21, 2013.  If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311.\r\nFor questions regarding this recall call 585-453-4224.",
      "product_quantity": "86,503 units (30,378 units domestically & 56,125 units internationally)",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, Germany, France, Italy, and Spain.",
      "openfda": {
        "k_number": [
          "K160570",
          "K820722",
          "K790015",
          "K810997",
          "K801668",
          "K800028",
          "K082226",
          "K872029",
          "K062128",
          "K861387",
          "K790088",
          "K945083",
          "K802521",
          "K781690",
          "K823211",
          "K803169",
          "K792023",
          "K780504",
          "K952250",
          "K981495",
          "K790343",
          "K854708",
          "K990434",
          "K781432",
          "K855188",
          "K900806",
          "K883223",
          "K790917",
          "K843007",
          "K102706",
          "K941584",
          "K790918",
          "K790916",
          "K903025",
          "K962126",
          "K813406",
          "K943024",
          "K990439"
        ],
        "registration_number": [
          "1219913",
          "9681753",
          "2432235",
          "1319681",
          "1319809",
          "2517506",
          "9610529",
          "9610126",
          "3003795116",
          "2250051",
          "3006198300"
        ],
        "fei_number": [
          "1219913",
          "2432235",
          "3002806559",
          "1000136573",
          "2517506",
          "3000303338",
          "3003795116",
          "3002806560",
          "2250051",
          "1000305840",
          "3006198300"
        ],
        "device_name": "Differential Rate Kinetic Method, Cpk Or Isoenzymes",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1215",
        "device_class": "2"
      }
    }
  ]
}