{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "114678",
      "product_res_number": "Z-0505-2013",
      "event_date_initiated": "2012-11-20",
      "event_date_posted": "2012-12-10",
      "recall_status": "Terminated",
      "event_date_terminated": "2015-08-25",
      "res_event_number": "63728",
      "product_code": "EOQ",
      "k_numbers": [
        "K914181"
      ],
      "product_description": "WANG Transbronchial Aspiration Needles Cat. No. MW-221,  WANG Cytology Needles, Central or Peripheral Regions, 21 g x 15 mm long needle. --- Common Name: Bronchoscope (flexible or rigid) and accessories.  The ConMed WANG Transbronchial Aspiration Needles are designed for use with a bronchoscope.  They are used to puncture the tracheobronchial wall and to aspirate sufficient tissue and/or cell specimens to stage bronchogenic carcinoma.   These are disposable sterile, single-use devices.",
      "code_info": "Beginning Lot Code 1209171, Ending Lot Code 1209201.  Lot codes on boxes and packaging contain a lot code in the following form 12 09 17 1 where 12 represents the year (2012), 09 represents the month (September),  17 is the day of the month, and 1 is the Manufacturing Shift Code (1, 2, 3, 4, 5, 6 or 9).",
      "firm_fei_number": "3001451571",
      "recalling_firm": "ConMed Corporation",
      "address_1": "525 French Rd",
      "city": "Utica",
      "state": "NY",
      "postal_code": "13502-5945",
      "reason_for_recall": "ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles would not insert properly through the instrument channel of a 2.0 mm bronchoscope.  If the device will not fit through the instrument channel of a 2.0 mm bronchoscope, it must be replaced with an alternate device and could result in minor inconvenience to the user.",
      "root_cause_description": "Component design/selection",
      "action": "ConMed Corporation sent an Urgent Field Safety Notice letter dated November 30, 2012 with Response Form , to all affected customers via USPS Priority Mail.  The letter identified the product, the problem, and the action to be taken by the customer.  Customers were asked to complete Attachment II and return it with any unused affected product to:\nConMed Corporation\n525 French Road, Utica, NY  13502\nUnited Sates of America\n\nFor information regarding this recall call 315-624-3225.",
      "product_quantity": "Total 452",
      "distribution_pattern": "Worldwide Distribution - USA including CA , FL, IA, IN, MA, MD, ME, MO, MS, NC, NM, OH, PA, TX.  Foreign distribution to Spain, Portugal, Japan and Hungary.",
      "openfda": {
        "k_number": [
          "K974642",
          "K944650",
          "K231107",
          "K022546",
          "K212886",
          "K040940",
          "K933433",
          "K921393",
          "K182253",
          "K242793",
          "K220299",
          "K163377",
          "K061313",
          "K171291",
          "K210379",
          "K931154",
          "K896146",
          "K213782",
          "K980046",
          "K232860",
          "K854581",
          "K952081",
          "K950786",
          "K211947",
          "K792267",
          "K211169",
          "K233671",
          "K962367",
          "K242416",
          "K223836",
          "K210883",
          "K962378",
          "K173554",
          "K253137",
          "K954707",
          "K944649",
          "K941264",
          "K071530",
          "K232984",
          "K233779",
          "K182188",
          "K024165",
          "K833483",
          "K934920",
          "K183516",
          "K222861",
          "K201758",
          "K822180",
          "K183654",
          "K243219",
          "K211030",
          "K832492",
          "K911922",
          "K252045",
          "K033225",
          "K202987",
          "K822844",
          "K951199",
          "K220399",
          "K192164",
          "K151315",
          "K230428",
          "K983124",
          "K242324",
          "K210928",
          "K940521",
          "K914181",
          "K963033",
          "K212048",
          "K191948",
          "K250862",
          "K193534",
          "K152319",
          "K122535",
          "K962555",
          "K832449",
          "K943700",
          "K060243",
          "K960880",
          "K251894",
          "K082720",
          "K071473",
          "K244017",
          "K260420",
          "K951826",
          "K260565",
          "K844834",
          "K022446",
          "K223144",
          "K213235",
          "K945268",
          "K081518",
          "K242070",
          "K163248",
          "K200397",
          "K242010",
          "K920025",
          "K831240",
          "K130845",
          "K223299",
          "K822136",
          "K981543",
          "K243541",
          "K251823",
          "K251731",
          "K142909",
          "K201146",
          "K870063",
          "K931442",
          "K090853",
          "K031966",
          "K202370",
          "K962901",
          "K141366",
          "K050907",
          "K193488",
          "K192367",
          "K770500",
          "K922204",
          "K200342",
          "K213635",
          "K220957",
          "K172955",
          "K963795",
          "K893006",
          "K250957",
          "K252528",
          "K202180",
          "K941161",
          "K211493",
          "K250232",
          "K950636",
          "K222014",
          "K202346",
          "K221340",
          "K941158",
          "K172726",
          "K201300",
          "K811059",
          "K221631",
          "K962736",
          "K232601",
          "K993900",
          "K234113",
          "K143727",
          "K212667",
          "K983122",
          "K183085",
          "K254037",
          "K260382",
          "K121959",
          "K252341",
          "K915816",
          "K012073",
          "K221302",
          "K173760",
          "K231521",
          "K212656",
          "K960791",
          "K895948",
          "K944976",
          "K953596",
          "K822257",
          "K982983",
          "K050220",
          "K142934",
          "K910423",
          "K013111",
          "K161656",
          "K222910",
          "K162611",
          "K230948",
          "K231970",
          "K954125",
          "K862835",
          "K200702",
          "K251752",
          "K922937",
          "K121035",
          "K013731",
          "K792342",
          "K021344",
          "K911412",
          "K152466",
          "K173727",
          "K910536",
          "K950965",
          "K961178",
          "K232582",
          "K163537",
          "K223782",
          "K131401",
          "K970107",
          "K963252",
          "K951051",
          "K171492",
          "K182825",
          "K954797",
          "K832010",
          "K040018",
          "K240289",
          "K023984",
          "K983878",
          "K962485",
          "K183256",
          "K811218",
          "K892687",
          "K854414",
          "K895415",
          "K241612",
          "K183419",
          "K142839",
          "K250210",
          "K950925",
          "K191828",
          "K240135",
          "K023376",
          "K942388",
          "K926430",
          "K923847",
          "K760171",
          "K131234",
          "K220606",
          "K183607",
          "K250863",
          "K011267",
          "K250118",
          "K243497",
          "K962380",
          "K990547",
          "K240075",
          "K131028",
          "K984157",
          "K131946",
          "K962533",
          "K150280",
          "K232038",
          "K811181",
          "K091250",
          "K091768",
          "K200678"
        ],
        "registration_number": [
          "",
          "3010432890",
          "9610617",
          "3005382983",
          "2029015",
          "3010707607",
          "3015224777",
          "3004837686",
          "3010280860",
          "3005099803",
          "2242464",
          "3013298431",
          "3014663814",
          "3023214418",
          "3007487007",
          "3017820152",
          "3042940623",
          "9610877",
          "3013528571",
          "9680515",
          "3003678543",
          "3010396739",
          "2124215",
          "3010273867",
          "3015925378",
          "3016607092",
          "3014491269",
          "1220850",
          "8040233",
          "3011050570",
          "3012448339",
          "3010684280",
          "3010157088",
          "1828132",
          "3016096100",
          "3013868968",
          "3010202439",
          "3016678045",
          "3013440400",
          "3007305485",
          "3020746799",
          "3006161756",
          "3014769442",
          "3010383847",
          "3011063460",
          "9615393",
          "3006174295",
          "3004644065",
          "3017425975",
          "3025349506",
          "3004065885",
          "3010493171",
          "9611680",
          "2011171",
          "3017367014",
          "3007084728",
          "1000121056",
          "3014246639",
          "1721686",
          "3018094310",
          "3012627460",
          "3033628477",
          "9681147",
          "2032112",
          "3004142795",
          "2938071",
          "3043226252",
          "9611997",
          "3016045249",
          "9710524",
          "3004111573",
          "3004962788",
          "3022472",
          "9611102",
          "3014605202",
          "1833700",
          "9614641",
          "3015139943",
          "3015970743",
          "3014342096",
          "3007738736",
          "3011416394",
          "1314417",
          "3013992977",
          "3015895145",
          "1220477",
          "3017616393",
          "3010041511",
          "3012843412",
          "3040391040",
          "3016555232",
          "3008309062",
          "1320894",
          "3025749544",
          "3020369549",
          "3021031769",
          "2020550",
          "3006984716",
          "3038182369",
          "3008040402",
          "3013665681",
          "3005369776",
          "3026787544",
          "3003637092",
          "3008729892",
          "9613846",
          "3004040397",
          "3012494290",
          "3009719225",
          "3038613760",
          "1000484830",
          "3014247261",
          "3010398867",
          "3010421104",
          "3012797688",
          "3006621103",
          "3006546082",
          "8030673",
          "3030504644",
          "3002808148",
          "3021325287",
          "3042225166",
          "3015136743",
          "3007695959",
          "3023102955",
          "9610691",
          "3008439484",
          "3023809956",
          "3016832577",
          "3009420598",
          "8010273",
          "3008386005",
          "3006786907",
          "3014635169",
          "3013666218",
          "3004083736",
          "3013557562",
          "1000393132",
          "1223925",
          "1650907",
          "3022631687",
          "3003183807",
          "3033509898",
          "3009039068",
          "3003761012",
          "3013283620",
          "3007867778",
          "3017140456",
          "3013077751",
          "1526611",
          "3019807891",
          "3005465238",
          "3020331900",
          "3005987240",
          "3029929134",
          "3003782610",
          "1037905",
          "3010220595",
          "2246552",
          "3035710892",
          "3024021261",
          "3036756245",
          "3012322232",
          "2244478",
          "3009171251",
          "2183744",
          "3025603301",
          "1319639",
          "3027205723",
          "3015612252",
          "3042172882",
          "1221485",
          "3015142721",
          "3033536312",
          "1220592",
          "3017619652",
          "3010273872",
          "3017859830",
          "2030624",
          "2243757",
          "9681260",
          "1037955",
          "3026606306",
          "3011137372",
          "8010047",
          "3010885402",
          "1221826",
          "3015316279",
          "9610595",
          "3001675293",
          "3015153469",
          "3023460764"
        ],
        "fei_number": [
          "",
          "3002807795",
          "3010432890",
          "3005382983",
          "1000138054",
          "3010707607",
          "3005747797",
          "3004837686",
          "3010280860",
          "3015224777",
          "3008185830",
          "3005099803",
          "2242464",
          "3013298431",
          "3014663814",
          "3023214418",
          "3007487007",
          "3017820152",
          "3042940623",
          "3013528571",
          "3003678543",
          "3010396739",
          "2124215",
          "3010273867",
          "3015925378",
          "3016607092",
          "1000491840",
          "1220850",
          "3002807969",
          "3011050570",
          "3012448339",
          "3010684280",
          "3010157088",
          "1828132",
          "3016096100",
          "3013868968",
          "3016678045",
          "3000206435",
          "3013440400",
          "3007305485",
          "3020746799",
          "3014769442",
          "3010383847",
          "3011063460",
          "3006174295",
          "3004644065",
          "3025349506",
          "3004065885",
          "3010493171",
          "3014447948",
          "2011171",
          "3006493107",
          "3003076831",
          "3017367014",
          "1000121056",
          "3014246639",
          "3007084728",
          "1721686",
          "3018094310",
          "3012627460",
          "3033628477",
          "3004142795",
          "2938071",
          "3043226252",
          "3016045249",
          "3004111573",
          "3004962788",
          "1000136269",
          "3022472",
          "3003499455",
          "3014605202",
          "3015139943",
          "3015970743",
          "3002806565",
          "1000305720",
          "3014342096",
          "3007738736",
          "3011416394",
          "3002964398",
          "1314417",
          "3003944071",
          "3013992977",
          "3015895145",
          "1220477",
          "3017616393",
          "3002806359",
          "3010041511",
          "3012843412",
          "3040391040",
          "3016555232",
          "3008309062",
          "3002681132",
          "3025749544",
          "3020369549",
          "3021031769",
          "2020550",
          "3006984716",
          "3038182369",
          "3008040402",
          "3013665681",
          "1000160458",
          "3005947796",
          "3005369776",
          "3026787544",
          "1000115586",
          "3003637092",
          "3004040397",
          "3005969827",
          "3012494290",
          "3009719225",
          "3038613760",
          "1000484830",
          "3014247261",
          "3010398867",
          "3010421104",
          "3012797688",
          "3006621103",
          "3006546082",
          "4486",
          "3030504644",
          "3007475226",
          "3002808148",
          "3021325287",
          "3042225166",
          "3001451571",
          "3015136743",
          "3007695959",
          "3023102955",
          "3008439484",
          "3023809956",
          "3016832577",
          "3009420598",
          "3008386005",
          "3002808268",
          "3006786907",
          "3014635169",
          "3002807769",
          "3013666218",
          "3004083736",
          "3013557562",
          "1000393132",
          "3022631687",
          "3003183807",
          "3033509898",
          "3009039068",
          "3003761012",
          "3013283620",
          "3003995201",
          "3007867778",
          "3017140456",
          "3013077751",
          "1526611",
          "3019807891",
          "3005465238",
          "3020331900",
          "3005987240",
          "3029929134",
          "3003782610",
          "3006759904",
          "1037905",
          "3010220595",
          "2246552",
          "3035710892",
          "3024021261",
          "3036756245",
          "3012322232",
          "2183744",
          "3025603301",
          "3027205723",
          "3015612252",
          "3042172882",
          "1221485",
          "3015142721",
          "3002808127",
          "3033536312",
          "1220592",
          "3017619652",
          "3010273872",
          "3004213967",
          "3017859830",
          "2243757",
          "1037955",
          "3026606306",
          "3010885402",
          "1221826",
          "3002807555",
          "3015316279",
          "3002808001",
          "3014491269",
          "3001675293",
          "3015153469",
          "3023460764"
        ],
        "device_name": "Bronchoscope (Flexible Or Rigid)",
        "medical_specialty_description": "Ear, Nose, Throat",
        "regulation_number": "874.4680",
        "device_class": "2"
      }
    }
  ]
}