{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "112615",
      "product_res_number": "Z-0011-2013",
      "event_date_initiated": "2012-09-11",
      "event_date_posted": "2012-10-04",
      "recall_status": "Terminated",
      "event_date_terminated": "2020-08-27",
      "res_event_number": "63116",
      "product_code": "MKJ",
      "k_numbers": [
        "K021453"
      ],
      "product_description": "Philips HeartStart XL M4735A Defibrillator/Monitors, used with the M3516A batteries \n\nThe Heart Start XL Defibrillator/Monitor is a fully featured, defibrillator intended for use by qualified medical personnel, trained in either Advanced Cardiac Life Support or in the operation of the device, in a hospital environment.",
      "code_info": "HeartStart XL battery M3516A labeled with Made in Taiwan and an R-2011-12 Date of Manufacture",
      "firm_fei_number": "1218950",
      "recalling_firm": "Philips Healthcare Inc.",
      "address_1": "3000 Minuteman Road",
      "city": "Andover",
      "state": "MA",
      "postal_code": "01810",
      "additional_info_contact": "Debra Levasseur\n978-687-1501",
      "reason_for_recall": "HeartStart XL Defibrillator/Monitor (M4735A) used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "The firm, Philips Healthcare, sent an \"URGENT-MEDICAL DEVICE RECALL/Field Safety Notice\" letter dated September 6, 2012 to its customers. The letter describes the product, problem and actions to be taken by the customer and  also identifies what action Philips plans to take to remedy the issue. The customers were instructed to operate the XL and AC power instead of battery power alone; use a battery from a different date of manufacture (must not have the R-2011-12 date of manufacture code) if they need to operate the XL on battery power alone; and use an alternative source for treatment or therapy (i.e., backup defibrillator/monitor) if unable to operate with AC power alone or locate a battery with a different date of manufacture than that identified.  Phillips will provide a replacement battery(s) to customers as soon as new production is available.\n\nIf you need any further information or support concerning this issue, contact your local Philips representative or call Quality Specialists,Recalls at 978-687-1501 or email: debra.levasseur@philips.com.",
      "product_quantity": "2500 units",
      "distribution_pattern": "Worldwide distribution: USA (Nationwide) and countries of: AUSTRALIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, REPUBLIC OF KUWAIT LEBANON, MACAU, MALAYSIA, MEXICO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, ROMANIA RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, and VIET NAM.",
      "openfda": {
        "k_number": [
          "K032439",
          "K971148",
          "K896919",
          "K992543",
          "K860676",
          "K983393",
          "K003565",
          "K040256",
          "K091561",
          "K031187",
          "K853564",
          "K030898",
          "K011144",
          "K103567",
          "K953204",
          "K955628",
          "K120406",
          "K111594",
          "K012766",
          "K002232",
          "K133484",
          "K150055",
          "K120536",
          "K993036",
          "K122600",
          "K020984",
          "K033771",
          "K993072",
          "K120907",
          "K060934",
          "K072430",
          "K110825",
          "K063510",
          "K041067",
          "K111693",
          "K130684",
          "K010214",
          "K063119",
          "K091265",
          "K123881",
          "K962359",
          "K932931",
          "K062177",
          "K942850",
          "K092598",
          "K943698",
          "K002445",
          "K142430",
          "K143714",
          "K955923",
          "K080421",
          "K040438",
          "K120350",
          "K935922",
          "K061707",
          "K012428",
          "K980857",
          "K003548",
          "K141774",
          "K040404",
          "K990762",
          "K112432",
          "K991556",
          "K954957",
          "K122839",
          "K100565",
          "K011901",
          "K052057",
          "K113787",
          "K040637",
          "K002806",
          "K081081",
          "K964991",
          "K041883",
          "K090989",
          "K072812",
          "K022732",
          "K934692",
          "K062537",
          "K062970",
          "K032051",
          "K133441",
          "K882067",
          "K082937",
          "K955828",
          "K133269",
          "K970391",
          "K992977",
          "K142915",
          "K942344",
          "K940445",
          "K001725",
          "K131494",
          "K991007",
          "K022210",
          "K973486",
          "K051134",
          "K041459",
          "K022929",
          "K990106",
          "K052161",
          "K002280",
          "K926172",
          "K081259",
          "K993745",
          "K955882",
          "K981802",
          "K991910",
          "K072233",
          "K033474",
          "K073716",
          "K011541",
          "K060559",
          "K954504",
          "K981737",
          "K991871",
          "K070455",
          "K972418",
          "K883344",
          "K955890",
          "K040479",
          "K003228",
          "K141795",
          "K954609",
          "K010634",
          "K012197",
          "K033642",
          "K062233",
          "K131550",
          "K040056",
          "K110154",
          "K925936",
          "K982710",
          "K972970",
          "K941262",
          "K142803",
          "K984582",
          "K883584",
          "K133239",
          "K970719",
          "K021168",
          "K052465",
          "K971149",
          "K131525",
          "K020743",
          "K952101",
          "K003819",
          "K972241",
          "K033275",
          "K002079",
          "K012030",
          "K051076",
          "K102972",
          "K014157",
          "K892005",
          "K072596",
          "K011388",
          "K112761",
          "K970741",
          "K965117",
          "K052316",
          "K010918",
          "K982264",
          "K051632",
          "K080903",
          "K141231",
          "K080232",
          "K110361",
          "K122758",
          "K870622",
          "K932028",
          "K010207",
          "K031987",
          "K010257",
          "K021453",
          "K020288",
          "K050004",
          "K020715",
          "K982992",
          "K071025",
          "K121853",
          "K983287",
          "K013896",
          "K091548",
          "K110978",
          "K081371",
          "K082090",
          "K130454",
          "K955854",
          "K982025",
          "K042404",
          "K040775",
          "K103186",
          "K892065",
          "K110742",
          "K133659",
          "K081828",
          "K063375",
          "K983307",
          "K072923",
          "K130153",
          "K110168",
          "K042007",
          "K092391",
          "K142709",
          "K041892",
          "K150198",
          "K033144",
          "K082861",
          "K020203",
          "K960329",
          "K042417",
          "K013425",
          "K091943",
          "K033896",
          "K941970",
          "K102496",
          "K110526",
          "K042088",
          "K073089",
          "K000638",
          "K032421",
          "K883341",
          "K984286",
          "K014209",
          "K023854",
          "K993533",
          "K051475"
        ],
        "pma_number": [
          "P160032",
          "P210015",
          "P200007",
          "P160033",
          "P160015",
          "P160022",
          "P160012",
          "P160008",
          "P190007",
          "P200004",
          "P190013",
          "P200005",
          "P160026",
          "P170018",
          "P160028",
          "P160034",
          "P180028"
        ],
        "registration_number": [
          "3030733800",
          "",
          "3026007685",
          "3003867964",
          "3013517171",
          "2132111",
          "1220908",
          "3020460367",
          "3010783192",
          "3003800159",
          "3010703925",
          "3012494290",
          "2112020",
          "3010839093",
          "3015997711",
          "1720929",
          "3004123209",
          "3017422354",
          "3004035727",
          "3026976873",
          "3005906887",
          "1317188",
          "3015134086",
          "3012104670",
          "3000126629",
          "3043226252",
          "3013500228",
          "1219103",
          "3004906872",
          "3003537036",
          "3014572985",
          "1423537",
          "3011823740",
          "1423662",
          "3021559257",
          "3010030355",
          "3012972516",
          "3005621729",
          "3038259592",
          "3029871637",
          "3007029079",
          "3013596024",
          "3008717264",
          "1930027",
          "3010317211",
          "3013826848",
          "3012607778",
          "3030446844",
          "3030677",
          "1218058",
          "3027815",
          "3002807314",
          "9681860",
          "3009077524",
          "9681191",
          "1320894",
          "3013557562",
          "3009886119",
          "1223925",
          "3003521780",
          "3002714378",
          "3006108746",
          "3014579161",
          "8020045",
          "1921846",
          "3012502601",
          "3015876"
        ],
        "fei_number": [
          "3030733800",
          "",
          "3026007685",
          "3003867964",
          "3013517171",
          "3007029079",
          "1220908",
          "3020460367",
          "3010783192",
          "3003800159",
          "3010703925",
          "3012494290",
          "2112020",
          "3010839093",
          "3001236905",
          "3015997711",
          "3004123209",
          "1720929",
          "3017422354",
          "3004035727",
          "3026976873",
          "3005906887",
          "1317188",
          "3015134086",
          "3012104670",
          "3000126629",
          "3043226252",
          "1000127185",
          "6736",
          "3013500228",
          "3003537036",
          "3004906872",
          "3014572985",
          "3011823740",
          "3021559257",
          "3010030355",
          "3012972516",
          "3005621729",
          "3038259592",
          "1000305720",
          "3029871637",
          "3013596024",
          "3008717264",
          "3003491392",
          "3010317211",
          "3013826848",
          "3012607778",
          "1930027",
          "1000140026",
          "3030446844",
          "3000215273",
          "1218058",
          "3002807314",
          "3009077524",
          "3013557562",
          "3009886119",
          "3003521780",
          "3002714378",
          "3006108746",
          "1000524572",
          "3014579161",
          "1921846",
          "3001451571",
          "1000148701",
          "3008207730",
          "3012502601",
          "3015876"
        ],
        "device_name": "Automated External Defibrillators (Non-Wearable)",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.5310",
        "device_class": "3"
      }
    }
  ]
}