  "results": [
    {
      "cfres_id": "111545",
      "product_res_number": "Z-0283-2013",
      "event_date_initiated": "2012-09-25",
      "event_date_posted": "2012-11-13",
      "recall_status": "Terminated",
      "event_date_terminated": "2013-07-11",
      "res_event_number": "62733",
      "product_code": "DTQ",
      "k_numbers": [
        "K915183"
      ],
      "product_description": "Perfusion System 8000, base 5 pump, 220V/240V safety monitor included\n\nThe Sarns\" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.",
      "code_info": "catalog number: 16406 serial numbers: 1001-1004,1007-1055,1057-1310, 1648, 1649, 1657, 1822, and 1823.",
      "firm_fei_number": "1828100",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Road",
      "city": "Ann Arbor",
      "state": "MI",
      "postal_code": "48103-9586",
      "additional_info_contact": "734-741-6173",
      "reason_for_recall": "Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the arterial and cardioplegia monitors for the Sarns Modular\nPerfusion System 8000. The malfunctions can be broadly categorized as follows monitor power malfunctions, inability to measure pressure and inaccurate and/or false pressure\nalarms.\nIf the malfunction of the arterial or cardio",
      "root_cause_description": "Device Design",
      "action": "The firm, Terumo CVS, sent an \"URGENT MEDICAL DEVICE RECALL\" notice/Safety Advisory Alert dated September 24, 2012 and an addendum to the Operator's Manual to alert all affected Customers. The notice described the product, problem and actions to be taken. The customers were instructed to review this medical device recall notice; assure that all users are aware to this notice; place the addendum in the Operator's Manual just before page 4.1 and complete and return the attached Customer Response Form via Fax to: 1-734-741-6149 or email to:www.termuo-cvs.com.\n\nIf you have any questions or concerns contact Terumo CVS Customer Service at 1-800-521-1818; Monday - Friday, 8 AM - 6 PM ET.",
      "product_quantity": "312",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including states of:  AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, and WY; and countries of:  ARGENTINA, AUSTRALIA, BANGLADESH, BRAZIL, CALI COLOMBIA, CALLO PERU, GERMANY, GUAYQUIL ECUADOR, HONG KONG, INDIA, INDONESIA, JAPAN, LEBANON, MADRID SPAIN, MEXICO, NEW ZEALAND, SANTIAGO, SANTIAGO CHILE, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SRI LANKA, TAIWAN, THAILAND, and UNITED ARAB EMIRATES (UAE).",
      "openfda": {
        "k_number": [
          "K870976",
          "K022947",
          "K221373",
          "K172220",
          "K942582",
          "K091007",
          "K133598",
          "K002116",
          "K915183",
          "K232291",
          "K801097",
          "K842138",
          "K093882",
          "K102726",
          "K822068",
          "K982014",
          "K863541",
          "K851615",
          "K080832",
          "K772001",
          "K163531",
          "K221895",
          "K770002",
          "K813034",
          "K210130",
          "K950990",
          "K871131",
          "K090006",
          "K103762",
          "K173834",
          "K002118",
          "K202125",
          "K070213",
          "K060053",
          "K062396",
          "K071318",
          "K021694",
          "K153376",
          "K020571",
          "K964017",
          "K030141",
          "K862973",
          "K131041",
          "K851614",
          "K011088",
          "K863960",
          "K861388",
          "K943803",
          "K052555",
          "K092463",
          "K151349",
          "K190282"
        ],
        "registration_number": [
          "9611109",
          "3019807891",
          "3002807090",
          "1828100",
          "3011554160",
          "3007427881",
          "8010762",
          "3042990265",
          "3009507273",
          "3010163695",
          "3006073153",
          "3001589655",
          "3010703925",
          "1218444",
          "3005941719",
          "2916596",
          "9680841",
          "2184009"
        ],
        "fei_number": [
          "3010703925",
          "3019807891",
          "1828100",
          "3011554160",
          "3007427881",
          "3009649876",
          "3042990265",
          "3002807111",
          "3009507273",
          "3010163695",
          "3005426550",
          "1000116158",
          "3006073153",
          "3001589655",
          "3006003617",
          "1218444",
          "2916596",
          "3002808267"
        ],
        "device_name": "Console, Heart-Lung Machine, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4220",
        "device_class": "2"
      }
    }
  ]
}