  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "105388",
      "product_res_number": "Z-0278-2012",
      "event_date_initiated": "2011-03-22",
      "event_date_posted": "2011-11-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-10-29",
      "res_event_number": "60430",
      "product_code": "LNX",
      "product_description": "SCC Soft Computer,  Softlab Laboratory information System with DORA versions: 2.0.0.4 with Hot Fix 1.2808, 1.2809, 1.2956, or 1.3476; 2.0.0.5; 2.0.0.6; 2.0.0.7 with Hot Fix 1.6595; and 2.1.0.3.",
      "code_info": "Softlab with DORA versions: 2.0.0.4 with Hot Fix 1.2808, 1.2809, 1.2956, or 1.3476; 2.0.0.5; 2.0.0.6; 2.0.0.7 with Hot Fix 1.6595; and 2.1.0.3.",
      "firm_fei_number": "1000307651",
      "recalling_firm": "SCC Soft Computer",
      "address_1": "5400 Tech Data Drive",
      "city": "Clearwater",
      "state": "FL",
      "postal_code": "33760",
      "additional_info_contact": "Kathryn Branca\n727-789-0100",
      "reason_for_recall": "Potential for patient data to be sent to the wrong patient record.  A software change affected how the Softlab handles recovery from certain types of table access conflicts. When such conflicts occur, this function has the potential to cause faulty writing to the database.",
      "root_cause_description": "Software change control",
      "action": "Clients were notified through SCC Soft Computer's proprietary Task Management System On 03/22/201. All clients have received a hot fix that corrected the issue or made the hosparam change that prevents the issue.",
      "product_quantity": "12",
      "distribution_pattern": "Sold throughout US and Canada.",
      "openfda": {
        "k_number": [
          "K921846"
        ]
      }
    }
  ]
}