  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "105363",
      "product_res_number": "Z-0362-2012",
      "event_date_initiated": "2011-10-25",
      "event_date_posted": "2011-12-02",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-03-16",
      "res_event_number": "60416",
      "product_code": "LNX",
      "product_description": "SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. \n\nSoftReports is a web-based report launching tool to be used in a hospital or laboratory environment by knowledgeable, trained, and experienced personnel. SoftReports enables the user to schedule and generate reports for the purpose of viewing and printing data stored in SCC product databases.",
      "code_info": "SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7.",
      "firm_fei_number": "1000307651",
      "recalling_firm": "SCC Soft Computer",
      "address_1": "5400 Tech Data Drive",
      "city": "Clearwater",
      "state": "FL",
      "postal_code": "33760",
      "additional_info_contact": "Kathryn Branca\n727-789-0100",
      "reason_for_recall": "On 10/25/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. When SoftReports Designer opened a layout, the system did not load the layout with the previously saved value in the NewPage property. This caused multiple patient records to be printed on the same page.  All affected clients were notified on the issue on",
      "root_cause_description": "Software design",
      "action": "SCC Soft Computer sent a \"Correction Notification\" letter dated October 25, 2011 to all affected customers.  The letter described the product, problem, and actions to be taken by the customers.  The hot fixes that were previously loaded which introduced this issue were corrected and delivered to all affected customers as a mandatory correction in hot fixes 1.7199.8 and 1.7200.8 and higher.",
      "product_quantity": "5",
      "distribution_pattern": "Nationwide Distribution.",
      "openfda": {
        "k_number": [
          "K921846"
        ]
      }
    }
  ]
}