  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "105265",
      "product_res_number": "Z-0364-2012",
      "event_date_initiated": "2011-10-31",
      "event_date_posted": "2011-12-02",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-07-25",
      "res_event_number": "60411",
      "product_code": "-IYE",
      "k_numbers": [
        "K111106"
      ],
      "product_description": "TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.",
      "code_info": "Model numbers: H191113, H191034, H191105, H191154, H191046, H191053, H191044, H191015, H191035, H191108, 191152, H191140, H191069, H191017, H191027, H191109, 191130, H191002, H191071, H191024, H191078, H191111, 191072, H191102, H191074, H191087, H191151, H191125, 191018, H191058, H191038, H191115, H191085, H191032, 191077, H191023, H191103, H191095, H191163, H191146, 191124, H191138, H191093, H191097, H191040, H191030, 191029, H191060, H191066, H191122, H191054, H191126, 191114, H191082, H191067, H191081, H191005, H191013, 191116, H191051, H191094, H191052, H191031, H191021, 191048, H191139, H191099, H191020, H191039, H191128, 191156, H191057, H191061, H191112, H191049, H191083, 191008, H191123, H191065, H191119, H191089, H191120, 191092, H191134, H191129, H191056, H191090, H191033, 191047, H191037, H191076, H191007, H191028, H191080, 191106, H191118, H191073, H191117, H191155, H191086, 191141, H191025, H191036, H191042, H191079, H191135, 191010, H191100, H191107, H191127, H191063, H191070, 191153, H191026, H191096, H191014, H191064, H191059, 191009, H191142, H191068, H191001, H191132, H191011, 191003, H191136, H191022, H191121, H191016, H191012, 191045, H191101, H191088, H191041, H191043, H191150, 191019, H191055, H191133, H191131, H191091, H191098, 191006 and H191110",
      "firm_fei_number": "2916710",
      "recalling_firm": "Varian Medical Systems, Inc. Oncology Systems",
      "address_1": "911 Hansen Way",
      "city": "Palo Alto",
      "state": "CA",
      "postal_code": "94304-1028",
      "additional_info_contact": "Peter J. Coronado\n650-424-5731",
      "reason_for_recall": "An anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.",
      "root_cause_description": "Device Design",
      "action": "On 11/04/2011, an Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps. Varian is developing a corrective action for this issue. A customer service representative will contact you when the correction is available to schedule installation on your system. Questions regarding the letter should be directed to the Varian Oncology Help Desk Contact Information at USA and Canada: 1-888-827- 4265 or Europe +41-41-749-8844.",
      "distribution_pattern": "Worldwide distribution including the US and Internationally.",