{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "104853",
      "product_res_number": "Z-0165-2012",
      "event_date_initiated": "2011-10-12",
      "event_date_posted": "2011-11-04",
      "recall_status": "Terminated",
      "event_date_terminated": "2013-05-01",
      "res_event_number": "60241",
      "product_code": "HNO",
      "k_numbers": [
        "K031960"
      ],
      "product_description": "Abbott brand IntraLase Femtosecond Laser System, Model 2; Catalog Numbers: 20003D/J/K, R20003D/J/K, 20005D/J/K; Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA \n\nProduct Usage:  Intended Use for: IntraLase Femtosecond (Model 2 and 3) and iFS Advanced Femtosecond Laser Systems. FS Laser Systems are ophthalmic surgical lasers used in certain patient populations: (1) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea; (2) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments; (3) In lamellar IEK [lntraLase Enabled Keratoplasty] and corneal harvesting ; (4) In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea; (5) In the creation of a lamellar cut/resection of the cornea for lamellar IEK for the creation of a penetrating cut/incision for penetrating IEK.",
      "code_info": "All units of these models (models 2, 3 and iFS) except model 1.",
      "firm_fei_number": "3006695864",
      "recalling_firm": "AMO Manufacturing USA, LLC",
      "address_1": "510 Cottonwood Dr",
      "city": "Milpitas",
      "state": "CA",
      "postal_code": "95035-7403",
      "additional_info_contact": "408-273-4100",
      "reason_for_recall": "The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or deeper) created using the Laser System.",
      "root_cause_description": "Process design",
      "action": "Abbott Medical Optics Inc. (AMO) issued an \"ADVISORY NOTICE\" letter dated October 12, 2011 to all affected customers. The letter identifies the affected products, the problem, and the actions needed to be taken.  Customers were instructed to exercise caution when making cuts that approach the corneal endothelium. The letter includes a 1-page Operator's Manual addendum that clarifies the recommendations.  The letter states that AMO will conduct a field correction to adjust the EOF on the affected FS Laser Systems.  The letter includes a response form to be completed and returned as per the instructions.  Customers were instructed to notify all appropriate staff/personnel of the Advisory Notice.  \nFor any further questions regarding the information or recommendations, contact an AMO representative at 1-800-266-3375",
      "product_quantity": "1226 total laser systems",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide)",
      "openfda": {
        "k_number": [
          "K021640",
          "K911062",
          "K022205",
          "K032910",
          "K022716",
          "K870807",
          "K760726",
          "K003820",
          "K980985",
          "K990900",
          "K904845",
          "K980510",
          "K082043",
          "K980508",
          "K984540",
          "K001176",
          "K011195",
          "K983317",
          "K043150",
          "K990227",
          "K043183",
          "K980675",
          "K941550",
          "K010260",
          "K014000",
          "K013337",
          "K993225",
          "K980924",
          "K972808",
          "K001144",
          "K042083",
          "K022982",
          "K033457",
          "K962661",
          "K973294",
          "K010872",
          "K042810",
          "K981742",
          "K992978",
          "K993190",
          "K032836",
          "K003143",
          "K960395",
          "K994015",
          "K172994",
          "K040204",
          "K002515",
          "K041110",
          "K033293",
          "K041206",
          "K935342",
          "K032297",
          "K032978",
          "K052891",
          "K051486",
          "K173371",
          "K001806",
          "K970377",
          "K972727",
          "K903912",
          "K001464",
          "K860001",
          "K033182",
          "K981554",
          "K072102",
          "K020122",
          "K974004",
          "K993122",
          "K002849",
          "K981155",
          "K011833",
          "K974598",
          "K864065",
          "K981071",
          "K030891",
          "K020079",
          "K954058",
          "K062465",
          "K020482",
          "K033236",
          "K031735",
          "K913697",
          "K030401"
        ],
        "registration_number": [
          "3007606649",
          "2085081",
          "1226074",
          "2029275",
          "9710340",
          "1043653",
          "3003288808",
          "3038356864",
          "2246552",
          "3000264417",
          "3010208880",
          "3016214537",
          "2028159",
          "1928237",
          "3004858034",
          "1920664",
          "1649518",
          "3009298127",
          "9615030",
          "8043792",
          "3010383847",
          "3006695864",
          "1644019",
          "3005012805",
          "3010291427",
          "3012236936",
          "1650907",
          "3005263901",
          "3010487482",
          "9615659",
          "3005941719"
        ],
        "fei_number": [
          "3002808269",
          "3001236277",
          "1000142667",
          "1000306431",
          "1043653",
          "3002483854",
          "3003288808",
          "3038356864",
          "2246552",
          "3000264417",
          "3010208880",
          "3002616700",
          "3016214537",
          "1928237",
          "1000125771",
          "3004858034",
          "3003289411",
          "1920664",
          "3001362763",
          "3009298127",
          "3010383847",
          "3006695864",
          "3005012805",
          "1000117501",
          "3010291427",
          "3002681132",
          "3012236936",
          "3005263901",
          "3010487482",
          "3002807111",
          "3003691824"
        ],
        "device_name": "Keratome, Ac-Powered",
        "medical_specialty_description": "Ophthalmic",
        "regulation_number": "886.4370",
        "device_class": "1"
      }
    }
  ]
}