  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "104239",
      "product_res_number": "Z-0105-2012",
      "event_date_initiated": "2011-09-26",
      "event_date_posted": "2011-10-24",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-11-04",
      "res_event_number": "60017",
      "product_code": "EPT",
      "product_description": "OPMI VARIO 700 Surgical Microscope.\n\nManufactured by: Carl Zeiss Meditec AG,\nOberkochen, Germany.\n\nProvide a magnified view of the surgical field.",
      "code_info": "Serial numbers: 6636101067; 6636101068; 6636101070.",
      "firm_fei_number": "2918630",
      "recalling_firm": "Carl Zeiss Meditec, Inc.",
      "address_1": "5160 Hacienda Dr",
      "city": "Dublin",
      "state": "CA",
      "postal_code": "94568-7562",
      "additional_info_contact": "Judith A. Brincombe, M.A.\n925-557-4616",
      "reason_for_recall": "An incorrect nut and bolt set was used in the axis three (3) spring assembly of the OPMI VARIO 700 surgical microscope distributed to three US locations.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "Carl Zeiss Meditec had Field Service Engineers directly contact all affected customers.  The Field Service Engineers will replace the defective component.  Questions regarding this Field Corrective Action should be directed to the Director of Clinical/Regulatory Affairs at 925-557-4616.",
      "product_quantity": "3",
      "distribution_pattern": "USA (nationwide) including the states of FL, GA, and LA .",
      "openfda": {
        "k_number": [
          "K931284",
          "K822014",
          "K933310",
          "K812234",
          "K802903",
          "K882269",
          "K860811",
          "K821689",
          "K823011",
          "K942233",
          "K832887",
          "K760696",
          "K822561",
          "K903412",
          "K844858",
          "K883581",
          "K802904",
          "K822009"
        ],
        "registration_number": [
          "3004001706",
          "3007123908",
          "3012322232",
          "3001644167",
          "3004091527",
          "8020986",
          "3003969012",
          "9611484",
          "3007050292",
          "3043007473",
          "3013403214",
          "3023816720",
          "2936485",
          "3003972337",
          "3023464160",
          "9615030",
          "1412854",
          "3010208880",
          "3011554142",
          "3013538161",
          "3038356864",
          "9611102",
          "3012075008",
          "3011110913",
          "1937051",
          "3008538162",
          "9610810",
          "1721685",
          "9617248",
          "3003974370",
          "1054713",
          "3015168749",
          "1725006",
          "3014342096",
          "1927430",
          "1836161",
          "8010047",
          "3016527960",
          "3008606424",
          "1000204638",
          "2028159",
          "3016875543",
          "3033162886",
          "3017861319",
          "3013360312",
          "3019252865",
          "3009746052",
          "2020550",
          "3009057691",
          "3010202439",
          "1926681",
          "3006546082",
          "3016017660",
          "3005216736",
          "3030860009",
          "3013576617",
          "3010160527",
          "3003713161",
          "3014214496",
          "8030607",
          "3016245408",
          "3014953747",
          "3007204745",
          "3003637092",
          "3012060713",
          "3010707607",
          "3005285078",
          "9610617",
          "3027305794",
          "3009768364",
          "3015136927",
          "3007532396",
          "3015488952",
          "9611036",
          "3015515531",
          "3032734659",
          "9615010"
        ],
        "fei_number": [
          "3002807040",
          "3004001706",
          "3007123908",
          "3012322232",
          "3001644167",
          "3004091527",
          "3003258642",
          "3005435924",
          "3003969012",
          "3007050292",
          "3043007473",
          "3013403214",
          "3023816720",
          "2936485",
          "3003972337",
          "1000125771",
          "3023464160",
          "1412854",
          "3002808268",
          "3010208880",
          "3011554142",
          "3013538161",
          "3038356864",
          "3012075008",
          "4196",
          "3011110913",
          "1937051",
          "3008538162",
          "1721685",
          "3003974370",
          "3002809599",
          "3015168749",
          "3014342096",
          "1927430",
          "1000114111",
          "3016527960",
          "3008606424",
          "3002750084",
          "1000204638",
          "3002807390",
          "3016875543",
          "3033162886",
          "3007475226",
          "3013360312",
          "3017861319",
          "3002808001",
          "3019252865",
          "3009746052",
          "2020550",
          "3009057691",
          "3002807375",
          "1926681",
          "3006546082",
          "3016017660",
          "3005216736",
          "3030860009",
          "3013576617",
          "3003713161",
          "3001362763",
          "3014214496",
          "3016245408",
          "3014953747",
          "3007204745",
          "3002807795",
          "3003481798",
          "3003637092",
          "3012060713",
          "3010707607",
          "3005285078",
          "3027305794",
          "3009768364",
          "3015136927",
          "3007532396",
          "3015488952",
          "3015515531",
          "3006509795",
          "3032734659"
        ],
        "device_name": "Microscope, Surgical",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4700",