  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "I",
      "report_to_fda": "*",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "",
      "manufacturer_contact_address_1": "",
      "manufacturer_contact_pcity": "",
      "event_type": "Injury",
      "report_number": "3009185973-2017-00538",
      "type_of_report": [
        "Initial submission"
      ],
      "product_problem_flag": "N",
      "date_received": "20170406",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K151359",
      "date_of_event": "20160913",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "",
      "reporter_occupation_code": "HEALTH PROFESSIONAL",
      "manufacturer_contact_plocal": "",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "",
      "source_type": [
        "COMPANY REPRESENTATIVE",
        "HEALTH"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "",
      "manufacturer_g1_state": "",
      "reporter_country_code": "US",
      "manufacturer_contact_area_code": "",
      "date_added": "20170406",
      "manufacturer_contact_f_name": "",
      "previous_use_code": "",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "244859",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20170406",
          "brand_name": "ROSA BRAIN",
          "generic_name": "COMPUTER-ASSISTED SURGICAL DEVICE",
          "manufacturer_d_name": "MEDTECH SA",
          "manufacturer_d_address_1": "ZAC EUREKA",
          "manufacturer_d_address_2": "900 RUE DU MAS DE VERCHANT",
          "manufacturer_d_city": "MONTPELLIER, LANGUEDOC-ROUSSILLON",
          "manufacturer_d_state": "",
          "manufacturer_d_zip_code": "34000",
          "manufacturer_d_zip_code_ext": "",
          "manufacturer_d_country": "FR",
          "manufacturer_d_postal_code": "34000",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "ROSA BRAIN 3.0",
          "catalog_number": "",
          "lot_number": "",
          "other_id_number": "",
          "device_availability": "Yes",
          "device_report_product_code": "HAW",
          "device_age_text": "",
          "device_evaluated_by_manufacturer": "",
          "combination_product_flag": "N",
          "udi_di": "",
          "udi_public": "",
          "openfda": {
            "device_name": "Neurological Stereotaxic Instrument",
            "medical_specialty_description": "Neurology",
            "regulation_number": "882.4560",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Imprecision",
        "Imprecision"
      ],
      "reporter_state_code": "PA",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "",
      "mfr_report_type": "Thirty-Day",
      "manufacturer_contact_country": "",
      "date_changed": "20250824",
      "health_professional": "",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20170406",
          "sequence_number_treatment": [
            ""
          ],
          "sequence_number_outcome": [
            "Other"
          ],
          "patient_age": "",
          "patient_sex": "",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "No Code Available",
            "No Code Available"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20170406",
      "initial_report_to_fda": "",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "",
      "manufacturer_contact_city": "",
      "mdr_report_key": "6463881",
      "removal_correction_number": "",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20160913",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        "Inspection"
      ],
      "manufacturer_g1_zip_code": "",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "",
      "adverse_event_flag": "Y",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "71881685",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "1",
          "text": "DEVICE BR 16 017 WAS INSPECTED FOR INVESTIGATION PURPOSES. ACCURACY TESTING CONFIRMED THE LASER WAS FULLY FUNCTIONAL AND OPERATING IN SPECIFICATION. NO PRODUCT DEFECT WAS IDENTIFIED. THE CAUSE OF THE DISCREPANCY WAS IDENTIFIED TO BE MOST LIKELY AN ALTERATION OF THE PATIENT ANATOMY BY THE HEAD HOLDER AND INTERFERENCE OF REFLECTIVE PRODUCT APPLIED TO THE SKIN."
        },
        {
          "mdr_text_key": "71881686",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "1",
          "text": "IT HAS WAS REPORTED THAT DURING A SURGERY THERE WAS A LASER INACCURACY OBSERVED DURING THE REGISTRATION PROCESS. THE REGISTRATION WAS REPEATED TWICE AND THE SURGEON DECIDED TO PROCEED WITH SURGERY. A DELAY OF 30 MINUTES WAS REPORTED."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "",
      "manufacturer_contact_exchange": "",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": ""
    }
  ]
}