  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "I",
      "report_to_fda": "*",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "",
      "manufacturer_contact_address_1": "",
      "manufacturer_contact_pcity": "",
      "event_type": "Injury",
      "report_number": "3009185973-2017-00100",
      "type_of_report": [
        "Initial submission"
      ],
      "product_problem_flag": "N",
      "date_received": "20170323",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K151511",
      "date_of_event": "20150512",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "",
      "reporter_occupation_code": "HEALTH PROFESSIONAL",
      "manufacturer_contact_plocal": "",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "",
      "source_type": [
        "HEALTH PROFESSIONAL"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "",
      "manufacturer_g1_state": "",
      "reporter_country_code": "FR",
      "manufacturer_contact_area_code": "",
      "date_added": "20170323",
      "manufacturer_contact_f_name": "",
      "previous_use_code": "",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "210201",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20170323",
          "brand_name": "ROSA SPINE",
          "generic_name": "COMPUTER-ASSISTED SURGICAL DEVICE",
          "manufacturer_d_name": "MEDTECH SA",
          "manufacturer_d_address_1": "ZAC EUREKA",
          "manufacturer_d_address_2": "900 RUE DU MAS DE VERCHANT",
          "manufacturer_d_city": "MONTPELLIER, LANGUEDOC-ROUSSILLON",
          "manufacturer_d_state": "",
          "manufacturer_d_zip_code": "34000",
          "manufacturer_d_zip_code_ext": "",
          "manufacturer_d_country": "FR",
          "manufacturer_d_postal_code": "34000",
          "device_operator": "I",
          "model_number": "ROSA SPINE 1.0.2",
          "catalog_number": "",
          "lot_number": "",
          "other_id_number": "",
          "device_availability": "Yes",
          "device_report_product_code": "OLO",
          "device_age_text": "",
          "device_evaluated_by_manufacturer": "",
          "combination_product_flag": "N",
          "udi_di": "",
          "udi_public": "",
          "openfda": {
            "device_name": "Orthopedic Stereotaxic Instrument",
            "medical_specialty_description": "Neurology",
            "regulation_number": "882.4560",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Device Displays Incorrect Message",
        "Device Displays Incorrect Message"
      ],
      "reporter_state_code": "",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "",
      "mfr_report_type": "Thirty-Day",
      "manufacturer_contact_country": "",
      "date_changed": "20250813",
      "health_professional": "",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20170323",
          "sequence_number_treatment": [
            ""
          ],
          "sequence_number_outcome": [
            "Other"
          ],
          "patient_age": "",
          "patient_sex": "",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "No Code Available",
            "No Code Available"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20170316",
      "initial_report_to_fda": "",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "",
      "manufacturer_contact_city": "",
      "mdr_report_key": "6427622",
      "removal_correction_number": "",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20150512",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        "Inspection"
      ],
      "manufacturer_g1_zip_code": "",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "",
      "adverse_event_flag": "Y",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "70686794",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "1",
          "text": "THE DEVICE SP 14 003 HAS BEEN INSPECTED FOR INVESTIGATION PURPOSE. THE TESTS PERFORMED CONFIRMED THAT A DESIGN DEFICIENCY CAUSED THE SOFTWARE FAILURE. THE INTERNAL COMPLAINT REFERENCE IS THE FOLLOWING: (B)(4)."
        },
        {
          "mdr_text_key": "70686795",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "1",
          "text": "IT HAS BEEN REPORTED THAT DURING A SURGERY, A SOFTWARE FAILURE WAS DETECTED. A COMMUNICATION ERROR HAS BEEN IDENTIFIED. A SURGERY DELAY HAS BEEN REPORTED BETWEEN 30 MINUTES AND 1 HOUR."
        }
      ],