{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "I",
      "report_to_fda": "*",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "",
      "manufacturer_contact_address_1": "",
      "manufacturer_contact_pcity": "",
      "event_type": "Injury",
      "report_number": "3005075696-2021-00091",
      "type_of_report": [
        "Initial submission"
      ],
      "product_problem_flag": "N",
      "date_received": "20210803",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K182077",
      "date_of_event": "20210512",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "",
      "reporter_occupation_code": "PHYSICIAN",
      "manufacturer_contact_plocal": "",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "",
      "source_type": [
        "HEALTH PROFESSIONAL",
        "LITERATUR"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "",
      "manufacturer_g1_state": "",
      "reporter_country_code": "US",
      "manufacturer_contact_area_code": "",
      "date_added": "20210803",
      "manufacturer_contact_f_name": "",
      "previous_use_code": "",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "1163843",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20210803",
          "brand_name": "MAZOR X SYSTEM",
          "generic_name": "ORTHOPEDIC STEREOTAXIC INSTRUMENT",
          "manufacturer_d_name": "MAZOR ROBOTICS LTD",
          "manufacturer_d_address_1": "5 SHACHAM STREET",
          "manufacturer_d_address_2": "P.O. BOX 3104",
          "manufacturer_d_city": "CAESAREA HEFA,IL",
          "manufacturer_d_state": "",
          "manufacturer_d_zip_code": "30795",
          "manufacturer_d_zip_code_ext": "67",
          "manufacturer_d_country": "IS",
          "manufacturer_d_postal_code": "3079567",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "TPL0059",
          "catalog_number": "TPL0059",
          "lot_number": "",
          "other_id_number": "",
          "device_availability": "No",
          "device_report_product_code": "OLO",
          "device_age_text": "",
          "device_evaluated_by_manufacturer": "",
          "combination_product_flag": "N",
          "udi_di": "",
          "udi_public": "",
          "openfda": {
            "device_name": "Orthopedic Stereotaxic Instrument",
            "medical_specialty_description": "Neurology",
            "regulation_number": "882.4560",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Adverse Event Without Identified Device or Use Problem",
        "Adverse Event Without Identified Device or Use Problem"
      ],
      "reporter_state_code": "NY",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "",
      "mfr_report_type": "Thirty-Day",
      "manufacturer_contact_country": "",
      "date_changed": "20250813",
      "health_professional": "",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20210803",
          "sequence_number_treatment": [
            ""
          ],
          "sequence_number_outcome": [
            "Required Intervention"
          ],
          "patient_age": "",
          "patient_sex": "",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "Erosion",
            "Cerebrospinal Fluid Leakage",
            "Unspecified Nervous System Problem",
            "Insufficient Information",
            "Erosion",
            "Cerebrospinal Fluid Leakage",
            "Unspecified Nervous System Problem",
            "Insufficient Information"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20210803",
      "initial_report_to_fda": "",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "",
      "manufacturer_contact_city": "",
      "mdr_report_key": "12269185",
      "removal_correction_number": "",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20210709",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        ""
      ],
      "manufacturer_g1_zip_code": "",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "",
      "adverse_event_flag": "Y",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "265244106",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "1",
          "text": "PLEASE NOTE THAT THIS DATE IS BASED OFF THE DATE OF PUBLICATION OF THE ARTICLE AS THE ACTUAL EVENT DATE WAS NOT PROVIDED. THE REPORTED EVENT WAS FROM THE FOLLOWING LITERATURE ARTICLE: LEE N, ZUCKERMAN S, BUCHANAN I, BODDAPATI V, MATHEW J, LEUNG E, PARK P, PHAM M, BUCHOLZ A, KHAN A, POLLINA J, MULLIN J, JAZINI E, HAINES C, SCHULER T, GOOD C, LOMBARDI J, LEHMAN R. IS THERE A DIFFERENCE BETWEEN NAVIGATED AND NON-NAVIGATED ROBOT COHORTS IN ROBOT-ASSISTED SPINE SURGERY? A MULTICENTER, PROPENSITY-MATCHED ANALYSIS OF 2,800 SCREWS AND 372 PATIENTS. THE SPINE JOURNAL. 2021. 1-9. HTTPS://DOI.ORG/10.1016/J.SPINEE.2021.05.015. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED."
        },
        {
          "mdr_text_key": "265244107",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "1",
          "text": "ABSTRACT: ROBOT-ASSISTED SPINE SURGERY CONTINUES TO RAPIDLY DEVELOP AS EVIDENCED BY THE GROWING LITERATURE IN RECENT YEARS. IN ADDITION TO DEMONSTRATING EXCELLENT PEDICLE SCREW ACCURACY, EARLY STUDIES HAVE EXPLORED THE IMPACT OF ROBOT-ASSISTED SPINE SURGERY ON REDUCING RADIATION TIME, LENGTH OF HOSPITAL STAY, OPERATIVE TIME, AND PERIOPERATIVE COMPLICATIONS IN COMPARISON TO CONVENTIONAL FREEHAND TECHNIQUE. RECENTLY, THE MAZOR X STEALTH EDITION WAS INTRODUCED IN 2018. THIS ROBOTIC SYSTEM INTEGRATES MEDTRONIC¿S STEALTH NAVIGATION TECHNOLOGY INTO THE MAZOR X PLATFORM, WHICH WAS INTRODUCED IN 2016. IT IS UNCLEAR WHAT THE IMPACT OF THESE ADVANCEMENTS HAVE MADE ON CLINICAL OUTCOMES. TO COMPARE THE OUTCOMES AND COMPLICATIONS BETWEEN THE MOST RECENT ITERATIONS OF THE MAZOR ROBOT SYSTEMS: MAZOR X AND MAZOR X STEALTH EDITION. AMONG FOUR DIFFERENT INSTITUTIONS, WE INCLUDED ADULT (=18 YEARS OLD) PATIENTS WHO UNDERWENT ROBOT-ASSISTED SPINE SURGERY WITH EITHER THE MAZOR X (NON-NAVIGATED ROBOT) OR STEALTH (NAVIGATED ROBOT) PLATFORMS. PRIMARY OUTCOMES INCLUDED ROBOT TIME PER SCREW, FLUOROSCOPIC RADIATION TIME, SCREW ACCURACY, ROBOT ABANDONMENT, AND CLINICAL OUTCOMES WITH A MINIMUM 90 DAY FOLLOW UP. A ONE-TO-ONE PROPENSITY-SCORE MATCHING ALGORITHM BASED ON PERIOPERATIVE FACTORS (E.G. DEMOGRAPHICS, COMORBIDITIES, PRIMARY DIAGNOSIS, OPEN VS. PERCUTANEOUS INSTRUMENTATION, PRIOR SPINE SURGERY, INSTRUMENTED LEVELS, PELVIC FIXATION, INTERBODY FUSION, NUMBER OF PLANNED ROBOT SCREWS) WAS EMPLOYED TO CONTROL FOR THE POTENTIAL SELECTION BIAS BETWEEN THE TWO ROBOTIC SYSTEMS. CHI-SQUARE/ FISHER EXACT TEST AND T-TEST/ANOVA WERE USED FOR CATEGORICAL AND CONTINUOUS VARIABLES, RESPECTIVELY. FROM A TOTAL OF 646 PATIENTS, A TOTAL OF 372 ADULT PATIENTS WERE INCLUDED IN THIS STUDY (X: 186, STEALTH: 186) AFTER PROPENSITY SCORE MATCHING. THE MEAN NUMBER OF INSTRUMENTED LEVELS WAS 4.3. THE MEAN NUMBER OF PLANNED ROBOT SCREWS WAS 7.8. SIMILAR TOTAL OPERATIVE TIME AND ROBOT TIME PER SCREW OCCURRED BETWEEN COHORTS (P>0.05). HOWEVER, STEALTH ACHIEVED SIGNIFICANTLY SHORTER FLUOROSCOPIC RADIATION TIME PER SCREW (STEALTH: 7.2 SECONDS VS. X: 10.4 SECONDS, P<(><<)>.001) THAN X. THE SCREW ACCURACY FOR BOTH ROBOTS WAS EXCELLENT (STEALTH: 99.6% VS. X: 99.1%, P=0.120). IN ADDITION, STEALTH ACHIEVED A SIGNIFICANTLY LOWER ROBOT ABANDONMENT RATE (STEALTH: 0% VS. X: 2.2%, P=0.044). FURTHERMORE, A LOWER BLOOD TRANSFUSION RATE WAS OBSERVED FOR STEALTH THAN X (STEALTH: 4.3% VS. X: 10.8%, P=0.018). NON-ROBOT RELATED COMPLICATIONS SUCH AS DURA TEAR, MOTOR/SENSORY DEFICITS, RETURN TO THE OPERATING ROOM DURING SAME ADMISSION, AND LENGTH OF STAY WAS SIMILAR BETWEEN ROBOTS (P>0.05). THE 90-DAY COMPLICATION RATES WERE LOW AND SIMILAR BETWEEN ROBOT COHORTS (STEALTH: 5.4% VS. X: 3.8%, P=0.456). IN THIS MULTICENTER STUDY, BOTH ROBOT SYSTEMS ACHIEVED EXCELLENT SCREW ACCURACY AND LOW ROBOT TIME PER SCREW. HOWEVER, USING STEALTH LED TO SIGNIFICANTLY LESS FLUOROSCOPIC RADIATION TIME, LOWER ROBOT ABANDONMENT RATES, AND REDUCED BLOOD TRANSFUSION RATES THAN MAZOR X. OTHER FACTORS INCLUDING LENGTH OF STAY, AND 90-DAY COMPLICATIONS WERE SIMILAR. REPORTED EVENT: 1. AFTER USE OF THE GUIDANCE SYSTEM DURING SPINAL PROCEDURES, THERE WERE FIVE OCCURRENCES OF INTRA-OPERATIVE DURAL TEARS. 2. DURING THE 90 DAY FOLLOW UP PERIOD AFTER THE USE OF THE GUIDANCE SYSTEM DURING SPINAL PROCEDURES, THERE WAS ONE OCCURRENCE OF WOUND COMPLICATIONS, TWO OCCURRENCES OF NEUROLOGIC DEFICIT, THREE OCCURRENCES OF IMPLANT FAILURE, TWO OCCURRENCES OF PERSISTENT SYMPTOMS REQUIRING DECOMPRESSION, ONE OCCURRENCE OF SCREW MALPOSITION AND TWO OCCURRENCES OF AN UNKNOWN SYMPTOM THAT REQUIRED REOPERATION."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "",
      "manufacturer_contact_exchange": "",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": ""
    }
  ]
}