  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "",
      "report_to_fda": "",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "CA",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "SANTA CLARA",
      "manufacturer_contact_address_1": "3410 CENTRAL EXPRESSWAY",
      "manufacturer_contact_pcity": "40852321",
      "event_type": "Injury",
      "report_number": "2955842-2025-34980",
      "type_of_report": [
        "Initial submission",
        "Followup"
      ],
      "product_problem_flag": "N",
      "date_received": "20250826",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K143132",
      "date_of_event": "20250728",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "95051",
      "reporter_occupation_code": "NURSE",
      "manufacturer_contact_plocal": "4085232100",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "NIELSON",
      "source_type": [
        "Other",
        "Foreign",
        "Health Professional"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "95051",
      "manufacturer_g1_state": "CA",
      "reporter_country_code": "CH",
      "manufacturer_contact_area_code": "408",
      "date_added": "20250826",
      "manufacturer_contact_f_name": "IZABEL",
      "device_date_of_manufacturer": "20240514",
      "previous_use_code": "I",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "2123793",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20250826",
          "brand_name": "DA VINCI",
          "generic_name": "HARMONIC ACE",
          "manufacturer_d_name": "INTUITIVE SURGICAL, INC",
          "manufacturer_d_address_1": "3410 CENTRAL EXPRESSWAY",
          "manufacturer_d_address_2": "",
          "manufacturer_d_city": "SANTA CLARA",
          "manufacturer_d_state": "CA",
          "manufacturer_d_zip_code": "95051",
          "manufacturer_d_zip_code_ext": "",
          "manufacturer_d_country": "US",
          "manufacturer_d_postal_code": "95051",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "480275-10",
          "catalog_number": "480275",
          "lot_number": "L80240514",
          "other_id_number": "",
          "device_availability": "Device was returned to manufacturer",
          "device_report_product_code": "NAY",
          "device_age_text": "DA",
          "device_evaluated_by_manufacturer": "Y",
          "combination_product_flag": "N",
          "udi_di": "10886874112738",
          "udi_public": "(01)10886874112738(17)260531(10)L80240514",
          "openfda": {
            "device_name": "System, Surgical, Computer Controlled Instrument",
            "medical_specialty_description": "Gastroenterology, Urology",
            "regulation_number": "876.1500",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Material Fragmentation",
        "Material Fragmentation"
      ],
      "reporter_state_code": "",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "09/22/2025",
      "mfr_report_type": "Follow-Up",
      "manufacturer_contact_country": "US",
      "date_changed": "20251103",
      "health_professional": "Y",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "4085",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20250826",
          "sequence_number_treatment": [
            "DA VINCI INSTRUMENTS AND ACCESSORIES"
          ],
          "sequence_number_outcome": [
            "Required Intervention"
          ],
          "patient_age": "NA",
          "patient_sex": "Unknown",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "No Clinical Signs, Symptoms or Conditions",
            "No Clinical Signs, Symptoms or Conditions"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20250731",
      "initial_report_to_fda": "No",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "US",
      "manufacturer_contact_city": "SANTA CLARA",
      "mdr_report_key": "22894074",
      "removal_correction_number": "N/A",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20250731",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        "Other"
      ],
      "manufacturer_g1_zip_code": "95051",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "INTUITIVE SURGICAL, INC",
      "adverse_event_flag": "N",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "Y",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "378530565",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "0",
          "text": "IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PANCREATICODUODENECTOMY PROCEDURE, THE CUSTOMER ENCOUNTERED A \"REDUCE TIP PRESSURE\" MESSAGE WHILE USING A HARMONIC ACE INSTRUMENT. A NURSE TOOK OUT THE INSTRUMENT, NOTED THAT THE INSTRUMENT WAS INTACT, AND THEN WIPED ADHERED TISSUE AT THE TIP OF THE INSTRUMENT WITH A GAUZE. AFTER REINSTALLING THE HARMONIC ACE INSTRUMENT, THE CUSTOMER NOTICED THAT THE TIP OF THE HARMONIC ACE BROKE AND A FRAGMENT HAD FALLEN INSIDE THE PATIENT. THE FRAGMENT WAS RETRIEVED DURING THE SAME SURGICAL PROCEDURE WHICH WAS COMPLETED ROBOTICALLY USING A BACKUP HARMONIC ACE INSTRUMENT."
        },
        {
          "mdr_text_key": "378530566",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "THE HARMONIC ACE INSTRUMENT HAS NOT BEEN RECEIVED FOR FAILURE ANALYSIS EVALUATION. A REVIEW OF AN IMAGE PROVIDED BY THE CUSTOMER WAS PERFORMED. THE IMAGE APPEARS TO SHOW A BROKEN CURVED BLADE ON A HARMONIC ACE INSTRUMENT."
        },
        {
          "mdr_text_key": "384087239",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "0",
          "text": "REFER TO H11 FOR FOLLOW-UP INFORMATION."
        },
        {
          "mdr_text_key": "384087240",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE HARMONIC ACE INSTRUMENT WAS ANALYZED. THE INSTRUMENT WAS FOUND TO HAVE THE CURVED BLADE BROKEN AT THE TIP OF THE BLADE. A PIECE APPROXIMATELY 0.456\" X 0.085\" WAS FOUND TO BE BROKEN OFF. THE BROKEN PIECE WAS NOT RETURNED. COMPONENTS ADJACENT TO THIS BROKEN BLADE DO NOT SHOW DAMAGE. ADDITIONAL OBSERVATION RELATED TO CUSTOMER REPORTED COMPLAINT WAS THAT THE INSTRUMENT WAS FOUND TO FAIL ENERGY RECOGNITION TEST. THE INSTRUMENT WAS PLACED ON AN INHOUSE SYSTEM. THE INSTRUMENT WAS CONNECTED TO AN IN-HOUSE ENERGY GENERATOR. A TORQUE WRENCH WAS USED TO TIGHTEN THE ASSEMBLY UNTIL IT WAS AS TIGHT AS IT COULD BE (CLICKING SOUNDS). THE INSTRUMENT FAILED THE ENERGY RECOGNITION TEST."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "3410 CENTRAL EXPRESSWAY",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "95051",
      "manufacturer_contact_exchange": "408",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": "10/09/2025"
    }
  ]
}