  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "",
      "report_to_fda": "",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "CA",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "SANTA CLARA",
      "manufacturer_contact_address_1": "3410 CENTRAL EXPRESSWAY",
      "manufacturer_contact_pcity": "40852321",
      "event_type": "Injury",
      "report_number": "2955842-2024-19759",
      "type_of_report": [
        "Initial submission"
      ],
      "product_problem_flag": "Y",
      "date_received": "20240924",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K143132",
      "date_of_event": "20240829",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "95051",
      "reporter_occupation_code": "003",
      "manufacturer_contact_plocal": "4085232100",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "NIELSON",
      "source_type": [
        "Other",
        "Foreign"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "95051",
      "manufacturer_g1_state": "CA",
      "reporter_country_code": "JA",
      "manufacturer_contact_area_code": "408",
      "date_added": "20240924",
      "manufacturer_contact_f_name": "IZABEL",
      "device_date_of_manufacturer": "20231102",
      "previous_use_code": "I",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "2105269",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20240924",
          "brand_name": "DA VINCI",
          "generic_name": "HARMONIC ACE",
          "manufacturer_d_name": "INTUITIVE SURGICAL, INC",
          "manufacturer_d_address_1": "3410 CENTRAL EXPRESSWAY",
          "manufacturer_d_address_2": "",
          "manufacturer_d_city": "SANTA CLARA",
          "manufacturer_d_state": "CA",
          "manufacturer_d_zip_code": "95051",
          "manufacturer_d_zip_code_ext": "",
          "manufacturer_d_country": "US",
          "manufacturer_d_postal_code": "95051",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "480275-08",
          "catalog_number": "480275",
          "lot_number": "L80231102 0194",
          "other_id_number": "",
          "device_availability": "Device was returned to manufacturer",
          "date_returned_to_manufacturer": "20240910",
          "device_report_product_code": "NAY",
          "device_age_text": "",
          "device_evaluated_by_manufacturer": "Y",
          "combination_product_flag": "N",
          "udi_di": "10886874112738",
          "udi_public": "(01)10886874112738(17)251130(10)L80231102",
          "openfda": {
            "device_name": "System, Surgical, Computer Controlled Instrument",
            "medical_specialty_description": "Gastroenterology, Urology",
            "regulation_number": "876.1500",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Detachment of Device or Device Component",
        "Detachment of Device or Device Component"
      ],
      "reporter_state_code": "",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "",
      "mfr_report_type": "Initial",
      "manufacturer_contact_country": "US",
      "date_changed": "20250821",
      "health_professional": "",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "4085",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20240924",
          "sequence_number_treatment": [
            "DA VINCI INSTRUMENTS AND ACCESSORIES"
          ],
          "sequence_number_outcome": [
            "Required Intervention"
          ],
          "patient_age": "NA",
          "patient_sex": "Prefer Not To Disclose",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "Foreign Body In Patient",
            "Foreign Body In Patient"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20240828",
      "initial_report_to_fda": "No",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "US",
      "manufacturer_contact_city": "SANTA CLARA",
      "mdr_report_key": "20294671",
      "removal_correction_number": "N/A",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20240828",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        "Other"
      ],
      "manufacturer_g1_zip_code": "95051",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "INTUITIVE SURGICAL, INC",
      "adverse_event_flag": "Y",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "Y",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "352540267",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE HARMONIC ACE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A DISLODGED TEFLON PAD ON THE CLAMPING ARM AT THE DISTAL END. THERE WAS NO MATERIAL MISSING. THE EVENT CAUSED PARTIALLY OF WHOLLY BY THE USER OF THE DEVICE INCLUDING SAMPLE HANDLING. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS."
        },
        {
          "mdr_text_key": "352540268",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "0",
          "text": "IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED DISTAL GASTRECTOMY SURGICAL PROCEDURE THE TISSUE PAD OF THE HARMONIC ACE INSTRUMENT BECAME DISLODGED DURING THE PROCEDURE. THE FRAGMENT WAS RETRIEVED DURING THE SAME PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A BACKUP INSTRUMENT. INTUITIVE RECEIVED THE FOLLOWING ADDITIONAL INFORMATION VIA FOLLOW-UP: THE FRAGMENT FELL WHEN THE SURGEON WAS TREATING THE GREATER CURVATURE SIDE OF THE STOMACH, DURING THE FIRST FIRE. IT REPORTEDLY FELT LIKE POWDER WAS FLYING AROUND. THE SURGEON PICKED UP THE FRAGMENT, WRAPPED IT IN GAUZE, AND REMOVED IT. THE HARMONIC ACE INSTRUMENT WAS IMMEDIATELY REPLACED WITH A NEW ONE, AND THE SURGERY WAS COMPLETED WITHOUT ANY PROBLEMS. THE INSTRUMENT DID NOT HAVE ANY COLLISIONS AND DID NOT APPEAR TO HAVE ANY OPERATIONAL PROBLEMS BEFORE THE REPORTED ISSUE OCCURRED. THE INSTRUMENT HAD NOT BEEN REMOVED PRIOR TO THE BREAKAGE. UPON FINAL REMOVAL OF THE INSTRUMENT, THE WRIST WAS STRAIGHTENED. NO RESISTANCE WAS FELT THROUGH THE CANNULA. NO DAMAGE WAS NOTICED TO THE CANNULA. NO ADDITIONAL INSTRUMENT DAMAGE WAS FOUND. THERE WERE NO ADDITIONAL PROCEDURES PERFORMED, NOR TESTING, DUE TO THE ISSUE. THE PATIENT HAS NOT RETURNED TO THE HOSPITAL DUE TO POST-SURGICAL COMPLICATIONS."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "3410 CENTRAL EXPRESSWAY",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "95051",
      "manufacturer_contact_exchange": "408",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": ""
    }
  ]
}