  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "",
      "report_to_fda": "",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "CA",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "SANTA CLARA",
      "manufacturer_contact_address_1": "3410 CENTRAL EXPRESSWAY",
      "manufacturer_contact_pcity": "40852321",
      "event_type": "Injury",
      "report_number": "2955842-2024-14736",
      "type_of_report": [
        "Initial submission",
        "Followup",
        "Followup"
      ],
      "product_problem_flag": "Y",
      "date_received": "20240520",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K191280",
      "date_of_event": "20240411",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "95051",
      "reporter_occupation_code": "003",
      "manufacturer_contact_plocal": "4085232100",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "NIELSON",
      "source_type": [
        "Other",
        "Foreign"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "95051",
      "manufacturer_g1_state": "CA",
      "reporter_country_code": "JA",
      "manufacturer_contact_area_code": "408",
      "date_added": "20240520",
      "manufacturer_contact_f_name": "IZABEL",
      "device_date_of_manufacturer": "20230831",
      "previous_use_code": "I",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "1146480",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20240520",
          "brand_name": "DA VINCI ENERGY",
          "generic_name": "SYNCHROSEAL",
          "manufacturer_d_name": "INTUITIVE SURGICAL, INC",
          "manufacturer_d_address_1": "3410 CENTRAL EXPRESSWAY",
          "manufacturer_d_address_2": "",
          "manufacturer_d_city": "SANTA CLARA",
          "manufacturer_d_state": "CA",
          "manufacturer_d_zip_code": "95051",
          "manufacturer_d_zip_code_ext": "",
          "manufacturer_d_country": "US",
          "manufacturer_d_postal_code": "95051",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "480440-06",
          "catalog_number": "480440",
          "lot_number": "L11230831 0139",
          "other_id_number": "",
          "device_availability": "Device was returned to manufacturer",
          "date_returned_to_manufacturer": "20240509",
          "device_report_product_code": "NAY",
          "device_age_text": "DA",
          "device_evaluated_by_manufacturer": "Y",
          "combination_product_flag": "N",
          "udi_di": "10886874117306",
          "udi_public": "(01)10886874117306(17)250831(10)L11230831",
          "openfda": {
            "device_name": "System, Surgical, Computer Controlled Instrument",
            "medical_specialty_description": "Gastroenterology, Urology",
            "regulation_number": "876.1500",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Detachment of Device or Device Component",
        "Detachment of Device or Device Component"
      ],
      "reporter_state_code": "",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "05/20/2024;07/09/2024",
      "mfr_report_type": "Initial",
      "manufacturer_contact_country": "US",
      "date_changed": "20250820",
      "health_professional": "",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "4085",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20240520",
          "sequence_number_treatment": [
            "DA VINCI INSTRUMENTS AND ACCESSORIES"
          ],
          "sequence_number_outcome": [
            "Required Intervention"
          ],
          "patient_age": "NA",
          "patient_sex": "Unknown",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "Foreign Body In Patient",
            "Foreign Body In Patient"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20240423",
      "initial_report_to_fda": "No",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "US",
      "manufacturer_contact_city": "SANTA CLARA",
      "mdr_report_key": "19350094",
      "removal_correction_number": "N/A",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20240423",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        "Other"
      ],
      "manufacturer_g1_zip_code": "95051",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "INTUITIVE SURGICAL, INC",
      "adverse_event_flag": "Y",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "Y",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "342865069",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE SYNCHROSEAL INSTRUMENT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS; HOWEVER, THE EVALUATION IS NOT YET COMPLETE. A FOLLOW-UP MDR WILL BE SUBMITTED ONCE THE PRODUCT IS EVALUATED AND/ OR IF ADDITIONAL INFORMATION IS RECEIVED."
        },
        {
          "mdr_text_key": "342865070",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "0",
          "text": "IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SYNCHROSEAL ¿SCREW¿ CAME OFF. THE FRAGMENT WAS RETRIEVED DURING THE SAME PROCEDURE. THE PROCEDURE WAS COMPLETING AS PLANNED. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE WITH NO ISSUE. THE SCREW FELL INTO THE PATIENT¿S ANATOMY WHEN THE INSTRUMENT WAS BEING USED. THE INSTRUMENT DID NOT GRASP THE TISSUE. THE SURGEON BELIEVED THAT MANUFACTURING WAS THE CAUSE OF THE ISSUE. THE INSTRUMENT DID NOT COLLIDE WITH ANY OTHER INSTRUMENT OR HARD MATERIAL. THE SURGEON USED A LAPAROSCOPIC FORCEPS TO RETRIEVE THE FRAGMENT. NO ADDITIONAL SURGICAL PROCEDURE WAS PERFORMED TO REMOVE THE FRAGMENT. UPON VISUAL INSPECTION, THE SURGEON CONFIRMED THAT THERE WAS NO REMAINING FRAGMENT. NO POST-OPERATIVE TESTS WERE PERFORMED. THE SURGEON STATED THAT NO PATIENT INJURY. THE PATIENT DID NOT RETURN TO THE HOSPITAL DUE TO ANY POST-SURGICAL COMPLICATIONS RELATED TO RETAINING A FOREIGN OBJECT. THE INSTRUMENT AND THE FRAGMENT WILL BE RETURNED."
        },
        {
          "mdr_text_key": "344677715",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "INTUITIVE SURGICAL, INC. (ISI) RECEIVED A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE SYNCHROSEAL INSTRUMENT WAS ANALYZED AND FOUND TO HAVE THE WASHER DISLODGED FROM ITS ORIGINAL POSITION AND NOT ATTACHED TO THE INSTRUMENT. THE INSTRUMENT WAS COMPARED TO AN IN-HOUSE SYNCHROSEAL, AND THE WASHER WAS FOUND TO BE MISSING. THE MISSING PIECE WAS NOT RETURNED WITH THE INSTRUMENT. NO FAILURES WERE FOUND IN THE LOGS. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS."
        },
        {
          "mdr_text_key": "344677716",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "0",
          "text": "REFER TO H10/H11 FOR FOLLOW-UP INFORMATION."
        },
        {
          "mdr_text_key": "349032408",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "THE SYNCHROSEAL INSTRUMENT WAS TRANSFERRED TO THE FAILURE ANALYSIS ENGINEER TEAM FOR FURTHER INVESTIGATION. INITIAL FAILURE ANALYSIS FINDINGS WERE CONFIRMED. THE JAW COVER WAS REMOVED, AND IT WAS NOTICED THAT THE END OF THE PIVOT PIN THAT'S SUPPOSED TO BE SWAGED WAS NOT SWAGED PROPERLY THAT LIKELY LED TO THE WASHER GETTING DISLODGED FROM THAT SIDE OF THE PIN."
        },
        {
          "mdr_text_key": "349032409",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "0",
          "text": "REFER TO H10/H11 FOR FOLLOW-UP INFORMATION."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "3410 CENTRAL EXPRESSWAY",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "95051",
      "manufacturer_contact_exchange": "408",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": "06/13/2024;08/04/2024"
    }
  ]
}