  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "",
      "report_to_fda": "",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "CA",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "SANTA CLARA",
      "manufacturer_contact_address_1": "3410 CENTRAL EXPRESSWAY",
      "manufacturer_contact_pcity": "40852321",
      "event_type": "Injury",
      "report_number": "2955842-2021-10665",
      "type_of_report": [
        "Initial submission",
        "Followup",
        "Followup"
      ],
      "product_problem_flag": "Y",
      "date_received": "20210616",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K143132",
      "date_of_event": "20210518",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "",
      "reporter_occupation_code": "NURSE",
      "manufacturer_contact_plocal": "4085232100",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "WANG",
      "source_type": [
        "Other",
        "Foreign",
        "Health Professional"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "",
      "manufacturer_g1_state": "CA",
      "reporter_country_code": "KS",
      "manufacturer_contact_area_code": "408",
      "date_added": "20210616",
      "manufacturer_contact_f_name": "DAVID",
      "device_date_of_manufacturer": "20201203",
      "previous_use_code": "I",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "903968",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20210616",
          "brand_name": "DA VINCI",
          "generic_name": "HARMONIC ACE",
          "manufacturer_d_name": "INTUITIVE SURGICAL, INC",
          "manufacturer_d_address_1": "3410 CENTRAL EXPRESSWAY",
          "manufacturer_d_address_2": "",
          "manufacturer_d_city": "SANTA CLARA",
          "manufacturer_d_state": "CA",
          "manufacturer_d_zip_code": "",
          "manufacturer_d_zip_code_ext": "",
          "manufacturer_d_country": "",
          "manufacturer_d_postal_code": "",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "480275-08",
          "catalog_number": "480275",
          "lot_number": "M90201209 0038",
          "other_id_number": "",
          "device_availability": "Device was returned to manufacturer",
          "date_returned_to_manufacturer": "20210625",
          "device_report_product_code": "NAY",
          "device_age_text": "DA",
          "device_evaluated_by_manufacturer": "Y",
          "combination_product_flag": "N",
          "udi_di": "10886874112738",
          "udi_public": "(01)10886874112738(10)NA",
          "openfda": {
            "device_name": "System, Surgical, Computer Controlled Instrument",
            "medical_specialty_description": "Gastroenterology, Urology",
            "regulation_number": "876.1500",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Detachment of Device or Device Component",
        "Detachment of Device or Device Component"
      ],
      "reporter_state_code": "",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "06/28/2021;05/24/2021",
      "mfr_report_type": "",
      "manufacturer_contact_country": "",
      "date_changed": "20250819",
      "health_professional": "Y",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "4085",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20210616",
          "sequence_number_treatment": [
            "DA VINCI INSTRUMENTS AND ACCESSORIES"
          ],
          "sequence_number_outcome": [
            ""
          ],
          "patient_age": "44 YR",
          "patient_sex": "Male",
          "patient_weight": "76 KG",
          "patient_ethnicity": "",
          "patient_race": "Asian",
          "patient_problems": [
            "No Clinical Signs, Symptoms or Conditions",
            "No Clinical Signs, Symptoms or Conditions"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20210524",
      "initial_report_to_fda": "No",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "",
      "manufacturer_contact_city": "SANTA CLARA",
      "mdr_report_key": "12009759",
      "removal_correction_number": "N/A",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20210524",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        "Other"
      ],
      "manufacturer_g1_zip_code": "",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "INTUITIVE SURGICAL, INC",
      "adverse_event_flag": "Y",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "Y",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "326260779",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "D11 - INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE HARMONIC ACE INSTRUMENT INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. FAILURE ANALYSIS INVESTIGATIONS CONFIRMED THE CUSTOMER REPORTED COMPLAINT OF \"CUSTOMER WAS USING A HARMONIC INSTRUMENT AND SUDDENLY FELT ABNORMAL MOVEMENT OF THE JAW AND FOUND BLADE HAS BEEN BROKE, SO THE CUSTOMER RETRIEVED THE FRAGMENT OUTSIDE OF THE PATIENT AND WAS LOOKING AT THE INSTRUMENT BUT COULDN'T FIND OTHER DAMAGES.\" FAILURE ANALYSIS FOUND THE INSTRUMENT TO HAVE A PRIMARY FAILURE OF BROKEN BLADE TO BE RELATED TO THE CUSTOMER REPORTED COMPLAINT. THE BLADE BROKE AT ROUGHLY 0.194¿ FROM THE BASE AND THE BROKEN PIECE WAS RETURNED. CRACKED OR BROKEN BLADES ARE TRIGGERED BY ANY INADVERTENT CONTACT WITH STAPLES, CLIPS, OR OTHER INSTRUMENTS WHILE THE DEVICE IS ACTIVATED. IN ADDITION, SCRATCHES ON THE BLADE TIP MAY ALSO LEAD TO PREMATURE BLADE FAILURE. BLADE DAMAGE MAY BE DETECTED BY THE GENERATOR WITH A SOLID TONE OR AN ERROR. THE ROOT CAUSE OF BROKEN INSTRUMENT BLADES IS TYPICALLY ATTRIBUTED TO MISHANDLING/MISUSE. ADDITIONAL OBSERVATION(S) NOT REPORTED BY SITE WAS THAT THE INSTRUMENT WAS FOUND TO HAVE BLADE DAMAGE. THE ROOT CAUSE OF MECHANICAL INDENTATIONS/BURRS ON THE INSTRUMENT BLADES IS TYPICALLY ATTRIBUTED TO MISHANDLING/MISUSE. A REVIEW OF THE INSTRUMENT LOG FOR THE HARMONIC ACE INSTRUMENT (PART # 480275-08/ (B)(6) ) ASSOCIATED WITH THIS EVENT HAS BEEN PERFORMED. PER LOGS, THE ENDOSCOPE WAS LAST USED ON (B)(6) 2021 ON SYSTEM (B)(6). THIS IS A SINGLE USE INSTRUMENT. IN ADDITION, A REVIEW OF THE SITE'S COMPLAINT HISTORY IDENTIFIED NO OTHER COMPLAINTS RELATED TO THE INSTRUMENT AND OR THIS EVENT. NO IMAGE OR VIDEO CLIP FOR THE REPORTED EVENT WAS SUBMITTED TO ISI FOR REVIEW. IN ADDITION, ISI BECAME AWARE THAT THE PROCEDURE RELATED TO THIS EVENT WAS AN ESOPHAGECTOMY TRANSTHORACIC-NECK ANASTOMOSIS."
        },
        {
          "mdr_text_key": "335712267",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "BASED ON A RE-EVALUATION OF THE COMPLAINT INFORMATION, THIS COMPLAINT HAS BEEN RECLASSIFIED AS AN ADVERSE EVENT AND PRODUCT PROBLEM RATHER THAN JUST A PRODUCT PROBLEM, AS PREVIOUSLY REPORTED. CORRECTED INFORMATION CAN BE FOUND THE FOLLOWING FIELD: B1, B2, AND H1. B1 UPDATED FROM \"PRODUCT PROBLEM\" TO \"ADVERSE EVENT AND PRODUCT PROBLEM\". B2 UPDATED TO \"REQUIRED INTERVENTION\". H1 UPDATED FROM \"MALFUNCTION\" TO \"SERIOUS INJURY\"."
        },
        {
          "mdr_text_key": "335712268",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "0",
          "text": "REFER TO H10/H11 FOR FOLLOW-UP INFORMATION."
        },
        {
          "mdr_text_key": "261989502",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "1",
          "text": "INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE HARMONIC ACE INSTRUMENT FOR EVALUATION. THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE HAS NOT BEEN DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/ OR IF ADDITIONAL INFORMATION IS RECEIVED. A REVIEW OF THE SITE'S COMPLAINT HISTORY DOES NOT SHOW ANY ADDITIONAL COMPLAINTS RELATED TO THIS PRODUCT AND/OR THIS EVENT. IN ADDITION, THE BATCH SEQUENCE NUMBER OF THE PRODUCT'S LOT NUMBER WAS NOT PROVIDED. THEREFORE, AN INSTRUMENT LOG REVIEW OF THE PRODUCT RELATED TO THE COMPLAINT CANNOT BE PERFORMED AT THIS TIME. NO IMAGE OR VIDEO CLIP FOR THE REPORTED EVENT WAS SUBMITTED TO ISI FOR REVIEW. THIS EVENT IS BEING REPORTED BECAUSE THE BLADE OF THE HARMONIC ACE INSTRUMENT REPORTEDLY BROKE AND FELL INSIDE THE PATIENT. ALL FRAGMENTS WERE RETRIEVED, AND NO ADDITIONAL SURGICAL INTERVENTION WAS REQUIRED. HOWEVER, UNINTENDED FRAGMENTS FALLING INSIDE THE PATIENT MAY REQUIRE SURGICAL INTERVENTION. AT THIS TIME, IT IS UNKNOWN WHAT CAUSED THE BREAKAGE TO OCCUR. WHILE THERE WAS NO HARM OR INJURY TO THE PATIENT, THE REPORTED FAILURE MODE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF IT WERE TO RECUR."
        },
        {
          "mdr_text_key": "261989503",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "1",
          "text": "IT WAS REPORTED THAT AFTER TESTING THE HARMONIC ACE DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER WAS USING THE INSTRUMENT AND SUDDENLY FELT \"ABNORMAL\" MOVEMENT OF THE JAW. THE INSTRUMENT BLADE BROKE AND FELL INSIDE THE PATIENT. THE FRAGMENT WAS RETRIEVED DURING THE SAME PROCEDURE. THE CUSTOMER FOUND NO OTHER DAMAGES ON THE INSTRUMENT. A SIMILAR BACKUP DA VINCI INSTRUMENT WAS USED. THE PROCEDURE WAS COMPLETED WITH NO REPORT OF PATIENT HARM, ADVERSE OUTCOME, OR INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE WITH NO DAMAGE FOUND. THE INSTRUMENT WAS IN USE FOR APPROXIMATELY 20 MINUTES AND BROKE WHILE IT WAS BEING USED FOR DISSECTING TISSUE. ALL THE FRAGMENTS WERE RETRIEVED DURING THE SAME PROCEDURE AND WAS CONFIRMED WITH VISUAL INSPECTION. IT WAS UNKNOWN WHAT CAUSED THE BREAKAGE AS THE INSTRUMENT DID NOT COLLIDE WITH ANY HARD MATERIALS OR OTHER INSTRUMENTS. THERE WAS NO PATIENT INJURY. THE SURGEON DID NOT NOTICE ANY ISSUES WITH THE FUNCTIONALITY OF THE INSTRUMENT. THE FRAGMENT(S) DID NOT FALL INSIDE THE PATIENT DURING AN INSTRUMENT TIP OR ACCESSORY COLLISION. THE INSTRUMENT WAS NOT REMOVED DURING THE PROCEDURE (PRIOR TO THE BREAKAGE). UPON FINAL REMOVAL OF THE INSTRUMENT, THE WRIST WAS STRAIGHTENED, AND THE SURGICAL STAFF DID NOT FEEL ANY RESISTANCE UPON REMOVAL OF THE INSTRUMENT THROUGH THE CANNULA. NO DAMAGE TO THE CANNULA NOR THE INSTRUMENT AFTER THE EVENT OCCURRED. NO POST-OPERATIVE TESTS LIKE AN X-RAY OR ULTRASOUND WERE PERFORMED TO CHECK FOR REMAINING FRAGMENTS. THE PATIENT HAS NOT RETURNED TO THE HOSPITAL DUE TO EXPERIENCING ANY POST-SURGICAL COMPLICATIONS RELATED TO RETAINING A FOREIGN OBJECT. THERE WERE NO PHOTOGRAPHIC IMAGES OF THE DEVICE(S) OR A VIDEO RECORDING OF THE PROCEDURE AVAILABLE FOR ISI REVIEW."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "3410 CENTRAL EXPRESSWAY",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "",
      "manufacturer_contact_exchange": "408",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": "07/23/2021;05/24/2023"
    }
  ]
}