  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "",
      "report_to_fda": "",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "CA",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "SANTA CLARA",
      "manufacturer_contact_address_1": "3410 CENTRAL EXPRESSWAY",
      "manufacturer_contact_pcity": "40852321",
      "event_type": "Injury",
      "report_number": "2955842-2021-10592",
      "type_of_report": [
        "Initial submission",
        "Followup"
      ],
      "product_problem_flag": "Y",
      "date_received": "20210531",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K143132",
      "date_of_event": "20210504",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "",
      "reporter_occupation_code": "NURSE",
      "manufacturer_contact_plocal": "4085232100",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "WANG",
      "source_type": [
        "Other",
        "Health Professional"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "",
      "manufacturer_g1_state": "CA",
      "reporter_country_code": "US",
      "manufacturer_contact_area_code": "408",
      "date_added": "20210531",
      "manufacturer_contact_f_name": "DAVID",
      "device_date_of_manufacturer": "20201008",
      "previous_use_code": "I",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "802534",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20210531",
          "brand_name": "DA VINCI",
          "generic_name": "HARMONIC ACE",
          "manufacturer_d_name": "INTUITIVE SURGICAL, INC",
          "manufacturer_d_address_1": "3410 CENTRAL EXPRESSWAY",
          "manufacturer_d_address_2": "",
          "manufacturer_d_city": "SANTA CLARA",
          "manufacturer_d_state": "CA",
          "manufacturer_d_zip_code": "",
          "manufacturer_d_zip_code_ext": "",
          "manufacturer_d_country": "",
          "manufacturer_d_postal_code": "",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "480275-08",
          "catalog_number": "480275",
          "lot_number": "M90201011 0127",
          "other_id_number": "",
          "device_availability": "No",
          "device_report_product_code": "NAY",
          "device_age_text": "DA",
          "device_evaluated_by_manufacturer": "R",
          "combination_product_flag": "N",
          "udi_di": "10886874112738",
          "udi_public": "(01)10886874112738(17)221031(10)M90201011",
          "openfda": {
            "device_name": "System, Surgical, Computer Controlled Instrument",
            "medical_specialty_description": "Gastroenterology, Urology",
            "regulation_number": "876.1500",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Detachment of Device or Device Component",
        "Detachment of Device or Device Component"
      ],
      "reporter_state_code": "OH",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "05/04/2021",
      "mfr_report_type": "Initial",
      "manufacturer_contact_country": "",
      "date_changed": "20250819",
      "health_professional": "Y",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "4085",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20210531",
          "sequence_number_treatment": [
            "DA VINCI INSTRUMENTS AND ACCESSORIES"
          ],
          "sequence_number_outcome": [
            ""
          ],
          "patient_age": "",
          "patient_sex": "Unknown",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "No Clinical Signs, Symptoms or Conditions",
            "No Clinical Signs, Symptoms or Conditions"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20210504",
      "initial_report_to_fda": "",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "",
      "manufacturer_contact_city": "SANTA CLARA",
      "mdr_report_key": "11906816",
      "removal_correction_number": "N/A",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20210504",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        "Other"
      ],
      "manufacturer_g1_zip_code": "",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "INTUITIVE SURGICAL, INC",
      "adverse_event_flag": "Y",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "Y",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "336071100",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "BASED ON A RE-EVALUATION OF THE COMPLAINT INFORMATION, THIS COMPLAINT HAS BEEN RECLASSIFIED AS AN ADVERSE EVENT AND PRODUCT PROBLEM RATHER THAN JUST A PRODUCT PROBLEM, AS PREVIOUSLY REPORTED."
        },
        {
          "mdr_text_key": "254825386",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "1",
          "text": "INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE HARMONIC ACE INSTRUMENT FOR EVALUATION. THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE HAS NOT BEEN DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/ OR IF ADDITIONAL INFORMATION IS RECEIVED. A REVIEW OF THE SITE'S COMPLAINT HISTORY DOES NOT SHOW ANY ADDITIONAL COMPLAINTS RELATED TO THIS PRODUCT. NO IMAGE OR VIDEO CLIP FOR THE REPORTED EVENT WAS SUBMITTED FOR REVIEW. PER THE LOG REVIEW, THE INSTRUMENT WAS LAST USED ON (B)(6) 2021 ON SYSTEM (B)(4), FOR 00:19:24 SEC. THE HARMONIC ACE INSTRUMENT IS A SINGLE-USE INSTRUMENT SO THEREFORE HAD NO USES REMAINING. THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, IT WAS ALLEGED THAT THE HARMONIC ACE SHEAR FELL INSIDE THE PATIENT. ALTHOUGH, THE BROKEN PIECE WAS RETRIEVED, NO PATIENT HARM, AND ADVERSE OUTCOME OR INJURY WAS REPORTED, IF THE FAILURE WERE TO RECUR, IT COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT."
        },
        {
          "mdr_text_key": "254825387",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "1",
          "text": "IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED HIATAL HERNIA-PARAESOPHAGEAL SURGICAL PROCEDURE, PART OF THE SHEAR OF THE HARMONIC ACE INSTRUMENT BROKE AND FELL INTO THE PATIENT. THE PIECE WAS RETRIEVED FROM THE ABDOMEN. THE CUSTOMER OPENED ANOTHER HARMONIC ACE INSTRUMENT TO COMPLETE THE CASE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "3410 CENTRAL EXPRESSWAY",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "",
      "manufacturer_contact_exchange": "408",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": "05/24/2023"
    }
  ]
}