  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "I",
      "report_to_fda": "N",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "",
      "manufacturer_contact_address_1": "",
      "manufacturer_contact_pcity": "",
      "event_type": "Malfunction",
      "report_number": "1040777-2017-00037",
      "type_of_report": [
        "Initial submission",
        "Followup"
      ],
      "product_problem_flag": "Y",
      "date_received": "20171129",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K972614",
      "date_of_event": "20171025",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "",
      "reporter_occupation_code": "MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE",
      "manufacturer_contact_plocal": "",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "",
      "source_type": [
        "USER FACILITY"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "",
      "manufacturer_g1_state": "",
      "reporter_country_code": "US",
      "manufacturer_contact_area_code": "",
      "date_added": "20171129",
      "manufacturer_contact_f_name": "",
      "previous_use_code": "",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "845427",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20171129",
          "brand_name": "MILLENNIUM OXYGEN CONCENTRATOR",
          "generic_name": "GENERATOR, OXYGEN, PORTABLE",
          "manufacturer_d_name": "RESPIRONICS, INC.",
          "manufacturer_d_address_1": "175 CHASTIAN MEADOWS COURT",
          "manufacturer_d_address_2": "",
          "manufacturer_d_city": "KENNESAW",
          "manufacturer_d_state": "GA",
          "manufacturer_d_zip_code": "30144",
          "manufacturer_d_zip_code_ext": "3724",
          "manufacturer_d_country": "US",
          "manufacturer_d_postal_code": "301443724",
          "device_operator": "LAY USER/PATIENT",
          "model_number": "M10605",
          "catalog_number": "M10605",
          "lot_number": "",
          "other_id_number": "",
          "device_availability": "Device was returned to manufacturer",
          "date_returned_to_manufacturer": "20171212",
          "device_report_product_code": "CAW",
          "device_age_text": "DA",
          "device_evaluated_by_manufacturer": "",
          "combination_product_flag": "N",
          "udi_di": "00606959015685",
          "udi_public": "00606959015685",
          "openfda": {
            "device_name": "Generator, Oxygen, Portable",
            "medical_specialty_description": "Anesthesiology",
            "regulation_number": "868.5440",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Fire"
      ],
      "reporter_state_code": "CA",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "02/05/2018",
      "mfr_report_type": "",
      "manufacturer_contact_country": "",
      "date_changed": "20250814",
      "health_professional": "",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20171129",
          "sequence_number_treatment": [
            ""
          ],
          "sequence_number_outcome": [
            ""
          ],
          "patient_age": "",
          "patient_sex": "",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "No Patient Involvement",
            "No Known Impact Or Consequence To Patient"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20171107",
      "initial_report_to_fda": "",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "",
      "manufacturer_contact_city": "",
      "mdr_report_key": "7067752",
      "removal_correction_number": "",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20171107",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        ""
      ],
      "manufacturer_g1_zip_code": "",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "",
      "adverse_event_flag": "N",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "93305265",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "1",
          "text": "THE MANUFACTURER RECEIVED INFORMATION ALLEGING A MILLENNIUM OXYGEN CONCENTRATOR CAUGHT ON FIRE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE HAS YET TO BE RETURNED TO THE MANUFACTURER'S SERVICE CENTER. A FOLLOW UP FINAL REPORT WILL BE FILED WHEN THE MANUFACTURER HAS COMPLETED THE INVESTIGATION."
        },
        {
          "mdr_text_key": "99332509",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "1",
          "text": "THE MANUFACTURER PREVIOUSLY REPORTED A MILLENNIUM OXYGEN CONCENTRATOR ALLEGEDLY CAUGHT ON FIRE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S SERVICE CENTER FOR EVALUATION. THE TECHNICIAN FOUND THE COMPRESSOR WIRES WERE CUT CAUSED BY THE MOUNTING BRACKETS OF THE COMPRESSOR. THERE WAS EVIDENCE OF THERMAL DAMAGE TO THE COMPRESSOR BRACKET. THE ENCLOSURE OF THE DEVICE WAS NOT COMPROMISED."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "",
      "manufacturer_contact_exchange": "",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": "02/08/2018"
    }
  ]
}