{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 132
    }
  },
  "results": [
    {
      "manufacturer_contact_zip_ext": "",
      "manufacturer_g1_address_2": "",
      "event_location": "I",
      "report_to_fda": "*",
      "manufacturer_contact_t_name": "",
      "manufacturer_contact_state": "",
      "manufacturer_link_flag": "Y",
      "manufacturer_contact_address_2": "",
      "manufacturer_g1_city": "",
      "manufacturer_contact_address_1": "",
      "manufacturer_contact_pcity": "",
      "event_type": "Malfunction",
      "report_number": "1717344-2020-00558",
      "type_of_report": [
        "Initial submission",
        "Followup"
      ],
      "product_problem_flag": "Y",
      "date_received": "20200601",
      "manufacturer_address_2": "",
      "pma_pmn_number": "K151649",
      "date_of_event": "20200511",
      "reprocessed_and_reused_flag": "N",
      "manufacturer_address_1": "",
      "exemption_number": "",
      "manufacturer_contact_zip_code": "",
      "reporter_occupation_code": "BIOMEDICAL ENGINEER",
      "manufacturer_contact_plocal": "",
      "noe_summarized": "1",
      "manufacturer_contact_l_name": "",
      "source_type": [
        "USER FACILITY"
      ],
      "distributor_zip_code_ext": "",
      "manufacturer_g1_postal_code": "",
      "manufacturer_g1_state": "",
      "reporter_country_code": "US",
      "manufacturer_contact_area_code": "",
      "date_added": "20200601",
      "manufacturer_contact_f_name": "",
      "previous_use_code": "",
      "device": [
        {
          "device_event_key": "",
          "implant_flag": "",
          "date_removed_flag": "",
          "device_sequence_number": "569440",
          "implant_date_year": "",
          "date_removed_year": "",
          "serviced_by_3rd_party_flag": "",
          "date_received": "20200601",
          "brand_name": "VALLEYLAB FT10",
          "generic_name": "ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES",
          "manufacturer_d_name": "COVIDIEN MFG DC BOULDER",
          "manufacturer_d_address_1": "5920 LONGBOW DR",
          "manufacturer_d_address_2": "",
          "manufacturer_d_city": "BOULDER",
          "manufacturer_d_state": "CO",
          "manufacturer_d_zip_code": "80301",
          "manufacturer_d_zip_code_ext": "3299",
          "manufacturer_d_country": "US",
          "manufacturer_d_postal_code": "803013299",
          "device_operator": "HEALTH PROFESSIONAL",
          "model_number": "VLFT10GEN",
          "catalog_number": "VLFT10GEN",
          "lot_number": "",
          "other_id_number": "",
          "device_availability": "Device was returned to manufacturer",
          "date_returned_to_manufacturer": "20200520",
          "device_report_product_code": "GEI",
          "device_age_text": "DA",
          "device_evaluated_by_manufacturer": "",
          "combination_product_flag": "N",
          "udi_di": "10884521516328",
          "udi_public": "10884521516328",
          "openfda": {
            "device_name": "Electrosurgical, Cutting & Coagulation & Accessories",
            "medical_specialty_description": "General, Plastic Surgery",
            "regulation_number": "878.4400",
            "device_class": "2"
          }
        }
      ],
      "product_problems": [
        "Self-Activation or Keying",
        "Self-Activation or Keying"
      ],
      "reporter_state_code": "TN",
      "manufacturer_zip_code": "",
      "suppl_dates_mfr_received": "08/04/2020",
      "mfr_report_type": "Thirty-Day",
      "manufacturer_contact_country": "",
      "date_changed": "20250812",
      "health_professional": "",
      "summary_report_flag": "N",
      "manufacturer_g1_zip_code_ext": "",
      "manufacturer_contact_extension": "",
      "manufacturer_city": "",
      "manufacturer_contact_phone_number": "",
      "patient": [
        {
          "patient_sequence_number": "1",
          "date_received": "20200601",
          "sequence_number_treatment": [
            ""
          ],
          "sequence_number_outcome": [
            ""
          ],
          "patient_age": "",
          "patient_sex": "",
          "patient_weight": "",
          "patient_ethnicity": "",
          "patient_race": "",
          "patient_problems": [
            "No Patient Involvement",
            "No Patient Involvement"
          ]
        }
      ],
      "distributor_city": "",
      "date_report": "20200827",
      "initial_report_to_fda": "",
      "distributor_state": "",
      "event_key": "",
      "manufacturer_g1_country": "",
      "manufacturer_contact_city": "",
      "mdr_report_key": "10105538",
      "removal_correction_number": "",
      "number_devices_in_event": "",
      "date_manufacturer_received": "20200511",
      "manufacturer_name": "",
      "report_source_code": "Manufacturer report",
      "remedial_action": [
        ""
      ],
      "manufacturer_g1_zip_code": "",
      "report_to_manufacturer": "",
      "manufacturer_zip_code_ext": "",
      "manufacturer_g1_name": "",
      "adverse_event_flag": "N",
      "distributor_address_1": "",
      "manufacturer_state": "",
      "distributor_address_2": "",
      "manufacturer_postal_code": "",
      "single_use_flag": "",
      "manufacturer_country": "",
      "mdr_text": [
        {
          "mdr_text_key": "193064856",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "1",
          "text": "IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED."
        },
        {
          "mdr_text_key": "193064857",
          "text_type_code": "Description of Event or Problem",
          "patient_sequence_number": "1",
          "text": "ACCORDING TO THE REPORTER, PRIOR TO USE, IT MISFIRED THE OUTPUT WITHOUT THE DOCTOR ACTIVATING THE UNIT. THERE WAS NO PATIENT INVOLVEMENT."
        },
        {
          "mdr_text_key": "319307587",
          "text_type_code": "Additional Manufacturer Narrative",
          "patient_sequence_number": "0",
          "text": "H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. IT WAS REPORTED THAT THE DEVICE ACTIVATED BY ITSELF WITHOUT USER INPUT. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY ROOT CAUSE COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. THE PRODUCT ANALYSIS DETECTED AN ADDITIONAL CONDITION THAT WAS NOT RELATED TO THE REPORTED ISSUE: THE UNIT HAD A CIRCUIT BOARD FAILURE. THE MOST LIKELY CAUSE FOR THE ADDITIONAL CONDITION IS COMPONENT FAILURE. THE PRODUCT ANALYSIS DETECTED ANOTHER CONDITION THAT WAS NOT RELATED TO THE REPORTED ISSUE: THE UNIT HAD A RECEPTACLE FAILURE. THE MOST LIKELY CAUSE FOR THE ADDITIONAL CONDITION IS DESIGN FAILURE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE ALSO THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED."
        }
      ],
      "number_patients_in_event": "",
      "distributor_name": "",
      "manufacturer_g1_address_1": "",
      "distributor_zip_code": "",
      "manufacturer_contact_postal_code": "",
      "manufacturer_contact_exchange": "",
      "manufacturer_contact_pcountry": "",
      "suppl_dates_fda_received": "08/27/2020"
    }
  ]
}