      "country": "United States",
      "city": "Newark",
      "address_1": "500 GBC Drive, Mailstop 514",
      "reason_for_recall": "Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085.      Quality Control (QC) materials have not exhibited low recovery of the same magnitude.    Internal method regression testing versus LC/MS has shown an average 25% low bias compared to LC/MS across the assay range.  Individual patients, particularly samples with lower TACR values",
      "address_2": "PO BOX 6101",
      "product_quantity": "2,976",
      "code_info": "Lot numbers FA3085, exp 2013-03-26; and FA3316, exp 2013-11-12",
      "center_classification_date": "20130315",
      "distribution_pattern": "Product was distributed within the United States to AL, AZ, CA, DC, FL, LA, MD, MI, MO, NJ, NM, NC, OH, OR, PA, TN, TX, UT, VA, and WA.",
      "state": "DE",
      "product_description": "Siemens Dimension(R) Tacrolimus Flex(R) reagent cartridge (DF107). The TACR method is used as an in vitro diagnostic test.",
      "report_date": "20130327",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-0957-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63767",
      "termination_date": "20140804",
      "recall_initiation_date": "20121127",
      "postal_code": "19714-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}