  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-05",
    "results": {
      "skip": 3200,
      "limit": 1,
      "total": 39677
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97539",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "3651 Birchwood Dr",
      "address_2": "",
      "postal_code": "60085-8337",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.",
      "recall_number": "Z-0208-2026",
      "product_description": "Salem Sump PVC Tubes:    Product Code/Product Name:  8888265108\tSalem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36  (91 cm)  8888265116\tSalem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48  (122 cm)  8888265124\tSalem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48  (122 cm)  8888265132\tSalem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48  (122 cm)  8888265140\tSalem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 18 Fr/Ch (6.0 mm), 48  (122 cm)  8888266213\tSalem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 10 Fr/Ch (3.3 mm) x 36  (91 cm)  8888266221\tSalem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 12 Fr/Ch (4.0 mm) x 48  (122 cm)  8888266247\tSalem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 14 Fr/Ch (4.7 mm) x 48  (122 cm)  8888266262\tSalem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 16 Fr/Ch (5.3 mm) x 48  (122 cm)  8888266288\tSalem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 18 Fr/Ch (6.0 mm) x 48  (122 cm)",
      "product_quantity": "all lots",
      "reason_for_recall": "Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.",
      "recall_initiation_date": "20250918",
      "center_classification_date": "20251017",
      "report_date": "20251029",
      "code_info": "Product Code/Product Name/UDI-DI:  8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36  (91 cm) 10884521000766 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48  (122 cm) 10884521020078 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48  (122 cm) 10884521020085 8888265132 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48  (122 cm) 10884521020092 8888265140 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 18 Fr/Ch (6.0 mm), 48  (122 cm) 10884521020108 8888266213 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 10 Fr/Ch (3.3 mm) x 36  (91 cm) 10192253012781 8888266221 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 12 Fr/Ch (4.0 mm) x 48  (122 cm) 10884521000803 8888266247 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 14 Fr/Ch (4.7 mm) x 48  (122 cm) 10884521000827 8888266262 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 16 Fr/Ch (5.3 mm) x 48  (122 cm) 10884521000841 8888266288 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 18 Fr/Ch (6.0 mm) x 48  (122 cm) 10884521000858  All Lots",
      "more_code_info": ""
    }
  ]
}