  "results": [
    {
      "status": "Ongoing",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96943",
      "recalling_firm": "Edwards Lifesciences, LLC",
      "address_1": "1 Edwards Way",
      "address_2": "N/A",
      "postal_code": "92614-5688",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA,  ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA,  FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY,  LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI,  MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO,  NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece,  Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta,  Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania,  Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan,  Thailand, United Arab Emirates (UAE), United Kingdom.",
      "recall_number": "Z-2083-2025",
      "product_description": "Brand Name: FEM-FLEX II  Product Name: Femoral Arterial Cannula  Model/Catalog Number: FEMII016AS  Software Version: N/A  Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours.  Component: No",
      "product_quantity": "847",
      "reason_for_recall": "due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.",
      "recall_initiation_date": "20250514",
      "center_classification_date": "20250710",
      "report_date": "20250716",
      "code_info": "Lot Code: Model No; FEMII016AS; UDI: 0060913168341; All lots up to: BSLC1689 Serial Numbers: FEMII016AS FEMII016AS FEMII016AS",
      "more_code_info": ""
    }
  ]
}