{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-19",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96943",
      "recalling_firm": "Edwards Lifesciences, LLC",
      "address_1": "1 Edwards Way",
      "address_2": "N/A",
      "postal_code": "92614-5688",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA,  ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA,  FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY,  LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI,  MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO,  NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece,  Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta,  Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania,  Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan,  Thailand, United Arab Emirates (UAE), United Kingdom.",
      "recall_number": "Z-2082-2025",
      "product_description": "Brand Name: FEM-FLEX II  Product Name: Femoral Arterial Cannula  Model/Catalog Number: FEMII016A  Software Version: N/A  Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours.  Component: No",
      "product_quantity": "6801",
      "reason_for_recall": "due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.",
      "recall_initiation_date": "20250514",
      "center_classification_date": "20250710",
      "report_date": "20250716",
      "code_info": "Lot Code: Model No: FEMII016A; UDI: 00690103031232; All lots up to: BSLC8179 Serial Numbers: 130544 130757 131105 131313 131511 131535 131888 131925 132504 BSLC0184 BSLC0386 BSLC0596 BSLC1311 BSLC3741 BSLC4162 BSLC5040 BSLC6487 BSLC8179",
      "more_code_info": ""
    }
  ]
}