  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-19",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96208",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.",
      "recall_number": "Z-1046-2025",
      "product_description": "Optima XR646 HD. X-Ray imaging system.",
      "product_quantity": "3149 (1697 US; 1452 OUS)",
      "reason_for_recall": "The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.",
      "recall_initiation_date": "20250128",
      "center_classification_date": "20250203",
      "report_date": "20250212",
      "code_info": "GTIN: 00195278526892",
      "more_code_info": ""
    }
  ]
}